Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Numbers: U010577, U016620, U019084
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
ulrich medical USA Inc
Reason for Recall:
Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Product Codes/Lot Numbers:

Lot Numbers: U010577, U016620, U019084

Distribution:

Distributed in: AZ, TX, GA, FL, AR, CA