Access Substrate, REF 81906, For use with the Access Immunoassay Systems
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI 15099590229986, Lot Numbers: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, 338538. UPDATED - ADDITIONAL LOT NUMBERS PER EXPANSION EMAIL DATED 10/30/2024: 234600, 234633, 234682, 234706, 234721, 234759, 234790, 234801, 234834, 234862, 337904, 337930, 337935, 337959, 337984, 338040, 338065, 338119, 338163, 338214, 338229, 338254, 338270, 338289, 338322, 338333, 338346, 338420, 338435, 33845, 33848, 338511, 338538, 338566, 338587, 338621, 338622, 338698, 338719, 338720, 338738, 338753, 338762, 338783, 338860, 338996, 339040, 339041, 339056, 339107, 339106, 339146, 439113, 439111, 439157, 439229, 439239, 439285¿.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter, Inc.
- Reason for Recall:
- On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman Coulter has identified that Access Substrate bottles in the identified lots were released from manufacturing without labels. Missing label may cause some patient results reporting delay but will not cause any errorreous results. However, because labs are required to have redundancy such that delay would not significantly impact clinical management and because the defect is easily recognized by the user, the probability of harm is remote.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Access Substrate, REF 81906, For use with the Access Immunoassay Systems
Product Codes/Lot Numbers:
UDI/DI 15099590229986, Lot Numbers: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, 338538. UPDATED - ADDITIONAL LOT NUMBERS PER EXPANSION EMAIL DATED 10/30/2024: 234600, 234633, 234682, 234706, 234721, 234759, 234790, 234801, 234834, 234862, 337904, 337930, 337935, 337959, 337984, 338040, 338065, 338119, 338163, 338214, 338229, 338254, 338270, 338289, 338322, 338333, 338346, 338420, 338435, 33845, 33848, 338511, 338538, 338566, 338587, 338621, 338622, 338698, 338719, 338720, 338738, 338753, 338762, 338783, 338860, 338996, 339040, 339041, 339056, 339107, 339106, 339146, 439113, 439111, 439157, 439229, 439239, 439285¿.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1376-2024
Related Recalls
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
MicroScan Neg MIC 3J REF C54814
Beckman Coulter
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics