Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.

Class I - Dangerous
🏥 Medical Devices Recalled: February 13, 2025 ZOLL Medical Infusion Pumps Nationwide

What Should You Do?

  1. Check if you have this product:
    Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO; UDI-DI: (1) 00812394021192, (2) 00812394021390; Serial Numbers: D00000003715, D00000003785, D00000003909, D00000003916, D00000003920, D00000003937, D00000003940, D00000003946, D00000003948, D00000003953, D00000003969, D00000003975, D00000004007, D00000004008, D00000004009, D00000004010, D00000004012, D00000004014, D00000004016, D00000004017, D00000004018, D00000004020, D00000004023, D00000004024, D00000004025, D00000004027, D00000004028, D00000004029, D00000004030, D00000004032, D00000004033, D00000004034, D00000004035, D00000004038, D00000004039, D00000004041, D00000004042, D00000004044, D00000004045, D00000004047, D00000004049, D00000004050, D00000004051, D00000004052, D00000004053, D00000004054, D00000004055, D00000004056, D00000004057, D00000004079, D00000004097, D00000004103, D00000004106, D00000004108, D00000004122, D00000004124, D00000004127, D00000004134, D00000004135, D00000004136, D00000004137, D00000004139, D00000004144, D00000004195, D00000004198, D00000004200, D00000004202, D00000004210, D00000004212, D00000004218, D00000004222, D00000004223, D00000004228, D00000004232, D00000004236, D00000004237, D00000004239, D00000004244, D00000004245, D00000004250, D00000004251, D00000004252, D00000004256, D00000004260, D00000004264, D00000004265, D00000004275, D00000004307, D00000004315, D00000004377, D00000004380, D00000004382, D00000004383, D00000004385, D00000004386, D00000004387, D00000004391, D00000004392, D00000004395, D00000004396, D00000004397, D00000004398, D00000004399, D00000004401, D00000004406, D00000004409, D00000004414, D00000004415, D00000004416, D00000004417, D00000004418, D00000004420, D00000004422, D00000004426, D00000004428, D00000004429, D00000004433, D00000004442, D00000004444, D00000004446, D00000004447, D00000004449, D00000004450, D00000004452, D00000004453, D00000004459, D00000004461, D00000004462, D00000004463, D00000004465, D00000004471, D00000004472, D00000004474, D00000004476, D00000004477, D00000004485, D00000004486, D00000004487, D00000004491, D00000004494, D00000004495, D00000004498, D00000004499, D00000004500, D00000004503, D00000004504, D00000004505, D00000004506, D00000004507, D00000004510, D00000004511, D00000004512, D00000004513, D00000004514, D00000004522, D00000004526, D00000004527, D00000004529, D00000004531, D00000004534, D00000004535, D00000004538, D00000004540, D00000004542, D00000004543, D00000004546, D00000004551, D00000004575, D00000004579, D00000004585, D00000004589, D00000004591, D00000004594, D00000004595, D00000004596, D00000004597, D00000004598, D00000004601, D00000004602, D00000004604, D00000004605, D00000004607, D00000004608, D00000004610, D00000004614, D00000004676, D00000004677, D00000004685, D00000004689, D00000004695, D00000004704, D00000004705, D00000004706, D00000004707, D00000004709, D00000004710, D00000004711, D00000004712, D00000004720, D00000004722, D00000004724, D00000004748, D00000004751, D00000004753, D00000004830, D00000005146, D00000005147, D00000005148, D00000005153, D00000005154, D00000005155, D00000005160, D00000005162, D00000005170, D00000005176, D00000005179, D00000005262, D00000005263, D00000005264, D00000005265, D00000005266, D00000005268, D00000005269, D00000005270, D00000005271, D00000005283, D00000005284, D00000005285, D00000005286, D00000005287, D00000005288, D00000005289, D00000005290, D00000005291, D00000005292, D00000005293, D00000005294, D00000005295, D00000005296, D00000005297, D00000005298, D00000005299, D00000005300, D00000005301, D00000005302, D00000005303, D00000005304, D00000005305, D00000005306, D00000005307, D00000005308, D00000005309, D00000005310, D00000005311, D00000005312, D00000005313, D00000005314, D00000005315, D00000005316, D00000005318, D00000005320, D00000005321, D00000005322, D00000005327, D00000005357, D00000006043, D00000006044, D00000006045, D00000006046, D00000006047, D00000006048, D00000006049, D00000006050, D00000006051, D00000006052, D00000006053, D00000006054, D00000006055, D00000006056, D00000006057, D00000006058, D00000006059, D00000006060, D00000006061, D00000006062, D00000006063, D00000006065, D00000006066, D00000006067, D00000006068, D00000006070, D00000006071, D00000006072, D00000006073, D00000006084, D00000006085, D00000006086, D00000006089, D00000006096, D00000006099, D00000006100, D00000006101, D00000006102, D00000006106, D00000006111, D00000006125, D00000006126, D00000006127, D00000006130, D00000006131, D00000006132, D00000006133, D00000006134, D00000006135, D00000006136, D00000006143, D00000006148, D00000006149, D00000006153, D00000006154, D00000006155, D00000006157, D00000006158, D00000006159, D00000006160, D00000006161, D00000006162, D00000006163, D00000006164, D00000006166, D00000006169, D00000006171, D00000006172, D00000006173, D00000006174, D00000006175, D00000006178, D00000006179, D00000006180, D00000006181, D00000006182, D00000006183, D00000006184, D00000006185, D00000006186, D00000006187, D00000006188, D00000006189, D00000006190, D00000006191, D00000006192, D00000006193, D00000006194, D00000006195, D00000006196, D00000006197, D00000006198, D00000006199, D00000006200, D00000006201, D00000006202, D00000006203, D00000006204, D00000006205, D00000006206, D00000006207, D00000006208, D00000006209, D00000006210, D00000006211, D00000006212, D00000006214, D00000006215, D00000006216, D00000006217, D00000006218, D00000006219, D00000006220, D00000006221, D00000006222, D00000006223, D00000006225, D00000006226, D00000006227, D00000006228, D00000006229, D00000006230, D00000006231, D00000006232, D00000006233, D00000006234, D00000006236, D00000006237, D00000006238, D00000006239, D00000006240, D00000006242, D00000006243, D00000006245, D00000006246, D00000006247, D00000006249, D00000006250, D00000006251, D00000006252, D00000006253, D00000006254, D00000006255, D00000006256, D00000006257, D00000006258, D00000006259, D00000006260, D00000006262, D00000006263, D00000006264, D00000006265, D00000006266, D00000006267, D00000006268, D00000006269, D00000006271, D00000006278, D00000006279, D00000006280, D00000006281, D00000006282, D00000006283, D00000006288, D00000006289, D00000006290, D00000006291, D00000006292, D00000006295, D00000006297, D00000006298, D00000006300, D00000006301, D00000006302, D00000006303, D00000006304, D00000006305, D00000006306, D00000006307, D00000006308, D00000006309, D00000006310, D00000006311, D00000006312, D00000006314, D00000006315, D00000006318, D00000006319, D00000006320, D00000006321, D00000006322, D00000006323, D00000006324, D00000006325, D00000006326, D00000006327, D00000006328, D00000006329, D00000006330, D00000006331, D00000006332, D00000006333, D00000006334, D00000006336, D00000006340, D00000006342, D00000006344, D00000006349, D00000006350, D00000006355, D00000006356, D00000006357, D00000006358, D00000006359, D00000006360, D00000006361, D00000006362, D00000006363, D00000006364, D00000006365, D00000006366, D00000006367, D00000006368, D00000006369, D00000006370, D00000006371, D00000006372, D00000006373, D00000006374, D00000006375, D00000006376, D00000006377, D00000006378, D00000006379, D00000006402, D00000006412, D00000006418, D00000006420, D00000006422, D00000006424, D00000006425, D00000006426, D00000006427, D00000006428, D00000006429, D00000006430, D00000006431, D00000006432, D00000006433, D00000006435, D00000006436, D00000006437, D00000006438, D00000006439, D00000006440, D00000006444, D00000006445, D00000006446, D00000006447, D00000006449, D00000006451, D00000006452, D00000006453, D00000006454, D00000006455, D00000006456, D00000006502, D00000006504, D00000006506, D00000006507, D00000006534, D00000006536, D00000006538, D00000006539, D00000006546, D00000006554, D00000006555, D00000006578, D00000006580, D00000006587, D00000006589, D00000006591, D00000006593, D00000006595, D00000006597, D00000006602, D00000006603, D00000006605, D00000006610, D00000006616, D00000006617, D00000006618, D00000006620, D00000006622, D00000006624, D00000006636, D00000006638, D00000006640, D00000006760, D00000