HeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigational Device Exemption (IDE) 2. HeartMate 3 Sealed Outflow Graft with Bend Relief, US 3. HeartMate 3 Sealed Outflow Graft with Bend Relief, OUS 4. HeartMate 3 LVAS Implant Kit, IDE 5. HeartMate 3 LVAS Implant Kit, US 6. HeartMate 3 LVAS Implant Kit, OUS HeartMate 3 Left Ventricular Assist System: The HeartMate 3 LVAS is designed to provide therapeutic benefit to those affected with advanced heart failure. The LVAS assumes some or all of the workload of the left ventricle. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus supporting equipment.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    HeartMate 3 LVAS Model Numbers: 1. 105581 (IDE, cannot be implanted without a Left Ventricular Assist Device (LVAD)) 2. 105581US (cannot be implanted without a LVAD) 3. 105581INT (cannot be implanted without a LVAD) 4. 106524 (IDE) 5. 106524US 6. 106524INT UDI-DI codes: HeartMate 3 LVAS: 1. N/A (IDE) 2. 00813024013266 3. 00813024011675 4. N/A (IDE) 5. 00813024013297 6. 00813024011712 Serial Numbers: For Model 106524 US MLP-000991 MLP-002058 MLP-002066 MLP-002069 MLP-002070 MLP-002076 MLP-002117 MLP-002126 MLP-002128 MLP-002133 MLP-002136 MLP-002152 MLP-002158 MLP-002159 MLP-002160 MLP-002164 MLP-002165 MLP-002191 MLP-002192 MLP-002197 MLP-002206 MLP-002208 MLP-002216 MLP-002217 MLP-002239 MLP-002241 MLP-002246 MLP-002251 MLP-002254 MLP-002259 MLP-002263 MLP-002268 MLP-002269 MLP-002278 MLP-002288 MLP-002289 MLP-002291 MLP-002294 MLP-002300 MLP-002304 MLP-002309 MLP-002310 MLP-002316 MLP-002318 MLP-002321 MLP-002324 MLP-002337 MLP-002342 MLP-002347 MLP-002355 MLP-002356 MLP-002361 MLP-002365 MLP-002450 MLP-002455 MLP-002456 MLP-002472 MLP-002475 MLP-002510 MLP-002514 MLP-002518 MLP-002519 MLP-002520 MLP-002530 MLP-002536 MLP-002538 MLP-002539 MLP-002543 MLP-002589 MLP-002597 MLP-002603 MLP-002633 MLP-002643 MLP-002646 MLP-002650 MLP-002654 MLP-002660 MLP-002663 MLP-002666 MLP-002805 MLP-002807 MLP-002808 MLP-002809 MLP-002827 MLP-002828 MLP-002838 MLP-002840 MLP-002842 MLP-002844 MLP-002850 MLP-002858 MLP-002957 MLP-003028 MLP-003033 MLP-003057 MLP-003058 MLP-003068 MLP-003070 MLP-003072 MLP-003074 MLP-003077 MLP-003085 MLP-003086 MLP-003089 MLP-003097 MLP-003110 MLP-003114 MLP-003125 MLP-003130 MLP-003131 MLP-003140 MLP-003148 MLP-003186 MLP-003237 MLP-003255 MLP-003257 MLP-003259 MLP-003277 MLP-003291 MLP-003299 MLP-003300 MLP-003302 MLP-003307 MLP-003308 MLP-003314 MLP-003317 MLP-003340 MLP-003347 MLP-003348 MLP-003351 MLP-003352 MLP-003384 MLP-003385 MLP-003388 MLP-003389 MLP-003390 MLP-003441 MLP-003442 MLP-003461 MLP-003466 MLP-003469 MLP-003471 MLP-003571 MLP-003572 MLP-003574 MLP-003577 MLP-003578 MLP-003582 MLP-003588 MLP-003615 MLP-003616 MLP-003617 MLP-003621 MLP-003629 MLP-003654 MLP-003659 MLP-003661 MLP-003689 MLP-003692 MLP-003738 MLP-003741 MLP-003828 MLP-003829 MLP-003858 MLP-003861 MLP-003862 MLP-003863 MLP-003867 MLP-004093 MLP-004110 MLP-004118 MLP-004119 MLP-004125 MLP-004129 MLP-004131 MLP-004137 MLP-004148 MLP-004151 MLP-004153 MLP-004156 MLP-004157 MLP-004165 MLP-004166 MLP-004171 MLP-004212 MLP-004242 MLP-004247 MLP-004267 MLP-004279 MLP-004281 MLP-004293 MLP-004298 MLP-004310 MLP-004384 MLP-004469 MLP-004486 MLP-004490 MLP-004525 MLP-004584 MLP-004591 MLP-004667 MLP-004675 MLP-004686 MLP-004697 MLP-004802 MLP-004805 MLP-004806 MLP-004866 MLP-004867 MLP-004869 MLP-004870 MLP-004874 MLP-004876 MLP-004877 MLP-004878 MLP-004884 MLP-004885 MLP-004887 MLP-004893 MLP-004954 MLP-004962 MLP-004963 MLP-004968 MLP-004969 MLP-004971 MLP-004983 MLP-004996 MLP-004998 MLP-005006 MLP-005032 MLP-005033 MLP-005036 MLP-005050 MLP-005052 MLP-005054 MLP-005083 MLP-005117 MLP-005122 MLP-005144 MLP-005145 MLP-005183 MLP-005186 MLP-005187 MLP-005193 MLP-005208 MLP-005212 MLP-005215 MLP-005219 MLP-005228 MLP-005258 MLP-005260 MLP-005261 MLP-005291 MLP-005292 MLP-005293 MLP-005298 MLP-005299 MLP-005320 MLP-005322 MLP-005323 MLP-005324 MLP-005328 MLP-005337 MLP-005391 MLP-005393 MLP-005396 MLP-005398 MLP-005402 MLP-005410 MLP-005411 MLP-005414 MLP-005416 MLP-005438 MLP-005440 MLP-005444 MLP-005460 MLP-005468 MLP-005483 MLP-005484 MLP-005485 MLP-005492 MLP-005494 MLP-005497 MLP-005498 MLP-005512 MLP-005531 MLP-005532 MLP-005539 MLP-005540 MLP-005541 MLP-005544 MLP-005552 MLP-005555 MLP-005559 MLP-005576 MLP-005594 MLP-005597 MLP-005600 MLP-005602 MLP-005613 MLP-005620 MLP-005636 MLP-005638 MLP-005641 MLP-005650 MLP-005653 MLP-005655 MLP-005663 MLP-005668 MLP-005680 MLP-005684 MLP-005703 MLP-005704 MLP-005706 MLP-005708 MLP-005729 MLP-005730 MLP-005746 MLP-005747 MLP-005752 MLP-005754 MLP-005769 MLP-005791 MLP-005793 MLP-005794 MLP-005795 MLP-005798 MLP-005812 MLP-005820 MLP-005821 MLP-005822 MLP-005825 MLP-005828 MLP-005831 MLP-005835 MLP-005836 MLP-005838 MLP-005853 MLP-005871 MLP-005879 MLP-005901 MLP-005902 MLP-005924 MLP-005934 MLP-005967 MLP-006005 MLP-006040 MLP-006041 MLP-006053 MLP-006067 MLP-006127 MLP-006128 MLP-006192 MLP-006193 MLP-006203 MLP-006204 MLP-006207 MLP-006216 MLP-006240 MLP-006245 MLP-006246 MLP-006249 MLP-006294 MLP-006303 MLP-006304 MLP-006390 MLP-006412 MLP-006413 MLP-006423 MLP-006431 MLP-006435 MLP-006437 MLP-006450 MLP-006452 MLP-006463 MLP-006470 MLP-006471 MLP-006477 MLP-006489 MLP-006490 MLP-006534 MLP-006535 MLP-006538 MLP-006539 MLP-006553 MLP-006571 MLP-006577 MLP-006578 MLP-006585 MLP-006587 MLP-006588 MLP-006595 MLP-006629 MLP-006631 MLP-006632 MLP-006633 MLP-006661 MLP-006663 MLP-006675 MLP-006677 MLP-006678 MLP-006688 MLP-006690 MLP-006710 MLP-006711 MLP-006715 MLP-006773 MLP-006801 MLP-006811 MLP-006827 MLP-006829 MLP-006831 MLP-006832 MLP-006846 MLP-006847 MLP-006848 MLP-006849 MLP-006851 MLP-006852 MLP-007250 MLP-007381 MLP-007388 MLP-007391 MLP-007396 MLP-007399 MLP-007400 MLP-007405 MLP-007408 MLP-007417 MLP-007420 MLP-007422 MLP-007428 MLP-007433 MLP-007444 MLP-007446 MLP-007447 MLP-007451 MLP-007459 MLP-007870 MLP-007871 MLP-007897 MLP-007905 MLP-007908 MLP-007921 MLP-007923 MLP-007926 MLP-007928 MLP-007929 MLP-007983 MLP-007985 MLP-007987 MLP-007988 MLP-007996 MLP-007997 MLP-007999 MLP-008005 MLP-008084 MLP-008085 MLP-008118 MLP-008119 MLP-008120 MLP-008161 MLP-008162 MLP-008165 MLP-008191 MLP-008194 MLP-008197 MLP-008199 MLP-008200 MLP-008250 MLP-008251 MLP-008252 MLP-008254 MLP-008255 MLP-008258 MLP-008266 MLP-008295 MLP-008296 MLP-008298 MLP-008302 MLP-008306 MLP-008308 MLP-008328 MLP-008329 MLP-008330 MLP-008385 MLP-008387 MLP-008388 MLP-008391 MLP-008392 MLP-008400 MLP-008401 MLP-008402 MLP-008403 MLP-008416 MLP-008417 MLP-008419 MLP-008425 MLP-008430 MLP-008445 MLP-008446 MLP-008524 MLP-008526 MLP-008528 MLP-008529 MLP-008550 MLP-008554 MLP-008556 MLP-008563 MLP-008565 MLP-008566 MLP-008576 MLP-008593 MLP-008606 MLP-008609 MLP-008651 MLP-008654 MLP-008656 MLP-008658 MLP-008659 MLP-008660 MLP-008661 MLP-008679 MLP-008681 MLP-008683 MLP-008809 MLP-008810 MLP-008811 MLP-008817 MLP-008819 MLP-008908 MLP-008911 MLP-008919 MLP-008928 MLP-008932 MLP-008933 MLP-008937 MLP-008939 MLP-009011 MLP-009012 MLP-009123 MLP-009124 MLP-009125 MLP-009126 MLP-009132 MLP-009153 MLP-009228 MLP-009259 MLP-009265 MLP-009273 MLP-009274 MLP-009275 MLP-009278 MLP-009280 MLP-009284 MLP-009285 MLP-009286 MLP-009287 MLP-009297 MLP-009308 MLP-009310 MLP-009311 MLP-009313 MLP-009315 MLP-009329 MLP-009330 MLP-009334 MLP-009335 MLP-009336 MLP-009341 MLP-009342 MLP-009357 MLP-009360 MLP-009364 MLP-009365 MLP-009368 MLP-009399 MLP-009413 MLP-009414 MLP-009416 MLP-009418 MLP-009423 MLP-009427 MLP-009429 MLP-009438 MLP-009442 MLP-009443 MLP-009448 MLP-009451 MLP-009453 MLP-009455 MLP-009456 MLP-009483 MLP-009491 MLP-009503 