Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Code: Software versions and UDIs 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206 12.1, UDI: (01)00842000100928(10)12.1(11)210122; 12.1.1, UDI: (01)00842000100928(10)12.1.1(11)211220; 12.2 UDI: (01)00842000100065(10)12.2.0(11)220713; 12.3, UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.1 UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.2 UDI: (01)00842000101109(10)12.3.2(11)230612; 12.3.3 UDI: (01)00842000101109(10)12.3.3(11)230913; 12.4 UDI: (01)00842000101123(10)12.4.0(11)231220; 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522; 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Merge Healthcare, Inc.
Reason for Recall:
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.

Product Codes/Lot Numbers:

Lot Code: Software versions and UDIs 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206 12.1, UDI: (01)00842000100928(10)12.1(11)210122; 12.1.1, UDI: (01)00842000100928(10)12.1.1(11)211220; 12.2 UDI: (01)00842000100065(10)12.2.0(11)220713; 12.3, UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.1 UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.2 UDI: (01)00842000101109(10)12.3.2(11)230612; 12.3.3 UDI: (01)00842000101109(10)12.3.3(11)230913; 12.4 UDI: (01)00842000101123(10)12.4.0(11)231220; 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522; 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206

Distribution:

Distributed in: US

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1402-2025

Related Recalls

Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.

Aug 11, 2025 Diagnostic Equipment Nationwide View Details →

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Feb 19, 2025 Infusion Pumps View Details →

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Feb 19, 2025 Infusion Pumps View Details →