da Vinci S / Si Surgical System IS2000 / IS3000, Instrument Arm Drape, 20 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product number 420015-03; da Vinci S / Si Instrument Arm Drape, 20 Pack. Lot Numbers for Cloudy/Waxy issue: ALL lot numbers Lot Numbers for Tear issue: D142175, D142535, D142895, D143255, D150135A, D150555, DA142695A, DA143465A, D142175A, D142535A, D142895A, D143255A, D150145, D150555A, DA142725, DA143515, D142185, D142545, D142905, D143285, D150145A, D150565, DA142745, DA150075, D142185A, D142545A, D142905A, D143285A, D150155, D150565A, DA142745A, DA150085, D142195, D142555, D142915, D143295, D150155A, D150575, DA142755, DA150085A, D142195A, D142555A, D142915A, D143295A, D150165, D150575A, DA142755A, DA150125, D142205, D142585, D142935, D143305, D150165A, D150625, DA142765, DA150125A, D142205A, D142585A, D142935A, D143305A, D150195, D150625A, DA142795, DA150135, D142215, D142595, D142945, D143315, D150195A, D150635, DA142795A, DA150135A, D142235, D142595A, D142945, D143315, D150205, D150635A, DA142825, DA150195, D142235A, D142605, D142955, D143325, D150205A, D150645, DA142835, DA150205, D142245, D142605A, D142955A, D143325A, D150215, D150645A, DA142835A, DA150205A, D142245A, D142615, D142965, D143335, D150215A, D150655, DA142885, DA150215, D142255, D142615A, D142965A, D143355, D150225, D150655A, DA142885A, DA150275, D142255A, D142625, D142975, D143355A, D150225A, D150685, DA142905, DA150275A, D142265, D142625A, D142975A, D143365, D150235, D150685A, DA142955, DA150285, D142265A, D142635, D142985, D143365A, D150235A, D150695, DA142965, DA150285A, D142275, D142655, D143005, D143375, D150275, D150695A, DA142965A, DA150295, D142275A, D142655A, D143005A, D143375A, D150275A, DA142185, DA142975, DA150335, D142305, D142665, D143015, D143385, D150285, DA142195, DA143005, DA150345, D142305A, D142665A, D143015A, D143385A, D150285A, DA142205, DA143025, DA150345A, D142315, D142675, D143025, D143395, D150295, DA142235, DA143025A, DA150365, D142315A, D142675A, D143025A, D143395A, D150295A, DA142235A, DA143035, DA150405, D142325, D142685, D143035, D143405, D150305, DA142245, DA143035A, DA150405A, D142325A, D142685A, D143035A, D143425, D150305A, DA142255, DA143085, DA150415, D142335, D142695, D143045, D143425A, D150335, DA142275, DA143095, DA150425, D142335A, D142695A, D143045A, D143435, D150335A, DA142305, DA143105, DA150435, D142345A, D142725, D143075, D143435A, D150345, DA142305A, DA143105A, DA150435A, D142375, D142725A, D143075A, D143445, D150345A, DA142315, DA143115, DA150475, D142375A, D142735, D143085, D143445A, D150355, DA142345, DA143155, DA150475A, D142385, D142735A, D143085A, D143455, D150355A, DA142345A, DA143155A, DA150495, D142385A, D142745, D143095, D143455A, D150375, DA142375, DA143165, DA150495A, D142395, D142745A, D143095A, D143465, D150405, DA142375A, DA143175, DA150515, D142395A, D142755, D143105, D143465A, D150405A, DA142385, DA143175A, DA150515A, D142405, D142755A, D143105A, D143495, D150415, DA142395, DA143185, DA150545, D142405A, D142765, D143115, D143495A, D150415A, DA142455, DA143215, DA150545A, D142415, D142765A, D143115A, D143505, D150425, DA142465, DA143235, DA150555, D142415A, D142795, D143155, D143505A, D150425A, DA142485, DA143235A, DA150555A, D142445, D142795A, D143155A, D143515, D150435, DA142515, DA143245, DA150635, D142445A, D142805, D143165, D143515A, D150435A, DA142515A, DA143295, DA150645, D142455, D142805A, D143165A, D150035, D150435A, DA142525, DA143295A, DA150645, D142455A, D142815, D143175, D150065, D150475, DA142535, DA143315, DA150645A, D142465, D142815A, D143175A, D150065A, D150475A, DA142535A, DA143315A, DA150655, D142465A, D142825, D143185, D150075, D150485, DA142585, DA143365, DA150685A, D142475, D142825A, D143185A, D150075A, D150485A, DA142585A, DA143365A, DA150685A, D142475A, D142835, D143215, D150085, D150495, DA142605, DA143385, DA150695, D142485, D142835A, D143215A, D150085A, D150495A, DA142615, DA143385A, DA150695A, D142485A, D142865, D143225, D150095, D150505, DA142615A, DA143395, D142515, D142875, D143225A, D150095A, D150505A, DA142625, DA143435, D142515A, D142875A, D143235, D150105, D150515, DA142655, DA143445, D142525, D142885, D143235A, D150125, D150515A, DA142665, DA143445A, D142525, D142885A, D143245, D150125A, D150545, DA142685, DA143455, D142525A, D142885A, D143245A, D150135, D150545A, DA142695, DA143465
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Intuitive Surgical, Inc.
- Reason for Recall:
- Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to tear.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
da Vinci S / Si Surgical System IS2000 / IS3000, Instrument Arm Drape, 20 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.