MLP-009504 MLP-009505 MLP-009506 MLP-009510 MLP-009521 MLP-009522 MLP-009528 MLP-009543 MLP-009647 MLP-009648 MLP-009649 MLP-009651 MLP-009652 MLP-009654 MLP-009655 MLP-009675 MLP-009677 MLP-009678 MLP-009681 MLP-009682 MLP-009695 MLP-009698 MLP-009699 MLP-009712 MLP-009732 MLP-009736 MLP-009741 MLP-009742 MLP-009751 MLP-009752 MLP-009753 MLP-009754 MLP-009755 MLP-009757 MLP-009758 MLP-009760 MLP-009765 MLP-009774 MLP-009778 MLP-009779 MLP-009783 MLP-009785 MLP-009790 MLP-009792 MLP-009796 MLP-009800 MLP-009803 MLP-009804 MLP-009810 MLP-009815 MLP-009816 MLP-009820 MLP-009822 MLP-009825 MLP-009826 MLP-009832 MLP-009837 MLP-009839 MLP-009865 MLP-009867 MLP-009868 MLP-009870 MLP-009872 MLP-009873 MLP-009910 MLP-009911 MLP-009915 MLP-009916 MLP-009921 MLP-009922 MLP-009925 MLP-009926 MLP-009927 MLP-009929 MLP-009980 MLP-009981 MLP-009995 MLP-009996 MLP-010001 MLP-010003 MLP-010006 MLP-010032 MLP-010034 MLP-010046 MLP-010047 MLP-010048 MLP-010065 MLP-010068 MLP-010105 MLP-010120 MLP-010121 MLP-010166 MLP-010167 MLP-010179 MLP-010180 MLP-010182 MLP-010183 MLP-010199 MLP-010276 MLP-010280 MLP-010288 MLP-010306 MLP-010317 MLP-010385 MLP-010386 MLP-010392 MLP-010406 MLP-010412 MLP-010424 MLP-010435 MLP-010437 MLP-010452 MLP-010453 MLP-010465 MLP-010473 MLP-010474 MLP-010485 MLP-010504 MLP-010573 MLP-010574 MLP-010577 MLP-010578 MLP-010609 MLP-010776 MLP-010847 MLP-010848 MLP-010849 MLP-010852 MLP-010853 MLP-010860 MLP-010861 MLP-010906 MLP-010924 MLP-010938 MLP-010939 MLP-010940 MLP-010941 MLP-010942 MLP-010945 MLP-010946 MLP-010952 MLP-010953 MLP-010955 MLP-010957 MLP-010970 MLP-011066 MLP-011070 MLP-011088 MLP-011117 MLP-011118 MLP-011120 MLP-011123 MLP-011124 MLP-011125 MLP-011131 MLP-011136 MLP-011137 MLP-011138 MLP-011140 MLP-011190 MLP-011191 MLP-011193 MLP-011336 MLP-011370 MLP-011371 MLP-011372 MLP-011542 MLP-011545 MLP-011547 MLP-011550 MLP-011552 MLP-011647 MLP-011654 MLP-011656 MLP-011657 MLP-011658 MLP-011660 MLP-011661 MLP-011701 MLP-011702 MLP-011704 MLP-012115 MLP-012117 MLP-012121 MLP-012123 MLP-012145 MLP-012160 MLP-012180 MLP-012182 MLP-012183 MLP-012184 MLP-012185 MLP-012189 MLP-012190 MLP-012191 MLP-012193 MLP-012199 MLP-012266 MLP-012267 MLP-012268 MLP-012303 MLP-012334 MLP-012338 MLP-012349 MLP-012377 MLP-012405 MLP-012407 MLP-012408 MLP-012410 MLP-012420 MLP-012421 MLP-012612 MLP-012618 MLP-012619 MLP-012629 MLP-012631 MLP-012633 MLP-012636 MLP-012667 MLP-012745 OUS MLP-002000 MLP-002003 MLP-002005 MLP-002015 MLP-002019 MLP-002022 MLP-002023 MLP-002024 MLP-002026 MLP-002027 MLP-002028 MLP-002043 MLP-002050 MLP-002053 MLP-002055 MLP-002064 MLP-002095 MLP-002097 MLP-002098 MLP-002101 MLP-002103 MLP-002105 MLP-002106 MLP-002111 MLP-002113 MLP-002114 MLP-002115 MLP-002116 MLP-002122 MLP-002143 MLP-002145 MLP-002147 MLP-002150 MLP-002153 MLP-002186 MLP-002190 MLP-002194 MLP-002230 MLP-002322 MLP-002323 MLP-002328 MLP-006360 MLP-006364 MLP-007409 MLP-007873 MLP-007917 MLP-007918 MLP-008122 MLP-008123 MLP-009290 MLP-009372 MLP-009420 For Model 106524US US MLP-006853 MLP-006869 MLP-007411 MLP-007412 MLP-007445 MLP-007465 MLP-007468 MLP-007474 MLP-007475 MLP-007504 MLP-007508 MLP-007510 MLP-007538 MLP-007584 MLP-007585 MLP-007590 MLP-007604 MLP-007611 MLP-007625 MLP-007628 MLP-007630 MLP-007631 MLP-007635 MLP-007637 MLP-007642 MLP-007643 MLP-007657 MLP-007658 MLP-007660 MLP-007662 MLP-007670 MLP-007672 MLP-007678 MLP-007679 MLP-007682 MLP-007685 MLP-007686 MLP-007687 MLP-007690 MLP-007702 MLP-007705 MLP-007711 MLP-007716 MLP-007717 MLP-007719 MLP-007765 MLP-007766 MLP-007767 MLP-007768 MLP-007773 MLP-007774 MLP-007775 MLP-007777 MLP-007780 MLP-007792 MLP-007801 MLP-007802 MLP-007803 MLP-007823 MLP-007826 MLP-007835 MLP-007837 MLP-007874 MLP-007876 MLP-007880 MLP-007883 MLP-007888 MLP-007889 MLP-007909 MLP-007912 MLP-007913 MLP-007915 MLP-007916 MLP-007936 MLP-007938 MLP-007943 MLP-008023 MLP-008024 MLP-008028 MLP-008046 MLP-008049 MLP-008052 MLP-008054 MLP-008068 MLP-008069 MLP-008079 MLP-008080 MLP-008081 MLP-008082 MLP-008089 MLP-008104 MLP-008110 MLP-008111 MLP-008113 MLP-008116 MLP-008117 MLP-008121 MLP-008129 MLP-008130 MLP-008131 MLP-008138 MLP-008140 MLP-008164 MLP-008167 MLP-008168 MLP-008201 MLP-008204 MLP-008205 MLP-008208 MLP-008210 MLP-008217 MLP-008220 MLP-008222 MLP-008247 MLP-008270 MLP-008277 MLP-008284 MLP-008303 MLP-008312 MLP-008314 MLP-008346 MLP-008350 MLP-008351 MLP-008353 MLP-008359 MLP-008363 MLP-008366 MLP-008380 MLP-008381 MLP-008426 MLP-008429 MLP-008442 MLP-008464 MLP-008469 MLP-008503 MLP-008506 MLP-008507 MLP-008510 MLP-008516 MLP-008517 MLP-008521 MLP-008535 MLP-008541 MLP-008542 MLP-008548 MLP-008552 MLP-008557 MLP-008559 MLP-008578 MLP-008582 MLP-008586 MLP-008588 MLP-008611 MLP-008615 MLP-008618 MLP-008619 MLP-008621 MLP-008623 MLP-008647 MLP-008650 MLP-008766 MLP-008773 MLP-008796 MLP-008799 MLP-008813 MLP-008815 MLP-008824 MLP-008827 MLP-008831 MLP-008832 MLP-008839 MLP-008842 MLP-008843 MLP-008846 MLP-008847 MLP-008879 MLP-008882 MLP-008885 MLP-008886 MLP-008888 MLP-008890 MLP-008898 MLP-008900 MLP-008902 MLP-008903 MLP-008914 MLP-008923 MLP-008929 MLP-008963 MLP-008964 MLP-008966 MLP-008970 MLP-008971 MLP-008973 MLP-008975 MLP-008976 MLP-008981 MLP-008982 MLP-008989 MLP-008991 MLP-008994 MLP-008996 MLP-009004 MLP-009014 MLP-009044 MLP-009046 MLP-009048 MLP-009061 MLP-009067 MLP-009068 MLP-009069 MLP-009071 MLP-009075 MLP-009076 MLP-009079 MLP-009080 MLP-009086 MLP-009089 MLP-009090 MLP-009091 MLP-009093 MLP-009105 MLP-009152 MLP-009156 MLP-009164 MLP-009166 MLP-009168 MLP-009170 MLP-009177 MLP-009178 MLP-009179 MLP-009188 MLP-009190 MLP-009192 MLP-009194 MLP-009225 MLP-009271 MLP-009272 MLP-009293 MLP-009296 MLP-009301 MLP-009302 MLP-009304 MLP-009316 MLP-009366 MLP-009385 MLP-009386 MLP-009388 MLP-009394 MLP-009400 MLP-009408 MLP-009424 MLP-009428 MLP-009431 MLP-009432 MLP-009490 MLP-009493 MLP-009501 MLP-009502 MLP-009511 MLP-009517 MLP-009524 MLP-009529 MLP-009534 MLP-009547 MLP-009549 MLP-009550 MLP-009566 MLP-009568 MLP-009589 MLP-009590 MLP-009591 MLP-009593 MLP-009598 MLP-009604 MLP-009610 MLP-009630 MLP-009632 MLP-009637 MLP-009640 MLP-009641 MLP-009685 MLP-009688 MLP-009691 MLP-009694 MLP-009700 MLP-009713 MLP-009714 MLP-009715 MLP-009746 MLP-009786 MLP-009791 MLP-009821 MLP-009831 MLP-009833 MLP-009838 MLP-009874 MLP-009896 MLP-009897 MLP-009900 MLP-009912 MLP-009917 MLP-009933 MLP-009938 MLP-009939 MLP-009940 MLP-009941 MLP-009943 MLP-009971 MLP-009974 MLP-009993 MLP-010011 MLP-010012 MLP-010025 MLP-010030 MLP-010053 MLP-010055 MLP-010057 MLP-010060 MLP-010066 MLP-010095 MLP-010103 MLP-010106 MLP-010119 MLP-010123 MLP-010125 MLP-010127 MLP-010130 MLP-010133 MLP-010134 MLP-010135 MLP-010136 MLP-010149 MLP-010173 MLP-010202 MLP-010204 MLP-010209 MLP-010211 MLP-010283 MLP-010286 MLP-010293 MLP-010305 MLP-010310 MLP-010312 MLP-010313 MLP-010321 MLP-010335 MLP-010338 MLP-010342 MLP-010344 MLP-010347 MLP-010348 MLP-010349 MLP-010350 MLP-010373 MLP-010374 MLP-010378 MLP-010380 MLP-010381 MLP-010384 MLP-010390 MLP-010391 MLP-010416 MLP-010419 MLP-010436 MLP-010445 MLP-010451 MLP-010468 MLP-010469 MLP-010472 MLP-010476 MLP-010498 MLP-010500 MLP-010505 MLP-010507 MLP-010548 MLP-010556 MLP-010583 MLP-010586 MLP-010591 MLP-010592 MLP-010611 MLP-010615 MLP-010618 MLP-010620 MLP-010621 MLP-010623 MLP-010624 MLP-010645 MLP-010646 MLP-010647 MLP-010650 MLP-010651 MLP-010657 MLP-010660 MLP-010663 MLP-010692 MLP-010767 MLP-010773 MLP-010774 MLP-010775 MLP-010779 MLP-010790 MLP-010796 MLP-010798 MLP-010800 MLP-010817 MLP-010819 MLP-010820 MLP-010844 MLP-010846 MLP-010855 MLP-010870 MLP-010872 MLP-010877 MLP-010878 MLP-010902 MLP-010908 MLP-010925 MLP-010933 MLP-010934 MLP-010935 MLP-010937 MLP-010964 MLP-010967 MLP-010968 MLP-010975 MLP-010976 MLP-010977 MLP-010981 MLP-010985 MLP-010988 MLP-011000 MLP-011001 MLP-011003 MLP-011004 MLP-011005 MLP-011009 MLP-011012 MLP-011013 MLP-011015 MLP-011016 MLP-011017 MLP-011018 MLP-011025 MLP-011027 MLP-011030 MLP-011115 MLP-011143 MLP-011181 MLP-011184 MLP-011185 MLP-011211 MLP-011213 MLP-011334 MLP-011341 MLP-011378 MLP-011454 MLP-011480 MLP-011481 MLP-011484 MLP-011515 MLP-011516 MLP-011517 MLP-011532 MLP-011544 MLP-011554 MLP-011556 MLP-011557 MLP-011559 MLP-011561 MLP-011563 MLP-011611 MLP-011644 MLP-011645 MLP-011646 MLP-011662 MLP-011673 MLP-011676 MLP-011677 MLP-011679 MLP-011687 MLP-011709 MLP-011710 MLP-011722 MLP-011883 MLP-011889 MLP-011901 MLP-011906 MLP-011907 MLP-011909 MLP-011928 MLP-011931 MLP-011933 MLP-011934 MLP-011935 MLP-011939 MLP-011941 MLP-011943 MLP-011951 MLP-011955 MLP-012063 MLP-012068 MLP-012076 MLP-012105 MLP-012116 MLP-012119 MLP-012120 MLP-012128 MLP-012133 MLP-012134 MLP-012138 MLP-012144 MLP-012159 MLP-012165 MLP-012181 MLP-012217 MLP-012230 MLP-012272 MLP-012273 MLP-012276 MLP-012278 MLP-012279 MLP-012281 MLP-012284 MLP-012285 MLP-012288 MLP-012293 MLP-012294 MLP-012308 MLP-012309 MLP-012311 MLP-012315 MLP-012330 MLP-012336 MLP-012339 MLP-012340 MLP-012341 MLP-012342 MLP-012348 MLP-012354 MLP-012357 MLP-012362 MLP-012364 MLP-012365 MLP-012367 MLP-012370 MLP-012381 MLP-012382 MLP-012383 MLP-012412 MLP-012422 MLP-012471 MLP-012473 MLP-012474 MLP-012477 MLP-012487 MLP-012489 MLP-012490 MLP-012491 MLP-012493 MLP-012494 MLP-012500 MLP-012501 MLP-012521 MLP-012533 MLP-012535 MLP-012536 MLP-012539 MLP-012546 MLP-012549 MLP-012550 MLP-012551 MLP-012552 MLP-012553 MLP-012555 MLP-012556 MLP-012557 MLP-012559 MLP-012560 MLP-012569 MLP-012574 MLP-012580 MLP-012584 MLP-012585 MLP-012586 MLP-012588 MLP-012589 MLP-012590 MLP-012591 MLP-012592 MLP-012604 MLP-012605 MLP-012614 MLP-012615 MLP-012616 MLP-012710 MLP-012723 MLP-012741 MLP-012749 MLP-012755 MLP-012756 MLP-012757 MLP-012760 MLP-012763 MLP-012766 MLP-012769 MLP-012770 MLP-012771 MLP-012773 MLP-012775 MLP-012776 MLP-012778 MLP-012781 MLP-012788 MLP-012789 MLP-012790 MLP-012863 MLP-012864 MLP-012866 MLP-012869 MLP-012871 MLP-012918 MLP-012919 MLP-012923 MLP-012925 MLP-012926 MLP-012929 MLP-012933 MLP-012963 MLP-012967 MLP-012968 MLP-012970 MLP-012973 MLP-012975 MLP-012976 MLP-012981 MLP-012983 MLP-012986 MLP-012991 MLP-012992 MLP-012994 MLP-012996 MLP-013006 MLP-013008 MLP-013009 MLP-013012 MLP-013017 MLP-013020 MLP-013021 MLP-013023 MLP-013024 MLP-013029 MLP-013033 MLP-013034 MLP-013035 MLP-013036 MLP-013041 MLP-013044 MLP-013046 MLP-013047 MLP-013064 MLP-013065 MLP-013066 MLP-013069 MLP-013070 MLP-013072 MLP-013073 MLP-013074 MLP-013076 MLP-013079 MLP-013081 MLP-013083 MLP-013084 MLP-013085 MLP-013086 MLP-013087 MLP-013089 MLP-013092 MLP-013103 MLP-013104 MLP-013106 MLP-013131 MLP-013133 MLP-013135 MLP-013136 MLP-013157 MLP-013158 MLP-013159 MLP-013168 MLP-013170 MLP-013176 MLP-013181 MLP-013187 MLP-013189 MLP-013191 MLP-013195 MLP-013212 MLP-013241 MLP-013251 MLP-013256 MLP-013257 MLP-013261 MLP-013262 MLP-013269 MLP-013271 MLP-013273 MLP-013274 MLP-013314 MLP-013324 MLP-013327 MLP-013400 MLP-013406 MLP-013417 MLP-013418 MLP-013420 MLP-013421 MLP-013422 MLP-013424 MLP-013425 MLP-013426 MLP-013429 MLP-013432 MLP-013434 MLP-013436 MLP-013446 MLP-013452 MLP-013464 MLP-013466 MLP-013477 MLP-013493 MLP-013502 MLP-013512 MLP-013514 MLP-013517 MLP-013518 MLP-013535 MLP-013542 MLP-013544 MLP-013545 MLP-013550 MLP-013558 MLP-013563 MLP-013565 MLP-013568 MLP-013573 MLP-013576 MLP-013586 MLP-013587 MLP-013589 MLP-013617 MLP-013625 MLP-013626 MLP-013627 MLP-013633 MLP-013635 MLP-013637 MLP-013641 MLP-013652 MLP-013657 MLP-013660 MLP-013662 MLP-013663 MLP-013664 MLP-013665 MLP-013669 MLP-013672 MLP-013679 MLP-013691 MLP-013707 MLP-013711 MLP-013712 MLP-013714 MLP-013737 MLP-013738 MLP-013746 MLP-013759 MLP-013761 MLP-013762 MLP-013763 MLP-013765 MLP-013768 MLP-013772 MLP-013776 MLP-013778 MLP-013783 MLP-013784 MLP-013788 MLP-013790 MLP-013791 MLP-013793 MLP-013798 MLP-013799 MLP-013800 MLP-013801 MLP-013802 MLP-013812 MLP-013817 MLP-013819 MLP-013824 MLP-013826 MLP-013827 MLP-013829 MLP-013830 MLP-013832 MLP-013852 MLP-013889 MLP-013890 MLP-013891 MLP-013893 MLP-013895 MLP-013897 MLP-013900 MLP-013906 MLP-013907 MLP-013913 MLP-013915 MLP-013921 MLP-013922 MLP-013933 MLP-013935 MLP-013938 MLP-013940 MLP-013942 MLP-013946 MLP-013947 MLP-013951 MLP-013953 MLP-013958 MLP-013960 MLP-013962 MLP-013963 MLP-013967 MLP-013968 MLP-013971 MLP-013973 MLP-013976 MLP-013988 MLP-013989 MLP-013992 MLP-013993 MLP-013997 MLP-013998 MLP-014002 MLP-014003 MLP-014004 MLP-014006 MLP-014007 MLP-014008 MLP-014009 MLP-014014 MLP-014018 MLP-014019 MLP-014021 MLP-014022 MLP-014027 MLP-014063 MLP-014070 MLP-014075 MLP-014076 MLP-014097 MLP-014099 MLP-014101 MLP-014104 MLP-014106 MLP-014111 MLP-014118 MLP-014119 MLP-014139 MLP-014140 MLP-014142 MLP-014148 MLP-014154 MLP-014161 MLP-014166 MLP-014169 MLP-014172 MLP-014174 MLP-014175 MLP-014176 MLP-014177 MLP-014178 MLP-014188 MLP-014189 MLP-014194 MLP-014195 MLP-014200 MLP-014201 MLP-014202 MLP-014203 MLP-014204 MLP-014206 MLP-014207 MLP-014209 MLP-014210 MLP-014211 MLP-014214 MLP-014219 MLP-014235 MLP-014243 MLP-014246 MLP-014253 MLP-014255 MLP-014275 MLP-014276 MLP-014278 MLP-014279 MLP-014281 MLP-014283 MLP-014327 MLP-014328 MLP-014333 MLP-014334 MLP-014337 MLP-014339 MLP-014342 MLP-014370 MLP-014371 MLP-014375 MLP-014376 MLP-014381 MLP-014383 MLP-014384 MLP-014385 MLP-014389 MLP-014409 MLP-014411 MLP-014412 MLP-014426 MLP-014428 MLP-014429 MLP-014430 MLP-014433 MLP-014434 MLP-014437 MLP-014438 MLP-014443 MLP-014477 MLP-014486 MLP-014488 MLP-014489 MLP-014493 MLP-014508 MLP-014509 MLP-014510 MLP-014512 MLP-014513 MLP-014520 MLP-014546 MLP-014548 MLP-014638 MLP-014640 MLP-014646 MLP-014650 MLP-014652 MLP-014654 MLP-014655 MLP-014657 MLP-014660 MLP-014664 MLP-014668 MLP-014673 MLP-014674 MLP-014677 MLP-014678 MLP-014679 MLP-014682 MLP-014684 MLP-014687 MLP-014688 MLP-014689 MLP-014700 MLP-014702 MLP-014705 MLP-014706 MLP-014735 MLP-014737 MLP-014738 MLP-014739 MLP-014741 MLP-014743 MLP-014749 MLP-014750 MLP-014755 MLP-014756 MLP-014769 MLP-014770 MLP-014772 MLP-014774 MLP-014777 MLP-014778 MLP-014779 MLP-014796 MLP-014798 MLP-014800 MLP-014801 MLP-014803 MLP-014804 MLP-014806 MLP-014807 MLP-014810 MLP-014811 MLP-014814 MLP-014818 MLP-014821 MLP-014822 MLP-014827 MLP-014830 MLP-014833 MLP-014835 MLP-014836 MLP-014839 MLP-014840 MLP-014841 MLP-014849 MLP-014866 MLP-014868 MLP-014870 MLP-014873 MLP-014874 MLP-014875 MLP-014876 MLP-014880 MLP-014883 MLP-014886 MLP-014895 MLP-014896 MLP-014898 MLP-014900 MLP-014902 MLP-014904 MLP-014906 MLP-014913 MLP-014915 MLP-014916 MLP-014917 MLP-014922 MLP-014924 MLP-014925 MLP-014930 MLP-014932 MLP-014933 MLP-014934 MLP-014943 MLP-014944 MLP-014947 MLP-014948 MLP-014952 MLP-014954 MLP-014958 MLP-014959 MLP-014960 MLP-014961 MLP-014963 MLP-014964 MLP-015015 MLP-015016 MLP-015018 MLP-015019 MLP-015022 MLP-015024 MLP-015025 MLP-015026 MLP-015028 MLP-015029 MLP-015030 MLP-015031 MLP-015035 MLP-015036 MLP-015038 MLP-015040 MLP-015041 MLP-015042 MLP-015056 MLP-015057 MLP-015058 MLP-015061 MLP-015062 MLP-015063 MLP-015067 MLP-015068 MLP-015069 MLP-015070 MLP-015071 MLP-015073 MLP-015074 MLP-015075 MLP-015076 MLP-015077 MLP-015080 MLP-015083 MLP-015084 MLP-015088 MLP-015089 MLP-015092 MLP-015093 MLP-015099 MLP-015102 MLP-015104 MLP-015105 MLP-015106 MLP-015107 MLP-015117 MLP-015120 MLP-015123 MLP-015126 MLP-015128 MLP-015129 MLP-015130 MLP-015131 MLP-015133 MLP-015134 MLP-015136 MLP-015137 MLP-015138 MLP-015139 MLP-015140 MLP-015142 MLP-015146 MLP-015147 MLP-015151 MLP-015156 MLP-015161 MLP-015162 MLP-015163 MLP-015165 MLP-015168 MLP-015169 MLP-015174 MLP-015177 MLP-015180 MLP-015181 MLP-015182 MLP-015183 MLP-015184 MLP-015186 MLP-015187 MLP-015205 MLP-015214 MLP-015215 MLP-015217 MLP-015218 MLP-015224 MLP-015242 MLP-015243 MLP-015244 MLP-015246 MLP-015248 MLP-015252 MLP-015255 MLP-015257 MLP-015261 MLP-015276 MLP-015281 MLP-015293 MLP-015299 MLP-015301 MLP-015305 MLP-015306 MLP-015307 MLP-015316 MLP-015320 MLP-015321 MLP-015324 MLP-015325 MLP-015343 MLP-015355 MLP-015358 MLP-015371 MLP-015373 MLP-015381 MLP-015383 MLP-015384 MLP-015386 MLP-015389 MLP-015392 MLP-015393 MLP-015395 MLP-015396 MLP-015399 MLP-015400 MLP-015401 MLP-015405 MLP-015413 MLP-015417 MLP-015422 MLP-015424 MLP-015434 MLP-015436 MLP-015443 MLP-015451 MLP-015463 MLP-015464 MLP-015475 MLP-015504 MLP-015505 MLP-015508 MLP-015509 MLP-015512 MLP-015513 MLP-015515 MLP-015517 MLP-015519 MLP-015521 MLP-015523 MLP-015535 MLP-015536 MLP-015540 MLP-015541 MLP-015547 MLP-015556 MLP-015558 MLP-015559 MLP-015566 MLP-015567 MLP-015570 MLP-015572 MLP-015573 MLP-015574 MLP-015580 MLP-015581 MLP-015582 MLP-015584 MLP-015585 MLP-015589 MLP-015590 MLP-015592 MLP-015594 MLP-015596 MLP-015598 MLP-015600 MLP-015602 MLP-015604 MLP-015605 MLP-015606 MLP-015607 MLP-015611 MLP-015631 MLP-015648 MLP-015650 MLP-015651 MLP-015652 MLP-015653 MLP-015654 MLP-015655 MLP-015657 MLP-015658 MLP-015660 MLP-015661 MLP-015674 MLP-015685 MLP-015687 MLP-015688 MLP-015690 MLP-015691 MLP-015696 MLP-015699 MLP-015706 MLP-015709 MLP-015712 MLP-015713 MLP-015716 MLP-015721 MLP-015722 MLP-015729 MLP-015748 MLP-015749 MLP-015750 MLP-015752 MLP-015753 MLP-015756 MLP-015758 MLP-015769 MLP-015774 MLP-015778 MLP-015811 MLP-015812 MLP-015818 MLP-015819 MLP-015824 MLP-015825 MLP-015827 MLP-015829 MLP-015841 MLP-015842 MLP-015843 MLP-015877 MLP-015879 MLP-015881 MLP-015883 MLP-015897 MLP-015902 MLP-015903 MLP-015904 MLP-015905 MLP-015906 MLP-015908 MLP-015909 MLP-015910 MLP-015915 MLP-015916 MLP-015920 MLP-015922 MLP-015931 MLP-015932 MLP-015933 MLP-015936 MLP-015939 MLP-015940 MLP-015941 MLP-015942 MLP-015946 MLP-015947 MLP-015949 MLP-015950 MLP-015967 MLP-015973 MLP-015976 MLP-015978 MLP-015980 MLP-015988 MLP-015991 MLP-015994 MLP-015998 MLP-015999 MLP-016000 MLP-016001 MLP-016002 MLP-016004 MLP-016011 MLP-016012 MLP-016014 MLP-016016 MLP-016018 MLP-016019 MLP-016020 MLP-016030 MLP-016033 MLP-016035 MLP-016036 MLP-016038 MLP-016040 MLP-016058 MLP-016067 MLP-016071 MLP-016078 MLP-016080 MLP-016081 MLP-016083 MLP-016084 MLP-016085 MLP-016102 MLP-016103 MLP-016104 MLP-016105 MLP-016107 MLP-016109 MLP-016110 MLP-016115 MLP-016118 MLP-016119 MLP-016121 MLP-016122 MLP-016124 MLP-016126 MLP-016130 MLP-016135 MLP-016142 MLP-016144 MLP-016157 MLP-016158 MLP-016160 MLP-016161 MLP-016162 MLP-016174 MLP-016175 MLP-016183 MLP-016187 MLP-016192 MLP-016193 MLP-016200 MLP-016202 MLP-016204 MLP-016216 MLP-016217 MLP-016218 MLP-016230 MLP-016231 MLP-016232 MLP-016233 MLP-016234 MLP-016241 MLP-016244 MLP-016246 MLP-016248 MLP-016249 MLP-016251 MLP-016253 MLP-016254 MLP-016255 MLP-016272 MLP-016273 MLP-016275 MLP-016289 MLP-016292 MLP-016293 MLP-016297 MLP-016298 MLP-016300 MLP-016302 MLP-016303 MLP-016305 MLP-016306 MLP-016307 MLP-016308 MLP-016325 MLP-016327 MLP-016328 MLP-016331 MLP-016335 MLP-016336 MLP-016338 MLP-016339 MLP-016341 MLP-016342 MLP-016349 MLP-016363 MLP-016364 MLP-016382 MLP-016383 MLP-016384 MLP-016387 MLP-016388 MLP-016404 MLP-016405 MLP-016407 MLP-016409 MLP-016411 MLP-016419 MLP-016421 MLP-016424 MLP-016425 MLP-016428 MLP-016429 MLP-016430 MLP-016443 MLP-016446 MLP-016457 MLP-016458 MLP-016459 MLP-016460 MLP-016462 MLP-016480 MLP-016481 MLP-016482 MLP-016483 MLP-016485 MLP-016486 MLP-016495 MLP-016497 MLP-016498 MLP-016499 MLP-016502 MLP-016508 MLP-016509 MLP-016510 MLP-016511 MLP-016512 MLP-016513 MLP-016520 MLP-016543 MLP-016551 MLP-016552 MLP-016553 MLP-016554 MLP-016564 MLP-016565 MLP-016566 MLP-016567 MLP-016568 MLP-016572 MLP-016573 MLP-016577 MLP-016579 MLP-016582 MLP-016583 MLP-016677 MLP-016682 MLP-016684 MLP-016686 MLP-016687 MLP-016692 MLP-016693 MLP-016696 MLP-016700 MLP-016703 MLP-016708 MLP-016709 MLP-016710 MLP-016713 MLP-016714 MLP-016716 MLP-016718 MLP-016725 MLP-016728 MLP-016731 MLP-016742 MLP-016743 MLP-016744 MLP-016749 MLP-016751 MLP-016755 MLP-016756 MLP-016760 MLP-016761 MLP-016765 MLP-016780 MLP-016792 MLP-016794 MLP-016795 MLP-016810 MLP-016813 MLP-016828 MLP-016833 MLP-016834 MLP-016838 MLP-016841 MLP-016844 MLP-016847 MLP-016849 MLP-016853 MLP-016856 MLP-016861 MLP-016862 MLP-016864 MLP-016865 MLP-016866 MLP-016868 MLP-016873 MLP-016874 MLP-016896 MLP-016897 MLP-016898 MLP-016899 MLP-016906 MLP-016907 MLP-016909 MLP-016928 MLP-016929 MLP-016931 MLP-016932 MLP-016938 MLP-016939 MLP-016940 MLP-016941 MLP-016942 MLP-016952 MLP-016955 MLP-016961 MLP-016984 MLP-016985 MLP-016986 MLP-016987 MLP-016988 MLP-016990 MLP-016992 MLP-016994 MLP-017004 MLP-017005 MLP-017007 MLP-017008 MLP-017009 MLP-017010 MLP-017014 MLP-017015 MLP-017016 MLP-017017 MLP-017018 MLP-017021 MLP-017023 MLP-017024 MLP-017025 MLP-017026 MLP-017029 MLP-017030 MLP-017031 MLP-017034 MLP-017035 MLP-017036 MLP-017038 MLP-017039 MLP-017041 MLP-017042 MLP-017043 MLP-017047 MLP-017053 MLP-017054 MLP-017056 MLP-017058 MLP-017062 MLP-017063 MLP-017064 MLP-017065 MLP-017074 MLP-017079 MLP-017080 MLP-017081 MLP-017082 MLP-017083 MLP-017084 MLP-017086 MLP-017087 MLP-017089 MLP-017092 MLP-017093 MLP-017095 MLP-017097 MLP-017098 MLP-017100 MLP-017101 MLP-017102 MLP-017112 MLP-017114 MLP-017115 MLP-017120 MLP-017125 MLP-017127 MLP-017128 MLP-017130 MLP-017133 MLP-017134 MLP-017155 MLP-017158 MLP-017173 MLP-017174 MLP-017176 MLP-017179 MLP-017181 MLP-017182 MLP-017198 MLP-017201 MLP-017203 MLP-017204 MLP-017210 MLP-017213 MLP-017214 MLP-017215 MLP-017218 MLP-017219 MLP-017224 MLP-017225 MLP-017226 MLP-017260 MLP-017261 MLP-017288 MLP-017290 MLP-017291 MLP-017309 MLP-017310 MLP-017311 MLP-017317 MLP-017319 MLP-017321 MLP-017324 MLP-017332 MLP-017333 MLP-017349 MLP-017351 MLP-017353 MLP-017366 MLP-017371 MLP-017375 MLP-017377 MLP-017384 MLP-017385 MLP-017387 MLP-017389 MLP-017393 MLP-017394 MLP-017401 MLP-017402 MLP-017404 MLP-017436 MLP-017437 MLP-017438 MLP-017447 MLP-017449 MLP-017450 MLP-017453 MLP-017455 MLP-017459 MLP-017461 MLP-017462 MLP-017463 MLP-017470 MLP-017472 MLP-017473 MLP-017474 MLP-017475 MLP-017515 MLP-017516 MLP-017517 MLP-017520 MLP-017534 MLP-017540 MLP-017547 MLP-017548 MLP-017549 MLP-017571 MLP-017572 MLP-017573 MLP-017574 MLP-017576 MLP-017579 MLP-017580 MLP-017582 MLP-017583 MLP-017598 MLP-017600 MLP-017602 MLP-017605 MLP-017610 MLP-017613 MLP-017615 MLP-017618 MLP-017619 MLP-017620 MLP-017624 MLP-017625 MLP-017626 MLP-017627 MLP-017628 MLP-017632 MLP-017635 MLP-017639 MLP-017640 MLP-017641 MLP-017642 MLP-017653 MLP-017663 MLP-017664 MLP-017665 MLP-017667 MLP-017684 MLP-017685 MLP-017687 MLP-017690 MLP-017691 MLP-017695 MLP-017697 MLP-017698 MLP-017699 MLP-017711 MLP-017712 MLP-017718 MLP-017719 MLP-017720 MLP-017729 MLP-017735 MLP-017742 MLP-017743 MLP-017746 MLP-017747 MLP-017748 MLP-017749 MLP-017750 MLP-017751 MLP-017752 MLP-017754 MLP-017767 MLP-017770 MLP-017776 MLP-017779 MLP-017784 MLP-017787 MLP-017788 MLP-017789 MLP-017790 MLP-017793 MLP-017799 MLP-017800 MLP-017802 MLP-017803 MLP-017804 MLP-017805 MLP-017807 MLP-017808 MLP-017809 MLP-017816 MLP-017817 MLP-017818 MLP-017833 MLP-017836 MLP-017837 MLP-017840 MLP-017841 MLP-017842 MLP-017843 MLP-017844 MLP-017845 MLP-017849 MLP-017855 MLP-017858 MLP-017860 MLP-017861 MLP-017862 MLP-017864 MLP-017865 MLP-017867 MLP-017872 MLP-017883 MLP-017884 MLP-017885 MLP-017886 MLP-017887 MLP-017888 MLP-017889 MLP-017904 MLP-017905 MLP-017907 MLP-017910 MLP-017912 MLP-017933 MLP-017947 MLP-017949 MLP-017950 MLP-017952 MLP-017959 MLP-017960 MLP-017961 MLP-017962 MLP-017977 MLP-017979 MLP-017980 MLP-017982 MLP-017984 MLP-017992 MLP-017993 MLP-017997 MLP-017999 MLP-018002 MLP-018003 MLP-018006 MLP-018007 MLP-018009 MLP-018010 MLP-018013 MLP-018014 MLP-018015 MLP-018016 MLP-018017 MLP-018018 MLP-018020 MLP-018023 MLP-018034 MLP-018035 MLP-018047 MLP-018049 MLP-018052 MLP-018054 MLP-018055 MLP-018057 MLP-018059 MLP-018060 MLP-018061 MLP-018062 MLP-018069 MLP-018071 MLP-018089 MLP-018091 MLP-018095 MLP-018097 MLP-018105 MLP-018108 MLP-018120 MLP-018122 MLP-018124 MLP-018132 MLP-018135 MLP-018137 MLP-018138 MLP-018139 MLP-018140 MLP-018143 MLP-018155 MLP-018158 MLP-018160 MLP-018161 MLP-018162 MLP-018175 MLP-018180 MLP-018183 MLP-018187 MLP-018188 MLP-018191 MLP-018194 MLP-018195 MLP-018197 MLP-018201 MLP-018205 MLP-018213 MLP-018215 MLP-018218 MLP-018220 MLP-018221 MLP-018223 MLP-018226 MLP-018235 MLP-018237 MLP-018240 MLP-018245 MLP-018246 MLP-018258 MLP-018259 MLP-018261 MLP-018262 MLP-018263 MLP-018264 MLP-018266 MLP-018269 MLP-018272 MLP-018274 MLP-018276 MLP-018277 MLP-018278 MLP-018299 MLP-018320 MLP-018323 MLP-018327 MLP-018332 MLP-018334 MLP-018337 MLP-018340 MLP-018341 MLP-018351 MLP-018359 MLP-018360 MLP-018364 MLP-018367 MLP-018370 MLP-018376 MLP-018377 MLP-018383 MLP-018386 MLP-018401 MLP-018408 MLP-018413 MLP-018415 MLP-018416 MLP-018418 MLP-018421 MLP-018439 MLP-018441 MLP-018443 MLP-018445 MLP-018446 MLP-018447 MLP-018465 MLP-018466 MLP-018467 MLP-018470 MLP-018483 MLP-018484 MLP-018485 MLP-018486 MLP-018487 MLP-018489 MLP-018490 MLP-018492 MLP-018493 MLP-018494 MLP-018495 MLP-018496 MLP-018498 MLP-018500 MLP-018501 MLP-018511 MLP-018513 MLP-018514 MLP-018515 MLP-018517 MLP-018523 MLP-018524 MLP-018525 MLP-018527 MLP-018528 MLP-018531 MLP-018536 MLP-018539 MLP-018541 MLP-018542 MLP-018543 MLP-018544 MLP-018546 MLP-018547 MLP-018549 MLP-018550 MLP-018551 MLP-018553 MLP-018555 MLP-018560 MLP-018563 MLP-018564 MLP-018565 MLP-018566 MLP-018567 MLP-018585 MLP-018587 MLP-018588 MLP-018589 MLP-018602 MLP-018604 MLP-018605 MLP-018606 MLP-018608 MLP-018610 MLP-018636 MLP-018637 MLP-018638 MLP-018644 MLP-018646 MLP-018649 MLP-018651 MLP-018652 MLP-018653 MLP-018654 MLP-018655 MLP-018656 MLP-018660 MLP-018663 MLP-018666 MLP-018667 MLP-018668 MLP-018675 MLP-018678 MLP-018679 MLP-018680 MLP-018683 MLP-018685 MLP-018686 MLP-018715 MLP-018718 MLP-018719 MLP-018722 MLP-018729 MLP-018748 MLP-018750 MLP-018751 MLP-018752 MLP-018756 MLP-018758 MLP-018760 MLP-018765 MLP-0187
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Thoratec Corp.
Reason for Recall:
Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

HeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigational Device Exemption (IDE) 2. HeartMate 3 Sealed Outflow Graft with Bend Relief, US 3. HeartMate 3 Sealed Outflow Graft with Bend Relief, OUS 4. HeartMate 3 LVAS Implant Kit, IDE 5. HeartMate 3 LVAS Implant Kit, US 6. HeartMate 3 LVAS Implant Kit, OUS HeartMate 3 Left Ventricular Assist System: The HeartMate 3 LVAS is designed to provide therapeutic benefit to those affected with advanced heart failure. The LVAS assumes some or all of the workload of the left ventricle. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus supporting equipment.

Product Codes/Lot Numbers:

HeartMate 3 LVAS Model Numbers: 1. 105581 (IDE, cannot be implanted without a Left Ventricular Assist Device (LVAD)) 2. 105581US (cannot be implanted without a LVAD) 3. 105581INT (cannot be implanted without a LVAD) 4. 106524 (IDE) 5. 106524US 6. 106524INT UDI-DI codes: HeartMate 3 LVAS: 1. N/A (IDE) 2. 00813024013266 3. 00813024011675 4. N/A (IDE) 5. 00813024013297 6. 00813024011712 Serial Numbers: For Model 106524 US MLP-000991 MLP-002058 MLP-002066 MLP-002069 MLP-002070 MLP-002076 MLP-002117 MLP-002126 MLP-002128 MLP-002133 MLP-002136 MLP-002152 MLP-002158 MLP-002159 MLP-002160 MLP-002164 MLP-002165 MLP-002191 MLP-002192 MLP-002197 MLP-002206 MLP-002208 MLP-002216 MLP-002217 MLP-002239 MLP-002241 MLP-002246 MLP-002251 MLP-002254 MLP-002259 MLP-002263 MLP-002268 MLP-002269 MLP-002278 MLP-002288 MLP-002289 MLP-002291 MLP-002294 MLP-002300 MLP-002304 MLP-002309 MLP-002310 MLP-002316 MLP-002318 MLP-002321 MLP-002324 MLP-002337 MLP-002342 MLP-002347 MLP-002355 MLP-002356 MLP-002361 MLP-002365 MLP-002450 MLP-002455 MLP-002456 MLP-002472 MLP-002475 MLP-002510 MLP-002514 MLP-002518 MLP-002519 MLP-002520 MLP-002530 MLP-002536 MLP-002538 MLP-002539 MLP-002543 MLP-002589 MLP-002597 MLP-002603 MLP-002633 MLP-002643 MLP-002646 MLP-002650 MLP-002654 MLP-002660 MLP-002663 MLP-002666 MLP-002805 MLP-002807 MLP-002808 MLP-002809 MLP-002827 MLP-002828 MLP-002838 MLP-002840 MLP-002842 MLP-002844 MLP-002850 MLP-002858 MLP-002957 MLP-003028 MLP-003033 MLP-003057 MLP-003058 MLP-003068 MLP-003070 MLP-003072 MLP-003074 MLP-003077 MLP-003085 MLP-003086 MLP-003089 MLP-003097 MLP-003110 MLP-003114 MLP-003125 MLP-003130 MLP-003131 MLP-003140 MLP-003148 MLP-003186 MLP-003237 MLP-003255 MLP-003257 MLP-003259 MLP-003277 MLP-003291 MLP-003299 MLP-003300 MLP-003302 MLP-003307 MLP-003308 MLP-003314 MLP-003317 MLP-003340 MLP-003347 MLP-003348 MLP-003351 MLP-003352 MLP-003384 MLP-003385 MLP-003388 MLP-003389 MLP-003390 MLP-003441 MLP-003442 MLP-003461 MLP-003466 MLP-003469 MLP-003471 MLP-003571 MLP-003572 MLP-003574 MLP-003577 MLP-003578 MLP-003582 MLP-003588 MLP-003615 MLP-003616 MLP-003617 MLP-003621 MLP-003629 MLP-003654 MLP-003659 MLP-003661 MLP-003689 MLP-003692 MLP-003738 MLP-003741 MLP-003828 MLP-003829 MLP-003858 MLP-003861 MLP-003862 MLP-003863 MLP-003867 MLP-004093 MLP-004110 MLP-004118 MLP-004119 MLP-004125 MLP-004129 MLP-004131 MLP-004137 MLP-004148 MLP-004151 MLP-004153 MLP-004156 MLP-004157 MLP-004165 MLP-004166 MLP-004171 MLP-004212 MLP-004242 MLP-004247 MLP-004267 MLP-004279 MLP-004281 MLP-004293 MLP-004298 MLP-004310 MLP-004384 MLP-004469 MLP-004486 MLP-004490 MLP-004525 MLP-004584 MLP-004591 MLP-004667 MLP-004675 MLP-004686 MLP-004697 MLP-004802 MLP-004805 MLP-004806 MLP-004866 MLP-004867 MLP-004869 MLP-004870 MLP-004874 MLP-004876 MLP-004877 MLP-004878 MLP-004884 MLP-004885 MLP-004887 MLP-004893 MLP-004954 MLP-004962 MLP-004963 MLP-004968 MLP-004969 MLP-004971 MLP-004983 MLP-004996 MLP-004998 MLP-005006 MLP-005032 MLP-005033 MLP-005036 MLP-005050 MLP-005052 MLP-005054 MLP-005083 MLP-005117 MLP-005122 MLP-005144 MLP-005145 MLP-005183 MLP-005186 MLP-005187 MLP-005193 MLP-005208 MLP-005212 MLP-005215 MLP-005219 MLP-005228 MLP-005258 MLP-005260 MLP-005261 MLP-005291 MLP-005292 MLP-005293 MLP-005298 MLP-005299 MLP-005320 MLP-005322 MLP-005323 MLP-005324 MLP-005328 MLP-005337 MLP-005391 MLP-005393 MLP-005396 MLP-005398 MLP-005402 MLP-005410 MLP-005411 MLP-005414 MLP-005416 MLP-005438 MLP-005440 MLP-005444 MLP-005460 MLP-005468 MLP-005483 MLP-005484 MLP-005485 MLP-005492 MLP-005494 MLP-005497 MLP-005498 MLP-005512 MLP-005531 MLP-005532 MLP-005539 MLP-005540 MLP-005541 MLP-005544 MLP-005552 MLP-005555 MLP-005559 MLP-005576 MLP-005594 MLP-005597 MLP-005600 MLP-005602 MLP-005613 MLP-005620 MLP-005636 MLP-005638 MLP-005641 MLP-005650 MLP-005653 MLP-005655 MLP-005663 MLP-005668 MLP-005680 MLP-005684 MLP-005703 MLP-005704 MLP-005706 MLP-005708 MLP-005729 MLP-005730 MLP-005746 MLP-005747 MLP-005752 MLP-005754 MLP-005769 MLP-005791 MLP-005793 MLP-005794 MLP-005795 MLP-005798 MLP-005812 MLP-005820 MLP-005821 MLP-005822 MLP-005825 MLP-005828 MLP-005831 MLP-005835 MLP-005836 MLP-005838 