NUVASIVE FREEDOM: End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm,¿ +0mm; Packaged End Cap 8.5/10.7mm,¿ +5mm; Packaged End Cap 8.5/10.7mm,¿ +10mm; Packaged End Cap 8.5/10.7mm,¿ +15mm; Packaged End Cap 8.5/10.7mm,¿ +20mm; Packaged End Cap 12.5mm,¿ +0mm; Packaged End Cap 12.5mm,¿ +5mm; Packaged End Cap 12.5mm,¿ +10mm; Packaged End Cap 12.5mm,¿ +15mm; Packaged End Cap 12.5mm,¿ +20mm; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 4.0 mm X 65 mm Locking Screw, Peg; 4.0 mm X 70 mm Locking Screw, Peg; 4.0 mm X 75 mm Locking Screw, Peg; 4.0 mm X 80 mm Locking Screw, Peg; 4.0 mm X 85 mm Locking Screw, Peg; 4.0 mm X 90 mm Locking Screw, Peg; 4.0 mm X 95 mm Locking Screw, Peg; 4.0 mm X 100 mm Locking Screw, Peg; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm X 80 mm Locking Screw, Peg; 5.0 mm X 85 mm Locking Screw, Peg; 5.0 mm X 90 mm Locking Screw, Peg; 5.0 mm X 95 mm Locking Screw, Peg; 5.0 mm X 100 mm Locking Screw, Peg; Orthopedic intramedullary limb lengthening rod, residual limb lengthener, 14 mm diameter, 130 mm length; Orthopedic surgical instrument - for use with the intramedullary limb lengthening rod, residual limb lengthener;

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Lots manufactured between April 2015 to February 2021. NUVASIVE Freedom Product Codes: End Caps: CPA1-000; CPA2-000; CPA2-005; CPA2-010; CPA2-015; CPA2-020; CPA3-000; CPA3-005; CPA3-010; CPA3-015; CPA3-020; Locking screws: LSB4-020; LSB4-025; LSB4-030; LSB4-035; LSB4-040; LSB4-045; LSB4-050; LSB4-055; LSB4-060; LSB5-020; LSB5-025; LSB5-030; LSB5-035; LSB5-040; LSB5-045; LSB5-050; LSB5-055; LSB5-060; LSB5-065; LSB5-070; LSB5-075; LSC4-020; LSC4-025; LSC4-030; LSC4-035; LSC4-040; LSC4-045; LSC4-050; LSC4-055; LSC4-060; LSC4-065; LSC4-070; LSC4-075; LSC4-080; LSC4-085; LSC4-090; LSC4-095; LSC4-100; LSC5-020; LSC5-025; LSC5-030; LSC5-035; LSC5-040; LSC5-045; LSC5-050; LSC5-055; LSC5-060; LSC5-065; LSC5-070; LSC5-075; LSC5-080; LSC5-085; LSC5-090; LSC5-095; LSC5-100; Limb lengthening intramedullary rod, residual limb lengthener: R14-100A130; Orthopedic surgical instrument - for use with the intramedullary limb lengthening rod, residual limb lengthener: RLL1-000; GTIN Numbers: 00812258020002; 00887517001139; 00887517001146; 00887517001153; 00887517001160; 00887517001177; 00887517001184; 00887517001191; 00887517001207; 00887517001214; 00887517001221; 00856719002763; 00856719002770; 00856719002787; 00856719002794; 00856719002800; 00856719002817; 00856719002824; 00856719002831; 00856719002848; 00856719002886; 00856719002893; 00856719002909; 00856719002916; 00856719002923; 00856719002930; 00856719002947; 00856719002954; 00856719002961; 00856719002978; 00856719002985; 00856719002992; 00812258021429; 00812258021436; 00812258021443; 00812258021450; 00812258021467; 00812258021474; 00812258021481; 00812258021498; 00812258021504; 00887517000958; 00887517000965; 00887517000972; 00887517000989; 00887517000996; 00887517001009; 00887517001016; 00887517001023; 00812258021511; 00812258021528; 00812258021535; 00812258021542; 00812258021559; 00812258021566; 00812258021573; 00812258021580; 00812258021597; 00812258021603; 00812258021610; 00812258021627; 00887517001030; 00887517001047; 00887517001054; 00887517001061; 00887517001078; 00812258024239; 00812258026479;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Nuvasive Specialized Orthopedics Inc
Reason for Recall:
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