Product Codes/Lot Numbers:
Product number 420015-03; da Vinci S / Si Instrument Arm Drape, 20 Pack. Lot Numbers for Cloudy/Waxy issue: ALL lot numbers Lot Numbers for Tear issue: D142175, D142535, D142895, D143255, D150135A, D150555, DA142695A, DA143465A, D142175A, D142535A, D142895A, D143255A, D150145, D150555A, DA142725, DA143515, D142185, D142545, D142905, D143285, D150145A, D150565, DA142745, DA150075, D142185A, D142545A, D142905A, D143285A, D150155, D150565A, DA142745A, DA150085, D142195, D142555, D142915, D143295, D150155A, D150575, DA142755, DA150085A, D142195A, D142555A, D142915A, D143295A, D150165, D150575A, DA142755A, DA150125, D142205, D142585, D142935, D143305, D150165A, D150625, DA142765, DA150125A, D142205A, D142585A, D142935A, D143305A, D150195, D150625A, DA142795, DA150135, D142215, D142595, D142945, D143315, D150195A, D150635, DA142795A, DA150135A, D142235, D142595A, D142945, D143315, D150205, D150635A, DA142825, DA150195, D142235A, D142605, D142955, D143325, D150205A, D150645, DA142835, DA150205, D142245, D142605A, D142955A, D143325A, D150215, D150645A, DA142835A, DA150205A, D142245A, D142615, D142965, D143335, D150215A, D150655, DA142885, DA150215, D142255, D142615A, D142965A, D143355, D150225, D150655A, DA142885A, DA150275, D142255A, D142625, D142975, D143355A, D150225A, D150685, DA142905, DA150275A, D142265, D142625A, D142975A, D143365, D150235, D150685A, DA142955, DA150285, D142265A, D142635, D142985, D143365A, D150235A, D150695, DA142965, DA150285A, D142275, D142655, D143005, D143375, D150275, D150695A, DA142965A, DA150295, D142275A, D142655A, D143005A, D143375A, D150275A, DA142185, DA142975, DA150335, D142305, D142665, D143015, D143385, D150285, DA142195, DA143005, DA150345, D142305A, D142665A, D143015A, D143385A, D150285A, DA142205, DA143025, DA150345A, D142315, D142675, D143025, D143395, D150295, DA142235, DA143025A, DA150365, D142315A, D142675A, D143025A, D143395A, D150295A, DA142235A, DA143035, DA150405, D142325, D142685, D143035, D143405, D150305, DA142245, DA143035A, DA150405A, D142325A, D142685A, D143035A, D143425, D150305A, DA142255, DA143085, DA150415, D142335, D142695, D143045, D143425A, D150335, DA142275, DA143095, DA150425, D142335A, D142695A, D143045A, D143435, D150335A, DA142305, DA143105, DA150435, D142345A, D142725, D143075, D143435A, D150345, DA142305A, DA143105A, DA150435A, D142375, D142725A, D143075A, D143445, D150345A, DA142315, DA143115, DA150475, D142375A, D142735, D143085, D143445A, D150355, DA142345, DA143155, DA150475A, D142385, D142735A, D143085A, D143455, D150355A, DA142345A, DA143155A, DA150495, D142385A, D142745, D143095, D143455A, D150375, DA142375, DA143165, DA150495A, D142395, D142745A, D143095A, D143465, D150405, DA142375A, DA143175, DA150515, D142395A, D142755, D143105, D143465A, D150405A, DA142385, DA143175A, DA150515A, D142405, D142755A, D143105A, D143495, D150415, DA142395, DA143185, DA150545, D142405A, D142765, D143115, D143495A, D150415A, DA142455, DA143215, DA150545A, D142415, D142765A, D143115A, D143505, D150425, DA142465, DA143235, DA150555, D142415A, D142795, D143155, D143505A, D150425A, DA142485, DA143235A, DA150555A, D142445, D142795A, D143155A, D143515, D150435, DA142515, DA143245, DA150635, D142445A, D142805, D143165, D143515A, D150435A, DA142515A, DA143295, DA150645, D142455, D142805A, D143165A, D150035, D150435A, DA142525, DA143295A, DA150645, D142455A, D142815, D143175, D150065, D150475, DA142535, DA143315, DA150645A, D142465, D142815A, D143175A, D150065A, D150475A, DA142535A, DA143315A, DA150655, D142465A, D142825, D143185, D150075, D150485, DA142585, DA143365, DA150685A, D142475, D142825A, D143185A, D150075A, D150485A, DA142585A, DA143365A, DA150685A, D142475A, D142835, D143215, D150085, D150495, DA142605, DA143385, DA150695, D142485, D142835A, D143215A, D150085A, D150495A, DA142615, DA143385A, DA150695A, D142485A, D142865, D143225, D150095, D150505, DA142615A, DA143395, D142515, D142875, D143225A, D150095A, D150505A, DA142625, DA143435, D142515A, D142875A, D143235, D150105, D150515, DA142655, DA143445, D142525, D142885, D143235A, D150125, D150515A, DA142665, DA143445A, D142525, D142885A, D143245, D150125A, D150545, DA142685, DA143455, D142525A, D142885A, D143245A, D150135, D150545A, DA142695, DA143465
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1387-2015
Related Recalls
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
8MM,MEGA SUTURECUT ND,IS4000 REF 471309
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.