CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 7.5MM(P) X 4MM(H) Reference Number: IEHA574

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Numbers: 1229061
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Biomet 3i, LLC
Reason for Recall:
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin contour misalignment
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 7.5MM(P) X 4MM(H) Reference Number: IEHA574

Product Codes/Lot Numbers:

Lot Numbers: 1229061

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1388-2020

Related Recalls

Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

Aug 31, 2020 Other Medical Devices View Details →

Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

Aug 31, 2020 Other Medical Devices View Details →

Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

Aug 31, 2020 Other Medical Devices View Details →