Implantable cardioverter defibrillators (ICD): 1) Evera family: Evera XT DR, Evera S DR, Evera MRI XT DR, Evera MRI S DR, Evera XT VR, Evera S VR and Evera MRI S VR models; 2) Visia family: Visia AF ICD, Visia AF S ICD, Visia AF MRI ICD, and Visia AF MRI S ICD models. Implantable Cardioverter Defibrillators (ICDs) are implantable medical devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1) Evera Family: MODEL NUMBER / NAME / GTIN: SERIAL NUMBERS: DDBB1D1 / ICD-DR DDBB1D1 EVERA XT US IS1/DF1 / 00643169018099: BWC231766H, BWC230184H, BWC230891H, BWC228489H, BWC232037H, BWC230770H, BWC231219H, BWC231215H, BWC231190H, BWC231202H, BWC231199H, BWC231780H, BWC231556H, BWC230255H, BWC230772H, BWC230906H, BWC231500H, BWC231572H, BWC231618H, BWC231655H, BWC231937H, BWC231942H, BWC230859H, BWC230949H, BWC231848H, BWC230918H, BWC231237H, BWC231580H, BWC230893H, BWC231576H, BWC230426H, BWC230475H, BWC230431H, BWC231494H, BWC230876H, BWC230776H, BWC228873H, BWC229321H, BWC230873H, BWC229698H, BWC230449H, BWC231452H, BWC230877H, BWC230720H, BWC231479H, BWC231482H, BWC231578H, BWC231498H, BWC229796H, BWC230863H, BWC230874H, BWC231239H, BWC230917H, BWC229337H, BWC230524H, BWC230887H, BWC231525H, BWC231240H, BWC230672H, BWC230769H, BWC232164H, BWC232196H, BWC230048H, BWC230520H, BWC231568H, BWC231671H, BWC231761H, BWC229598H, BWC230335H, BWC230889H, BWC231570H, BWC228859H, BWC230914H, BWC231346H, BWC230884H, BWC231501H, BWC229058H, BWC229697H, BWC228813H, BWC232165H, BWC230870H, BWC230901H, BWC231259H, BWC230872H, BWC231106H, BWC230195H, BWC230911H, BWC228448H, BWC229342H, BWC230535H, BWC230655H, BWC230919H, BWC231269H, BWC231480H, BWC231668H, BWC230044H, BWC228426H, BWC230910H, BWC227091H, BWC231456H, BWC229099H, BWC229101H, BWC231874H, BWC228336H, BWC230869H, BWC230730H, BWC230630H, BWC231872H, BWC231343H, BWC228800H, BWC228656H, BWC228666H, BWC229097H, BWC231944H, BWC226887H, BWC230885H, BWC231768H, BWC232133H, BWC227119H, BWC228318H, BWC229963H, BWC230899H, BWC228760H, BWC230159H, BWC230882H, BWC230381H, BWC228544H, BWC230098H, BWC231400H, BWC228205H, BWC228858H, BWC226830H, BWC230454H, BWC228816H, BWC230608H, BWC228988H, BWC227088H, BWC228586H, BWC226816H, BWC227609H, BWC228180H, BWC228449H, BWC229289H, BWC229600H, BWC230118H, BWC230732H, BWC232122H, BWC231241H, BWC228818H, BWC229317H, BWC229344H, BWC229118H, BWC230071H, BWC227808H, BWC232126H, BWC232191H, BWC232170H, BWC227125H, BWC227610H, BWC227114H, BWC230530H, BWC227067H, BWC227782H, BWC228871H, BWC228399H, BWC228427H, BWC231659H, BWC228044H, BWC228820H, BWC228206H, BWC229013H, BWC231037H, BWC231809H, BWC226906H, BWC229163H, BWC227324H, BWC228320H, BWC227017H, BWC227077H, BWC227034H, BWC227374H, BWC228469H, BWC231485H, BWC225637H, BWC227113H, BWC228867H, BWC231257H, BWC227047H, BWC227071H, BWC224560H, BWC226819H, BWC226824H, BWC226960H, BWC227129H, BWC228864H, BWC230020H, BWC231332H, BWC231458H, BWC226403H, BWC226833H, BWC227056H, BWC227062H, BWC227126H, BWC228869H, BWC226939H, BWC227326H, BWC230083H, BWC231865H, BWC229346H, BWC231669H, BWC227323H, BWC230503H, BWC231049H, BWC228832H, BWC229536H, BWC226805H, BWC226822H, BWC226914H, BWC231406H, BWC228248H, BWC225448H, BWC226799H, BWC227109H, BWC228132H, BWC229168H, BWC231459H, BWC231673H, BWC226801H, BWC226817H, BWC228812H, BWC231043H, BWC231192H, BWC229055H, BWC226987H, BWC231173H, BWC229988H, BWC226820H, BWC226831H, BWC230642H, BWC226798H, BWC229516H, BWC229886H, BWC231781H, BWC226832H, BWC225279H, BWC226809H, BWC226812H, BWC227130H, BWC228297H, BWC229806H, BWC226664H, BWC230704H, BWC231684H, BWC226998H, BWC230322H, BWC225786H, BWC226298H, BWC227032H, BWC230010H, BWC231324H, BWC224500H, BWC229757H, BWC230573H, BWC227008H, BWC230091H, BWC231550H, BWC225592H, BWC226988H, BWC227163H, BWC227322H, BWC224540H, BWC230309H, BWC232166H, BWC227025H, BWC230791H, BWC231036H, BWC227845H, BWC230784H, BWC232050H, BWC226416H, BWC226966H, BWC228943H, BWC228960H, BWC228978H, BWC231908H, BWC228014H, BWC228050H, BWC229537H, BWC226984H, BWC227330H, BWC231176H, BWC231412H, BWC229009H, BWC230481H, BWC226584H, BWC229898H, BWC225730H, BWC226415H, BWC226924H, BWC227316H, BWC227321H, BWC230411H, BWC226578H, BWC229981H, BWC230279H, BWC224159H, BWC225056H, BWC230137H, BWC229606H, BWC231774H, BWC225810H, BWC226471H, BWC227919H, BWC228962H, BWC229743H, BWC230565H, BWC228743H, BWC231660H, BWC225960H, BWC228522H, BWC232106H, BWC232113H, BWC225555H, BWC227536H, BWC228919H, BWC228974H, BWC208701H, BWC231047H, BWC226182H, BWC231976H, BWC226398H, BWC226639H, BWC229914H, BWC230318H, BWC230916H, BWC230937H, BWC231395H, BWC232174H, BWC226053H, BWC228270H, BWC225617H, BWC226923H, BWC228131H, BWC229142H, BWC230648H, BWC231137H, BWC224556H, BWC224377H, BWC225611H, BWC226587H, BWC228342H, BWC229758H, BWC230247H, BWC230404H, BWC224492H, BWC228171H, BWC228363H, BWC228582H, BWC229921H, BWC230590H, BWC230691H, BWC231024H, BWC231250H, BWC231392H, BWC232118H, BWC232181H, BWC222232H, BWC223575H, BWC226139H, BWC226703H, BWC227005H, BWC227196H, BWC228392H, BWC228969H, BWC229071H, BWC229414H, BWC230028H, BWC230181H, BWC231490H, BWC231729H, BWC231864H, BWC231997H, BWC232155H, BWC232206H, BWC225667H, BWC226056H, BWC226643H, BWC231184H, BWC232115H, BWC226299H, BWC227028H, BWC228060H, BWC227023H, BWC228445H, BWC229075H, BWC229787H, BWC232001H, BWC220805H, BWC225084H, BWC227612H, BWC227882H, BWC228414H, BWC230410H, BWC230774H, BWC231403H, BWC231549H, BWC232090H, BWC228497H, BWC229396H, BWC229590H, BWC231105H, BWC224761H, BWC225054H, BWC225527H, BWC226180H, BWC226451H, BWC230337H, BWC230701H, BWC230934H, BWC231189H, BWC227871H, BWC228545H, BWC230505H, BWC231039H, BWC231705H, BWC231845H, BWC224849H, BWC226278H, BWC229634H, BWC231477H, BWC231528H, BWC232116H, BWC232145H, BWC224591H, BWC226782H, BWC228310H, BWC229104H, BWC230740H, BWC226791H, BWC228968H, BWC229868H, BWC229920H, BWC231175H, BWC231436H, BWC230117H, BWC231198H, BWC231496H, BWC231970H, BWC230302H, BWC231274H, BWC223441H, BWC228258H, BWC231905H, BWC227947H, BWC228058H, BWC228068H, BWC229039H, BWC229098H, BWC229248H, BWC229618H, BWC232032H, BWC225186H, BWC225677H, BWC226274H, BWC228083H, BWC229573H, BWC232046H, BWC226964H, BWC230055H, BWC231147H, BWC231478H, BWC224493H, BWC227552H, BWC228446H, BWC230284H, BWC231021H, BWC231465H, BWC231916H, BWC228096H, BWC229960H, BWC231214H, BWC224572H, BWC224596H, BWC224602H, BWC224615H, BWC226736H, BWC228147H, BWC228277H, BWC229145H, BWC230330H, BWC231211H, BWC231364H, BWC231923H, BWC224335H, BWC224579H, BWC228824H, BWC230447H, BWC230745H, BWC231114H, BWC231296H, BWC231830H, BWC223169H, BWC225062H, BWC225458H, BWC226225H, BWC230960H, BWC231138H, BWC223551H, BWC223978H, BWC229236H, BWC229699H, BWC230644H, BWC231495H, BWC231888H, BWC231894H, BWC231947H, BWC225001H, BWC225950H, BWC229477H, BWC230561H, BWC230953H, BWC231404H, BWC224769H, BWC226650H, BWC228191H, BWC228926H, BWC229822H, BWC230690H, BWC225929H, BWC229204H, BWC229888H, BWC230094H, BWC230499H, BWC230640H, BWC230789H, BWC225802H, BWC229925H, BWC221930H, BWC224327H, BWC224563H, BWC224609H, BWC230018H, BWC230164H, BWC230692H, BWC231281H, BWC231597H, BWC231755H, BWC224716H, BWC228165H, BWC228670H, BWC229350H, BWC229912H, BWC230068H, BWC230477H, BWC230987H, BWC231328H, BWC232192H, BWC224809H, BWC227258H, BWC228264H, BWC228387H, BWC230143H, BWC230216H, BWC230654H, BWC230697H, BWC224568H, BWC226949H, BWC228223H, BWC230115H, BWC230136H, BWC230656H, BWC231209H, BWC231464H, BWC225157H, BWC228394H, BWC228673H, BWC229585H, BWC230206H, BWC230317H, BWC231881H, BWC224768H, BWC224838H, BWC227072H, BWC229990H, BWC230467H, BWC230502H, BWC231443H, BWC232024H, BWC232167H, BWC221901H, BWC224605H, BWC224683H, BWC230658H, BWC231933H, BWC232120H, BWC228161H, BWC229230H, BWC229576H, BWC230380H, BWC230722H, BWC230744H, BWC223304H, BWC224573H, BWC224976H, BWC225998H, BWC228654H, BWC229490H, BWC229933H, BWC231431H, BWC231596H, BWC231734H, BWC228554H, BWC228588H, BWC230668H, BWC230792H, BWC231444H, BWC229546H, BWC222703H, BWC224052H, BWC224571H, BWC225153H, BWC227927H, BWC228309H, BWC229775H, BWC230078H, BWC230315H, BWC230729H, BWC231276H, BWC224517H, BWC224960H, BWC228214H, BWC228579H, BWC229117H, BWC229691H, BWC232020H, BWC226111H, BWC230089H, BWC227370H, BWC230053H, BWC230193H, BWC230443H, BWC230700H, BWC231777H, BWC224597H, BWC227185H, BWC228003H, BWC228196H, BWC229377H, BWC229485H, BWC229519H, BWC230316H, BWC230816H, BWC230832H, BWC231409H, BWC231413H, BWC224581H, BWC221774H, BWC223806H, BWC224504H, BWC225399H, BWC225791H, BWC229059H, BWC230534H, BWC230749H, BWC231125H, BWC231909H, BWC231934H, BWC224490H, BWC227688H, BWC232019H, BWC223259H, BWC224559H, BWC225344H, BWC226698H, BWC230173H, BWC230295H, BWC230728H, BWC231212H, BWC232128H, BWC223821H, BWC223973H, BWC227073H, BWC232073H, BWC232175H, BWC232179H, BWC218949H, BWC223948H, BWC224594H, BWC224786H, BWC224811H, BWC226026H, BWC227939H, BWC228680H, BWC228970H, BWC230202H, BWC230986H, BWC231307H, BWC231337H, BWC231754H, BWC231910H, BWC232169H, BWC223161H, BWC225308H, BWC227572H, BWC227913H, BWC230022H, BWC230417H, BWC222829H, BWC223122H, BWC223979H, BWC224616H, BWC230200H, BWC230259H, BWC232102H, BWC224607H, BWC224963H, BWC230604H, BWC230696H, BWC231311H, BWC231312H, BWC231960H, BWC222823H, BWC224576H, BWC224800H, BWC227530H, BWC228763H, BWC229755H, BWC230497H, BWC230800H, BWC231347H, BWC231434H, BWC232054H, BWC232190H, BWC223372H, BWC225684H, BWC230019H, BWC230554H, BWC230736H, BWC231255H, BWC231435H, BWC222777H, BWC222865H, BWC230015H, BWC230190H, BWC231136H, BWC221087H, BWC222263H, BWC223494H, BWC225738H, BWC225913H, BWC230008H, BWC230170H, BWC230513H, BWC230546H, BWC231264H, BWC232017H, BWC232018H, BWC221622H, BWC222804H, BWC224588H, BWC227295H, BWC228130H, BWC228384H, BWC230237H, BWC231249H, BWC232003H, BWC221917H, BWC228744H, BWC229432H, BWC229735H, BWC231355H, BWC225110H, BWC225135H, BWC227986H, BWC227991H, BWC229447H, BWC230042H, BWC230308H, BWC231749H, BWC221936H, BWC222564H, BWC225761H, BWC227044H, BWC229777H, BWC230354H, BWC230372H, BWC230485H, BWC231028H, BWC232168H, BWC222392H, BWC223370H, BWC226823H, BWC228240H, BWC229186H, BWC229966H, BWC232161H, BWC221611H, BWC221761H, BWC223079H, BWC225401H, BWC225906H, BWC227049H, BWC229007H, BWC221773H, BWC228305H, BWC229756H, BWC229838H, BWC230296H, BWC232004H, BWC232045H, BWC232083H, BWC226747H, BWC230359H, BWC231124H, BWC231375H, BWC232007H, BWC232184H, BWC230326H, BWC231949H, BWC219373H, BWC226828H, BWC227874H, BWC228136H, BWC228471H, BWC228927H, BWC229265H, BWC229826H, BWC230299H, BWC230441H, BWC230570H, BWC230585H, BWC230767H, BWC231305H, BWC231407H, BWC231763H, BWC231861H, BWC231897H, BWC232103H, BWC219794H, BWC222825H, BWC225926H, BWC228275H, BWC228353H, BWC228939H, BWC230157H, BWC230221H, BWC230331H, BWC230444H, BWC230900H, BWC230938H, BWC231156H, BWC231360H, BWC231687H, BWC231879H, BWC232178H, BWC221006H, BWC222840H, BWC223041H, BWC229192H, BWC229893H, BWC230516H, BWC230605H, BWC230657H, BWC230797H, BWC231031H, BWC231873H, BWC220956H, BWC222763H, BWC223302H, BWC223598H, BWC227170H, BWC227678H, BWC227804H, BWC229004H, BWC229752H, BWC229907H, BWC230139H, BWC230760H, BWC231100H, BWC231530H, BWC231857H, BWC232076H, BWC232124H, BWC221539H, BWC225159H, BWC225673H, BWC229213H, BWC229357H, BWC230538H, BWC230614H, BWC230686H, BWC231351H, BWC231504H, BWC231832H, BWC231849H, BWC221448H, BWC221549H, BWC221864H, BWC223893H, BWC228866H, BWC229056H, BWC229115H, BWC229273H, BWC229849H, BWC230483H, BWC230922H, BWC231965H, BWC220490H, BWC220673H, BWC221847H, BWC222048H, BWC223384H, BWC224835H, BWC225321H, BWC226930H, BWC227485H, BWC227984H, BWC228082H, BWC228120H, BWC230026H, BWC230040H, BWC230063H, BWC230504H, BWC230543H, BWC231422H, BWC231971H, BWC221002H, BWC229985H, BWC230057H, BWC230583H, BWC230669H, BWC230727H, BWC230936H, BWC231023H, BWC232052H, BWC232114H, BWC220153H, BWC221545H, BWC224371H, BWC225024H, BWC225848H, BWC226019H, BWC228644H, BWC228712H, BWC229149H, BWC229369H, BWC230168H, BWC230537H, BWC230626H, BWC230670H, BWC231925H, BWC231991H, BWC232000H, BWC232029H, BWC219394H, BWC225316H, BWC226121H, BWC226583H, BWC226933H, BWC227993H, BWC228220H, BWC228556H, BWC229629H, BWC230041H, BWC230356H, BWC230521H, BWC230689H, BWC230779H, BWC231058H, BWC231172H, BWC231718H, BWC231922H, BWC232153H, BWC231919H, BWC232142H, BWC223121H, BWC223378H, BWC227733H, BWC227973H, BWC229381H, BWC229649H, BWC229662H, BWC229770H, BWC231235H, BWC231326H, BWC221102H, BWC221531H, BWC222136H, BWC222780H, BWC222984H, BWC224911H, BWC225451H, BWC225758H, BWC228837H, BWC229614H, BWC230188H, BWC231833H, BWC220869H, BWC221892H, BWC221908H, BWC224894H, BWC228383H, BWC228389H, BWC229776H, BWC231834H, BWC231904H, BWC231994H, BWC222819H, BWC227302H, BWC229157H, BWC230100H, BWC230397H, BWC230473H, BWC230674H, BWC231090H, BWC231928H, BWC232022H, BWC224225H, BWC225465H, BWC227237H, BWC227501H, BWC228440H, BWC229132H, BWC230217H, BWC230258H, BWC230509H, BWC230579H, BWC230673H, BWC230981H, BWC231038H, BWC231318H, BWC231336H, BWC231835H, BWC230176H, BWC231526H, BWC221051H, BWC221378H, BWC226994H, BWC231309H, BWC231448H, BWC231954H, BWC231978H, BWC221606H, BWC224824H, BWC224877H, BWC228089H, BWC229006H, BWC229360H, BWC229679H, BWC229953H, BWC230185H, BWC230362H, BWC230435H, BWC230612H, BWC230803H, BWC231226H, BWC231252H, BWC231348H, BWC231531H, BWC231837H, BWC231856H, BWC231958H, BWC232141H, BWC221775H, BWC222003H, BWC225649H, BWC226762H, BWC227296H, BWC227494H, BWC227562H, BWC228355H, BWC228601H, BWC229217H, BWC229259H, BWC229748H, BWC229918H, BWC230388H, BWC230693H, BWC230935H, BWC231299H, BWC231906H, BWC231980H, BWC220857H, BWC224813H, BWC225474H, BWC225859H, BWC227608H, BWC229162H, BWC230904H, BWC231119H, BWC231253H, BWC231315H, BWC231340H, BWC231614H, BWC231871H, BWC224821H, BWC226066H, BWC226466H, BWC229715H, BWC229767H, BWC230263H, BWC230578H, BWC231283H, BWC231593H, BWC231733H, BWC232038H, BWC232146H, BWC221796H, BWC223280H, BWC224883H, BWC224933H, BWC225182H, BWC228009H, BWC228848H, BWC230371H, BWC230526H, BWC230621H, BWC231050H, BWC231938H, BWC232084H, BWC225403H, BWC225731H, BWC227652H, BWC228273H, BWC229426H, BWC229722H, BWC230273H, BWC230383H, BWC230398H, BWC230507H, BWC230705H, BWC231643H, BWC231700H, BWC220741H, BWC222941H, BWC223104H, BWC226052H, BWC226205H, BWC228362H, BWC229388H, BWC229658H, BWC229782H, BWC230229H, BWC230366H, BWC230568H, BWC230582H, BWC231685H, BWC231831H, BWC231889H, BWC232013H, BWC232111H, BWC220461H, BWC221129H, BWC229465H, BWC229798H, BWC230562H, BWC231917H, BWC232078H, BWC232095H, BWC220648H, BWC229065H, BWC231072H, BWC231930H, BWC221996H, BWC224705H, BWC227692H, BWC227779H, BWC228185H, BWC229450H, BWC229544H, BWC230205H, BWC230392H, BWC230482H, BWC230665H, BWC231350H, BWC231373H, BWC231752H, BWC231877H, BWC222061H, BWC222295H, BWC223242H, BWC227229H, BWC227526H, BWC228661H, BWC228996H, BWC229821H, BWC230082H, BWC230541H, BWC231132H, BWC231134H, BWC231445H, BWC231529H, BWC232026H, BWC232066H, BWC218954H, BWC219280H, BWC220239H, BWC221167H, BWC221340H, BWC226421H, BWC227538H, BWC227885H, BWC229123H, BWC229478H, BWC229564H, BWC230569H, BWC230928H, BWC230970H, BWC231019H, BWC231195H, BWC231638H, BWC231950H, BWC231973H, BWC232015H, BWC232071H, BWC232180H, BWC219852H, BWC223071H, BWC225553H, BWC225813H, BWC227413H, BWC227869H, BWC228018H, BWC228286H, BWC229143H, BWC229364H, BWC229652H, BWC229779H, BWC230268H, BWC230393H, BWC231146H, BWC232129H, BWC221314H, BWC227206H, BWC227715H, BWC231402H, BWC231985H, BWC231990H, BWC232127H, BWC231065H, BWC221584H, BWC223982H, BWC225388H, BWC228022H, BWC228457H, BWC230087H, BWC230236H, BWC230515H, BWC230681H, BWC231875H, BWC231972H, BWC232025H, BWC232172H, BWC218966H, BWC221149H, BWC224829H, BWC227849H, BWC229254H, BWC229854H, BWC230539H, BWC230663H, BWC230915H, BWC231527H, BWC231900H, BWC231975H, BWC232099H, BWC232160H, BWC221700H, BWC222838H, BWC227111H, BWC227412H, BWC227468H, BWC229067H, BWC229359H, BWC229718H, BWC229841H, BWC229910H, BWC230211H, BWC230931H, BWC231758H, BWC231920H, BWC232005H, BWC218454H, BWC220603H, BWC221163H, BWC221550H, BWC221859H, BWC227622H, BWC228677H, BWC228736H, BWC228775H, BWC229291H, BWC229764H, BWC229877H, BWC230047H, BWC230076H, BWC230099H, BWC230470H, BWC230982H, BWC231063H, BWC231112H, BWC231365H, BWC231677H, BWC231852H, BWC231903H, BWC223009H, BWC223382H, BWC223883H, BWC224507H, BWC225310H, BWC226179H, BWC228027H, BWC230203H, BWC230710H, BWC230810H, BWC230933H, BWC231208H, BWC231331H, BWC231913H, BWC231974H, BWC231999H, BWC232006H, BWC232014H, BWC232109H, BWC232140H, BWC225721H, BWC226071H, BWC226166H, BWC226990H, BWC228418H, BWC230017H, BWC230067H, BWC230080H, BWC230269H, BWC230572H, BWC231201H, BWC231243H, BWC231385H, BWC231586H, BWC231982H, BWC232053H, BWC232094H, BWC232162H, BWC220270H, BWC222148H, BWC228793H, BWC230158H, BWC230238H, BWC230464H, BWC230471H, BWC231088H, BWC231196H, BWC231339H, BWC