TLC-2000 Therapeutic Medical Laser System
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 3245 to 3336 inclusive
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Theralase Inc.
- Reason for Recall:
- 905 nm laser diodes may be below rated power output.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
TLC-2000 Therapeutic Medical Laser System
Product Codes/Lot Numbers:
Serial Numbers: 3245 to 3336 inclusive
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1354-2020
Related Recalls
Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards
660 nm laser diodes may be below rated power output.
Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)