GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP Endo-Exo Medullary System is used for the treatment of fractures or correction of deformities in the femur, tibia and humerus of pediatric patients, who have reached skeletal maturity. The GAP Nail is an intramedullary nail. It is the main component of the GAP Endo-Exo Medullary System. The rest of the components of the system are screws, plates and caps that are assembled to the nail during the surgical procedure. The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surgically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP System is used for pediatric patients (child and adolescent) ages 2 to 21. It can be used to correct the following conditions: -Diaphyseal fracture of the femur, tibia and humerus -Fractures of the femoral neck -Subtrochanteric, intertrochanteric and combination fractures -Correction of deformities (OI, Coxa vara, Coxa valga) -Nonunions and malunions
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot numbers: 110913-001; 110913-002; 110913-004; 110913-005;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Pega Medical Inc.
- Reason for Recall:
- Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred in patients with tibial pseudoarthrosis (CPT) or Osteogenesis imperfecta (OI). Higher failure rates have been detected in tibias with distal fractures that are close to the distal locking holes.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP Endo-Exo Medullary System is used for the treatment of fractures or correction of deformities in the femur, tibia and humerus of pediatric patients, who have reached skeletal maturity. The GAP Nail is an intramedullary nail. It is the main component of the GAP Endo-Exo Medullary System. The rest of the components of the system are screws, plates and caps that are assembled to the nail during the surgical procedure. The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surgically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP System is used for pediatric patients (child and adolescent) ages 2 to 21. It can be used to correct the following conditions: -Diaphyseal fracture of the femur, tibia and humerus -Fractures of the femoral neck -Subtrochanteric, intertrochanteric and combination fractures -Correction of deformities (OI, Coxa vara, Coxa valga) -Nonunions and malunions
Product Codes/Lot Numbers:
Lot numbers: 110913-001; 110913-002; 110913-004; 110913-005;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1362-2015
Related Recalls
The lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.
SLIM Handle - catalogue #: SLM-HND100 a component of Simple Locking Intramedullary (SLIM) System
Pega Medical
Handle could jam with the Driver due to the detachment of an internal Connector Ring.