PRECICE PLATE - Product Usage: intended for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Lots affected. Below are: Primary DI Number(GTIN)/Product Code (REF)/Product Description: 00887517021915 LP4.5-DLF30 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 0-Hole; 00887517021908 LP4.5-DLF31 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 1-Hole; 00887517021892 LP4.5-DLF32 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 2-Hole; 00887517021885 LP4.5-DLF3B PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, Bump; 00887517021878 LP4.5-DLF40 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 0-Hole; 00887517021861 LP4.5-DLF41 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 1-Hole; 00887517021854 LP4.5-DLF42 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 2-Hole; 00887517021847 LP4.5-DLF4B PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, Bump; 00887517021830 LP4.5-DRF30 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 0-Hole; 00887517021823 LP4.5-DRF31 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 1-Hole; 00887517021816 LP4.5-DRF32 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 2-Hole; 00887517021809 LP4.5-DRF3B PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, Bump; 00887517021793 LP4.5-DRF40 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 0-Hole; 00887517021786 LP4.5-DRF41 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 1-Hole; 00887517021779 LP4.5-DRF42 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 2-Hole; 00887517021762 LP4.5-DRF4B PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, Bump; 00887517021755 LP4.5-MLT30 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 0-Hole; 00887517021748 LP4.5-MLT31 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 1-Hole; 00887517021731 LP4.5-MLT32 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 2-Hole; 00887517021724 LP4.5-MLT3B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, Bump; 00887517021717 LP4.5-MLT40 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 0-Hole; 00887517021700 LP4.5-MLT41 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 1-Hole; 00887517021694 LP4.5-MLT42 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 2-Hole; 00887517021687 LP4.5-MLT4B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, Bump; 00887517021670 LP4.5-MRT30 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 0-Hole; 00887517021663 LP4.5-MRT31 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 1-Hole; 00887517021656 LP4.5-MRT32 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 2-Hole; 00887517021649 LP4.5-MRT3B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, Bump; 00887517021632 LP4.5-MRT40 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 0-Hole; 00887517021625 LP4.5-MRT41 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 1-Hole; 00887517021618 LP4.5-MRT42 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 2-Hole; 00887517021601 LP4.5-MRT4B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, Bump; 00887517021595 LP4.5-PLF30 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 0-Hole; 00887517021588 LP4.5-PLF31 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 1-Hole; 00887517021571 LP4.5-PLF32 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 2-Hole; 00887517021564 LP4.5-PLF3B PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, Bump; 00887517021557 LP4.5-PLF40 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 0-Hole; 00887517021540 LP4.5-PLF41 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 1-Hole; 00887517021533 LP4.5-PLF42 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 2-Hole; 00887517021526 LP4.5-PLF4B PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, Bump; 00887517021519 LP4.5-PLT30 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 0-Hole; 00887517021502 LP4.5-PLT31 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 1-Hole; 00887517021496 LP4.5-PLT32 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 2-Hole; 00887517021489 LP4.5-PLT3B PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, Bump; 00887517021472 LP4.5-PLT40 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 0-Hole; 00887517021465 LP4.5-PLT41 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 1-Hole; 00887517021458 LP4.5-PLT42 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 2-Hole; 00887517021441 LP4.5-PLT4B PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, Bump; 00887517021434 LP4.5-PRF30 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 0-Hole; 00887517021427 LP4.5-PRF31 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 1-Hole; 00887517021410 LP4.5-PRF32 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 2-Hole; 00887517021403 LP4.5-PRF3B PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, Bump; 00887517021397 LP4.5-PRF40 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 0-Hole; 00887517021380 LP4.5-PRF41 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 1-Hole; 00887517021373 LP4.5-PRF42 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 2-Hole; 00887517021366 LP4.5-PRF4B PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, Bump; 00887517021359 LP4.5-PRT30 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 0-Hole; 00887517021342 LP4.5-PRT31 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 1-Hole; 00887517021335 LP4.5-PRT32 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 2-Hole; 00887517021328 LP4.5-PRT3B PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, Bump; 00887517021311 LP4.5-PRT40 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 0-Hole; 00887517021304 LP4.5-PRT41 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 1-Hole; 00887517021298 LP4.5-PRT42 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 2-Hole; 00887517021281 LP4.5-PRT4B PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, Bump; 00887517021274 LP4.5-SLF30 PRECICE Plate 4.5mm Shaft Plate, Left, 35mm, Long Magnet;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Nuvasive Specialized Orthopedics Inc
Reason for Recall:
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

PRECICE PLATE - Product Usage: intended for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

Product Codes/Lot Numbers:

