Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: M9461091010. Lot Numbers: 2316701, 2232601, 23030, 2228501, 2327802, 22152
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Think Surgical, Inc.
Reason for Recall:
There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Product Codes/Lot Numbers:

UDI-DI: M9461091010. Lot Numbers: 2316701, 2232601, 23030, 2228501, 2327802, 22152

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1318-2024

Related Recalls