006768, D00000006771, D00000006773, D00000006777, D00000006788, D00000006789, D00000006790, D00000006792, D00000006793, D00000006797, D00000006799, D00000006800, D00000006801, D00000006804, D00000006805, D00000006807, D00000006809, D00000006810, D00000006811, D00000006812, D00000006814, D00000006817, D00000006818, D00000006820, D00000006821, D00000006823, D00000006825, D00000006826, D00000006832, D00000006833, D00000006834, D00000006836, D00000006843, D00000006844, D00000006845, D00000006846, D00000006885, D00000006886, D00000006965, D00000006967, D00000006969, D00000006971, D00000006972, D00000006973, D00000006978, D00000006979, D00000006981, D00000006982, D00000006983, D00000006984, D00000006985, D00000006986, D00000006987, D00000006988, D00000007000, D00000007001, D00000007003, D00000007051, D00000007056, D00000007071, D00000007092, D00000007132, D00000007133, D00000007134, D00000007136, D00000007137, D00000007140, D00000007142, D00000007144, D00000007147, D00000007148, D00000007149, D00000007151, D00000007154, D00000007155, D00000007159, D00000007167, D00000007169, D00000007170, D00000007171, D00000007172, D00000007173, D00000007174, D00000007176, D00000007177, D00000007178, D00000007179, D00000007180, D00000007181, D00000007182, D00000007183, D00000007184, D00000007185, D00000007186, D00000007187, D00000007188, D00000007189, D00000007190, D00000007191, D00000007192, D00000007193, D00000007194, D00000007195, D00000007196, D00000007197, D00000007198, D00000007199, D00000007200, D00000007201, D00000007203, D00000007204, D00000007205, D00000007206, D00000007207, D00000007208, D00000007209, D00000007210, D00000007214, D00000007215, D00000007216, D00000007217, D00000007218, D00000007220, D00000007221, D00000007222, D00000007223, D00000007224, D00000007225, D00000007226, D00000007227, D00000007228, D00000007230, D00000007231, D00000007232, D00000007233, D00000007234, D00000007235, D00000007236, D00000007240, D00000007246, D00000007247, D00000007248, D00000007250, D00000007251, D00000007253, D00000007261, D00000007268, D00000007269, D00000007274, D00000007276, D00000007280, D00000007286, D00000007287, D00000007289, D00000007291, D00000007293, D00000007297, D00000007302, D00000007305, D00000007311, D00000007312, D00000007313, D00000007314, D00000007317, D00000007322, D00000007323, D00000007324, D00000007325, D00000007326, D00000007327, D00000007328, D00000007329, D00000007330, D00000007334, D00000007337, D00000007338, D00000007339, D00000007340, D00000007343, D00000007344, D00000007345, D00000007346, D00000007347, D00000007348, D00000007350, D00000007351, D00000007353, D00000007356, D00000007358, D00000007359, D00000007361, D00000007363, D00000007365, D00000007373, D00000007375, D00000007376, D00000007378, D00000007385, D00000007386, D00000007388, D00000007391, D00000007393, D00000007394, D00000007395, D00000007396, D00000007397, D00000007398, D00000007399, D00000007401, D00000007402, D00000007403, D00000007406, D00000007407, D00000007408, D00000007409, D00000007410, D00000007412, D00000007413, D00000007414, D00000007415, D00000007416, D00000007417, D00000007418, D00000007419, D00000007420, D00000007421, D00000007422, D00000007424, D00000007425, D00000007426, D00000007427, D00000007431, D00000007432, D00000007433, D00000007435, D00000007436, D00000007437, D00000007438, D00000007440, D00000007444, D00000007446, D00000007460, D00000007467, D00000007471, D00000007472, D00000007473, D00000007475, D00000007477, D00000007478, D00000007479, D00000007483, D00000007484, D00000007507, D00000007510, D00000007529, D00000007531, D00000007534, D00000007536, D00000007539, D00000007540, D00000007541, D00000007542, D00000007543, D00000007549, D00000007558, D00000007572, D00000007606, D00000007611, D00000007614, D00000007620, D00000007624, D00000007625, D00000007627, D00000007629, D00000007630, D00000007631, D00000007634, D00000007636, D00000007637, D00000007641, D00000007643, D00000007644, D00000007645, D00000007648, D00000007652, D00000007653, D00000007654, D00000007655, D00000007656, D00000007657, D00000007663, D00000007665, D00000007666, D00000007668, D00000007670, D00000007671, D00000007672, D00000007673, D00000007675, D00000007676, D00000007677, D00000007678, D00000007679, D00000007680, D00000007681, D00000007682, D00000007683, D00000007684, D00000007685, D00000007686, D00000007687, D00000007688, D00000007689, D00000007690, D00000007691, D00000007692, D00000007693, D00000007694, D00000007695, D00000007696, D00000007697, D00000007698, D00000007699, D00000007700, D00000007701, D00000007703, D00000007704, D00000007705, D00000007706, D00000007708, D00000007709, D00000007710, D00000007711, D00000007712, D00000007713, D00000007714, D00000007715, D00000007716, D00000007717, D00000007718, D00000007719, D00000007720, D00000007722, D00000007723, D00000007724, D00000007725, D00000007726, D00000007727, D00000007728, D00000007729, D00000007730, D00000007731, D00000007732, D00000007734, D00000007735, D00000007736, D00000007738, D00000007739, D00000007741, D00000007742, D00000007743, D00000007745, D00000007746, D00000007748, D00000007749, D00000007751, D00000007752, D00000007753, D00000007754, D00000007755, D00000007756, D00000007757, D00000007758, D00000007759, D00000007760, D00000007761, D00000007762, D00000007764, D00000007765, D00000007766, D00000007767, D00000007768, D00000007769, D00000007770, D00000007771, D00000007772, D00000007773, D00000007774, D00000007775, D00000007776, D00000007777, D00000007778, D00000007779, D00000007781, D00000007782, D00000007783, D00000007785, D00000007786, D00000007787, D00000007788, D00000007789, D00000007790, D00000007791, D00000007792, D00000007793, D00000007794, D00000007795, D00000007796, D00000007797, D00000007798, D00000007799, D00000007800, D00000007801, D00000007802, D00000007804, D00000007805, D00000007807, D00000007808, D00000007809, D00000007811, D00000007812, D00000007813, D00000007814, D00000007815, D00000007816, D00000007817, D00000007818, D00000007819, D00000007820, D00000007821, D00000007822, D00000007823, D00000007824, D00000007825, D00000007826, D00000007827, D00000007828, D00000007829, D00000007830, D00000007831, D00000007832, D00000007833, D00000007834, D00000007839, D00000007840, D00000007841, D00000007842, D00000007843, D00000007844, D00000007846, D00000007847, D00000007848, D00000007849, D00000007851, D00000007852, D00000007853, D00000007854, D00000007856, D00000007862, D00000007881, D00000007907, D00000007908, D00000007909, D00000007910, D00000007911, D00000007912, D00000007913, D00000007914, D00000007915, D00000007916, D00000007917, D00000007918, D00000007921, D00000007922, D00000007924, D00000007934, D00000007940, D00000007943, D00000007944, D00000007946, D00000007948, D00000007950, D00000007951, D00000007952, D00000007954, D00000007955, D00000007956, D00000007958, D00000007960, D00000007963, D00000007966, D00000007969, D00000007971, D00000007973, D00000007979, D00000008010, D00000008014, D00000008034, D00000008044, D00000008046, D00000008051, D00000008064, D00000008066, D00000008069, D00000008084, D00000008116, D00000008133, D00000008151, D00000008234, D00000008239, D00000008245, D00000008265, D00000008266, D00000008267, D00000008268, D00000008269, D00000008270, D00000008271, D00000008272, D00000008273, D00000008274, D00000008275, D00000008276, D00000008277, D00000008278, D00000008279, D00000008280, D00000008281, D00000008282, D00000008283, D00000008284, D00000008285, D00000008286, D00000008287, D00000008288, D00000008289, D00000008290, D00000008291, D00000008292, D00000008293, D00000008294, D00000008295, D00000008296, D00000008297, D00000008298, D00000008299, D00000008300, D00000008301, D00000008302, D00000008303, D00000008304, D00000008305, D00000008306, D00000008307, D00000008309, D00000008311, D00000008315, D00000008316, D00000008317, D00000008319, D00000008320, D00000008322, D00000008323, D00000008324, D00000008326, D00000008327, D00000008328, D00000008329, D00000008331, D00000008332, D00000008333, D00000008334, D00000008335, D00000008336, D00000008337, D00000008338, D00000008339, D00000008340, D00000008341, D00000008342, D00000008343, D00000008344, D00000008345, D00000008346, D00000008347, D00000008348, D00000008349, D00000008350, D00000008351, D00000