MLP-005853 MLP-005871 MLP-005879 MLP-005901 MLP-005902 MLP-005924 MLP-005934 MLP-005967 MLP-006005 MLP-006040 MLP-006041 MLP-006053 MLP-006067 MLP-006127 MLP-006128 MLP-006192 MLP-006193 MLP-006203 MLP-006204 MLP-006207 MLP-006216 MLP-006240 MLP-006245 MLP-006246 MLP-006249 MLP-006294 MLP-006303 MLP-006304 MLP-006390 MLP-006412 MLP-006413 MLP-006423 MLP-006431 MLP-006435 MLP-006437 MLP-006450 MLP-006452 MLP-006463 MLP-006470 MLP-006471 MLP-006477 MLP-006489 MLP-006490 MLP-006534 MLP-006535 MLP-006538 MLP-006539 MLP-006553 MLP-006571 MLP-006577 MLP-006578 MLP-006585 MLP-006587 MLP-006588 MLP-006595 MLP-006629 MLP-006631 MLP-006632 MLP-006633 MLP-006661 MLP-006663 MLP-006675 MLP-006677 MLP-006678 MLP-006688 MLP-006690 MLP-006710 MLP-006711 MLP-006715 MLP-006773 MLP-006801 MLP-006811 MLP-006827 MLP-006829 MLP-006831 MLP-006832 MLP-006846 MLP-006847 MLP-006848 MLP-006849 MLP-006851 MLP-006852 MLP-007250 MLP-007381 MLP-007388 MLP-007391 MLP-007396 MLP-007399 MLP-007400 MLP-007405 MLP-007408 MLP-007417 MLP-007420 MLP-007422 MLP-007428 MLP-007433 MLP-007444 MLP-007446 MLP-007447 MLP-007451 MLP-007459 MLP-007870 MLP-007871 MLP-007897 MLP-007905 MLP-007908 MLP-007921 MLP-007923 MLP-007926 MLP-007928 MLP-007929 MLP-007983 MLP-007985 MLP-007987 MLP-007988 MLP-007996 MLP-007997 MLP-007999 MLP-008005 MLP-008084 MLP-008085 MLP-008118 MLP-008119 MLP-008120 MLP-008161 MLP-008162 MLP-008165 MLP-008191 MLP-008194 MLP-008197 MLP-008199 MLP-008200 MLP-008250 MLP-008251 MLP-008252 MLP-008254 MLP-008255 MLP-008258 MLP-008266 MLP-008295 MLP-008296 MLP-008298 MLP-008302 MLP-008306 MLP-008308 MLP-008328 MLP-008329 MLP-008330 MLP-008385 MLP-008387 MLP-008388 MLP-008391 MLP-008392 MLP-008400 MLP-008401 MLP-008402 MLP-008403 MLP-008416 MLP-008417 MLP-008419 MLP-008425 MLP-008430 MLP-008445 MLP-008446 MLP-008524 MLP-008526 MLP-008528 MLP-008529 MLP-008550 MLP-008554 MLP-008556 MLP-008563 MLP-008565 MLP-008566 MLP-008576 MLP-008593 MLP-008606 MLP-008609 MLP-008651 MLP-008654 MLP-008656 MLP-008658 MLP-008659 MLP-008660 MLP-008661 MLP-008679 MLP-008681 MLP-008683 MLP-008809 MLP-008810 MLP-008811 MLP-008817 MLP-008819 MLP-008908 MLP-008911 MLP-008919 MLP-008928 MLP-008932 MLP-008933 MLP-008937 MLP-008939 MLP-009011 MLP-009012 MLP-009123 MLP-009124 MLP-009125 MLP-009126 MLP-009132 MLP-009153 MLP-009228 MLP-009259 MLP-009265 MLP-009273 MLP-009274 MLP-009275 MLP-009278 MLP-009280 MLP-009284 MLP-009285 MLP-009286 MLP-009287 MLP-009297 MLP-009308 MLP-009310 MLP-009311 MLP-009313 MLP-009315 MLP-009329 MLP-009330 MLP-009334 MLP-009335 MLP-009336 MLP-009341 MLP-009342 MLP-009357 MLP-009360 MLP-009364 MLP-009365 MLP-009368 MLP-009399 MLP-009413 MLP-009414 MLP-009416 MLP-009418 MLP-009423 MLP-009427 MLP-009429 MLP-009438 MLP-009442 MLP-009443 MLP-009448 MLP-009451 MLP-009453 MLP-009455 MLP-009456 MLP-009483 MLP-009491 MLP-009503 MLP-009504 MLP-009505 MLP-009506 MLP-009510 MLP-009521 MLP-009522 MLP-009528 MLP-009543 MLP-009647 MLP-009648 MLP-009649 MLP-009651 MLP-009652 MLP-009654 MLP-009655 MLP-009675 MLP-009677 MLP-009678 MLP-009681 MLP-009682 MLP-009695 MLP-009698 MLP-009699 MLP-009712 MLP-009732 MLP-009736 MLP-009741 MLP-009742 MLP-009751 MLP-009752 MLP-009753 MLP-009754 MLP-009755 MLP-009757 MLP-009758 MLP-009760 MLP-009765 MLP-009774 MLP-009778 MLP-009779 MLP-009783 MLP-009785 MLP-009790 MLP-009792 MLP-009796 MLP-009800 MLP-009803 MLP-009804 MLP-009810 MLP-009815 MLP-009816 MLP-009820 MLP-009822 MLP-009825 MLP-009826 MLP-009832 MLP-009837 MLP-009839 MLP-009865 MLP-009867 MLP-009868 MLP-009870 MLP-009872 MLP-009873 MLP-009910 MLP-009911 MLP-009915 MLP-009916 MLP-009921 MLP-009922 MLP-009925 MLP-009926 MLP-009927 MLP-009929 MLP-009980 MLP-009981 MLP-009995 MLP-009996 MLP-010001 MLP-010003 MLP-010006 MLP-010032 MLP-010034 MLP-010046 MLP-010047 MLP-010048 MLP-010065 MLP-010068 MLP-010105 MLP-010120 MLP-010121 MLP-010166 MLP-010167 MLP-010179 MLP-010180 MLP-010182 MLP-010183 MLP-010199 MLP-010276 MLP-010280 MLP-010288 MLP-010306 MLP-010317 MLP-010385 MLP-010386 MLP-010392 MLP-010406 MLP-010412 MLP-010424 MLP-010435 MLP-010437 MLP-010452 MLP-010453 MLP-010465 MLP-010473 MLP-010474 MLP-010485 MLP-010504 MLP-010573 MLP-010574 MLP-010577 MLP-010578 MLP-010609 MLP-010776 MLP-010847 MLP-010848 MLP-010849 MLP-010852 MLP-010853 MLP-010860 MLP-010861 MLP-010906 MLP-010924 MLP-010938 MLP-010939 MLP-010940 MLP-010941 MLP-010942 MLP-010945 MLP-010946 MLP-010952 MLP-010953 MLP-010955 MLP-010957 MLP-010970 MLP-011066 MLP-011070 MLP-011088 MLP-011117 MLP-011118 MLP-011120 MLP-011123 MLP-011124 MLP-011125 MLP-011131 MLP-011136 MLP-011137 MLP-011138 MLP-011140 MLP-011190 MLP-011191 MLP-011193 MLP-011336 MLP-011370 MLP-011371 MLP-011372 MLP-011542 MLP-011545 MLP-011547 MLP-011550 MLP-011552 MLP-011647 MLP-011654 MLP-011656 MLP-011657 MLP-011658 MLP-011660 MLP-011661 MLP-011701 MLP-011702 MLP-011704 MLP-012115 MLP-012117 MLP-012121 MLP-012123 MLP-012145 MLP-012160 MLP-012180 MLP-012182 MLP-012183 MLP-012184 MLP-012185 MLP-012189 MLP-012190 MLP-012191 MLP-012193 MLP-012199 MLP-012266 MLP-012267 MLP-012268 MLP-012303 MLP-012334 MLP-012338 MLP-012349 MLP-012377 MLP-012405 MLP-012407 MLP-012408 MLP-012410 MLP-012420 MLP-012421 MLP-012612 MLP-012618 MLP-012619 MLP-012629 MLP-012631 MLP-012633 MLP-012636 MLP-012667 MLP-012745 OUS MLP-002000 MLP-002003 MLP-002005 MLP-002015 MLP-002019 MLP-002022 MLP-002023 MLP-002024 MLP-002026 MLP-002027 MLP-002028 MLP-002043 MLP-002050 MLP-002053 MLP-002055 MLP-002064 MLP-002095 MLP-002097 MLP-002098 MLP-002101 MLP-002103 MLP-002105 MLP-002106 MLP-002111 MLP-002113 MLP-002114 MLP-002115 MLP-002116 MLP-002122 MLP-002143 MLP-002145 MLP-002147 MLP-002150 MLP-002153 MLP-002186 MLP-002190 MLP-002194 MLP-002230 MLP-002322 MLP-002323 MLP-002328 MLP-006360 MLP-006364 MLP-007409 MLP-007873 MLP-007917 MLP-007918 MLP-008122 MLP-008123 MLP-009290 MLP-009372 MLP-009420 For Model 106524US US MLP-006853 MLP-006869 MLP-007411 MLP-007412 MLP-007445 MLP-007465 MLP-007468 MLP-007474 MLP-007475 MLP-007504 MLP-007508 MLP-007510 MLP-007538 MLP-007584 MLP-007585 MLP-007590 MLP-007604 MLP-007611 MLP-007625 MLP-007628 MLP-007630 MLP-007631 MLP-007635 MLP-007637 MLP-007642 MLP-007643 MLP-007657 MLP-007658 MLP-007660 MLP-007662 MLP-007670 MLP-007672 MLP-007678 MLP-007679 MLP-007682 MLP-007685 MLP-007686 MLP-007687 MLP-007690 MLP-007702 MLP-007705 MLP-007711 MLP-007716 MLP-007717 MLP-007719 MLP-007765 MLP-007766 MLP-007767 MLP-007768 MLP-007773 MLP-007774 MLP-007775 MLP-007777 MLP-007780 MLP-007792 MLP-007801 MLP-007802 MLP-007803 MLP-007823 MLP-007826 MLP-007835 MLP-007837 MLP-007874 MLP-007876 MLP-007880 MLP-007883 MLP-007888 MLP-007889 MLP-007909 MLP-007912 MLP-007913 MLP-007915 MLP-007916 MLP-007936 MLP-007938 MLP-007943 MLP-008023 MLP-008024 MLP-008028 MLP-008046 MLP-008049 MLP-008052 MLP-008054 MLP-008068 MLP-008069 MLP-008079 MLP-008080 MLP-008081 MLP-008082 MLP-008089 MLP-008104 MLP-008110 MLP-008111 MLP-008113 MLP-008116 MLP-008117 MLP-008121 MLP-008129 MLP-008130 MLP-008131 MLP-008138 MLP-008140 MLP-008164 MLP-008167 MLP-008168 MLP-008201 MLP-008204 MLP-008205 MLP-008208 MLP-008210 MLP-008217 MLP-008220 MLP-008222 MLP-008247 MLP-008270 MLP-008277 MLP-008284 MLP-008303 MLP-008312 MLP-008314 MLP-008346 MLP-008350 MLP-008351 MLP-008353 MLP-008359 MLP-008363 MLP-008366 MLP-008380 MLP-008381 MLP-008426 MLP-008429 MLP-008442 MLP-008464 MLP-008469 MLP-008503 MLP-008506 MLP-008507 MLP-008510 MLP-008516 MLP-008517 MLP-008521 MLP-008535 MLP-008541 MLP-008542 MLP-008548 MLP-008552 MLP-008557 MLP-008559 MLP-008578 MLP-008582 MLP-008586 MLP-008588 MLP-008611 MLP-008615 MLP-008618 MLP-008619 MLP-008621 MLP-008623 MLP-008647 MLP-008650 MLP-008766 MLP-008773 