231378H, BWC231854H, BWC232058H, BWC224405H, BWC228221H, BWC228479H, BWC228753H, BWC228872H, BWC229110H, BWC229155H, BWC229609H, BWC229784H, BWC229874H, BWC230106H, BWC230396H, BWC231335H, BWC231369H, BWC231379H, BWC231408H, BWC231935H, BWC231966H, BWC232119H, BWC232193H, BWC220069H, BWC221992H, BWC222868H, BWC228746H, BWC229170H, BWC229508H, BWC230303H, BWC230773H, BWC230841H, BWC230890H, BWC231354H, BWC232034H, BWC232183H, BWC232188H, BWC232207H, BWC226935H, BWC227182H, BWC228339H, BWC229633H, BWC229742H, BWC230210H, BWC230566H, BWC230698H, BWC231004H, BWC231181H, BWC231929H, BWC231987H, BWC231989H, BWC232039H, BWC232075H, BWC220362H, BWC221937H, BWC223477H, BWC225034H, BWC226032H, BWC227372H, BWC228197H, BWC228405H, BWC229713H, BWC229819H, BWC229961H, BWC230145H, BWC230430H, BWC230523H, BWC231001H, BWC231187H, BWC231405H, BWC231840H, BWC231853H, BWC231863H, BWC231945H, BWC232042H, BWC232085H, BWC221527H, BWC222005H, BWC222783H, BWC223582H, BWC224442H, BWC226390H, BWC226787H, BWC227424H, BWC227690H, BWC229215H, BWC229754H, BWC229977H, BWC230160H, BWC230433H, BWC230798H, BWC230843H, BWC231030H, BWC231358H, BWC231566H, BWC231867H, BWC231915H, BWC232030H, BWC223995H, BWC224969H, BWC225740H, BWC225971H, BWC227381H, BWC227583H, BWC228539H, BWC228598H, BWC229135H, BWC229524H, BWC229526H, BWC229545H, BWC229554H, BWC230288H, BWC230351H, BWC230487H, BWC230511H, BWC230688H, BWC230924H, BWC231127H, BWC231357H, BWC231739H, BWC231884H, BWC231912H, BWC231956H, BWC231962H, BWC232088H, BWC220357H, BWC220604H, BWC220905H, BWC221564H, BWC221623H, BWC224834H, BWC225626H, BWC226941H, BWC227724H, BWC229383H, BWC229492H, BWC229646H, BWC229739H, BWC230544H, BWC230743H, BWC231282H, BWC231776H, BWC232065H, BWC232098H, BWC232194H, BWC219691H, BWC225536H, BWC226113H, BWC227139H, BWC227618H, BWC227780H, BWC228193H, BWC228526H, BWC229410H, BWC229638H, BWC229869H, BWC230081H, BWC230305H, BWC230813H, BWC230994H, BWC231169H, BWC231206H, BWC231319H, BWC231926H, BWC232121H, BWC223495H, BWC227691H, BWC227895H, BWC228940H, BWC229922H, BWC230257H, BWC230292H, BWC230714H, BWC230721H, BWC230750H, BWC231453H, BWC231963H, BWC232043H, BWC232044H, BWC232104H, BWC232117H, BWC218881H, BWC220978H, BWC223884H, BWC224078H, BWC226920H, BWC227963H, BWC228583H, BWC229420H, BWC229889H, BWC230102H, BWC230152H, BWC230174H, BWC230310H, BWC230452H, BWC230799H, BWC230940H, BWC231744H, BWC231898H, BWC231946H, BWC232028H, BWC219728H, BWC220059H, BWC220642H, BWC221211H, BWC226459H, BWC227411H, BWC227493H, BWC228063H, BWC228315H, BWC228890H, BWC229076H, BWC229169H, BWC229176H, BWC229177H, BWC229201H, BWC229355H, BWC229362H, BWC229528H, BWC229932H, BWC229958H, BWC230155H, BWC230456H, BWC230486H, BWC230560H, BWC230567H, BWC230716H, BWC230726H, BWC230831H, BWC231087H, BWC231203H, BWC231207H, BWC231295H, BWC231297H, BWC231300H, BWC231338H, BWC231359H, BWC231397H, BWC231410H, BWC231418H, BWC231423H, BWC231952H, BWC231993H, BWC231996H, BWC232040H, BWC220484H, BWC221155H, BWC223224H, BWC225161H, BWC225554H, BWC228078H, BWC229464H, BWC230103H, BWC230285H, BWC230563H, BWC231092H, BWC231421H, BWC231714H, BWC231838H, BWC231882H, BWC231981H, BWC232176H, BWC232185H, BWC232186H, BWC224487H, BWC225870H, BWC228348H, BWC228371H, BWC229373H, BWC229809H, BWC229859H, BWC229908H, BWC229929H, BWC229978H, BWC230194H, BWC230286H, BWC230995H, BWC231041H, BWC231171H, BWC231177H, BWC231907H, BWC231957H, BWC231979H, BWC232049H, BWC232067H, BWC218848H, BWC219795H, BWC221767H, BWC222630H, BWC225806H, BWC226083H, BWC226419H, BWC228404H, BWC229022H, BWC229532H, BWC229883H, BWC230075H, BWC230400H, BWC230438H, BWC230607H, BWC231053H, BWC231322H, BWC231374H, BWC231847H, BWC232008H, BWC232068H, BWC232086H, BWC232091H, BWC226216H, BWC228616H, BWC228776H, BWC229147H, BWC229676H, BWC229729H, BWC230251H, BWC230270H, BWC230394H, BWC230399H, BWC230420H, BWC230536H, BWC230558H, BWC230580H, BWC230615H, BWC231141H, BWC231433H, BWC231901H, BWC231951H, BWC231984H, BWC231995H, BWC232197H, BWC223517H, BWC225265H, BWC225976H, BWC226945H, BWC227834H, BWC229239H, BWC229540H, BWC229706H, BWC230153H, BWC230163H, BWC230201H, BWC230376H, BWC230500H, BWC231026H, BWC231102H, BWC231157H, BWC231182H, BWC231183H, BWC231398H, BWC231765H, BWC231924H, BWC231932H, BWC232107H, BWC220973H, BWC222601H, BWC225504H, BWC226384H, BWC228088H, BWC228261H, BWC228768H, BWC229341H, BWC229795H, BWC229894H, BWC230274H, BWC230277H, BWC230365H, BWC230613H, BWC230616H, BWC230631H, BWC230825H, BWC231059H, BWC231532H, BWC231772H, BWC231851H, BWC231940H, BWC231959H, BWC232033H, BWC232150H, BWC232152H, BWC230272H, BWC218592H, BWC218885H, BWC219739H, BWC221073H, BWC224804H, BWC225394H, BWC227960H, BWC229178H, BWC229232H, BWC229384H, BWC229510H, BWC229574H, BWC229771H, BWC229810H, BWC229926H, BWC229943H, BWC229957H, BWC230086H, BWC230198H, BWC230283H, BWC230333H, BWC230364H, BWC230419H, BWC230556H, BWC231227H, BWC231401H, BWC232023H, BWC232064H, BWC220341H, BWC223544H, BWC226300H, BWC227480H, BWC228716H, BWC229746H, BWC229823H, BWC229839H, BWC229924H, BWC230132H, BWC230514H, BWC230598H, BWC230711H, BWC230766H, BWC230777H, BWC231126H, BWC231151H, BWC231344H, BWC231447H, BWC231725H, BWC231955H, BWC231967H, BWC231968H, BWC231986H, BWC232092H, BWC232093H, BWC221513H, BWC221663H, BWC223877H, BWC224088H, BWC226602H, BWC226614H, BWC226821H, BWC228592H, BWC228990H, BWC229249H, BWC229442H, BWC229566H, BWC230156H, BWC230327H, BWC230990H, BWC230993H, BWC231033H, BWC231062H, BWC231144H, BWC231302H, BWC231442H, BWC231732H, BWC231850H, BWC232055H, BWC232057H, BWC232059H, BWC232060H, BWC232061H, BWC232062H, BWC222810H, BWC224682H, BWC224825H, BWC225568H, BWC227689H, BWC227894H, BWC228439H, BWC228623H, BWC229763H, BWC229993H, BWC230249H, BWC231368H, BWC231440H, BWC231843H, BWC225947H, BWC229406H, BWC229789H, BWC229845H, BWC230033H, BWC230035H, BWC230061H, BWC230406H, BWC230494H, BWC230545H, BWC230559H, BWC230576H, BWC230706H, BWC230739H, BWC230927H, BWC231166H, BWC231197H, BWC231284H, BWC231289H, BWC231885H, BWC231891H, BWC231893H, BWC231679H, BWC220334H, BWC221925H, BWC223675H, BWC226843H, BWC229962H, BWC230177H, BWC230224H, BWC230709H, BWC230943H, BWC231118H, BWC231142H, BWC231217H, BWC231304H, BWC231858H, BWC231914H, BWC232195H, BWC222667H, BWC224949H, BWC228587H, BWC228679H, BWC229088H, BWC229412H, BWC230336H, BWC230425H, BWC230542H, BWC231054H, BWC231290H, BWC231713H, BWC231883H, BWC220335H, BWC221928H, BWC224652H, BWC229467H, BWC229852H, BWC229857H, BWC230116H, BWC230386H, BWC230540H, BWC230955H, BWC231216H, BWC231911H, BWC231953H, BWC222126H, BWC225808H, BWC228966H, BWC230264H, BWC230363H, BWC230923H, BWC230932H, BWC231191H, BWC231349H, BWC231454H, BWC231886H, BWC231899H, BWC219618H, BWC220555H, BWC221209H, BWC224043H, BWC225674H, BWC225821H, BWC226326H, BWC226420H, BWC226963H, BWC226975H, BWC227537H, BWC229311H, BWC229808H, BWC229824H, BWC229942H, BWC230011H, BWC230113H, BWC230166H, BWC230304H, BWC230321H, BWC230517H, BWC230617H, BWC230703H, BWC230747H, BWC230826H, BWC230827H, BWC230829H, BWC230834H, BWC231425H, BWC223271H, BWC228333H, BWC229034H, BWC219708H, BWC220343H, BWC225413H, BWC225735H, BWC228070H, BWC228119H, BWC228175H, BWC228549H, BWC229209H, BWC229226H, BWC229316H, BWC229947H, BWC229949H, BWC230218H, BWC230222H, BWC230240H, BWC230280H, BWC230339H, BWC230378H, BWC230451H, BWC230468H, BWC230489H, BWC230491H, BWC230571H, BWC230603H, BWC230715H, BWC230819H, BWC230820H, BWC230821H, BWC230823H, BWC230881H, BWC230894H, BWC230909H, BWC230912H, BWC230913H, BWC230920H, BWC230944H, BWC230945H, BWC230950H, BWC231055H, BWC231085H, BWC231086H, BWC231116H, BWC231165H, BWC231178H, BWC231342H, BWC231345H, BWC231534H, BWC231535H, BWC231536H, BWC231537H, BWC231538H, BWC231539H, BWC231540H, BWC231541H, BWC231542H, BWC231543H, BWC231544H, BWC231545H, BWC231547H, BWC231548H, BWC231551H, BWC231552H, BWC231554H, BWC231557H, BWC231558H, BWC231561H, BWC231562H, BWC231563H, BWC231564H, BWC231565H, BWC231573H, BWC231575H, BWC231577H, BWC231579H, BWC231581H, BWC231583H, BWC231584H, BWC231585H, BWC231587H, BWC231588H, BWC231594H, BWC231595H, BWC231599H, BWC231600H, BWC231601H, BWC231602H, BWC231603H, BWC231606H, BWC231607H, BWC231608H, BWC231612H, BWC231613H, BWC231615H, BWC231617H, BWC231619H, BWC231620H, BWC231622H, BWC231623H, BWC231624H, BWC231625H, BWC231626H, BWC231627H, BWC231629H, BWC231630H, BWC231632H, BWC231639H, BWC231641H, BWC231642H, BWC231644H, BWC231645H, BWC231648H, BWC231649H, BWC231650H, BWC231654H, BWC231662H, BWC231667H, BWC231672H, BWC231675H, BWC231680H, BWC231681H, BWC231689H, BWC231696H, BWC231699H, BWC231701H, BWC231706H, BWC231707H, BWC231708H, BWC231716H, BWC231717H, BWC231719H, BWC231720H, BWC231721H, BWC231723H, BWC231724H, BWC231726H, BWC231742H, BWC231743H, BWC231745H, BWC231746H, BWC231748H, BWC231764H, BWC231769H, BWC231771H, BWC231779H, BWC231782H, BWC231783H, BWC231785H, BWC231787H, BWC231788H, BWC231789H, BWC231790H, BWC231792H, BWC231793H, BWC231794H, BWC231796H, BWC231798H, BWC231799H, BWC231800H, BWC231801H, BWC231802H, BWC231803H, BWC231804H, BWC231805H, BWC231806H, BWC231808H, BWC231810H, BWC231811H, BWC231812H, BWC231813H, BWC231814H, BWC231815H, BWC231817H, BWC231818H, BWC231820H, BWC231822H, BWC231823H, BWC231824H, BWC231825H, BWC231827H, BWC231828H, BWC231870H, BWC219624H, BWC220085H, BWC220346H, BWC220722H, BWC220959H, BWC222034H, BWC222745H, BWC224720H, BWC225362H, BWC225426H, BWC227305H, BWC227669H, BWC228468H, BWC228668H, BWC229298H, BWC229563H, BWC229847H, BWC230023H, BWC230289H, BWC230334H, BWC230466H, BWC230522H, BWC230533H, BWC230717H, BWC230763H, BWC230896H, BWC230921H, BWC230929H, BWC230930H, BWC231122H, BWC231277H, BWC231278H, BWC231279H, BWC231280H, BWC231294H, BWC231301H, BWC231306H, BWC231316H, BWC231317H, BWC231330H, BWC231341H, BWC231367H, BWC231371H, BWC231372H, BWC231377H, BWC231380H, BWC231381H, BWC231383H, BWC231384H, BWC231386H, BWC231393H, BWC231396H, BWC231414H, BWC231415H, BWC231416H, BWC231420H, BWC231424H, BWC231427H, BWC231428H, BWC231429H, BWC231430H, BWC231432H, BWC231437H, BWC231439H, BWC231450H, BWC231460H, BWC231462H, BWC231463H, BWC231467H, BWC231468H, BWC231469H, BWC231470H, BWC231471H, BWC231472H, BWC231473H, BWC231474H, BWC231475H, BWC231476H, BWC231483H, BWC231487H, BWC231489H, BWC231492H, BWC231493H, BWC231497H, BWC231499H, BWC231502H, BWC231503H, BWC231505H, BWC231507H, BWC231508H, BWC231509H, BWC231510H, BWC231511H, BWC231512H, BWC231514H, BWC231515H, BWC231516H, BWC231517H, BWC231518H, BWC231519H, BWC231520H, BWC231521H, BWC231523H, BWC231738H, BWC232148H, BWC232159H, BWC232163H, BWC232173H, BWC232198H, BWC232199H, BWC232200H, BWC232201H, BWC232203H, BWC232205H, BWC220520H, BWC220639H, BWC221534H, BWC224996H, BWC225836H, BWC226462H, BWC226644H, BWC227570H, BWC227739H, BWC227778H, BWC227974H, BWC227994H, BWC228156H, BWC228215H, BWC228590H, BWC228665H, BWC229085H, BWC229547H, BWC229751H, BWC229944H, BWC230038H, BWC230054H, BWC230088H, BWC230248H, BWC230267H, BWC230332H, BWC230488H, BWC230512H, BWC230557H, BWC230926H, BWC230942H, BWC230954H, BWC230977H, BWC230978H, BWC230983H, BWC230988H, BWC230991H, BWC230992H, BWC230996H, BWC230998H, BWC231002H, BWC231010H, BWC231017H, BWC231025H, BWC231027H, BWC231029H, BWC231032H, BWC231034H, BWC231045H, BWC231048H, BWC231069H, BWC231071H, BWC231089H, BWC231113H, BWC231154H, BWC231159H, BWC231186H, BWC231193H, BWC231205H, BWC231220H, BWC231221H, BWC231223H, BWC231224H, BWC231234H, BWC231260H, BWC231268H, BWC231271H, BWC231356H, BWC231582H, BWC231590H, BWC231628H, BWC231634H, BWC231636H, BWC231652H, BWC231657H, BWC231661H, BWC231664H, BWC231666H, BWC231674H, BWC231683H, BWC231691H, BWC231693H, BWC232021H, BWC232051H, BWC232056H, BWC232082H, BWC232108H, BWC232131H, BWC232132H, BWC232134H, BWC232135H, BWC232138H, BWC232139H, BWC232143H, BWC232147H, BWC232154H, BWC232156H, BWC232171H, BWC218886H, BWC219549H, BWC219658H, BWC220727H, BWC221931H, BWC223390H, BWC223542H, BWC223710H, BWC226324H, BWC226742H, BWC226940H, BWC226943H, BWC227891H, BWC229072H, BWC229120H, BWC229137H, BWC229336H, BWC229513H, BWC229734H, BWC229753H, BWC229812H, BWC229851H, BWC229901H, BWC229902H, BWC229945H, BWC230021H, BWC230314H, BWC230414H, BWC230768H, BWC230771H, BWC230781H, BWC230783H, BWC230785H, BWC230788H, BWC230790H, BWC230795H, BWC230796H, BWC230804H, BWC230805H, BWC230808H, BWC230836H, BWC230838H, BWC230839H, BWC230842H, BWC230844H, BWC230846H, BWC230848H, BWC230850H, BWC230851H, BWC230853H, BWC230856H, BWC230902H, BWC231210H, BWC231757H, BWC231773H, BWC231775H, BWC219343H, BWC221137H, BWC226031H, BWC226158H, BWC226315H, BWC227829H, BWC227858H, BWC228005H, BWC228368H, BWC228415H, BWC228517H, BWC228551H, BWC228694H, BWC229015H, BWC229080H, BWC229256H, BWC229281H, BWC229425H, BWC229587H, BWC229737H, BWC230079H, BWC230140H, BWC230319H, BWC230469H, BWC230661H, BWC230680H, BWC230708H, BWC230734H, BWC230735H, BWC230748H, BWC230753H, BWC230754H, BWC230757H, BWC230758H, BWC230759H, BWC231006H, BWC231016H, BWC231070H, BWC231081H, BWC231129H, BWC231167H, BWC231285H, BWC231353H, BWC231366H, BWC231694H, BWC231712H, BWC231727H, BWC231728H, BWC231731H, BWC231736H, BWC231737H, BWC231740H, BWC231741H, BWC231747H, BWC231751H, BWC231753H, BWC219819H, BWC221195H, BWC222852H, BWC223262H, BWC224700H, BWC224884H, BWC226345H, BWC226430H, BWC227975H, BWC227998H, BWC228137H, BWC228530H, BWC228645H, BWC228950H, BWC229514H, BWC229515H, BWC229879H, BWC230005H, BWC230215H, BWC230328H, BWC230360H, BWC230375H, BWC230427H, BWC230445H, BWC230501H, BWC230518H, BWC230564H, BWC230599H, BWC230600H, BWC230601H, BWC230602H, BWC230618H, BWC230623H, BWC230624H, BWC230625H, BWC230627H, BWC230629H, BWC230632H, BWC230633H, BWC230634H, BWC230635H, BWC230637H, BWC230638H, BWC230649H, BWC230650H, BWC230651H, BWC230653H, BWC230660H, BWC230664H, BWC230666H, BWC230667H, BWC230671H, BWC230676H, BWC230679H, BWC230682H, BWC230684H, BWC230845H, BWC230860H, BWC230862H, BWC230864H, BWC230866H, BWC230868H, BWC230871H, BWC230875H, BWC230883H, BWC230886H, BWC231008H, BWC231066H, BWC231153H, BWC231222H, BWC231286H, BWC231334H, BWC231387H, BWC231419H, BWC231610H, BWC231631H, BWC231635H, BWC231637H, BWC231665H, BWC231670H, BWC231676H, BWC231682H, BWC231695H, BWC231698H, BWC231702H, BWC231703H, BWC231711H, BWC231715H, BWC231730H, BWC231735H, BWC231759H, BWC217843H, BWC220810H, BWC225085H, BWC226373H, BWC226573H, BWC227959H, BWC228294H, BWC229023H, BWC229105H, BWC229174H, BWC229303H, BWC229663H, BWC229875H, BWC230178H, BWC230180H, BWC230324H, BWC230432H, BWC230476H, BWC230591H, BWC230592H, BWC230593H, BWC230594H, BWC230595H, BWC230782H, BWC231117H, BWC231163H, BWC231204H, BWC231213H, BWC231258H, BWC217131H, BWC220371H, BWC220696H, BWC221798H, BWC225798H, BWC226417H, BWC226837H, BWC226857H, BWC227694H, BWC227809H, BWC227889H, BWC229109H, BWC229415H, BWC229482H, BWC229731H, BWC229820H, BWC229825H, BWC229976H, BWC230004H, BWC230050H, BWC230059H, BWC230150H, BWC230186H, BWC230241H, BWC230282H, BWC230448H, BWC230484H, BWC230574H, BWC230793H, BWC231014H, BWC231149H, BWC231158H, BWC231162H, BWC231180H, BWC231230H, BWC231291H, BWC231313H, BWC231391H, BWC231399H, BWC231449H, BWC231451H, BWC231546H, BWC221572H, BWC222371H, BWC222433H, BWC222585H, BWC223393H, BWC223438H, BWC224773H, BWC225018H, BWC225552H, BWC226730H, BWC227662H, BWC228064H, BWC228425H, BWC228460H, BWC228861H, BWC228955H, BWC229010H, BWC229111H, BWC229322H, BWC229994H, BWC230077H, BWC230199H, BWC230213H, BWC230377H, BWC230385H, BWC230453H, BWC230532H, BWC230547H, BWC230548H, BWC230550H, BWC230551H, BWC230801H, BWC231097H, BWC231120H, BWC231273H, BWC231325H, BWC231417H, BWC221033H, BWC222004H, BWC223597H, BWC223868H, BWC225766H, BWC225916H, BWC226450H, BWC227631H, BWC228169H, BWC228279H, BWC228459H, BWC229429H, BWC229630H, BWC229645H, BWC229843H, BWC229855H, BWC229938H, BWC229989H, BWC230007H, BWC230242H, BWC230253H, BWC230265H, BWC230301H, BWC230368H, BWC230373H, BWC230519H, BWC230528H, BWC230529H, BWC230835H, BWC230867H, BWC231080H, BWC231109H, BWC231155H, BWC231389H, BWC228397H, BWC223261H, BWC225405H, BWC227465H, BWC228143H, BWC228393H, BWC229258H, BWC229476H, BWC229761H, BWC229919H, BWC229991H, BWC230032H, BWC230120H, BWC230230H, BWC230260H, BWC230261H, BWC230307H, BWC230490H, BWC230510H, BWC230961H, BWC231145H, BWC231270H, BWC231298H, BWC231333H, BWC231352H, BWC231362H, BWC232182H, BWC232187H,
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Reason for Recall:
A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Implantable cardioverter defibrillators (ICD): 1) Evera family: Evera XT DR, Evera S DR, Evera MRI XT DR, Evera MRI S DR, Evera XT VR, Evera S VR and Evera MRI S VR models; 2) Visia family: Visia AF ICD, Visia AF S ICD, Visia AF MRI ICD, and Visia AF MRI S ICD models. Implantable Cardioverter Defibrillators (ICDs) are implantable medical devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia.

Product Codes/Lot Numbers:

1) Evera Family: MODEL NUMBER / NAME / GTIN: SERIAL NUMBERS: DDBB1D1 / ICD-DR DDBB1D1 EVERA XT US IS1/DF1 / 00643169018099: BWC231766H, BWC230184H, BWC230891H, BWC228489H, BWC232037H, BWC230770H, BWC231219H, BWC231215H, BWC231190H, BWC231202H, BWC231199H, BWC231780H, BWC231556H, BWC230255H, BWC230772H, BWC230906H, BWC231500H, BWC231572H, BWC231618H, BWC231655H, BWC231937H, BWC231942H, BWC230859H, BWC230949H, BWC231848H, BWC230918H, BWC231237H, BWC231580H, BWC230893H, BWC231576H, BWC230426H, BWC230475H, BWC230431H, BWC231494H, BWC230876H, BWC230776H, BWC228873H, BWC229321H, BWC230873H, BWC229698H, BWC230449H, BWC231452H, BWC230877H, BWC230720H, BWC231479H, BWC231482H, BWC231578H, BWC231498H, BWC229796H, BWC230863H, BWC230874H, BWC231239H, BWC230917H, BWC229337H, BWC230524H, BWC230887H, BWC231525H, BWC231240H, BWC230672H, BWC230769H, BWC232164H, BWC232196H, BWC230048H, BWC230520H, BWC231568H, BWC231671H, BWC231761H, BWC229598H, BWC230335H, BWC230889H, BWC231570H, BWC228859H, BWC230914H, BWC231346H, BWC230884H, BWC231501H, BWC229058H, BWC229697H, BWC228813H, BWC232165H, BWC230870H, BWC230901H, BWC231259H, BWC230872H, BWC231106H, BWC230195H, BWC230911H, BWC228448H, BWC229342H, BWC230535H, BWC230655H, BWC230919H, BWC231269H, BWC231480H, BWC231668H, BWC230044H, BWC228426H, BWC230910H, BWC227091H, BWC231456H, BWC229099H, BWC229101H, BWC231874H, BWC228336H, BWC230869H, BWC230730H, BWC230630H, BWC231872H, BWC231343H, BWC228800H, BWC228656H, BWC228666H, BWC229097H, BWC231944H, BWC226887H, BWC230885H, BWC231768H, BWC232133H, BWC227119H, BWC228318H, BWC229963H, BWC230899H, BWC228760H, BWC230159H, BWC230882H, BWC230381H, BWC228544H, BWC230098H, BWC231400H, BWC228205H, BWC228858H, BWC226830H, BWC230454H, BWC228816H, BWC230608H, BWC228988H, BWC227088H, BWC228586H, BWC226816H, BWC227609H, BWC228180H, BWC228449H, BWC229289H, BWC229600H, BWC230118H, BWC230732H, BWC232122H, BWC231241H, BWC228818H, BWC229317H, BWC229344H, BWC229118H, BWC230071H, BWC227808H, BWC232126H, BWC232191H, BWC232170H, BWC227125H, BWC227610H, BWC227114H, BWC230530H, BWC227067H, BWC227782H, BWC228871H, BWC228399H, BWC228427H, BWC231659H, BWC228044H, BWC228820H, BWC228206H, BWC229013H, BWC231037H, BWC231809H, BWC226906H, BWC229163H, BWC227324H, BWC228320H, BWC227017H, BWC227077H, BWC227034H, BWC227374H, BWC228469H, BWC231485H, BWC225637H, BWC227113H, BWC228867H, BWC231257H, BWC227047H, BWC227071H, BWC224560H, BWC226819H, BWC226824H, BWC226960H, BWC227129H, BWC228864H, BWC230020H, BWC231332H, BWC231458H, BWC226403H, BWC226833H, BWC227056H, BWC227062H, BWC227126H, BWC228869H, BWC226939H, BWC227326H, BWC230083H, BWC231865H, BWC229346H, BWC231669H, BWC227323H, BWC230503H, BWC231049H, BWC228832H, BWC229536H, BWC226805H, BWC226822H, BWC226914H, BWC231406H, BWC228248H, BWC225448H, BWC226799H, BWC227109H, BWC228132H, BWC229168H, BWC231459H, BWC231673H, BWC226801H, BWC226817H, BWC228812H, BWC231043H, BWC231192H, BWC229055H, BWC226987H, BWC231173H, BWC229988H, BWC226820H, BWC226831H, BWC230642H, BWC226798H, BWC229516H, BWC229886H, BWC231781H, BWC226832H, BWC225279H, BWC226809H, BWC226812H, BWC227130H, BWC228297H, BWC229806H, BWC226664H, BWC230704H, BWC231684H, BWC226998H, BWC230322H, BWC225786H, BWC226298H, BWC227032H, BWC230010H, BWC231324H, BWC224500H, BWC229757H, BWC230573H, BWC227008H, BWC230091H, BWC231550H, BWC225592H, BWC226988H, BWC227163H, BWC227322H, BWC224540H, BWC230309H, BWC232166H, BWC227025H, BWC230791H, BWC231036H, BWC227845H, BWC230784H, BWC232050H, BWC226416H, BWC226966H, BWC228943H, BWC228960H, BWC228978H, BWC231908H, BWC228014H, BWC228050H, BWC229537H, BWC226984H, BWC227330H, BWC231176H, BWC231412H, BWC229009H, BWC230481H, BWC226584H, BWC229898H, BWC225730H, BWC226415H, BWC226924H, BWC227316H, BWC227321H, BWC230411H, BWC226578H, BWC229981H, BWC230279H, BWC224159H, BWC225056H, BWC230137H, BWC229606H, BWC231774H, BWC225810H, BWC226471H, BWC227919H, BWC228962H, BWC229743H, BWC230565H, BWC228743H, BWC231660H, BWC225960H, BWC228522H, BWC232106H, BWC232113H, BWC225555H, BWC227536H, BWC228919H, BWC228974H, BWC208701H, BWC231047H, BWC226182H, BWC231976H, BWC226398H, BWC226639H, BWC229914H, BWC230318H, BWC230916H, BWC230937H, BWC231395H, BWC232174H, BWC226053H, BWC228270H, BWC225617H, BWC226923H, BWC228131H, BWC229142H, BWC230648H, BWC231137H, BWC224556H, BWC224377H, BWC225611H, BWC226587H, BWC228342H, BWC229758H, BWC230247H, BWC230404H, BWC224492H, BWC228171H, BWC228363H, BWC228582H, BWC229921H, BWC230590H, BWC230691H, BWC231024H, BWC231250H, BWC231392H, BWC232118H, BWC232181H, BWC222232H, BWC223575H, BWC226139H, BWC226703H, BWC227005H, BWC227196H, BWC228392H, BWC228969H, BWC229071H, BWC229414H, BWC230028H, BWC230181H, BWC231490H, BWC231729H, BWC231864H, BWC231997H, BWC232155H, BWC232206H, BWC225667H, BWC226056H, BWC226643H, BWC231184H, BWC232115H, BWC226299H, BWC227028H, BWC228060H, BWC227023H, BWC228445H, BWC229075H, BWC229787H, BWC232001H, BWC220805H, BWC225084H, BWC227612H, BWC227882H, BWC228414H, BWC230410H, BWC230774H, BWC231403H, BWC231549H, BWC232090H, BWC228497H, BWC229396H, BWC229590H, BWC231105H, BWC224761H, BWC225054H, BWC225527H, BWC226180H, BWC226451H, BWC230337H, BWC230701H, BWC230934H, BWC231189H, BWC227871H, BWC228545H, BWC230505H, BWC231039H, BWC231705H, BWC231845H, BWC224849H, BWC226278H, BWC229634H, BWC231477H, BWC231528H, BWC232116H, BWC232145H, BWC224591H, BWC226782H, BWC228310H, BWC229104H, BWC230740H, BWC226791H, BWC228968H, BWC229868H, BWC229920H, BWC231175H, BWC231436H, BWC230117H, BWC231198H, BWC231496H, BWC231970H, BWC230302H, BWC231274H, BWC223441H, BWC228258H, BWC231905H, BWC227947H, BWC228058H, BWC228068H, BWC229039H, BWC229098H, BWC229248H, BWC229618H, BWC232032H, BWC225186H, BWC225677H, BWC226274H, BWC228083H, BWC229573H, BWC232046H, BWC226964H, BWC230055H, BWC231147H, BWC231478H, BWC224493H, BWC227552H, BWC228446H, BWC230284H, BWC231021H, BWC231465H, BWC231916H, BWC228096H, BWC229960H, BWC231214H, BWC224572H, BWC224596H, BWC224602H, BWC224615H, BWC226736H, BWC228147H, BWC228277H, BWC229145H, BWC230330H, BWC231211H, BWC231364H, BWC231923H, BWC224335H, BWC224579H, BWC228824H, BWC230447H, BWC230745H, BWC231114H, BWC231296H, BWC231830H, BWC223169H, BWC225062H, BWC225458H, BWC226225H, BWC230960H, BWC231138H, BWC223551H, BWC223978H, BWC229236H, BWC229699H, BWC230644H, BWC231495H, BWC231888H, BWC231894H, BWC231947H, BWC225001H, BWC225950H, BWC229477H, BWC230561H, BWC230953H, BWC231404H, BWC224769H, BWC226650H, BWC228191H, BWC228926H, BWC229822H, BWC230690H, BWC225929H, BWC229204H, BWC229888H, BWC230094H, BWC230499H, BWC230640H, BWC230789H, BWC225802H, BWC229925H, BWC221930H, BWC224327H, BWC224563H, BWC224609H, BWC230018H, BWC230164H, BWC230692H, BWC231281H, BWC231597H, BWC231755H, BWC224716H, BWC228165H, BWC228670H, BWC229350H, BWC229912H, BWC230068H, BWC230477H, BWC230987H, BWC231328H, BWC232192H, BWC224809H, BWC227258H, BWC228264H, BWC228387H, BWC230143H, BWC230216H, BWC230654H, BWC230697H, BWC224568H, BWC226949H, BWC228223H, BWC230115H, BWC230136H, BWC230656H, BWC231209H, BWC231464H, BWC225157H, BWC228394H, BWC228673H, BWC229585H, BWC230206H, BWC230317H, BWC231881H, BWC224768H, BWC224838H, BWC227072H, BWC229990H, BWC230467H, BWC230502H, BWC231443H, BWC232024H, BWC232167H, BWC221901H, BWC224605H, BWC224683H, BWC230658H, BWC231933H, BWC232120H, BWC228161H, BWC229230H, BWC229576H, BWC230380H, BWC230722H, BWC230744H, BWC223304H, BWC224573H, BWC224976H, BWC225998H, BWC228654H, BWC229490H, BWC229933H, BWC231431H, BWC231596H, BWC231734H, BWC228554H, BWC228588H, BWC230668H, BWC230792H, BWC231444H, BWC229546H, BWC222703H, BWC224052H, BWC224571H, BWC225153H, BWC227927H, BWC228309H, BWC229775H, BWC230078H, BWC230315H, BWC230729H, BWC231276H, BWC224517H, BWC224960H, BWC228214H, BWC228579H, BWC229117H, BWC229691H, BWC232020H, BWC226111H, BWC230089H, BWC227370H, BWC230053H, BWC230193H, BWC230443H, BWC230700H, BWC231777H, BWC224597H, BWC227185H, BWC228003H, BWC228196H, BWC229377H, BWC229485H, BWC229519H, BWC230316H, BWC230816H, BWC230832H, BWC231409H, BWC231413H, BWC224581H, BWC221774H, BWC223806H, BWC224504H, BWC225399H, BWC225791H, BWC229059H, BWC230534H, BWC230749H, BWC231125H, BWC231909H, BWC231934H, BWC224490H, BWC227688H, BWC232019H, BWC223259H, BWC224559H, BWC225344H, BWC226698H, BWC230173H, BWC230295H, BWC230728H, BWC231212H, BWC232128H, BWC223821H, BWC223973H, BWC227073H, BWC232073H, BWC232175H, BWC232179H, BWC218949H, BWC223948H, BWC224594H, BWC224786H, BWC224811H, BWC226026H, BWC227939H, BWC228680H, BWC228970H, BWC230202H, BWC230986H, BWC231307H, BWC231337H, BWC231754H, BWC231910H, BWC232169H, BWC223161H, BWC225308H, BWC227572H, BWC227913H, BWC230022H, BWC230417H, BWC222829H, BWC223122H, BWC223979H, BWC224616H, BWC230200H, BWC230259H, BWC232102H, BWC224607H, BWC224963H, BWC230604H, BWC230696H, BWC231311H, BWC231312H, BWC231960H, BWC222823H, BWC224576H, BWC224800H, BWC227530H, BWC228763H, BWC229755H, BWC230497H, BWC230800H, BWC231347H, BWC231434H, BWC232054H, BWC232190H, BWC223372H, BWC225684H, BWC230019H, BWC230554H, BWC230736H, BWC231255H, BWC231435H, BWC222777H, BWC222865H, BWC230015H, BWC230190H, BWC231136H, BWC221087H, BWC222263H, BWC223494H, BWC225738H, BWC225913H, BWC230008H, BWC230170H, BWC230513H, BWC230546H, BWC231264H, BWC232017H, BWC232018H, BWC221622H, BWC222804H, BWC224588H, BWC227295H, BWC228130H, BWC228384H, BWC230237H, BWC231249H, BWC232003H, BWC221917H, BWC228744H, BWC229432H, BWC229735H, BWC231355H, BWC225110H, BWC225135H, BWC227986H, BWC227991H, BWC229447H, BWC230042H, BWC230308H, BWC231749H, BWC221936H, BWC222564H, BWC225761H, BWC227044H, BWC229777H, BWC230354H, BWC230372H, BWC230485H, BWC231028H, BWC232168H, BWC222392H, BWC223370H, BWC226823H, BWC228240H, BWC229186H, BWC229966H, BWC232161H, BWC221611H, BWC221761H, BWC223079H, BWC225401H, BWC225906H, BWC227049H, BWC229007H, BWC221773H, BWC228305H, BWC229756H, BWC229838H, BWC230296H, BWC232004H, BWC232045H, BWC232083H, BWC226747H, BWC230359H, BWC231124H, BWC231375H, BWC232007H, BWC232184H, BWC230326H, BWC231949H, BWC219373H, BWC226828H, BWC227874H, BWC228136H, BWC228471H, BWC228927H, BWC229265H, BWC229826H, BWC230299H, BWC230441H, BWC230570H, BWC230585H, BWC230767H, BWC231305H, BWC231407H, BWC231763H, BWC231861H, BWC231897H, BWC232103H, BWC219794H, BWC222825H, BWC225926H, BWC228275H, BWC228353H, BWC228939H, BWC230157H, BWC230221H, BWC230331H, BWC230444H, BWC230900H, BWC230938H, BWC231156H, BWC231360H, BWC231687H, BWC231879H, BWC232178H, BWC221006H, BWC222840H, BWC223041H, BWC229192H, BWC229893H, BWC230516H, BWC230605H, BWC230657H, BWC230797H, BWC231031H, BWC231873H, BWC220956H, BWC222763H, BWC223302H, BWC223598H, BWC227170H, BWC227678H, BWC227804H, BWC229004H, BWC229752H, BWC229907H, BWC230139H, BWC230760H, BWC231100H, BWC231530H, BWC231857H, BWC232076H, BWC232124H, BWC221539H, BWC225159H, BWC225673H, BWC229213H, BWC229357H, BWC230538H, BWC230614H, BWC230686H, BWC231351H, BWC231504H, BWC231832H, BWC231849H, BWC221448H, BWC221549H, BWC221864H, BWC223893H, BWC228866H, BWC229056H, BWC229115H, BWC229273H, BWC229849H, BWC230483H, BWC230922H, BWC231965H, BWC220490H, BWC220673H, BWC221847H, BWC222048H, BWC223384H, BWC224835H, BWC225321H, BWC226930H, BWC227485H, BWC227984H, BWC228082H, BWC228120H, BWC230026H, BWC230040H, BWC230063H, BWC230504H, BWC230543H, BWC231422H, BWC231971H, BWC221002H, BWC229985H, BWC230057H, BWC230583H, BWC230669H, BWC230727H, BWC230936H, BWC231023H, BWC232052H, BWC232114H, BWC220153H, BWC221545H, BWC224371H, BWC225024H, BWC225848H, BWC226019H, BWC228644H, BWC228712H, BWC229149H, BWC229369H, BWC230168H, BWC230537H, BWC230626H, BWC230670H, BWC231925H, BWC231991H, BWC232000H, BWC232029H, BWC219394H, BWC225316H, BWC226121H, BWC226583H, BWC226933H, BWC227993H, BWC228220H, BWC228556H, BWC229629H, BWC230041H, BWC230356H, BWC230521H, BWC230689H, BWC230779H, BWC231058H, BWC231172H, BWC231718H, BWC231922H, BWC232153H, BWC231919H, BWC232142H, BWC223121H, BWC223378H, BWC227733H, BWC227973H, BWC229381H, BWC229649H, BWC229662H, BWC229770H, BWC231235H, BWC231326H, BWC221102H, BWC221531H, BWC222136H, BWC222780H, BWC222984H, BWC224911H, BWC225451H, BWC225758H, BWC228837H, BWC229614H, BWC230188H, BWC231833H, BWC220869H, BWC221892H, BWC221908H, BWC224894H, BWC228383H, BWC228389H, BWC229776H, BWC231834H, BWC231904H, BWC231994H, BWC222819H, BWC227302H, BWC229157H, BWC230100H, BWC230397H, BWC230473H, BWC230674H, BWC231090H, BWC231928H, BWC232022H, BWC224225H, BWC225465H, BWC227237H, BWC227501H, BWC228440H, BWC229132H, BWC230217H, BWC230258H, BWC230509H, BWC230579H, BWC230673H, BWC230981H, BWC231038H, BWC231318H, BWC231336H, BWC231835H, BWC230176H, BWC231526H, BWC221051H, BWC221378H, BWC226994H, BWC231309H, BWC231448H, BWC231954H, BWC231978H, BWC221606H, BWC224824H, BWC224877H, BWC228089H, BWC229006H, BWC229360H, BWC229679H, BWC229953H, BWC230185H, BWC230362H, BWC230435H, BWC230612H, BWC230803H, BWC231226H, BWC231252H, BWC231348H, BWC231531H, BWC231837H, BWC231856H, BWC231958H, BWC232141H, BWC221775H, BWC222003H, BWC225649H, BWC226762H, BWC227296H, BWC227494H, BWC227562H, BWC228355H, BWC228601H, BWC229217H, BWC229259H, BWC229748H, BWC229918H, BWC230388H, BWC230693H, BWC230935H, BWC231299H, BWC231906H, BWC231980H, BWC220857H, BWC224813H, BWC225474H, BWC225859H, BWC227608H, BWC229162H, BWC230904H, BWC231119H, BWC231253H, BWC231315H, BWC231340H, BWC231614H, BWC231871H, BWC224821H, BWC226066H, BWC226466H, BWC229715H, BWC229767H, BWC230263H, BWC230578H, BWC231283H, BWC231593H, BWC231733H, BWC232038H, BWC232146H, BWC221796H, BWC223280H, BWC224883H, BWC224933H, BWC225182H, BWC228009H, BWC228848H, BWC230371H, BWC230526H, BWC230621H, BWC231050H, BWC231938H, BWC232084H, BWC225403H, BWC225731H, BWC227652H, BWC228273H, BWC229426H, BWC229722H, BWC230273H, BWC230383H, BWC230398H, BWC230507H, BWC230705H, BWC231643H, BWC231700H, BWC220741H, BWC222941H, BWC223104H, BWC226052H, BWC226205H, BWC228362H, BWC229388H, BWC229658H, BWC229782H, BWC230229H, BWC230366H, BWC230568H, BWC230582H, BWC231685H, BWC231831H, BWC231889H, BWC232013H, BWC232111H, BWC220461H, BWC221129H, BWC229465H, BWC229798H, BWC230562H, BWC231917H, BWC232078H, BWC232095H, BWC220648H, BWC229065H, BWC231072H, BWC231930H, BWC221996H, BWC224705H, BWC227692H, BWC227779H, BWC228185H, BWC229450H, BWC229544H, BWC230205H, BWC230392H, BWC230482H, BWC230665H, BWC231350H, BWC231373H, BWC231752H, BWC231877H, BWC222061H, BWC222295H, BWC223242H, BWC227229H, BWC227526H, BWC228661H, BWC228996H, BWC229821H, BWC230082H, BWC230541H, BWC231132H, BWC231134H, BWC231445H, BWC231529H, BWC232026H, BWC232066H, BWC218954H, BWC219280H, BWC220239H, BWC221167H, BWC221340H, BWC226421H, BWC227538H, BWC227885H, BWC229123H, BWC229478H, BWC229564H, BWC230569H, BWC230928H, BWC230970H, BWC231019H, BWC231195H, BWC231638H, BWC231950H, BWC231973H, BWC232015H, BWC232071H, BWC232180H, BWC219852H, BWC223071H, BWC225553H, BWC225813H, BWC227413H, BWC227869H, BWC228018H, BWC228286H, BWC229143H, BWC229364H, BWC229652H, BWC229779H, BWC230268H, BWC230393H, BWC231146H, BWC232129H, BWC221314H, BWC227206H, BWC227715H, BWC231402H, BWC231985H, BWC231990H, BWC232127H, BWC231065H, BWC221584H, BWC223982H, BWC225388H, BWC228022H, BWC228457H, BWC230087H, BWC230236H, BWC230515H, BWC230681H, BWC231875H, BWC231972H, BWC232025H, BWC232172H, BWC218966H, BWC221149H, BWC224829H, BWC227849H, BWC229254H, BWC229854H, BWC230539H, BWC230663H, BWC230915H, BWC231527H, BWC231900H, BWC231975H, BWC232099H, BWC232160H, BWC221700H, BWC222838H, BWC227111H, BWC227412H, BWC227468H, BWC229067H, BWC229359H, BWC229718H, BWC229841H, BWC229910H, BWC230211H, BWC230931H, BWC231758H, BWC231920H, BWC232005H, BWC218454H, BWC220603H, BWC221163H, BWC221550H, BWC221859H, BWC227622H, BWC228677H, BWC228736H, BWC228775H, BWC229291H, BWC229764H, BWC229877H, BWC230047H, BWC230076H, BWC230099H, BWC230470H, BWC230982H, BWC231063H, BWC231112H, BWC231365H, BWC231677H, BWC231852H, BWC231903H, BWC223009H, BWC223382H, BWC223883H, BWC224507H, BWC225310H, BWC226179H, BWC228027H, BWC230203H, BWC230710H, BWC230810H, BWC230933H, BWC231208H, BWC231331H, BWC231913H, BWC231974H, BWC231999H, BWC232006H, BWC232014H, BWC232109H, BWC232140H, BWC225721H, BWC226071H, BWC226166H, BWC226990H, BWC228418H, BWC230017H, BWC230067H, BWC230080H, BWC230269H, BWC230572H, BWC231201H, BWC231243H, BWC231385H, BWC231586H, BWC231982H, BWC232053H, BWC232094H, BWC232162H, BWC220270H, BWC222148H, BWC228793H, BWC230158H, BWC230238H, BWC230464H, BWC230471H, BWC231088H, BWC231196H, BWC231339H, BWC231378H, BWC231854H, BWC232058H, BWC224405H, BWC228221H, BWC228479H, BWC228753H, BWC228872H, BWC229110H, BWC229155H, BWC229609H, BWC229784H, BWC229874H, BWC230106H, BWC