Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Elekta, Inc.
- Reason for Recall:
- Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Product Codes/Lot Numbers:
325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1402-2013
Related Recalls
Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.