MLP-008796 MLP-008799 MLP-008813 MLP-008815 MLP-008824 MLP-008827 MLP-008831 MLP-008832 MLP-008839 MLP-008842 MLP-008843 MLP-008846 MLP-008847 MLP-008879 MLP-008882 MLP-008885 MLP-008886 MLP-008888 MLP-008890 MLP-008898 MLP-008900 MLP-008902 MLP-008903 MLP-008914 MLP-008923 MLP-008929 MLP-008963 MLP-008964 MLP-008966 MLP-008970 MLP-008971 MLP-008973 MLP-008975 MLP-008976 MLP-008981 MLP-008982 MLP-008989 MLP-008991 MLP-008994 MLP-008996 MLP-009004 MLP-009014 MLP-009044 MLP-009046 MLP-009048 MLP-009061 MLP-009067 MLP-009068 MLP-009069 MLP-009071 MLP-009075 MLP-009076 MLP-009079 MLP-009080 MLP-009086 MLP-009089 MLP-009090 MLP-009091 MLP-009093 MLP-009105 MLP-009152 MLP-009156 MLP-009164 MLP-009166 MLP-009168 MLP-009170 MLP-009177 MLP-009178 MLP-009179 MLP-009188 MLP-009190 MLP-009192 MLP-009194 MLP-009225 MLP-009271 MLP-009272 MLP-009293 MLP-009296 MLP-009301 MLP-009302 MLP-009304 MLP-009316 MLP-009366 MLP-009385 MLP-009386 MLP-009388 MLP-009394 MLP-009400 MLP-009408 MLP-009424 MLP-009428 MLP-009431 MLP-009432 MLP-009490 MLP-009493 MLP-009501 MLP-009502 MLP-009511 MLP-009517 MLP-009524 MLP-009529 MLP-009534 MLP-009547 MLP-009549 MLP-009550 MLP-009566 MLP-009568 MLP-009589 MLP-009590 MLP-009591 MLP-009593 MLP-009598 MLP-009604 MLP-009610 MLP-009630 MLP-009632 MLP-009637 MLP-009640 MLP-009641 MLP-009685 MLP-009688 MLP-009691 MLP-009694 MLP-009700 MLP-009713 MLP-009714 MLP-009715 MLP-009746 MLP-009786 MLP-009791 MLP-009821 MLP-009831 MLP-009833 MLP-009838 MLP-009874 MLP-009896 MLP-009897 MLP-009900 MLP-009912 MLP-009917 MLP-009933 MLP-009938 MLP-009939 MLP-009940 MLP-009941 MLP-009943 MLP-009971 MLP-009974 MLP-009993 MLP-010011 MLP-010012 MLP-010025 MLP-010030 MLP-010053 MLP-010055 MLP-010057 MLP-010060 MLP-010066 MLP-010095 MLP-010103 MLP-010106 MLP-010119 MLP-010123 MLP-010125 MLP-010127 MLP-010130 MLP-010133 MLP-010134 MLP-010135 MLP-010136 MLP-010149 MLP-010173 MLP-010202 MLP-010204 MLP-010209 MLP-010211 MLP-010283 MLP-010286 MLP-010293 MLP-010305 MLP-010310 MLP-010312 MLP-010313 MLP-010321 MLP-010335 MLP-010338 MLP-010342 MLP-010344 MLP-010347 MLP-010348 MLP-010349 MLP-010350 MLP-010373 MLP-010374 MLP-010378 MLP-010380 MLP-010381 MLP-010384 MLP-010390 MLP-010391 MLP-010416 MLP-010419 MLP-010436 MLP-010445 MLP-010451 MLP-010468 MLP-010469 MLP-010472 MLP-010476 MLP-010498 MLP-010500 MLP-010505 MLP-010507 MLP-010548 MLP-010556 MLP-010583 MLP-010586 MLP-010591 MLP-010592 MLP-010611 MLP-010615 MLP-010618 MLP-010620 MLP-010621 MLP-010623 MLP-010624 MLP-010645 MLP-010646 MLP-010647 MLP-010650 MLP-010651 MLP-010657 MLP-010660 MLP-010663 MLP-010692 MLP-010767 MLP-010773 MLP-010774 MLP-010775 MLP-010779 MLP-010790 MLP-010796 MLP-010798 MLP-010800 MLP-010817 MLP-010819 MLP-010820 MLP-010844 MLP-010846 MLP-010855 MLP-010870 MLP-010872 MLP-010877 MLP-010878 MLP-010902 MLP-010908 MLP-010925 MLP-010933 MLP-010934 MLP-010935 MLP-010937 MLP-010964 MLP-010967 MLP-010968 MLP-010975 MLP-010976 MLP-010977 MLP-010981 MLP-010985 MLP-010988 MLP-011000 MLP-011001 MLP-011003 MLP-011004 MLP-011005 MLP-011009 MLP-011012 MLP-011013 MLP-011015 MLP-011016 MLP-011017 MLP-011018 MLP-011025 MLP-011027 MLP-011030 MLP-011115 MLP-011143 MLP-011181 MLP-011184 MLP-011185 MLP-011211 MLP-011213 MLP-011334 MLP-011341 MLP-011378 MLP-011454 MLP-011480 MLP-011481 MLP-011484 MLP-011515 MLP-011516 MLP-011517 MLP-011532 MLP-011544 MLP-011554 MLP-011556 MLP-011557 MLP-011559 MLP-011561 MLP-011563 MLP-011611 MLP-011644 MLP-011645 MLP-011646 MLP-011662 MLP-011673 MLP-011676 MLP-011677 MLP-011679 MLP-011687 MLP-011709 MLP-011710 MLP-011722 MLP-011883 MLP-011889 MLP-011901 MLP-011906 MLP-011907 MLP-011909 MLP-011928 MLP-011931 MLP-011933 MLP-011934 MLP-011935 MLP-011939 MLP-011941 MLP-011943 MLP-011951 MLP-011955 MLP-012063 MLP-012068 MLP-012076 MLP-012105 MLP-012116 MLP-012119 MLP-012120 MLP-012128 MLP-012133 MLP-012134 MLP-012138 MLP-012144 MLP-012159 MLP-012165 MLP-012181 MLP-012217 MLP-012230 MLP-012272 MLP-012273 MLP-012276 MLP-012278 MLP-012279 MLP-012281 MLP-012284 MLP-012285 MLP-012288 MLP-012293 MLP-012294 MLP-012308 MLP-012309 MLP-012311 MLP-012315 MLP-012330 MLP-012336 MLP-012339 MLP-012340 MLP-012341 MLP-012342 MLP-012348 MLP-012354 MLP-012357 MLP-012362 MLP-012364 MLP-012365 MLP-012367 MLP-012370 MLP-012381 MLP-012382 MLP-012383 MLP-012412 MLP-012422 MLP-012471 MLP-012473 MLP-012474 MLP-012477 MLP-012487 MLP-012489 MLP-012490 MLP-012491 MLP-012493 MLP-012494 MLP-012500 MLP-012501 MLP-012521 MLP-012533 MLP-012535 MLP-012536 MLP-012539 MLP-012546 MLP-012549 MLP-012550 MLP-012551 MLP-012552 MLP-012553 MLP-012555 MLP-012556 MLP-012557 MLP-012559 MLP-012560 MLP-012569 MLP-012574 MLP-012580 MLP-012584 MLP-012585 MLP-012586 MLP-012588 MLP-012589 MLP-012590 MLP-012591 MLP-012592 MLP-012604 MLP-012605 MLP-012614 MLP-012615 MLP-012616 MLP-012710 MLP-012723 MLP-012741 MLP-012749 MLP-012755 MLP-012756 MLP-012757 MLP-012760 MLP-012763 MLP-012766 MLP-012769 MLP-012770 MLP-012771 MLP-012773 MLP-012775 MLP-012776 MLP-012778 MLP-012781 MLP-012788 MLP-012789 MLP-012790 MLP-012863 MLP-012864 MLP-012866 MLP-012869 MLP-012871 MLP-012918 MLP-012919 MLP-012923 MLP-012925 MLP-012926 MLP-012929 MLP-012933 MLP-012963 MLP-012967 MLP-012968 MLP-012970 MLP-012973 MLP-012975 MLP-012976 MLP-012981 MLP-012983 MLP-012986 MLP-012991 MLP-012992 MLP-012994 MLP-012996 MLP-013006 MLP-013008 MLP-013009 MLP-013012 MLP-013017 MLP-013020 MLP-013021 MLP-013023 MLP-013024 MLP-013029 MLP-013033 MLP-013034 MLP-013035 MLP-013036 MLP-013041 MLP-013044 MLP-013046 MLP-013047 MLP-013064 MLP-013065 MLP-013066 MLP-013069 MLP-013070 MLP-013072 MLP-013073 MLP-013074 MLP-013076 MLP-013079 MLP-013081 MLP-013083 MLP-013084 MLP-013085 MLP-013086 MLP-013087 MLP-013089 MLP-013092 MLP-013103 MLP-013104 MLP-013106 MLP-013131 MLP-013133 MLP-013135 MLP-013136 MLP-013157 MLP-013158 MLP-013159 MLP-013168 MLP-013170 MLP-013176 MLP-013181 MLP-013187 MLP-013189 MLP-013191 MLP-013195 MLP-013212 MLP-013241 MLP-013251 MLP-013256 MLP-013257 MLP-013261 MLP-013262 MLP-013269 MLP-013271 MLP-013273 MLP-013274 MLP-013314 MLP-013324 MLP-013327 MLP-013400 MLP-013406 MLP-013417 MLP-013418 MLP-013420 MLP-013421 MLP-013422 MLP-013424 MLP-013425 MLP-013426 MLP-013429 MLP-013432 MLP-013434 MLP-013436 MLP-013446 MLP-013452 MLP-013464 MLP-013466 MLP-013477 MLP-013493 MLP-013502 MLP-013512 MLP-013514 MLP-013517 MLP-013518 MLP-013535 MLP-013542 MLP-013544 MLP-013545 MLP-013550 MLP-013558 MLP-013563 MLP-013565 MLP-013568 MLP-013573 MLP-013576 MLP-013586 MLP-013587 MLP-013589 MLP-013617 MLP-013625 MLP-013626 MLP-013627 MLP-013633 MLP-013635 MLP-013637 MLP-013641 MLP-013652 MLP-013657 MLP-013660 MLP-013662 MLP-013663 MLP-013664 MLP-013665 MLP-013669 MLP-013672 MLP-013679 MLP-013691 MLP-013707 MLP-013711 MLP-013712 MLP-013714 MLP-013737 MLP-013738 MLP-013746 MLP-013759 MLP-013761 MLP-013762 MLP-013763 MLP-013765 MLP-013768 MLP-013772 MLP-013776 MLP-013778 MLP-013783 MLP-013784 MLP-013788 MLP-013790 MLP-013791 MLP-013793 MLP-013798 MLP-013799 MLP-013800 MLP-013801 MLP-013802 MLP-013812 MLP-013817 MLP-013819 MLP-013824 MLP-013826 MLP-013827 MLP-013829 MLP-013830 MLP-013832 MLP-013852 MLP-013889 MLP-013890 MLP-013891 MLP-013893 MLP-013895 MLP-013897 MLP-013900 MLP-013906 MLP-013907 MLP-013913 MLP-013915 MLP-013921 MLP-013922 MLP-013933 MLP-013935 MLP-013938 MLP-013940 MLP-013942 MLP-013946 MLP-013947 MLP-013951 MLP-013953 MLP-013958 MLP-013960 MLP-013962 MLP-013963 MLP-013967 MLP-013968 MLP-013971 MLP-013973 MLP-013976 MLP-013988 MLP-013989 MLP-013992 MLP-013993 MLP-013997 MLP-013998 MLP-014002 MLP-014003 MLP-014004 MLP-014006 MLP-014007 MLP-014008 MLP-014009 MLP-014014 MLP-014018 MLP-014019 MLP-014021 MLP-014022 MLP-014027 