230396H, BWC231335H, BWC231369H, BWC231379H, BWC231408H, BWC231935H, BWC231966H, BWC232119H, BWC232193H, BWC220069H, BWC221992H, BWC222868H, BWC228746H, BWC229170H, BWC229508H, BWC230303H, BWC230773H, BWC230841H, BWC230890H, BWC231354H, BWC232034H, BWC232183H, BWC232188H, BWC232207H, BWC226935H, BWC227182H, BWC228339H, BWC229633H, BWC229742H, BWC230210H, BWC230566H, BWC230698H, BWC231004H, BWC231181H, BWC231929H, BWC231987H, BWC231989H, BWC232039H, BWC232075H, BWC220362H, BWC221937H, BWC223477H, BWC225034H, BWC226032H, BWC227372H, BWC228197H, BWC228405H, BWC229713H, BWC229819H, BWC229961H, BWC230145H, BWC230430H, BWC230523H, BWC231001H, BWC231187H, BWC231405H, BWC231840H, BWC231853H, BWC231863H, BWC231945H, BWC232042H, BWC232085H, BWC221527H, BWC222005H, BWC222783H, BWC223582H, BWC224442H, BWC226390H, BWC226787H, BWC227424H, BWC227690H, BWC229215H, BWC229754H, BWC229977H, BWC230160H, BWC230433H, BWC230798H, BWC230843H, BWC231030H, BWC231358H, BWC231566H, BWC231867H, BWC231915H, BWC232030H, BWC223995H, BWC224969H, BWC225740H, BWC225971H, BWC227381H, BWC227583H, BWC228539H, BWC228598H, BWC229135H, BWC229524H, BWC229526H, BWC229545H, BWC229554H, BWC230288H, BWC230351H, BWC230487H, BWC230511H, BWC230688H, BWC230924H, BWC231127H, BWC231357H, BWC231739H, BWC231884H, BWC231912H, BWC231956H, BWC231962H, BWC232088H, BWC220357H, BWC220604H, BWC220905H, BWC221564H, BWC221623H, BWC224834H, BWC225626H, BWC226941H, BWC227724H, BWC229383H, BWC229492H, BWC229646H, BWC229739H, BWC230544H, BWC230743H, BWC231282H, BWC231776H, BWC232065H, BWC232098H, BWC232194H, BWC219691H, BWC225536H, BWC226113H, BWC227139H, BWC227618H, BWC227780H, BWC228193H, BWC228526H, BWC229410H, BWC229638H, BWC229869H, BWC230081H, BWC230305H, BWC230813H, BWC230994H, BWC231169H, BWC231206H, BWC231319H, BWC231926H, BWC232121H, BWC223495H, BWC227691H, BWC227895H, BWC228940H, BWC229922H, BWC230257H, BWC230292H, BWC230714H, BWC230721H, BWC230750H, BWC231453H, BWC231963H, BWC232043H, BWC232044H, BWC232104H, BWC232117H, BWC218881H, BWC220978H, BWC223884H, BWC224078H, BWC226920H, BWC227963H, BWC228583H, BWC229420H, BWC229889H, BWC230102H, BWC230152H, BWC230174H, BWC230310H, BWC230452H, BWC230799H, BWC230940H, BWC231744H, BWC231898H, BWC231946H, BWC232028H, BWC219728H, BWC220059H, BWC220642H, BWC221211H, BWC226459H, BWC227411H, BWC227493H, BWC228063H, BWC228315H, BWC228890H, BWC229076H, BWC229169H, BWC229176H, BWC229177H, BWC229201H, BWC229355H, BWC229362H, BWC229528H, BWC229932H, BWC229958H, BWC230155H, BWC230456H, BWC230486H, BWC230560H, BWC230567H, BWC230716H, BWC230726H, BWC230831H, BWC231087H, BWC231203H, BWC231207H, BWC231295H, BWC231297H, BWC231300H, BWC231338H, BWC231359H, BWC231397H, BWC231410H, BWC231418H, BWC231423H, BWC231952H, BWC231993H, BWC231996H, BWC232040H, BWC220484H, BWC221155H, BWC223224H, BWC225161H, BWC225554H, BWC228078H, BWC229464H, BWC230103H, BWC230285H, BWC230563H, BWC231092H, BWC231421H, BWC231714H, BWC231838H, BWC231882H, BWC231981H, BWC232176H, BWC232185H, BWC232186H, BWC224487H, BWC225870H, BWC228348H, BWC228371H, BWC229373H, BWC229809H, BWC229859H, BWC229908H, BWC229929H, BWC229978H, BWC230194H, BWC230286H, BWC230995H, BWC231041H, BWC231171H, BWC231177H, BWC231907H, BWC231957H, BWC231979H, BWC232049H, BWC232067H, BWC218848H, BWC219795H, BWC221767H, BWC222630H, BWC225806H, BWC226083H, BWC226419H, BWC228404H, BWC229022H, BWC229532H, BWC229883H, BWC230075H, BWC230400H, BWC230438H, BWC230607H, BWC231053H, BWC231322H, BWC231374H, BWC231847H, BWC232008H, BWC232068H, BWC232086H, BWC232091H, BWC226216H, BWC228616H, BWC228776H, BWC229147H, BWC229676H, BWC229729H, BWC230251H, BWC230270H, BWC230394H, BWC230399H, BWC230420H, BWC230536H, BWC230558H, BWC230580H, BWC230615H, BWC231141H, BWC231433H, BWC231901H, BWC231951H, BWC231984H, BWC231995H, BWC232197H, BWC223517H, BWC225265H, BWC225976H, BWC226945H, BWC227834H, BWC229239H, BWC229540H, BWC229706H, BWC230153H, BWC230163H, BWC230201H, BWC230376H, BWC230500H, BWC231026H, BWC231102H, BWC231157H, BWC231182H, BWC231183H, BWC231398H, BWC231765H, BWC231924H, BWC231932H, BWC232107H, BWC220973H, BWC222601H, BWC225504H, BWC226384H, BWC228088H, BWC228261H, BWC228768H, BWC229341H, BWC229795H, BWC229894H, BWC230274H, BWC230277H, BWC230365H, BWC230613H, BWC230616H, BWC230631H, BWC230825H, BWC231059H, BWC231532H, BWC231772H, BWC231851H, BWC231940H, BWC231959H, BWC232033H, BWC232150H, BWC232152H, BWC230272H, BWC218592H, BWC218885H, BWC219739H, BWC221073H, BWC224804H, BWC225394H, BWC227960H, BWC229178H, BWC229232H, BWC229384H, BWC229510H, BWC229574H, BWC229771H, BWC229810H, BWC229926H, BWC229943H, BWC229957H, BWC230086H, BWC230198H, BWC230283H, BWC230333H, BWC230364H, BWC230419H, BWC230556H, BWC231227H, BWC231401H, BWC232023H, BWC232064H, BWC220341H, BWC223544H, BWC226300H, BWC227480H, BWC228716H, BWC229746H, BWC229823H, BWC229839H, BWC229924H, BWC230132H, BWC230514H, BWC230598H, BWC230711H, BWC230766H, BWC230777H, BWC231126H, BWC231151H, BWC231344H, BWC231447H, BWC231725H, BWC231955H, BWC231967H, BWC231968H, BWC231986H, BWC232092H, BWC232093H, BWC221513H, BWC221663H, BWC223877H, BWC224088H, BWC226602H, BWC226614H, BWC226821H, BWC228592H, BWC228990H, BWC229249H, BWC229442H, BWC229566H, BWC230156H, BWC230327H, BWC230990H, BWC230993H, BWC231033H, BWC231062H, BWC231144H, BWC231302H, BWC231442H, BWC231732H, BWC231850H, BWC232055H, BWC232057H, BWC232059H, BWC232060H, BWC232061H, BWC232062H, BWC222810H, BWC224682H, BWC224825H, BWC225568H, BWC227689H, BWC227894H, BWC228439H, BWC228623H, BWC229763H, BWC229993H, BWC230249H, BWC231368H, BWC231440H, BWC231843H, BWC225947H, BWC229406H, BWC229789H, BWC229845H, BWC230033H, BWC230035H, BWC230061H, BWC230406H, BWC230494H, BWC230545H, BWC230559H, BWC230576H, BWC230706H, BWC230739H, BWC230927H, BWC231166H, BWC231197H, BWC231284H, BWC231289H, BWC231885H, BWC231891H, BWC231893H, BWC231679H, BWC220334H, BWC221925H, BWC223675H, BWC226843H, BWC229962H, BWC230177H, BWC230224H, BWC230709H, BWC230943H, BWC231118H, BWC231142H, BWC231217H, BWC231304H, BWC231858H, BWC231914H, BWC232195H, BWC222667H, BWC224949H, BWC228587H, BWC228679H, BWC229088H, BWC229412H, BWC230336H, BWC230425H, BWC230542H, BWC231054H, BWC231290H, BWC231713H, BWC231883H, BWC220335H, BWC221928H, BWC224652H, BWC229467H, BWC229852H, BWC229857H, BWC230116H, BWC230386H, BWC230540H, BWC230955H, BWC231216H, BWC231911H, BWC231953H, BWC222126H, BWC225808H, BWC228966H, BWC230264H, BWC230363H, BWC230923H, BWC230932H, BWC231191H, BWC231349H, BWC231454H, BWC231886H, BWC231899H, BWC219618H, BWC220555H, BWC221209H, BWC224043H, BWC225674H, BWC225821H, BWC226326H, BWC226420H, BWC226963H, BWC226975H, BWC227537H, BWC229311H, BWC229808H, BWC229824H, BWC229942H, BWC230011H, BWC230113H, BWC230166H, BWC230304H, BWC230321H, BWC230517H, BWC230617H, BWC230703H, BWC230747H, BWC230826H, BWC230827H, BWC230829H, BWC230834H, BWC231425H, BWC223271H, BWC228333H, BWC229034H, BWC219708H, BWC220343H, BWC225413H, BWC225735H, BWC228070H, BWC228119H, BWC228175H, BWC228549H, BWC229209H, BWC229226H, BWC229316H, BWC229947H, BWC229949H, BWC230218H, BWC230222H, BWC230240H, BWC230280H, BWC230339H, BWC230378H, BWC230451H, BWC230468H, BWC230489H, BWC230491H, BWC230571H, BWC230603H, BWC230715H, BWC230819H, BWC230820H, BWC230821H, BWC230823H, BWC230881H, BWC230894H, BWC230909H, BWC230912H, BWC230913H, BWC230920H, BWC230944H, BWC230945H, BWC230950H, BWC231055H, BWC231085H, BWC231086H, BWC231116H, BWC231165H, BWC231178H, BWC231342H, BWC231345H, BWC231534H, BWC231535H, BWC231536H, BWC231537H, BWC231538H, BWC231539H, BWC231540H, BWC231541H, BWC231542H, BWC231543H, BWC231544H, BWC231545H, BWC231547H, BWC231548H, BWC231551H, BWC231552H, BWC231554H, BWC231557H, BWC231558H, BWC231561H, BWC231562H, BWC231563H, BWC231564H, BWC231565H, BWC231573H, BWC231575H, BWC231577H, BWC231579H, BWC231581H, BWC231583H, BWC231584H, BWC231585H, BWC231587H, BWC231588H, BWC231594H, BWC231595H, BWC231599H, BWC231600H, BWC231601H, BWC231602H, BWC231603H, BWC231606H, BWC231607H, BWC231608H, BWC231612H, BWC231613H, BWC231615H, BWC231617H, BWC231619H, BWC231620H, BWC231622H, BWC231623H, BWC231624H, BWC231625H, BWC231626H, BWC231627H, BWC231629H, BWC231630H, BWC231632H, BWC231639H, BWC231641H, BWC231642H, BWC231644H, BWC231645H, BWC231648H, BWC231649H, BWC231650H, BWC231654H, BWC231662H, BWC231667H, BWC231672H, BWC