ATTUNE¿ FIXED BEARING KNEE - Product Usage: composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot#: 9558300, Product Code: 150670003, GTIN 10603295492054 Lot#: 9555958, Product Code: 150670004, GTIN 10603295492061 Lot#: 9554006, Product Code: 150670005, GTIN 10603295492078 Lot#: 9556397, Product Code: 150670005, GTIN 10603295492078 Lot#: 9558618, Product Code: 150670006, GTIN 10603295492085 Lot#: 9557992, Product Code: 150670007, GTIN 10603295492092 Lot#: 9558238, Product Code: 150670008, GTIN 10603295492108 Lot#: 9558239, Product Code: 150670008, GTIN 10603295492108 Lot#: 9555613, Product Code: 150670009, GTIN 10603295492115
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- DePuy Orthopaedics, Inc.
- Reason for Recall:
- Globally the company is removing 37 lots affected lots for MBT & ATTUNE (S+) Tibial Bases due to potential damaged packaging. These implant components are utilized in Total Knee Arthroplasty. The affected tibial bases are designed to be used in cemented applications.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ATTUNE¿ FIXED BEARING KNEE - Product Usage: composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design.
Product Codes/Lot Numbers:
Lot#: 9558300, Product Code: 150670003, GTIN 10603295492054 Lot#: 9555958, Product Code: 150670004, GTIN 10603295492061 Lot#: 9554006, Product Code: 150670005, GTIN 10603295492078 Lot#: 9556397, Product Code: 150670005, GTIN 10603295492078 Lot#: 9558618, Product Code: 150670006, GTIN 10603295492085 Lot#: 9557992, Product Code: 150670007, GTIN 10603295492092 Lot#: 9558238, Product Code: 150670008, GTIN 10603295492108 Lot#: 9558239, Product Code: 150670008, GTIN 10603295492108 Lot#: 9555613, Product Code: 150670009, GTIN 10603295492115
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1314-2021
Related Recalls
ATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total Knee Replacement System.
DePuy Orthopaedics
Product incorrectly labelled.
Attune Revision Limb Preservation System (LPS) Tibial Insert XXSM 18MM, Part Number 151760118
DePuy Orthopaedics
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Attune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312
DePuy Orthopaedics
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.