NUVASIVE FREEDOM: End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm,¿ +0mm; Packaged End Cap 8.5/10.7mm,¿ +5mm; Packaged End Cap 8.5/10.7mm,¿ +10mm; Packaged End Cap 8.5/10.7mm,¿ +15mm; Packaged End Cap 8.5/10.7mm,¿ +20mm; Packaged End Cap 12.5mm,¿ +0mm; Packaged End Cap 12.5mm,¿ +5mm; Packaged End Cap 12.5mm,¿ +10mm; Packaged End Cap 12.5mm,¿ +15mm; Packaged End Cap 12.5mm,¿ +20mm; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 4.0 mm X 65 mm Locking Screw, Peg; 4.0 mm X 70 mm Locking Screw, Peg; 4.0 mm X 75 mm Locking Screw, Peg; 4.0 mm X 80 mm Locking Screw, Peg; 4.0 mm X 85 mm Locking Screw, Peg; 4.0 mm X 90 mm Locking Screw, Peg; 4.0 mm X 95 mm Locking Screw, Peg; 4.0 mm X 100 mm Locking Screw, Peg; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm X 80 mm Locking Screw, Peg; 5.0 mm X 85 mm Locking Screw, Peg; 5.0 mm X 90 mm Locking Screw, Peg; 5.0 mm X 95 mm Locking Screw, Peg; 5.0 mm X 100 mm Locking Screw, Peg; Orthopedic intramedullary limb lengthening rod, residual limb lengthener, 14 mm diameter, 130 mm length; Orthopedic surgical instrument - for use with the intramedullary limb lengthening rod, residual limb lengthener;

Product Codes/Lot Numbers:

All Lots manufactured between April 2015 to February 2021. NUVASIVE Freedom Product Codes: End Caps: CPA1-000; CPA2-000; CPA2-005; CPA2-010; CPA2-015; CPA2-020; CPA3-000; CPA3-005; CPA3-010; CPA3-015; CPA3-020; Locking screws: LSB4-020; LSB4-025; LSB4-030; LSB4-035; LSB4-040; LSB4-045; LSB4-050; LSB4-055; LSB4-060; LSB5-020; LSB5-025; LSB5-030; LSB5-035; LSB5-040; LSB5-045; LSB5-050; LSB5-055; LSB5-060; LSB5-065; LSB5-070; LSB5-075; LSC4-020; LSC4-025; LSC4-030; LSC4-035; LSC4-040; LSC4-045; LSC4-050; LSC4-055; LSC4-060; LSC4-065; LSC4-070; LSC4-075; LSC4-080; LSC4-085; LSC4-090; LSC4-095; LSC4-100; LSC5-020; LSC5-025; LSC5-030; LSC5-035; LSC5-040; LSC5-045; LSC5-050; LSC5-055; LSC5-060; LSC5-065; LSC5-070; LSC5-075; LSC5-080; LSC5-085; LSC5-090; LSC5-095; LSC5-100; Limb lengthening intramedullary rod, residual limb lengthener: R14-100A130; Orthopedic surgical instrument - for use with the intramedullary limb lengthening rod, residual limb lengthener: RLL1-000; GTIN Numbers: 00812258020002; 00887517001139; 00887517001146; 00887517001153; 00887517001160; 00887517001177; 00887517001184; 00887517001191; 00887517001207; 00887517001214; 00887517001221; 00856719002763; 00856719002770; 00856719002787; 00856719002794; 00856719002800; 00856719002817; 00856719002824; 00856719002831; 00856719002848; 00856719002886; 00856719002893; 00856719002909; 00856719002916; 00856719002923; 00856719002930; 00856719002947; 00856719002954; 00856719002961; 00856719002978; 00856719002985; 00856719002992; 00812258021429; 00812258021436; 00812258021443; 00812258021450; 00812258021467; 00812258021474; 00812258021481; 00812258021498; 00812258021504; 00887517000958; 00887517000965; 00887517000972; 00887517000989; 00887517000996; 00887517001009; 00887517001016; 00887517001023; 00812258021511; 00812258021528; 00812258021535; 00812258021542; 00812258021559; 00812258021566; 00812258021573; 00812258021580; 00812258021597; 00812258021603; 00812258021610; 00812258021627; 00887517001030; 00887517001047; 00887517001054; 00887517001061; 00887517001078; 00812258024239; 00812258026479;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1351-2021

Related Recalls

Precice Max REF: PM10.0-80B265 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 265mm PM10.0-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 280mm PM10.0-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 305mm PM10.0-80B335 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 335mm PM10.0-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 365mm PM10.0-80TJ265 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 265mm PM10.0-80TJ280 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 280mm PM10.0-80TJ305 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 305mm PM11.5-80B265 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 265mm PM11.5-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 280mm PM11.5-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 305mm PM11.5-80B335 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 335mm PM11.5-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 365mm PM11.5-80TJ265 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 265mm PM11.5-80TJ280 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 280mm PM11.5-80TJ305 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 305mm PM13.0-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 280mm PM13.0-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 305mm PM13.0-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 365mm The Precice Max Nail is intended to stabilize fractures, nonunions, and malunions of the femur, tibia, and humerus. In addition, the Precice Nail is intended to provide magnetically controlled lengthening of long bones and/or transporting of bone segments in the femur, tibia, and humerus.

NuVasive Specialized Orthopedics

Class I - Dangerous

Due to potential issues with design, manufacturing and documentation issues related to the validation of the affected products/devices.

Sep 30, 2024 Other Medical Devices View Details →

During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

Nov 1, 2022 Infusion Pumps Nationwide View Details →

MAGEC 1 System

NuVasive Specialized Orthopedics

Class I - Dangerous

Safety Alert to all surgeons informing them of post-implantation fractures of an internal metallic component for device manufactured prior to March 26, 2015.

Jun 25, 2019 Other Medical Devices View Details →