008352, D00000008353, D00000008354, D00000008355, D00000008356, D00000008357, D00000008358, D00000008359, D00000008360, D00000008361, D00000008363, D00000008364, D00000008365, D00000008366, D00000008367, D00000008368, D00000008369, D00000008370, D00000008371, D00000008372, D00000008373, D00000008374, D00000008375, D00000008376, D00000008377, D00000008378, D00000008379, D00000008380, D00000008381, D00000008382, D00000008383, D00000008384, D00000008385, D00000008386, D00000008387, D00000008388, D00000008389, D00000008390, D00000008391, D00000008393, D00000008394, D00000008395, D00000008396, D00000008397, D00000008398, D00000008399, D00000008400, D00000008401, D00000008402, D00000008403, D00000008404, D00000008405, D00000008406, D00000008407, D00000008408, D00000008409, D00000008410, D00000008411, D00000008412, D00000008413, D00000008414, D00000008416, D00000008417, D00000008418, D00000008419, D00000008421, D00000008422, D00000008424, D00000008425, D00000008426, D00000008427, D00000008428, D00000008429, D00000008430, D00000008431, D00000008432, D00000008433, D00000008434, D00000008435, D00000008436, D00000008437, D00000008438, D00000008441, D00000008442, D00000008443, D00000008444, D00000008445, D00000008446, D00000008447, D00000008449, D00000008450, D00000008451, D00000008452, D00000008459, D00000008461, D00000008462, D00000008463, D00000008465, D00000008469, D00000008476, D00000008625, D00000008628, D00000008671, D00000008675, D00000008685, D00000008686, D00000008688, D00000008698, D00000008701, D00000008702, D00000008703, D00000008706, D00000008708, D00000008709, D00000008713, D00000008714, D00000008716, D00000008717, D00000008718, D00000008719, D00000008720, D00000008721, D00000008722, D00000008723, D00000008724, D00000008725, D00000008726, D00000008727, D00000008729, D00000008730, D00000008731, D00000008733, D00000008737, D00000008744, D00000008745, D00000008746, D00000008747, D00000008750, D00000008751, D00000008752, D00000008753, D00000008754, D00000008755, D00000008756, D00000008757, D00000008758, D00000008759, D00000008760, D00000008761, D00000008762, D00000008763, D00000008764, D00000008765, D00000008766, D00000008767, D00000008768, D00000008769, D00000008770, D00000008771, D00000008772, D00000008773, D00000008774, D00000008775, D00000008776, D00000008777, D00000008778, D00000008779, D00000008780, D00000008781, D00000008782, D00000008783, D00000008784, D00000008785, D00000008786, D00000008787, D00000008788, D00000008789, D00000008790, D00000008791, D00000008792, D00000008793, D00000008794, D00000008795, D00000008796, D00000008797, D00000008798, D00000008799, D00000008800, D00000008801, D00000008802, D00000008803, D00000008804, D00000008805, D00000008806, D00000008807, D00000008808, D00000008809, D00000008810, D00000008811, D00000008812, D00000008813, D00000008814, D00000008815, D00000008816, D00000008817, D00000008824, D00000008825, D00000008826, D00000008827, D00000008829, D00000008830, D00000008831, D00000008834, D00000008837, D00000008838, D00000008840, D00000008841, D00000008842, D00000008843, D00000008844, D00000008845, D00000008846, D00000008847, D00000008849, D00000008850, D00000008860, D00000008861, D00000008865, D00000008866, D00000008868, D00000008869, D00000008870, D00000008871, D00000008872, D00000008874, D00000008875, D00000008879, D00000008881, D00000008882, D00000008886, D00000008894, D00000008895, D00000008897, D00000008900, D00000008905, D00000008907, D00000008908, D00000008909, D00000008911, D00000008912, D00000008913, D00000008914, D00000008916, D00000008917, D00000008918, D00000008919, D00000008921, D00000008922, D00000008923, D00000008924, D00000008925, D00000008926, D00000008927, D00000008928, D00000008929, D00000008930, D00000008931, D00000008932, D00000008933, D00000008934, D00000008935, D00000008936, D00000008937, D00000008938, D00000008939, D00000008940, D00000008941, D00000008942, D00000008944, D00000008945, D00000008946, D00000008947, D00000008948, D00000008949, D00000008950, D00000008951, D00000008952, D00000008953, D00000008954, D00000008955, D00000008956, D00000008957, D00000008958, D00000008959, D00000008960, D00000008961, D00000008962, D00000008963, D00000008964, D00000008965, D00000008966, D00000008967, D00000008968, D00000008969, D00000008970, D00000008971, D00000008973, D00000008975, D00000008976, D00000008978, D00000008980, D00000008983, D00000008985, D00000008989, D00000008992, D00000008993, D00000008994, D00000008995, D00000008996, D00000008998, D00000008999, D00000009000, D00000009001, D00000009002, D00000009003, D00000009004, D00000009005, D00000009006, D00000009007, D00000009008, D00000009009, D00000009010, D00000009011, D00000009012, D00000009013, D00000009014, D00000009015, D00000009016, D00000009017, D00000009018, D00000009019, D00000009020, D00000009021, D00000009022, D00000009023, D00000009024, D00000009025, D00000009026, D00000009027, D00000009028, D00000009029, D00000009030, D00000009031, D00000009032, D00000009033, D00000009034, D00000009035, D00000009036, D00000009037, D00000009038, D00000009039, D00000009040, D00000009041, D00000009042, D00000009043, D00000009044, D00000009045, D00000009046, D00000009047, D00000009048, D00000009049, D00000009050, D00000009051, D00000009052, D00000009053, D00000009054, D00000009055, D00000009056, D00000009057, D00000009058, D00000009059, D00000009060, D00000009061, D00000009062, D00000009063, D00000009064, D00000009065, D00000009066, D00000009067, D00000009068, D00000009069, D00000009070, D00000009071, D00000009072, D00000009073, D00000009074, D00000009075, D00000009076, D00000009077, D00000009078, D00000009079, D00000009080, D00000009081, D00000009083, D00000009084, D00000009085, D00000009086, D00000009087, D00000009089, D00000009090, D00000009091, D00000009092, D00000009093, D00000009095, D00000009096, D00000009097, D00000009098, D00000009100, D00000009101, D00000009105, D00000009106, D00000009136, D00000009137, D00000009138, D00000009139, D00000009140, D00000009141, D00000009142, D00000009144, D00000009146, D00000009147, D00000009149, D00000009150, D00000009151, D00000009152, D00000009153, D00000009155, D00000009156, D00000009157, D00000009202, D00000009203, D00000009205, D00000009253, D00000009254, D00000009273, D00000009275, D00000009322, D00000009327, D00000009329, D00000009330, D00000009332, D00000009333, D00000009334, D00000009335, D00000009336, D00000009337, D00000009340, D00000009341, D00000009343, D00000009344, D00000009345, D00000009346, D00000009347, D00000009348, D00000009349, D00000009352, D00000009353, D00000009354, D00000009358, D00000009400, D00000009405, D00000009413, D00000009414, D00000009415, D00000009417, D00000009421, D00000009424, D00000009427, D00000009428, D00000009429, D00000009431, D00000009438, D00000009444, D00000009445, D00000009446, D00000009450, D00000009455, D00000009457, D00000009462, D00000009463, D00000009471, D00000009475, D00000009476, D00000009477, D00000009478, D00000009479, D00000009482, D00000009484, D00000009485, D00000009490, D00000009491, D00000009492, D00000009495, D00000009496, D00000009498, D00000009499, D00000009500, D00000009501, D00000009502, D00000009503, D00000009504, D00000009505, D00000009506, D00000009507, D00000009508, D00000009510, D00000009511, D00000009512, D00000009513, D00000009514, D00000009515, D00000009521, D00000009523, D00000009524, D00000009525, D00000009528, D00000009530, D00000009531, D00000009532, D00000009533, D00000009534, D00000009535, D00000009536, D00000009537, D00000009538, D00000009539, D00000009540, D00000009541, D00000009542, D00000009544, D00000009545, D00000009546, D00000009547, D00000009548, D00000009549, D00000009550, D00000009551, D00000009552, D00000009553, D00000009554, D00000009555, D00000009557, D00000009558, D00000009559, D00000009560, D00000009561, D00000009562, D00000009563, D00000009565, D00000009566, D00000009568, D00000009569, D00000009570, D00000009571, D00000009572, D00000009573, D00000009574, D00000009575, D00000009576, D00000009578, D00000009579, D00000009580, D00000009581, D00000009582, D00000009584, D00000009585, D00000009586, D00000009587, D00000009588, D00000009590, D00000009591, D00000009592, D00000009593, D00000009594, D00000009595, D00000009597, D00000009598, D00000009599, D00000009601, D00000009603, D00000009604, D00000009605, D00000009606, D00000009607, D00000009608, D00000009609, D00000009610, D00000009611, D00000009612, D00000009613, D00000009614, D00000009615, D00000009616, D00000009617, D00000009618, D00000009619, D00000009620, D00000009621, D00000009622, D00000009623, D00000009624, D00000009625, D00000009628, D00000009629, D00000009630, D00000009632, D00000009634, D00000009636, D00000009637, D00000009639, D00000009640, D00000009641, D00000009642, D00000009643, D00000009644, D00000009645, D00000009646, D00000009647, D00000009648, D00000009649, D00000009650, D00000009651, D00000009652, D00000009653, D00000009654, D00000009655, D00000009656, D00000010132, D00000010139, D00000010141, D00000010143, D00000010144, D00000010145, D00000010146, D00000010147, D00000010148, D00000010149, D00000010150, D00000010151, D00000010152, D00000010153, D00000010154, D00000010155, D00000010156, D00000010157, D00000010158, D00000010160, D00000010161, D00000010162, D00000010164, D00000010181, D00000010184, D00000010187, D00000010188, D00000010189, D00000010190, D00000010191, D00000010192, D00000010193, D00000010194, D00000010195, D00000010196, D00000010197, D00000010198, D00000010199, D00000010200, D00000010201, D00000010202, D00000010203, D00000010204, D00000010205, D00000010206, D00000010207, D00000010208, D00000010209, D00000010210, D00000010211, D00000010213, D00000010215, D00000010216, D00000010217, D00000010218, D00000010219, D00000010220, D00000010222, D00000010223, D00000010224, D00000010225, D00000010226, D00000010228, D00000010229, D00000010230, D00000010231, D00000010232, D00000010233, D00000010234, D00000010236, D00000010237, D00000010239, D00000010240, D00000010241, D00000010242, D00000010243, D00000010244, D00000010245, D00000010247, D00000010250, D00000010318, D00000010320, D00000010321, D00000010322, D00000010323, D00000010324, D00000010325, D00000010326, D00000010327, D00000010328, D00000010329, D00000010330, D00000010331, D00000010332, D00000010333, D00000010334, D00000010335, D00000010336, D00000010337, D00000010338, D00000010339, D00000010340, D00000010341, D00000010342, D00000010343, D00000010344, D00000010345, D00000010346, D00000010347, D00000010348, D00000010349, D00000010350, D00000010351, D00000010355, D00000010356, D00000010357, D00000010358, D00000010359, D00000010360, D00000010362, D00000010363, D00000010364, D00000010365, D00000010366, D00000010367, D00000010368, D00000010369, D00000010370, D00000010371, D00000010372, D00000010373, D00000010374, D00000010375, D00000010376, D00000010377, D00000010378, D00000010379, D00000010380, D00000010381, D00000010382, D00000010383, D00000010384, D00000010385, D00000010386, D00000010387, D00000010388, D00000010390, D00000010391, D00000010392, D00000010393, D00000010395, D00000010396, D00000010397, D00000010398, D00000010399, D00000010409, D00000010410, D00000010414, D00000010415, D00000010416, D00000010418, D00000010421, D00000010422, D00000010424, D00000010425, D00000010426, D00000010428, D00000010429, D00000010430, D00000010431, D00000010432, D00000010433, D00000010434, D00000010435, D00000010437, D00000010439, D00000010441, D00000010442, D00000010443, D00000010444, D00000010445, D00000010446, D00000010447, D00000010448, D00000010449, D00000010450, D00000010451, D00000010452, D00000010453, D00000010454, D00000010455, D00000010457, D00000010458, D00000010459, D00000010460, D00000010461, D00000010462, D00000010463, D00000010464, D00000010465, D00000010466, D00000010467, D00000010468, D00000010469, D00000010470, D00000010471, D00000010472, D00000010473, D00000010474, D00000010475, D00000010477, D00000010478, D00000010479, D00000010480, D00000010482, D00000010483, D00000010484, D00000010485, D00000010486, D00000010487, D00000010488, D00000010489, D00000010490, D00000010491, D00000010492, D00000010493, D00000010494, D00000010495, D00000010499, D00000010500, D00000010501, D00000010502, D00000010503, D00000010504, D00000010506, D00000010507, D00000010508, D00000010509, D00000010510, D00000010511, D00000010512, D00000010513, D00000010514, D00000010515, D00000010516, D00000010517, D00000010518, D00000010519, D00000010520, D00000010521, D00000010522, D00000010523, D00000010524, D00000010525, D00000010526, D00000010527, D00000010528, D00000010529, D00000010530, D00000010531, D00000010532, D00000010533, D00000010534, D00000010535, D00000010536, D00000010537, D00000010538, D00000010540, D00000010541, D00000010542, D00000010543, D00000010544, D00000010546, D00000010547, D00000010548, D00000010549, D00000010550, D00000010551, D00000010552, D00000010553, D00000010555, D00000010556, D00000010557, D00000010558, D00000010559, D00000010560, D00000010561, D00000010562, D00000010563, D00000010564, D00000010565, D00000010566, D00000010567, D00000010568, D00000010569, D00000010570, D00000010571, D00000010572, D00000010573, D00000010574, D00000010575, D00000010576, D00000010577, D00000010578, D00000010579, D00000010580, D00000010581, D00000010582, D00000010583, D00000010584, D00000010585, D00000010586, D00000010587, D00000010588, D00000010590, D00000010591, D00000010592, D00000010593, D00000010596, D00000010597, D00000010598, D00000010599, D00000010600, D00000010601, D00000010602, D00000010603, D00000010605, D00000010606, D00000010607, D00000010611, D00000010612, D00000010615, D00000010616, D00000010622, D00000010627, D00000010630, D00000010631, D00000010632, D00000010634, D00000010640, D00000010641, D00000010647, D00000010653, D00000010655, D00000010657, D00000010661, D00000010670, D00000010671, D00000010678, D00000010685, D00000010687, D00000010689, D00000010691, D00000010693, D00000010695, D00000010696, D00000010697, D00000010698, D00000010700, D00000010702, D00000010703, D00000010704, D00000010705, D00000010706, D00000010707, D00000010709, D00000010710, D00000010711, D00000010712, D00000010715, D00000010716, D00000010718, D00000010720, D00000010721, D00000010724, D00000010732, D00000010739, D00000010756, D00000010761, D00000010762, D00000010766, D00000010776, D00000010778, D00000010782, D00000010783, D00000010788, D00000010795, D00000010797, D00000010804, D00000010807, D00000010821, D00000010866, D00000010883, D00000010885, D00000010888, D00000010889, D00000010890, D00000010895, D00000010898, D00000010902, D00000010903, D00000010904, D00000010905, D00000010907, D00000010908, D00000010909, D00000010911, D00000010912, D00000010920, D00000010930, D00000010934, D00000010935, D00000010936, D00000010937, D00000010938, D00000010939, D00000010940, D00000010943, D00000010944, D00000010977, D00000010993, D00000011033, D00000011053, D00000011062, D00000011063, D00000011068, D00000011069, D00000011073, D00000011075, D00000011077, D00000011078, D00000011079, D00000011080, D00000011081, D00000011082, D00000011083, D00000011084, D00000011085, D00000011086, D00000011087, D00000011088, D00000011089, D00000011090, D00000011091, D00000011092, D00000011093, D00000011094, D00000011095, D00000011096, D00000011097, D00000011098, D00000011099, D00000011100, D00000011101, D00000011102, D00000011103, D00000011104, D00000011105, D00000011106, D00000011107, D00000011108, D00000011109, D00000011110, D00000011111, D00000011112, D00000011115, D00000011116, D00000011117, D00000011118, D00000011119, D00000011120, D00000011122, D00000011123, D00000011124, D00000011126, D00000011128, D00000011129, D00000011135, D00000011136, D00000011137, D00000011140, D00000011141, D00000011142, D00000011143, D00000011144, D00000011145, D00000011146, D00000011147, D00000011148, D00000011149, D00000011150, D00000011151, D00000011152, D00000011153, D00000011154, D00000011155, D00000011156, D00000011157, D00000011158, D00000011159, D00000011160, D00000011161, D00000011162, D00000011163, D00000011164, D00000011165, D00000011166, D00000011167, D00000011168, D00000011169, D00000011170, D00000011171, D00000011172, D00000011173, D00000011174, D00000011175, D00000011176, D00000011177, D00000011178, D00000011179, D00000011180, D00000011182, D00000011183, D00000011184, D00000011185, D00000011188, D00000011189, D00000011190, D00000011191, D00000011192, D000
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
ZOLL Medical Corporation
Reason for Recall:
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.

