CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Class I - DangerousWhat Should You Do?
- Check if you have this product: (domestic only) SN18061848 SN17071689 SN18051832 SN190251 SN17111759 SN17111752 SN18011793 10000020 SN17121783 SN17121785 SN18021816 SN18101890 SN19031950 SN17071688 SN18101903 SN17071683 SN17101722 SN17101729 SN17111758 SN17111763 SN17031622 SN17071692 SN17041642 SN18101886 SN18011794 SN19021934 SN19021943 SN17111770 SN18011802 SN19021942 SN17121779 SN17031618 SN17101733 SN17021613 SN17021613
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems Gmbh, DMC
- Reason for Recall:
- When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Product Codes/Lot Numbers:
(domestic only) SN18061848 SN17071689 SN18051832 SN190251 SN17111759 SN17111752 SN18011793 10000020 SN17121783 SN17121785 SN18021816 SN18101890 SN19031950 SN17071688 SN18101903 SN17071683 SN17101722 SN17101729 SN17111758 SN17111763 SN17031622 SN17071692 SN17041642 SN18101886 SN18011794 SN19021934 SN19021943 SN17111770 SN18011802 SN19021942 SN17121779 SN17031618 SN17101733 SN17021613 SN17021613
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1328-2020
Related Recalls
CombiDiagnost R90 Software Version R1.0 and R1.1
Philips Medical Systems Gmbh, DMC
Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.
CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Philips Medical Systems Gmbh, DMC
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.
Stationary Fluoroscopic X-ray System CombiDiagnost R90, Model 709030 Stationary Fluoroscopic X-ray System
Philips Medical Systems Gmbh, DMC
kV/mA lockin not functioning as specified which may result in the patient receiving an increased radiation dose.