MAKO RIO THA Application User Guides, for total hip arthroplasty, orthopedic surgery.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Item #: 204863 - THA 1.0 Application User Guide Part Item #: 207116 - THA 2.0 Application User Guide Part Item #: 208692 - THA 2.1 Application User Guide Part Item #: 209711 - MAKOplasty THA Application User Guide Part Item #: 210555 - MAKIplast THA 3.1 Application User Guide Part Item#: 212026 - THA Application User Guide 2.1-2.1.1 THA 3.0, THA 3.1, 3.1.1 and 3.1.1.1 Part Item # 204863 THA 1.0 Application User Guide Part Item # 207116 THA 2.0 Application User Guide Part Item # 208692 THA 2.1 Application User Guide Part Item # 209711 MAKOplasty THA Application User Guide Part Item # 210555 MAKIplast THA 3.1 Application User Guide Part Item # 212026 THA Application User Guide
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Mako Surgical Corporation
- Reason for Recall:
- Five (5) AFMEA Risk Control Measures are missing from the MAKO RIO THA Application User Guides
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
MAKO RIO THA Application User Guides, for total hip arthroplasty, orthopedic surgery.
Product Codes/Lot Numbers:
Part Item #: 204863 - THA 1.0 Application User Guide Part Item #: 207116 - THA 2.0 Application User Guide Part Item #: 208692 - THA 2.1 Application User Guide Part Item #: 209711 - MAKOplasty THA Application User Guide Part Item #: 210555 - MAKIplast THA 3.1 Application User Guide Part Item#: 212026 - THA Application User Guide 2.1-2.1.1 THA 3.0, THA 3.1, 3.1.1 and 3.1.1.1 Part Item # 204863 THA 1.0 Application User Guide Part Item # 207116 THA 2.0 Application User Guide Part Item # 208692 THA 2.1 Application User Guide Part Item # 209711 MAKOplasty THA Application User Guide Part Item # 210555 MAKIplast THA 3.1 Application User Guide Part Item # 212026 THA Application User Guide
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1411-2017
Related Recalls
Stryker USB Converter, Polaris Spectra Camera, Catalog #201148, a component of the Mako Robotic-Arm Assisted Surgery System.
Mako Surgical
Certain units have the potential to exhibit infant failures whereby units either arrive non-functional, have the potential to become non-functional, or flash an error light while in use.
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.
Characterization issue associated with the product. This issue may cause a discrepancy during bone preparation in certain Mako Total Hip procedures.