Product Codes/Lot Numbers:

Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO; UDI-DI: (1) 00812394021192, (2) 00812394021390; Serial Numbers: D00000003715, D00000003785, D00000003909, D00000003916, D00000003920, D00000003937, D00000003940, D00000003946, D00000003948, D00000003953, D00000003969, D00000003975, D00000004007, D00000004008, D00000004009, D00000004010, D00000004012, D00000004014, D00000004016, D00000004017, D00000004018, D00000004020, D00000004023, D00000004024, D00000004025, D00000004027, D00000004028, D00000004029, D00000004030, D00000004032, D00000004033, D00000004034, D00000004035, D00000004038, D00000004039, D00000004041, D00000004042, D00000004044, D00000004045, D00000004047, D00000004049, D00000004050, D00000004051, D00000004052, D00000004053, D00000004054, D00000004055, D00000004056, D00000004057, D00000004079, D00000004097, D00000004103, D00000004106, D00000004108, D00000004122, D00000004124, D00000004127, D00000004134, D00000004135, D00000004136, D00000004137, D00000004139, D00000004144, D00000004195, D00000004198, D00000004200, D00000004202, D00000004210, D00000004212, D00000004218, D00000004222, D00000004223, D00000004228, D00000004232, D00000004236, D00000004237, D00000004239, D00000004244, D00000004245, D00000004250, D00000004251, D00000004252, D00000004256, D00000004260, D00000004264, D00000004265, D00000004275, D00000004307, D00000004315, D00000004377, D00000004380, D00000004382, D00000004383, D00000004385, D00000004386, D00000004387, D00000004391, D00000004392, D00000004395, D00000004396, D00000004397, D00000004398, D00000004399, D00000004401, D00000004406, D00000004409, D00000004414, D00000004415, D00000004416, D00000004417, D00000004418, D00000004420, D00000004422, D00000004426, D00000004428, D00000004429, D00000004433, D00000004442, D00000004444, D00000004446, D00000004447, D00000004449, D00000004450, D00000004452, D00000004453, D00000004459, D00000004461, D00000004462, D00000004463, D00000004465, D00000004471, D00000004472, D00000004474, D00000004476, D00000004477, D00000004485, D00000004486, D00000004487, D00000004491, D00000004494, D00000004495, D00000004498, D00000004499, D00000004500, D00000004503, D00000004504, D00000004505, D00000004506, D00000004507, D00000004510, D00000004511, D00000004512, D00000004513, D00000004514, D00000004522, D00000004526, D00000004527, D00000004529, D00000004531, D00000004534, D00000004535, D00000004538, D00000004540, D00000004542, D00000004543, D00000004546, D00000004551, D00000004575, D00000004579, D00000004585, D00000004589, D00000004591, D00000004594, D00000004595, D00000004596, D00000004597, D00000004598, D00000004601, D00000004602, D00000004604, D00000004605, D00000004607, D00000004608, D00000004610, D00000004614, D00000004676, D00000004677, D00000004685, D00000004689, D00000004695, D00000004704, D00000004705, D00000004706, D00000004707, D00000004709, D00000004710, D00000004711, D00000004712, D00000004720, D00000004722, D00000004724, D00000004748, D00000004751, D00000004753, D00000004830, D00000005146, D00000005147, D00000005148, D00000005153, D00000005154, D00000005155, D00000005160, D00000005162, D00000005170, D00000005176, D00000005179, D00000005262, D00000005263, D00000005264, D00000005265, D00000005266, D00000005268, D00000005269, D00000005270, D00000005271, D00000005283, D00000005284, D00000005285, D00000005286, D00000005287, D00000005288, D00000005289, D00000005290, D00000005291, D00000005292, D00000005293, D00000005294, D00000005295, D00000005296, D00000005297, D00000005298, D00000005299, D00000005300, D00000005301, D00000005302, D00000005303, D00000005304, D00000005305, D00000005306, D00000005307, D00000005308, D00000005309, D00000005310, D00000005311, D00000005312, D00000005313, D00000005314, D00000005315, D00000005316, D00000005318, D00000005320, D00000005321, D00000005322, D00000005327, D00000005357, D00000006043, D00000006044, D00000006045, D00000006046, D00000006047, D00000006048, D00000006049, D00000006050, D00000006051, D00000006052, D00000006053, D00000006054, D00000006055, D00000006056, D00000006057, D00000006058, D00000006059, D00000006060, D00000006061, D00000006062, D00000006063, D00000006065, D00000006066, D00000006067, D00000006068, D00000006070, D00000006071, D00000006072, D00000006073, D00000006084, D00000006085, D00000006086, D00000006089, D00000006096, D00000006099, D00000006100, D00000006101, D00000006102, D00000006106, D00000006111, D00000006125, D00000006126, D00000006127, D00000006130, D00000006131, D00000006132, D00000006133, D00000006134, D00000006135, D00000006136, D00000006143, D00000006148, D00000006149, D00000006153, D00000006154, D00000006155, D00000006157, D00000006158, D00000006159, D00000006160, D00000006161, D00000006162, D00000006163, D00000006164, D00000006166, D00000006169, D00000006171, D00000006172, D00000006173, D00000006174, D00000006175, D00000006178, D00000006179, D00000006180, D00000006181, D00000006182, D00000006183, D00000006184, D00000006185, D00000006186, D00000006187, D00000006188, D00000006189, D00000006190, D00000006191, D00000006192, D00000006193, D00000006194, D00000006195, D00000006196, D00000006197, D00000006198, D00000006199, D00000006200, D00000006201, D00000006202, D00000006203, D00000006204, D00000006205, D00000006206, D00000006207, D00000006208, D00000006209, D00000006210, D00000006211, D00000006212, D00000006214, D00000006215, D00000006216, D00000006217, D00000006218, D00000006219, D00000006220, D00000006221, D00000006222, D00000006223, D00000006225, D00000006226, D00000006227, D00000006228, D00000006229, D00000006230, D00000006231, D00000006232, D00000006233, D00000006234, D00000006236, D00000006237, D00000006238, D00000006239, D00000006240, D00000006242, D00000006243, D00000006245, D00000006246, D00000006247, D00000006249, D00000006250, D00000006251, D00000006252, D00000006253, D00000006254, D00000006255, D00000006256, D00000006257, D00000006258, D00000006259, D00000006260, D00000006262, D00000006263, D00000006264, D00000006265, D00000006266, D00000006267, D00000006268, D00000006269, D00000006271, D00000006278, D00000006279, D00000006280, D00000006281, D00000006282, D00000006283, D00000006288, D00000006289, D00000006290, D00000006291, D00000006292, D00000006295, D00000006297, D00000006298, D00000006300, D00000006301, D00000006302, D00000006303, D00000006304, D00000006305, D00000006306, D00000006307, D00000006308, D00000006309, D00000006310, D00000006311, D00000006312, D00000006314, D00000006315, D00000006318, D00000006319, D00000006320, D00000006321, D00000006322, D00000006323, D00000006324, D00000006325, D00000006326, D00000006327, D00000006328, D00000006329, D00000006330, D00000006331, D00000006332, D00000006333, D00000006334, D00000006336, D00000006340, D00000006342, D00000006344, D00000006349, D00000006350, D00000006355, D00000006356, D00000006357, D00000006358, D00000006359, D00000006360, D00000006361, D00000006362, D00000006363, D00000006364, D00000006365, D00000006366, D00000006367, D00000006368, D00000006369, D00000006370, D00000006371, D00000006372, D00000006373, D00000006374, D00000006375, D00000006376, D00000006377, D00000006378, D00000006379, D00000006402, D00000006412, D00000006418, D00000006420, D00000006422, D00000006424, D00000006425, D00000006426, D00000006427, D00000006428, D00000006429, D00000006430, D00000006431, D00000006432, D00000006433, D00000006435, D00000006436, D00000006437, D00000006438, D00000006439, D00000006440, D00000006444, D00000006445, D00000006446, D00000006447, D00000006449, D00000006451, D00000006452, D00000006453, D00000006454, D00000006455, D00000006456, D00000006502, D00000006504, D00000006506, D00000006507, D00000006534, D00000006536, D00000006538, D00000006539, D00000006546, D00000006554, D00000006555, D00000006578, D00000006580, D00000006587, D00000006589, D00000006591, D00000006593, D00000006595, D00000006597, D00000006602, D00000006603, D00000006605, D00000006610, D00000006616, D00000006617, D00000006618, D00000006620, D00000006622, D00000006624, D00000006636, D00000006638, D00000006640, D00000006760, D00000006768, D00000006771, D00000006773, D00000006777, D00000006788, D00000006789, D00000006790, D00000006792, D00000006793, D00000006797, D00000006799, D00000006800, D00000006801, D00000006804, D00000006805, D00000