VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.
Class I - DangerousWhat Should You Do?
- Check if you have this product: VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. (System: VITEK 2 60, Reference Number: 27225, Catalog Number: C20, UDI: 03573026162351); (System: VITEK 2 60, Reference Number: 27225R, Catalog Number: C20, UDI: 03573026206765); (System: VITEK 2 XL, Reference Number: 27227, Catalog Number: C20, UDI: 03573026162375); (System: VITEK 2 XL, Reference Number: 27227R, Catalog Number: C20, UDI: 03573026206789). Each instrument has unique serial number.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Biomerieux Inc
- Reason for Recall:
- bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK¿ 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK¿ 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.
Product Codes/Lot Numbers:
VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. (System: VITEK 2 60, Reference Number: 27225, Catalog Number: C20, UDI: 03573026162351); (System: VITEK 2 60, Reference Number: 27225R, Catalog Number: C20, UDI: 03573026206765); (System: VITEK 2 XL, Reference Number: 27227, Catalog Number: C20, UDI: 03573026162375); (System: VITEK 2 XL, Reference Number: 27227R, Catalog Number: C20, UDI: 03573026206789). Each instrument has unique serial number.
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1307-2021
Related Recalls
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.