Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-503 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Numbers: WAA090094, WAA090112, WAA090409, WAA090521, WAA090757, WAA090805, WAA091170, WAA091288, WAA091292, WAA091454, WAA091460 WAA091463, WAA091477. UDI Code 07340153700048
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Maquet Cardiovascular Us Sales, Llc
Reason for Recall:
Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-503 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.

Product Codes/Lot Numbers:

Serial Numbers: WAA090094, WAA090112, WAA090409, WAA090521, WAA090757, WAA090805, WAA091170, WAA091288, WAA091292, WAA091454, WAA091460 WAA091463, WAA091477. UDI Code 07340153700048

Distribution:

Distributed in: CA, CO, IA, MN, NY, SC, VA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1391-2020

Related Recalls

The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.

Sep 30, 2020 Diagnostic Equipment Nationwide View Details →