006807, D00000006809, D00000006810, D00000006811, D00000006812, D00000006814, D00000006817, D00000006818, D00000006820, D00000006821, D00000006823, D00000006825, D00000006826, D00000006832, D00000006833, D00000006834, D00000006836, D00000006843, D00000006844, D00000006845, D00000006846, D00000006885, D00000006886, D00000006965, D00000006967, D00000006969, D00000006971, D00000006972, D00000006973, D00000006978, D00000006979, D00000006981, D00000006982, D00000006983, D00000006984, D00000006985, D00000006986, D00000006987, D00000006988, D00000007000, D00000007001, D00000007003, D00000007051, D00000007056, D00000007071, D00000007092, D00000007132, D00000007133, D00000007134, D00000007136, D00000007137, D00000007140, D00000007142, D00000007144, D00000007147, D00000007148, D00000007149, D00000007151, D00000007154, D00000007155, D00000007159, D00000007167, D00000007169, D00000007170, D00000007171, D00000007172, D00000007173, D00000007174, D00000007176, D00000007177, D00000007178, D00000007179, D00000007180, D00000007181, D00000007182, D00000007183, D00000007184, D00000007185, D00000007186, D00000007187, D00000007188, D00000007189, D00000007190, D00000007191, D00000007192, D00000007193, D00000007194, D00000007195, D00000007196, D00000007197, D00000007198, D00000007199, D00000007200, D00000007201, D00000007203, D00000007204, D00000007205, D00000007206, D00000007207, D00000007208, D00000007209, D00000007210, D00000007214, D00000007215, D00000007216, D00000007217, D00000007218, D00000007220, D00000007221, D00000007222, D00000007223, D00000007224, D00000007225, D00000007226, D00000007227, D00000007228, D00000007230, D00000007231, D00000007232, D00000007233, D00000007234, D00000007235, D00000007236, D00000007240, D00000007246, D00000007247, D00000007248, D00000007250, D00000007251, D00000007253, D00000007261, D00000007268, D00000007269, D00000007274, D00000007276, D00000007280, D00000007286, D00000007287, D00000007289, D00000007291, D00000007293, D00000007297, D00000007302, D00000007305, D00000007311, D00000007312, D00000007313, D00000007314, D00000007317, D00000007322, D00000007323, D00000007324, D00000007325, D00000007326, D00000007327, D00000007328, D00000007329, D00000007330, D00000007334, D00000007337, D00000007338, D00000007339, D00000007340, D00000007343, D00000007344, D00000007345, D00000007346, D00000007347, D00000007348, D00000007350, D00000007351, D00000007353, D00000007356, D00000007358, D00000007359, D00000007361, D00000007363, D00000007365, D00000007373, D00000007375, D00000007376, D00000007378, D00000007385, D00000007386, D00000007388, D00000007391, D00000007393, D00000007394, D00000007395, D00000007396, D00000007397, D00000007398, D00000007399, D00000007401, D00000007402, D00000007403, D00000007406, D00000007407, D00000007408, D00000007409, D00000007410, D00000007412, D00000007413, D00000007414, D00000007415, D00000007416, D00000007417, D00000007418, D00000007419, D00000007420, D00000007421, D00000007422, D00000007424, D00000007425, D00000007426, D00000007427, D00000007431, D00000007432, D00000007433, D00000007435, D00000007436, D00000007437, D00000007438, D00000007440, D00000007444, D00000007446, D00000007460, D00000007467, D00000007471, D00000007472, D00000007473, D00000007475, D00000007477, D00000007478, D00000007479, D00000007483, D00000007484, D00000007507, D00000007510, D00000007529, D00000007531, D00000007534, D00000007536, D00000007539, D00000007540, D00000007541, D00000007542, D00000007543, D00000007549, D00000007558, D00000007572, D00000007606, D00000007611, D00000007614, D00000007620, D00000007624, D00000007625, D00000007627, D00000007629, D00000007630, D00000007631, D00000007634, D00000007636, D00000007637, D00000007641, D00000007643, D00000007644, D00000007645, D00000007648, D00000007652, D00000007653, D00000007654, D00000007655, D00000007656, D00000007657, D00000007663, D00000007665, D00000007666, D00000007668, D00000007670, D00000007671, D00000007672, D00000007673, D00000007675, D00000007676, D00000007677, D00000007678, D00000007679, D00000007680, D00000007681, D00000007682, D00000007683, D00000007684, D00000007685, D00000007686, D00000007687, D00000007688, D00000007689, D00000007690, D00000007691, D00000007692, D00000007693, D00000007694, D00000007695, D00000007696, D00000007697, D00000007698, D00000007699, D00000007700, D00000007701, D00000007703, D00000007704, D00000007705, D00000007706, D00000007708, D00000007709, D00000007710, D00000007711, D00000007712, D00000007713, D00000007714, D00000007715, D00000007716, D00000007717, D00000007718, D00000007719, D00000007720, D00000007722, D00000007723, D00000007724, D00000007725, D00000007726, D00000007727, D00000007728, D00000007729, D00000007730, D00000007731, D00000007732, D00000007734, D00000007735, D00000007736, D00000007738, D00000007739, D00000007741, D00000007742, D00000007743, D00000007745, D00000007746, D00000007748, D00000007749, D00000007751, D00000007752, D00000007753, D00000007754, D00000007755, D00000007756, D00000007757, D00000007758, D00000007759, D00000007760, D00000007761, D00000007762, D00000007764, D00000007765, D00000007766, D00000007767, D00000007768, D00000007769, D00000007770, D00000007771, D00000007772, D00000007773, D00000007774, D00000007775, D00000007776, D00000007777, D00000007778, D00000007779, D00000007781, D00000007782, D00000007783, D00000007785, D00000007786, D00000007787, D00000007788, D00000007789, D00000007790, D00000007791, D00000007792, D00000007793, D00000007794, D00000007795, D00000007796, D00000007797, D00000007798, D00000007799, D00000007800, D00000007801, D00000007802, D00000007804, D00000007805, D00000007807, D00000007808, D00000007809, D00000007811, D00000007812, D00000007813, D00000007814, D00000007815, D00000007816, D00000007817, D00000007818, D00000007819, D00000007820, D00000007821, D00000007822, D00000007823, D00000007824, D00000007825, D00000007826, D00000007827, D00000007828, D00000007829, D00000007830, D00000007831, D00000007832, D00000007833, D00000007834, D00000007839, D00000007840, D00000007841, D00000007842, D00000007843, D00000007844, D00000007846, D00000007847, D00000007848, D00000007849, D00000007851, D00000007852, D00000007853, D00000007854, D00000007856, D00000007862, D00000007881, D00000007907, D00000007908, D00000007909, D00000007910, D00000007911, D00000007912, D00000007913, D00000007914, D00000007915, D00000007916, D00000007917, D00000007918, D00000007921, D00000007922, D00000007924, D00000007934, D00000007940, D00000007943, D00000007944, D00000007946, D00000007948, D00000007950, D00000007951, D00000007952, D00000007954, D00000007955, D00000007956, D00000007958, D00000007960, D00000007963, D00000007966, D00000007969, D00000007971, D00000007973, D00000007979, D00000008010, D00000008014, D00000008034, D00000008044, D00000008046, D00000008051, D00000008064, D00000008066, D00000008069, D00000008084, D00000008116, D00000008133, D00000008151, D00000008234, D00000008239, D00000008245, D00000008265, D00000008266, D00000008267, D00000008268, D00000008269, D00000008270, D00000008271, D00000008272, D00000008273, D00000008274, D00000008275, D00000008276, D00000008277, D00000008278, D00000008279, D00000008280, D00000008281, D00000008282, D00000008283, D00000008284, D00000008285, D00000008286, D00000008287, D00000008288, D00000008289, D00000008290, D00000008291, D00000008292, D00000008293, D00000008294, D00000008295, D00000008296, D00000008297, D00000008298, D00000008299, D00000008300, D00000008301, D00000008302, D00000008303, D00000008304, D00000008305, D00000008306, D00000008307, D00000008309, D00000008311, D00000008315, D00000008316, D00000008317, D00000008319, D00000008320, D00000008322, D00000008323, D00000008324, D00000008326, D00000008327, D00000008328, D00000008329, D00000008331, D00000008332, D00000008333, D00000008334, D00000008335, D00000008336, D00000008337, D00000008338, D00000008339, D00000008340, D00000008341, D00000008342, D00000008343, D00000008344, D00000008345, D00000008346, D00000008347, D00000008348, D00000008349, D00000008350, D00000008351, D00000008352, D00000008353, D00000008354, D00000008355, D00000008356, D00000008357, D00000008358, D00000008359, D00000008360, D00000008361, D00000008363, D00000008364, D00000008365, D00000008366, D00000008367, D00000008368, D00000