PadPro(R) Adult Radiotransparent Multifunction Electrodes, REF/Catalog No. 2001M-C. The PadPro(R) Adult Radiotransparent Multifunction Electrodes with Carbon Leadwires and Medtronic(R) Connector are intended to be used in external pacing, defibrillation and monitoring applications
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Code Y092012-21, Manufacture Date 10/10/2012; Lot Code Y102612-12, Manufacture Date 11/17/2012; Lot Code Y110912-20, Manufacture Date 11/30/2012; Lot Code Y112912-10, Manufacture Date 12/22/2012; Lot Code Y121312-19, Manufacture Date 01/03/2013; Lot Code Y011013-13, Manufacture Date 01/29/2013; Lot Code Y012513-02, Manufacture Date 02/12/2013; Lot Code Y020713-05, Manufacture Date 02/26/2013; Lot Code Y021413-03, Manufacture Date 03/07/2013; Lot Code Y030113-04, Manufacture Date 03/23/2013
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ConMed Corporation
- Reason for Recall:
- ConMed was notified by the contract manufacturer that their wire/connector supplier reported damage to the connector tooling which could pinch or compress the lead wire filaments and potentially render the device incapable of delivering the appropriate energy or shock to the patient and/or could cause potential injury to user or patient.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
PadPro(R) Adult Radiotransparent Multifunction Electrodes, REF/Catalog No. 2001M-C. The PadPro(R) Adult Radiotransparent Multifunction Electrodes with Carbon Leadwires and Medtronic(R) Connector are intended to be used in external pacing, defibrillation and monitoring applications
Product Codes/Lot Numbers:
Lot Code Y092012-21, Manufacture Date 10/10/2012; Lot Code Y102612-12, Manufacture Date 11/17/2012; Lot Code Y110912-20, Manufacture Date 11/30/2012; Lot Code Y112912-10, Manufacture Date 12/22/2012; Lot Code Y121312-19, Manufacture Date 01/03/2013; Lot Code Y011013-13, Manufacture Date 01/29/2013; Lot Code Y012513-02, Manufacture Date 02/12/2013; Lot Code Y020713-05, Manufacture Date 02/26/2013; Lot Code Y021413-03, Manufacture Date 03/07/2013; Lot Code Y030113-04, Manufacture Date 03/23/2013
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1317-2013
Related Recalls
Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could cause a delay in procedure and could require another device to be used.
Tip Guides are potentially misaligned laterally which could affect the accuracy of the guide system and could cause a delay in procedure and may require another device or alternate surgical method to be used
Tip Guides are potentially misaligned laterally which could affect the accuracy of the guide system and could cause a delay in procedure and may require another device or alternate surgical method to be used