All Lots affected. Below are: Primary DI Number(GTIN)/Product Code (REF)/Product Description: 00887517021915 LP4.5-DLF30 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 0-Hole; 00887517021908 LP4.5-DLF31 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 1-Hole; 00887517021892 LP4.5-DLF32 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 2-Hole; 00887517021885 LP4.5-DLF3B PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, Bump; 00887517021878 LP4.5-DLF40 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 0-Hole; 00887517021861 LP4.5-DLF41 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 1-Hole; 00887517021854 LP4.5-DLF42 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 2-Hole; 00887517021847 LP4.5-DLF4B PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, Bump; 00887517021830 LP4.5-DRF30 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 0-Hole; 00887517021823 LP4.5-DRF31 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 1-Hole; 00887517021816 LP4.5-DRF32 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 2-Hole; 00887517021809 LP4.5-DRF3B PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, Bump; 00887517021793 LP4.5-DRF40 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 0-Hole; 00887517021786 LP4.5-DRF41 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 1-Hole; 00887517021779 LP4.5-DRF42 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 2-Hole; 00887517021762 LP4.5-DRF4B PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, Bump; 00887517021755 LP4.5-MLT30 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 0-Hole; 00887517021748 LP4.5-MLT31 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 1-Hole; 00887517021731 LP4.5-MLT32 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 2-Hole; 00887517021724 LP4.5-MLT3B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, Bump; 00887517021717 LP4.5-MLT40 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 0-Hole; 00887517021700 LP4.5-MLT41 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 1-Hole; 00887517021694 LP4.5-MLT42 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 2-Hole; 00887517021687 LP4.5-MLT4B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, Bump; 00887517021670 LP4.5-MRT30 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 0-Hole; 00887517021663 LP4.5-MRT31 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 1-Hole; 00887517021656 LP4.5-MRT32 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 2-Hole; 00887517021649 LP4.5-MRT3B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, Bump; 00887517021632 LP4.5-MRT40 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 0-Hole; 00887517021625 LP4.5-MRT41 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 1-Hole; 00887517021618 LP4.5-MRT42 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 2-Hole; 00887517021601 LP4.5-MRT4B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, Bump; 00887517021595 LP4.5-PLF30 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 0-Hole; 00887517021588 LP4.5-PLF31 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 1-Hole; 00887517021571 LP4.5-PLF32 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 2-Hole; 00887517021564 LP4.5-PLF3B PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, Bump; 00887517021557 LP4.5-PLF40 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 0-Hole; 00887517021540 LP4.5-PLF41 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 1-Hole; 00887517021533 LP4.5-PLF42 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 2-Hole; 00887517021526 LP4.5-PLF4B PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, Bump; 00887517021519 LP4.5-PLT30 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 0-Hole; 00887517021502 LP4.5-PLT31 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 1-Hole; 00887517021496 LP4.5-PLT32 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 2-Hole; 00887517021489 LP4.5-PLT3B PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, Bump; 00887517021472 LP4.5-PLT40 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 0-Hole; 00887517021465 LP4.5-PLT41 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 1-Hole; 00887517021458 LP4.5-PLT42 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 2-Hole; 00887517021441 LP4.5-PLT4B PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, Bump; 00887517021434 LP4.5-PRF30 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 0-Hole; 00887517021427 LP4.5-PRF31 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 1-Hole; 00887517021410 LP4.5-PRF32 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 2-Hole; 00887517021403 LP4.5-PRF3B PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, Bump; 00887517021397 LP4.5-PRF40 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 0-Hole; 00887517021380 LP4.5-PRF41 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 1-Hole; 00887517021373 LP4.5-PRF42 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 2-Hole; 00887517021366 LP4.5-PRF4B PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, Bump; 00887517021359 LP4.5-PRT30 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 0-Hole; 00887517021342 LP4.5-PRT31 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 1-Hole; 00887517021335 LP4.5-PRT32 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 2-Hole; 00887517021328 LP4.5-PRT3B PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, Bump; 00887517021311 LP4.5-PRT40 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 0-Hole; 00887517021304 LP4.5-PRT41 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 1-Hole; 00887517021298 LP4.5-PRT42 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 2-Hole; 00887517021281 LP4.5-PRT4B PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, Bump; 00887517021274 LP4.5-SLF30 PRECICE Plate 4.5mm Shaft Plate, Left, 35mm, Long Magnet;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1325-2021

Related Recalls

Precice Max REF: PM10.0-80B265 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 265mm PM10.0-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 280mm PM10.0-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 305mm PM10.0-80B335 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 335mm PM10.0-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 10.0mm x 365mm PM10.0-80TJ265 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 265mm PM10.0-80TJ280 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 280mm PM10.0-80TJ305 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 10.0mm x 305mm PM11.5-80B265 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 265mm PM11.5-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 280mm PM11.5-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 305mm PM11.5-80B335 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 335mm PM11.5-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 11.5mm x 365mm PM11.5-80TJ265 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 265mm PM11.5-80TJ280 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 280mm PM11.5-80TJ305 PRECICE MAX Antegrade Tibia 10¿ Bend 3 Hole 11.5mm x 305mm PM13.0-80B280 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 280mm PM13.0-80B305 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 305mm PM13.0-80B365 PRECICE MAX Antegrade Femur Piriformis Straight 13.0mm x 365mm The Precice Max Nail is intended to stabilize fractures, nonunions, and malunions of the femur, tibia, and humerus. In addition, the Precice Nail is intended to provide magnetically controlled lengthening of long bones and/or transporting of bone segments in the femur, tibia, and humerus.

NuVasive Specialized Orthopedics

Class I - Dangerous

Due to potential issues with design, manufacturing and documentation issues related to the validation of the affected products/devices.

Sep 30, 2024 Other Medical Devices View Details →

During accelerated shelf-life extension studies it was noted that the final packaging was experiencing a potential of failure of the sterile packaging barrier which could result in a hazardous situation where the device is no longer sterile.

Nov 1, 2022 Infusion Pumps Nationwide View Details →

MAGEC 1 System

NuVasive Specialized Orthopedics

Class I - Dangerous

Safety Alert to all surgeons informing them of post-implantation fractures of an internal metallic component for device manufactured prior to March 26, 2015.

Jun 25, 2019 Other Medical Devices View Details →