008369, D00000008370, D00000008371, D00000008372, D00000008373, D00000008374, D00000008375, D00000008376, D00000008377, D00000008378, D00000008379, D00000008380, D00000008381, D00000008382, D00000008383, D00000008384, D00000008385, D00000008386, D00000008387, D00000008388, D00000008389, D00000008390, D00000008391, D00000008393, D00000008394, D00000008395, D00000008396, D00000008397, D00000008398, D00000008399, D00000008400, D00000008401, D00000008402, D00000008403, D00000008404, D00000008405, D00000008406, D00000008407, D00000008408, D00000008409, D00000008410, D00000008411, D00000008412, D00000008413, D00000008414, D00000008416, D00000008417, D00000008418, D00000008419, D00000008421, D00000008422, D00000008424, D00000008425, D00000008426, D00000008427, D00000008428, D00000008429, D00000008430, D00000008431, D00000008432, D00000008433, D00000008434, D00000008435, D00000008436, D00000008437, D00000008438, D00000008441, D00000008442, D00000008443, D00000008444, D00000008445, D00000008446, D00000008447, D00000008449, D00000008450, D00000008451, D00000008452, D00000008459, D00000008461, D00000008462, D00000008463, D00000008465, D00000008469, D00000008476, D00000008625, D00000008628, D00000008671, D00000008675, D00000008685, D00000008686, D00000008688, D00000008698, D00000008701, D00000008702, D00000008703, D00000008706, D00000008708, D00000008709, D00000008713, D00000008714, D00000008716, D00000008717, D00000008718, D00000008719, D00000008720, D00000008721, D00000008722, D00000008723, D00000008724, D00000008725, D00000008726, D00000008727, D00000008729, D00000008730, D00000008731, D00000008733, D00000008737, D00000008744, D00000008745, D00000008746, D00000008747, D00000008750, D00000008751, D00000008752, D00000008753, D00000008754, D00000008755, D00000008756, D00000008757, D00000008758, D00000008759, D00000008760, D00000008761, D00000008762, D00000008763, D00000008764, D00000008765, D00000008766, D00000008767, D00000008768, D00000008769, D00000008770, D00000008771, D00000008772, D00000008773, D00000008774, D00000008775, D00000008776, D00000008777, D00000008778, D00000008779, D00000008780, D00000008781, D00000008782, D00000008783, D00000008784, D00000008785, D00000008786, D00000008787, D00000008788, D00000008789, D00000008790, D00000008791, D00000008792, D00000008793, D00000008794, D00000008795, D00000008796, D00000008797, D00000008798, D00000008799, D00000008800, D00000008801, D00000008802, D00000008803, D00000008804, D00000008805, D00000008806, D00000008807, D00000008808, D00000008809, D00000008810, D00000008811, D00000008812, D00000008813, D00000008814, D00000008815, D00000008816, D00000008817, D00000008824, D00000008825, D00000008826, D00000008827, D00000008829, D00000008830, D00000008831, D00000008834, D00000008837, D00000008838, D00000008840, D00000008841, D00000008842, D00000008843, D00000008844, D00000008845, D00000008846, D00000008847, D00000008849, D00000008850, D00000008860, D00000008861, D00000008865, D00000008866, D00000008868, D00000008869, D00000008870, D00000008871, D00000008872, D00000008874, D00000008875, D00000008879, D00000008881, D00000008882, D00000008886, D00000008894, D00000008895, D00000008897, D00000008900, D00000008905, D00000008907, D00000008908, D00000008909, D00000008911, D00000008912, D00000008913, D00000008914, D00000008916, D00000008917, D00000008918, D00000008919, D00000008921, D00000008922, D00000008923, D00000008924, D00000008925, D00000008926, D00000008927, D00000008928, D00000008929, D00000008930, D00000008931, D00000008932, D00000008933, D00000008934, D00000008935, D00000008936, D00000008937, D00000008938, D00000008939, D00000008940, D00000008941, D00000008942, D00000008944, D00000008945, D00000008946, D00000008947, D00000008948, D00000008949, D00000008950, D00000008951, D00000008952, D00000008953, D00000008954, D00000008955, D00000008956, D00000008957, D00000008958, D00000008959, D00000008960, D00000008961, D00000008962, D00000008963, D00000008964, D00000008965, D00000008966, D00000008967, D00000008968, D00000008969, D00000008970, D00000008971, D00000008973, D00000008975, D00000008976, D00000008978, D00000008980, D00000008983, D00000008985, D00000008989, D00000008992, D00000008993, D00000008994, D00000008995, D00000008996, D00000008998, D00000008999, D00000009000, D00000009001, D00000009002, D00000009003, D00000009004, D00000009005, D00000009006, D00000009007, D00000009008, D00000009009, D00000009010, D00000009011, D00000009012, D00000009013, D00000009014, D00000009015, D00000009016, D00000009017, D00000009018, D00000009019, D00000009020, D00000009021, D00000009022, D00000009023, D00000009024, D00000009025, D00000009026, D00000009027, D00000009028, D00000009029, D00000009030, D00000009031, D00000009032, D00000009033, D00000009034, D00000009035, D00000009036, D00000009037, D00000009038, D00000009039, D00000009040, D00000009041, D00000009042, D00000009043, D00000009044, D00000009045, D00000009046, D00000009047, D00000009048, D00000009049, D00000009050, D00000009051, D00000009052, D00000009053, D00000009054, D00000009055, D00000009056, D00000009057, D00000009058, D00000009059, D00000009060, D00000009061, D00000009062, D00000009063, D00000009064, D00000009065, D00000009066, D00000009067, D00000009068, D00000009069, D00000009070, D00000009071, D00000009072, D00000009073, D00000009074, D00000009075, D00000009076, D00000009077, D00000009078, D00000009079, D00000009080, D00000009081, D00000009083, D00000009084, D00000009085, D00000009086, D00000009087, D00000009089, D00000009090, D00000009091, D00000009092, D00000009093, D00000009095, D00000009096, D00000009097, D00000009098, D00000009100, D00000009101, D00000009105, D00000009106, D00000009136, D00000009137, D00000009138, D00000009139, D00000009140, D00000009141, D00000009142, D00000009144, D00000009146, D00000009147, D00000009149, D00000009150, D00000009151, D00000009152, D00000009153, D00000009155, D00000009156, D00000009157, D00000009202, D00000009203, D00000009205, D00000009253, D00000009254, D00000009273, D00000009275, D00000009322, D00000009327, D00000009329, D00000009330, D00000009332, D00000009333, D00000009334, D00000009335, D00000009336, D00000009337, D00000009340, D00000009341, D00000009343, D00000009344, D00000009345, D00000009346, D00000009347, D00000009348, D00000009349, D00000009352, D00000009353, D00000009354, D00000009358, D00000009400, D00000009405, D00000009413, D00000009414, D00000009415, D00000009417, D00000009421, D00000009424, D00000009427, D00000009428, D00000009429, D00000009431, D00000009438, D00000009444, D00000009445, D00000009446, D00000009450, D00000009455, D00000009457, D00000009462, D00000009463, D00000009471, D00000009475, D00000009476, D00000009477, D00000009478, D00000009479, D00000009482, D00000009484, D00000009485, D00000009490, D00000009491, D00000009492, D00000009495, D00000009496, D00000009498, D00000009499, D00000009500, D00000009501, D00000009502, D00000009503, D00000009504, D00000009505, D00000009506, D00000009507, D00000009508, D00000009510, D00000009511, D00000009512, D00000009513, D00000009514, D00000009515, D00000009521, D00000009523, D00000009524, D00000009525, D00000009528, D00000009530, D00000009531, D00000009532, D00000009533, D00000009534, D00000009535, D00000009536, D00000009537, D00000009538, D00000009539, D00000009540, D00000009541, D00000009542, D00000009544, D00000009545, D00000009546, D00000009547, D00000009548, D00000009549, D00000009550, D00000009551, D00000009552, D00000009553, D00000009554, D00000009555, D00000009557, D00000009558, D00000009559, D00000009560, D00000009561, D00000009562, D00000009563, D00000009565, D00000009566, D00000009568, D00000009569, D00000009570, D00000009571, D00000009572, D00000009573, D00000009574, D00000009575, D00000009576, D00000009578, D00000009579, D00000009580, D00000009581, D00000009582, D00000009584, D00000009585, D00000009586, D00000009587, D00000009588, D00000009590, D00000009591, D00000009592, D00000009593, D00000009594, D00000009595, D00000009597, D00000009598, D00000009599, D00000009601, D00000009603, D00000009604, D00000009605, D00000009606, D00000009607, D00000009608, D00000009609, D00000009610, D00000009611, D00000009612, D00000009613, D00000009614, D00000009615, D00000009616, D00000009617, D00000009618, D00000009619, D00