231675H, BWC231680H, BWC231681H, BWC231689H, BWC231696H, BWC231699H, BWC231701H, BWC231706H, BWC231707H, BWC231708H, BWC231716H, BWC231717H, BWC231719H, BWC231720H, BWC231721H, BWC231723H, BWC231724H, BWC231726H, BWC231742H, BWC231743H, BWC231745H, BWC231746H, BWC231748H, BWC231764H, BWC231769H, BWC231771H, BWC231779H, BWC231782H, BWC231783H, BWC231785H, BWC231787H, BWC231788H, BWC231789H, BWC231790H, BWC231792H, BWC231793H, BWC231794H, BWC231796H, BWC231798H, BWC231799H, BWC231800H, BWC231801H, BWC231802H, BWC231803H, BWC231804H, BWC231805H, BWC231806H, BWC231808H, BWC231810H, BWC231811H, BWC231812H, BWC231813H, BWC231814H, BWC231815H, BWC231817H, BWC231818H, BWC231820H, BWC231822H, BWC231823H, BWC231824H, BWC231825H, BWC231827H, BWC231828H, BWC231870H, BWC219624H, BWC220085H, BWC220346H, BWC220722H, BWC220959H, BWC222034H, BWC222745H, BWC224720H, BWC225362H, BWC225426H, BWC227305H, BWC227669H, BWC228468H, BWC228668H, BWC229298H, BWC229563H, BWC229847H, BWC230023H, BWC230289H, BWC230334H, BWC230466H, BWC230522H, BWC230533H, BWC230717H, BWC230763H, BWC230896H, BWC230921H, BWC230929H, BWC230930H, BWC231122H, BWC231277H, BWC231278H, BWC231279H, BWC231280H, BWC231294H, BWC231301H, BWC231306H, BWC231316H, BWC231317H, BWC231330H, BWC231341H, BWC231367H, BWC231371H, BWC231372H, BWC231377H, BWC231380H, BWC231381H, BWC231383H, BWC231384H, BWC231386H, BWC231393H, BWC231396H, BWC231414H, BWC231415H, BWC231416H, BWC231420H, BWC231424H, BWC231427H, BWC231428H, BWC231429H, BWC231430H, BWC231432H, BWC231437H, BWC231439H, BWC231450H, BWC231460H, BWC231462H, BWC231463H, BWC231467H, BWC231468H, BWC231469H, BWC231470H, BWC231471H, BWC231472H, BWC231473H, BWC231474H, BWC231475H, BWC231476H, BWC231483H, BWC231487H, BWC231489H, BWC231492H, BWC231493H, BWC231497H, BWC231499H, BWC231502H, BWC231503H, BWC231505H, BWC231507H, BWC231508H, BWC231509H, BWC231510H, BWC231511H, BWC231512H, BWC231514H, BWC231515H, BWC231516H, BWC231517H, BWC231518H, BWC231519H, BWC231520H, BWC231521H, BWC231523H, BWC231738H, BWC232148H, BWC232159H, BWC232163H, BWC232173H, BWC232198H, BWC232199H, BWC232200H, BWC232201H, BWC232203H, BWC232205H, BWC220520H, BWC220639H, BWC221534H, BWC224996H, BWC225836H, BWC226462H, BWC226644H, BWC227570H, BWC227739H, BWC227778H, BWC227974H, BWC227994H, BWC228156H, BWC228215H, BWC228590H, BWC228665H, BWC229085H, BWC229547H, BWC229751H, BWC229944H, BWC230038H, BWC230054H, BWC230088H, BWC230248H, BWC230267H, BWC230332H, BWC230488H, BWC230512H, BWC230557H, BWC230926H, BWC230942H, BWC230954H, BWC230977H, BWC230978H, BWC230983H, BWC230988H, BWC230991H, BWC230992H, BWC230996H, BWC230998H, BWC231002H, BWC231010H, BWC231017H, BWC231025H, BWC231027H, BWC231029H, BWC231032H, BWC231034H, BWC231045H, BWC231048H, BWC231069H, BWC231071H, BWC231089H, BWC231113H, BWC231154H, BWC231159H, BWC231186H, BWC231193H, BWC231205H, BWC231220H, BWC231221H, BWC231223H, BWC231224H, BWC231234H, BWC231260H, BWC231268H, BWC231271H, BWC231356H, BWC231582H, BWC231590H, BWC231628H, BWC231634H, BWC231636H, BWC231652H, BWC231657H, BWC231661H, BWC231664H, BWC231666H, BWC231674H, BWC231683H, BWC231691H, BWC231693H, BWC232021H, BWC232051H, BWC232056H, BWC232082H, BWC232108H, BWC232131H, BWC232132H, BWC232134H, BWC232135H, BWC232138H, BWC232139H, BWC232143H, BWC232147H, BWC232154H, BWC232156H, BWC232171H, BWC218886H, BWC219549H, BWC219658H, BWC220727H, BWC221931H, BWC223390H, BWC223542H, BWC223710H, BWC226324H, BWC226742H, BWC226940H, BWC226943H, BWC227891H, BWC229072H, BWC229120H, BWC229137H, BWC229336H, BWC229513H, BWC229734H, BWC229753H, BWC229812H, BWC229851H, BWC229901H, BWC229902H, BWC229945H, BWC230021H, BWC230314H, BWC230414H, BWC230768H, BWC230771H, BWC230781H, BWC230783H, BWC230785H, BWC230788H, BWC230790H, BWC230795H, BWC230796H, BWC230804H, BWC230805H, BWC230808H, BWC230836H, BWC230838H, BWC230839H, BWC230842H, BWC230844H, BWC230846H, BWC230848H, BWC230850H, BWC230851H, BWC230853H, BWC230856H, BWC230902H, BWC231210H, BWC231757H, BWC231773H, BWC231775H, BWC219343H, BWC221137H, BWC226031H, BWC226158H, BWC226315H, BWC227829H, BWC227858H, BWC228005H, BWC228368H, BWC228415H, BWC228517H, BWC228551H, BWC228694H, BWC229015H, BWC229080H, BWC229256H, BWC229281H, BWC229425H, BWC229587H, BWC229737H, BWC230079H, BWC230140H, BWC230319H, BWC230469H, BWC230661H, BWC230680H, BWC230708H, BWC230734H, BWC230735H, BWC230748H, BWC230753H, BWC230754H, BWC230757H, BWC230758H, BWC230759H, BWC231006H, BWC231016H, BWC231070H, BWC231081H, BWC231129H, BWC231167H, BWC231285H, BWC231353H, BWC231366H, BWC231694H, BWC231712H, BWC231727H, BWC231728H, BWC231731H, BWC231736H, BWC231737H, BWC231740H, BWC231741H, BWC231747H, BWC231751H, BWC231753H, BWC219819H, BWC221195H, BWC222852H, BWC223262H, BWC224700H, BWC224884H, BWC226345H, BWC226430H, BWC227975H, BWC227998H, BWC228137H, BWC228530H, BWC228645H, BWC228950H, BWC229514H, BWC229515H, BWC229879H, BWC230005H, BWC230215H, BWC230328H, BWC230360H, BWC230375H, BWC230427H, BWC230445H, BWC230501H, BWC230518H, BWC230564H, BWC230599H, BWC230600H, BWC230601H, BWC230602H, BWC230618H, BWC230623H, BWC230624H, BWC230625H, BWC230627H, BWC230629H, BWC230632H, BWC230633H, BWC230634H, BWC230635H, BWC230637H, BWC230638H, BWC230649H, BWC230650H, BWC230651H, BWC230653H, BWC230660H, BWC230664H, BWC230666H, BWC230667H, BWC230671H, BWC230676H, BWC230679H, BWC230682H, BWC230684H, BWC230845H, BWC230860H, BWC230862H, BWC230864H, BWC230866H, BWC230868H, BWC230871H, BWC230875H, BWC230883H, BWC230886H, BWC231008H, BWC231066H, BWC231153H, BWC231222H, BWC231286H, BWC231334H, BWC231387H, BWC231419H, BWC231610H, BWC231631H, BWC231635H, BWC231637H, BWC231665H, BWC231670H, BWC231676H, BWC231682H, BWC231695H, BWC231698H, BWC231702H, BWC231703H, BWC231711H, BWC231715H, BWC231730H, BWC231735H, BWC231759H, BWC217843H, BWC220810H, BWC225085H, BWC226373H, BWC226573H, BWC227959H, BWC228294H, BWC229023H, BWC229105H, BWC229174H, BWC229303H, BWC229663H, BWC229875H, BWC230178H, BWC230180H, BWC230324H, BWC230432H, BWC230476H, BWC230591H, BWC230592H, BWC230593H, BWC230594H, BWC230595H, BWC230782H, BWC231117H, BWC231163H, BWC231204H, BWC231213H, BWC231258H, BWC217131H, BWC220371H, BWC220696H, BWC221798H, BWC225798H, BWC226417H, BWC226837H, BWC226857H, BWC227694H, BWC227809H, BWC227889H, BWC229109H, BWC229415H, BWC229482H, BWC229731H, BWC229820H, BWC229825H, BWC229976H, BWC230004H, BWC230050H, BWC230059H, BWC230150H, BWC230186H, BWC230241H, BWC230282H, BWC230448H, BWC230484H, BWC230574H, BWC230793H, BWC231014H, BWC231149H, BWC231158H, BWC231162H, BWC231180H, BWC231230H, BWC231291H, BWC231313H, BWC231391H, BWC231399H, BWC231449H, BWC231451H, BWC231546H, BWC221572H, BWC222371H, BWC222433H, BWC222585H, BWC223393H, BWC223438H, BWC224773H, BWC225018H, BWC225552H, BWC226730H, BWC227662H, BWC228064H, BWC228425H, BWC228460H, BWC228861H, BWC228955H, BWC229010H, BWC229111H, BWC229322H, BWC229994H, BWC230077H, BWC230199H, BWC230213H, BWC230377H, BWC230385H, BWC230453H, BWC230532H, BWC230547H, BWC230548H, BWC230550H, BWC230551H, BWC230801H, BWC231097H, BWC231120H, BWC231273H, BWC231325H, BWC231417H, BWC221033H, BWC222004H, BWC223597H, BWC223868H, BWC225766H, BWC225916H, BWC226450H, BWC227631H, BWC228169H, BWC228279H, BWC228459H, BWC229429H, BWC229630H, BWC229645H, BWC229843H, BWC229855H, BWC229938H, BWC229989H, BWC230007H, BWC230242H, BWC230253H, BWC230265H, BWC230301H, BWC230368H, BWC230373H, BWC230519H, BWC230528H, BWC230529H, BWC230835H, BWC230867H, BWC231080H, BWC231109H, BWC231155H, BWC231389H, BWC228397H, BWC223261H, BWC225405H, BWC227465H, BWC228143H, BWC228393H, BWC229258H, BWC229476H, BWC229761H, BWC229919H, BWC229991H, BWC230032H, BWC230120H, BWC230230H, BWC230260H, BWC230261H, BWC230307H, BWC230490H, BWC230510H, BWC230961H, BWC231145H, BWC231270H, BWC231298H, BWC231333H, BWC231352H, BWC231362H, BWC232182H, BWC232187H,

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1305-2021

Related Recalls

Class I - Dangerous

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

May 10, 2023 Implants & Prosthetics Nationwide View Details →
Class I - Dangerous

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

May 10, 2023 Implants & Prosthetics Nationwide View Details →
Class I - Dangerous

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

May 10, 2023 Implants & Prosthetics Nationwide View Details →