000009620, D00000009621, D00000009622, D00000009623, D00000009624, D00000009625, D00000009628, D00000009629, D00000009630, D00000009632, D00000009634, D00000009636, D00000009637, D00000009639, D00000009640, D00000009641, D00000009642, D00000009643, D00000009644, D00000009645, D00000009646, D00000009647, D00000009648, D00000009649, D00000009650, D00000009651, D00000009652, D00000009653, D00000009654, D00000009655, D00000009656, D00000010132, D00000010139, D00000010141, D00000010143, D00000010144, D00000010145, D00000010146, D00000010147, D00000010148, D00000010149, D00000010150, D00000010151, D00000010152, D00000010153, D00000010154, D00000010155, D00000010156, D00000010157, D00000010158, D00000010160, D00000010161, D00000010162, D00000010164, D00000010181, D00000010184, D00000010187, D00000010188, D00000010189, D00000010190, D00000010191, D00000010192, D00000010193, D00000010194, D00000010195, D00000010196, D00000010197, D00000010198, D00000010199, D00000010200, D00000010201, D00000010202, D00000010203, D00000010204, D00000010205, D00000010206, D00000010207, D00000010208, D00000010209, D00000010210, D00000010211, D00000010213, D00000010215, D00000010216, D00000010217, D00000010218, D00000010219, D00000010220, D00000010222, D00000010223, D00000010224, D00000010225, D00000010226, D00000010228, D00000010229, D00000010230, D00000010231, D00000010232, D00000010233, D00000010234, D00000010236, D00000010237, D00000010239, D00000010240, D00000010241, D00000010242, D00000010243, D00000010244, D00000010245, D00000010247, D00000010250, D00000010318, D00000010320, D00000010321, D00000010322, D00000010323, D00000010324, D00000010325, D00000010326, D00000010327, D00000010328, D00000010329, D00000010330, D00000010331, D00000010332, D00000010333, D00000010334, D00000010335, D00000010336, D00000010337, D00000010338, D00000010339, D00000010340, D00000010341, D00000010342, D00000010343, D00000010344, D00000010345, D00000010346, D00000010347, D00000010348, D00000010349, D00000010350, D00000010351, D00000010355, D00000010356, D00000010357, D00000010358, D00000010359, D00000010360, D00000010362, D00000010363, D00000010364, D00000010365, D00000010366, D00000010367, D00000010368, D00000010369, D00000010370, D00000010371, D00000010372, D00000010373, D00000010374, D00000010375, D00000010376, D00000010377, D00000010378, D00000010379, D00000010380, D00000010381, D00000010382, D00000010383, D00000010384, D00000010385, D00000010386, D00000010387, D00000010388, D00000010390, D00000010391, D00000010392, D00000010393, D00000010395, D00000010396, D00000010397, D00000010398, D00000010399, D00000010409, D00000010410, D00000010414, D00000010415, D00000010416, D00000010418, D00000010421, D00000010422, D00000010424, D00000010425, D00000010426, D00000010428, D00000010429, D00000010430, D00000010431, D00000010432, D00000010433, D00000010434, D00000010435, D00000010437, D00000010439, D00000010441, D00000010442, D00000010443, D00000010444, D00000010445, D00000010446, D00000010447, D00000010448, D00000010449, D00000010450, D00000010451, D00000010452, D00000010453, D00000010454, D00000010455, D00000010457, D00000010458, D00000010459, D00000010460, D00000010461, D00000010462, D00000010463, D00000010464, D00000010465, D00000010466, D00000010467, D00000010468, D00000010469, D00000010470, D00000010471, D00000010472, D00000010473, D00000010474, D00000010475, D00000010477, D00000010478, D00000010479, D00000010480, D00000010482, D00000010483, D00000010484, D00000010485, D00000010486, D00000010487, D00000010488, D00000010489, D00000010490, D00000010491, D00000010492, D00000010493, D00000010494, D00000010495, D00000010499, D00000010500, D00000010501, D00000010502, D00000010503, D00000010504, D00000010506, D00000010507, D00000010508, D00000010509, D00000010510, D00000010511, D00000010512, D00000010513, D00000010514, D00000010515, D00000010516, D00000010517, D00000010518, D00000010519, D00000010520, D00000010521, D00000010522, D00000010523, D00000010524, D00000010525, D00000010526, D00000010527, D00000010528, D00000010529, D00000010530, D00000010531, D00000010532, D00000010533, D00000010534, D00000010535, D00000010536, D00000010537, D00000010538, D00000010540, D00000010541, D00000010542, D00000010543, D00000010544, D00000010546, D00000010547, D00000010548, D00000010549, D00000010550, D00000010551, D00000010552, D00000010553, D00000010555, D00000010556, D00000010557, D00000010558, D00000010559, D00000010560, D00000010561, D00000010562, D00000010563, D00000010564, D00000010565, D00000010566, D00000010567, D00000010568, D00000010569, D00000010570, D00000010571, D00000010572, D00000010573, D00000010574, D00000010575, D00000010576, D00000010577, D00000010578, D00000010579, D00000010580, D00000010581, D00000010582, D00000010583, D00000010584, D00000010585, D00000010586, D00000010587, D00000010588, D00000010590, D00000010591, D00000010592, D00000010593, D00000010596, D00000010597, D00000010598, D00000010599, D00000010600, D00000010601, D00000010602, D00000010603, D00000010605, D00000010606, D00000010607, D00000010611, D00000010612, D00000010615, D00000010616, D00000010622, D00000010627, D00000010630, D00000010631, D00000010632, D00000010634, D00000010640, D00000010641, D00000010647, D00000010653, D00000010655, D00000010657, D00000010661, D00000010670, D00000010671, D00000010678, D00000010685, D00000010687, D00000010689, D00000010691, D00000010693, D00000010695, D00000010696, D00000010697, D00000010698, D00000010700, D00000010702, D00000010703, D00000010704, D00000010705, D00000010706, D00000010707, D00000010709, D00000010710, D00000010711, D00000010712, D00000010715, D00000010716, D00000010718, D00000010720, D00000010721, D00000010724, D00000010732, D00000010739, D00000010756, D00000010761, D00000010762, D00000010766, D00000010776, D00000010778, D00000010782, D00000010783, D00000010788, D00000010795, D00000010797, D00000010804, D00000010807, D00000010821, D00000010866, D00000010883, D00000010885, D00000010888, D00000010889, D00000010890, D00000010895, D00000010898, D00000010902, D00000010903, D00000010904, D00000010905, D00000010907, D00000010908, D00000010909, D00000010911, D00000010912, D00000010920, D00000010930, D00000010934, D00000010935, D00000010936, D00000010937, D00000010938, D00000010939, D00000010940, D00000010943, D00000010944, D00000010977, D00000010993, D00000011033, D00000011053, D00000011062, D00000011063, D00000011068, D00000011069, D00000011073, D00000011075, D00000011077, D00000011078, D00000011079, D00000011080, D00000011081, D00000011082, D00000011083, D00000011084, D00000011085, D00000011086, D00000011087, D00000011088, D00000011089, D00000011090, D00000011091, D00000011092, D00000011093, D00000011094, D00000011095, D00000011096, D00000011097, D00000011098, D00000011099, D00000011100, D00000011101, D00000011102, D00000011103, D00000011104, D00000011105, D00000011106, D00000011107, D00000011108, D00000011109, D00000011110, D00000011111, D00000011112, D00000011115, D00000011116, D00000011117, D00000011118, D00000011119, D00000011120, D00000011122, D00000011123, D00000011124, D00000011126, D00000011128, D00000011129, D00000011135, D00000011136, D00000011137, D00000011140, D00000011141, D00000011142, D00000011143, D00000011144, D00000011145, D00000011146, D00000011147, D00000011148, D00000011149, D00000011150, D00000011151, D00000011152, D00000011153, D00000011154, D00000011155, D00000011156, D00000011157, D00000011158, D00000011159, D00000011160, D00000011161, D00000011162, D00000011163, D00000011164, D00000011165, D00000011166, D00000011167, D00000011168, D00000011169, D00000011170, D00000011171, D00000011172, D00000011173, D00000011174, D00000011175, D00000011176, D00000011177, D00000011178, D00000011179, D00000011180, D00000011182, D00000011183, D00000011184, D00000011185, D00000011188, D00000011189, D00000011190, D00000011191, D00000011192, D000

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1317-2025

Related Recalls

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Feb 13, 2025 Infusion Pumps Nationwide View Details →

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Feb 13, 2025 Infusion Pumps Nationwide View Details →

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Feb 13, 2025 Infusion Pumps Nationwide View Details →