MLP-014063 MLP-014070 MLP-014075 MLP-014076 MLP-014097 MLP-014099 MLP-014101 MLP-014104 MLP-014106 MLP-014111 MLP-014118 MLP-014119 MLP-014139 MLP-014140 MLP-014142 MLP-014148 MLP-014154 MLP-014161 MLP-014166 MLP-014169 MLP-014172 MLP-014174 MLP-014175 MLP-014176 MLP-014177 MLP-014178 MLP-014188 MLP-014189 MLP-014194 MLP-014195 MLP-014200 MLP-014201 MLP-014202 MLP-014203 MLP-014204 MLP-014206 MLP-014207 MLP-014209 MLP-014210 MLP-014211 MLP-014214 MLP-014219 MLP-014235 MLP-014243 MLP-014246 MLP-014253 MLP-014255 MLP-014275 MLP-014276 MLP-014278 MLP-014279 MLP-014281 MLP-014283 MLP-014327 MLP-014328 MLP-014333 MLP-014334 MLP-014337 MLP-014339 MLP-014342 MLP-014370 MLP-014371 MLP-014375 MLP-014376 MLP-014381 MLP-014383 MLP-014384 MLP-014385 MLP-014389 MLP-014409 MLP-014411 MLP-014412 MLP-014426 MLP-014428 MLP-014429 MLP-014430 MLP-014433 MLP-014434 MLP-014437 MLP-014438 MLP-014443 MLP-014477 MLP-014486 MLP-014488 MLP-014489 MLP-014493 MLP-014508 MLP-014509 MLP-014510 MLP-014512 MLP-014513 MLP-014520 MLP-014546 MLP-014548 MLP-014638 MLP-014640 MLP-014646 MLP-014650 MLP-014652 MLP-014654 MLP-014655 MLP-014657 MLP-014660 MLP-014664 MLP-014668 MLP-014673 MLP-014674 MLP-014677 MLP-014678 MLP-014679 MLP-014682 MLP-014684 MLP-014687 MLP-014688 MLP-014689 MLP-014700 MLP-014702 MLP-014705 MLP-014706 MLP-014735 MLP-014737 MLP-014738 MLP-014739 MLP-014741 MLP-014743 MLP-014749 MLP-014750 MLP-014755 MLP-014756 MLP-014769 MLP-014770 MLP-014772 MLP-014774 MLP-014777 MLP-014778 MLP-014779 MLP-014796 MLP-014798 MLP-014800 MLP-014801 MLP-014803 MLP-014804 MLP-014806 MLP-014807 MLP-014810 MLP-014811 MLP-014814 MLP-014818 MLP-014821 MLP-014822 MLP-014827 MLP-014830 MLP-014833 MLP-014835 MLP-014836 MLP-014839 MLP-014840 MLP-014841 MLP-014849 MLP-014866 MLP-014868 MLP-014870 MLP-014873 MLP-014874 MLP-014875 MLP-014876 MLP-014880 MLP-014883 MLP-014886 MLP-014895 MLP-014896 MLP-014898 MLP-014900 MLP-014902 MLP-014904 MLP-014906 MLP-014913 MLP-014915 MLP-014916 MLP-014917 MLP-014922 MLP-014924 MLP-014925 MLP-014930 MLP-014932 MLP-014933 MLP-014934 MLP-014943 MLP-014944 MLP-014947 MLP-014948 MLP-014952 MLP-014954 MLP-014958 MLP-014959 MLP-014960 MLP-014961 MLP-014963 MLP-014964 MLP-015015 MLP-015016 MLP-015018 MLP-015019 MLP-015022 MLP-015024 MLP-015025 MLP-015026 MLP-015028 MLP-015029 MLP-015030 MLP-015031 MLP-015035 MLP-015036 MLP-015038 MLP-015040 MLP-015041 MLP-015042 MLP-015056 MLP-015057 MLP-015058 MLP-015061 MLP-015062 MLP-015063 MLP-015067 MLP-015068 MLP-015069 MLP-015070 MLP-015071 MLP-015073 MLP-015074 MLP-015075 MLP-015076 MLP-015077 MLP-015080 MLP-015083 MLP-015084 MLP-015088 MLP-015089 MLP-015092 MLP-015093 MLP-015099 MLP-015102 MLP-015104 MLP-015105 MLP-015106 MLP-015107 MLP-015117 MLP-015120 MLP-015123 MLP-015126 MLP-015128 MLP-015129 MLP-015130 MLP-015131 MLP-015133 MLP-015134 MLP-015136 MLP-015137 MLP-015138 MLP-015139 MLP-015140 MLP-015142 MLP-015146 MLP-015147 MLP-015151 MLP-015156 MLP-015161 MLP-015162 MLP-015163 MLP-015165 MLP-015168 MLP-015169 MLP-015174 MLP-015177 MLP-015180 MLP-015181 MLP-015182 MLP-015183 MLP-015184 MLP-015186 MLP-015187 MLP-015205 MLP-015214 MLP-015215 MLP-015217 MLP-015218 MLP-015224 MLP-015242 MLP-015243 MLP-015244 MLP-015246 MLP-015248 MLP-015252 MLP-015255 MLP-015257 MLP-015261 MLP-015276 MLP-015281 MLP-015293 MLP-015299 MLP-015301 MLP-015305 MLP-015306 MLP-015307 MLP-015316 MLP-015320 MLP-015321 MLP-015324 MLP-015325 MLP-015343 MLP-015355 MLP-015358 MLP-015371 MLP-015373 MLP-015381 MLP-015383 MLP-015384 MLP-015386 MLP-015389 MLP-015392 MLP-015393 MLP-015395 MLP-015396 MLP-015399 MLP-015400 MLP-015401 MLP-015405 MLP-015413 MLP-015417 MLP-015422 MLP-015424 MLP-015434 MLP-015436 MLP-015443 MLP-015451 MLP-015463 MLP-015464 MLP-015475 MLP-015504 MLP-015505 MLP-015508 MLP-015509 MLP-015512 MLP-015513 MLP-015515 MLP-015517 MLP-015519 MLP-015521 MLP-015523 MLP-015535 MLP-015536 MLP-015540 MLP-015541 MLP-015547 MLP-015556 MLP-015558 MLP-015559 MLP-015566 MLP-015567 MLP-015570 MLP-015572 MLP-015573 MLP-015574 MLP-015580 MLP-015581 MLP-015582 MLP-015584 MLP-015585 MLP-015589 MLP-015590 MLP-015592 MLP-015594 MLP-015596 MLP-015598 MLP-015600 MLP-015602 MLP-015604 MLP-015605 MLP-015606 MLP-015607 MLP-015611 MLP-015631 MLP-015648 MLP-015650 MLP-015651 MLP-015652 MLP-015653 MLP-015654 MLP-015655 MLP-015657 MLP-015658 MLP-015660 MLP-015661 MLP-015674 MLP-015685 MLP-015687 MLP-015688 MLP-015690 MLP-015691 MLP-015696 MLP-015699 MLP-015706 MLP-015709 MLP-015712 MLP-015713 MLP-015716 MLP-015721 MLP-015722 MLP-015729 MLP-015748 MLP-015749 MLP-015750 MLP-015752 MLP-015753 MLP-015756 MLP-015758 MLP-015769 MLP-015774 MLP-015778 MLP-015811 MLP-015812 MLP-015818 MLP-015819 MLP-015824 MLP-015825 MLP-015827 MLP-015829 MLP-015841 MLP-015842 MLP-015843 MLP-015877 MLP-015879 MLP-015881 MLP-015883 MLP-015897 MLP-015902 MLP-015903 MLP-015904 MLP-015905 MLP-015906 MLP-015908 MLP-015909 MLP-015910 MLP-015915 MLP-015916 MLP-015920 MLP-015922 MLP-015931 MLP-015932 MLP-015933 MLP-015936 MLP-015939 MLP-015940 MLP-015941 MLP-015942 MLP-015946 MLP-015947 MLP-015949 MLP-015950 MLP-015967 MLP-015973 MLP-015976 MLP-015978 MLP-015980 MLP-015988 MLP-015991 MLP-015994 MLP-015998 MLP-015999 MLP-016000 MLP-016001 MLP-016002 MLP-016004 MLP-016011 MLP-016012 MLP-016014 MLP-016016 MLP-016018 MLP-016019 MLP-016020 MLP-016030 MLP-016033 MLP-016035 MLP-016036 MLP-016038 MLP-016040 MLP-016058 MLP-016067 MLP-016071 MLP-016078 MLP-016080 MLP-016081 MLP-016083 MLP-016084 MLP-016085 MLP-016102 MLP-016103 MLP-016104 MLP-016105 MLP-016107 MLP-016109 MLP-016110 MLP-016115 MLP-016118 MLP-016119 MLP-016121 MLP-016122 MLP-016124 MLP-016126 MLP-016130 MLP-016135 MLP-016142 MLP-016144 MLP-016157 MLP-016158 MLP-016160 MLP-016161 MLP-016162 MLP-016174 MLP-016175 MLP-016183 MLP-016187 MLP-016192 MLP-016193 MLP-016200 MLP-016202 MLP-016204 MLP-016216 MLP-016217 MLP-016218 MLP-016230 MLP-016231 MLP-016232 MLP-016233 MLP-016234 MLP-016241 MLP-016244 MLP-016246 MLP-016248 MLP-016249 MLP-016251 MLP-016253 MLP-016254 MLP-016255 MLP-016272 MLP-016273 MLP-016275 MLP-016289 MLP-016292 MLP-016293 MLP-016297 MLP-016298 MLP-016300 MLP-016302 MLP-016303 MLP-016305 MLP-016306 MLP-016307 MLP-016308 MLP-016325 MLP-016327 MLP-016328 MLP-016331 MLP-016335 MLP-016336 MLP-016338 MLP-016339 MLP-016341 MLP-016342 MLP-016349 MLP-016363 MLP-016364 MLP-016382 MLP-016383 MLP-016384 MLP-016387 MLP-016388 MLP-016404 MLP-016405 MLP-016407 MLP-016409 MLP-016411 MLP-016419 MLP-016421 MLP-016424 MLP-016425 MLP-016428 MLP-016429 MLP-016430 MLP-016443 MLP-016446 MLP-016457 MLP-016458 MLP-016459 MLP-016460 MLP-016462 MLP-016480 MLP-016481 MLP-016482 MLP-016483 MLP-016485 MLP-016486 MLP-016495 MLP-016497 MLP-016498 MLP-016499 MLP-016502 MLP-016508 MLP-016509 MLP-016510 MLP-016511 MLP-016512 MLP-016513 MLP-016520 MLP-016543 MLP-016551 MLP-016552 MLP-016553 MLP-016554 MLP-016564 MLP-016565 MLP-016566 MLP-016567 MLP-016568 MLP-016572 MLP-016573 MLP-016577 MLP-016579 MLP-016582 MLP-016583 MLP-016677 MLP-016682 MLP-016684 MLP-016686 MLP-016687 MLP-016692 MLP-016693 MLP-016696 MLP-016700 MLP-016703 MLP-016708 MLP-016709 MLP-016710 MLP-016713 MLP-016714 MLP-016716 MLP-016718 MLP-016725 MLP-016728 MLP-016731 MLP-016742 MLP-016743 MLP-016744 MLP-016749 MLP-016751 MLP-016755 MLP-016756 MLP-016760 MLP-016761 MLP-016765 MLP-016780 MLP-016792 MLP-016794 MLP-016795 MLP-016810 MLP-016813 MLP-016828 MLP-016833 MLP-016834 MLP-016838 MLP-016841 MLP-016844 MLP-016847 MLP-016849 MLP-016853 MLP-016856 MLP-016861 MLP-016862 MLP-016864 MLP-016865 MLP-016866 MLP-016868 MLP-016873 MLP-016874 MLP-016896 MLP-016897 MLP-016898 MLP-016899 MLP-016906 MLP-016907 MLP-016909 MLP-016928 MLP-016929 MLP-016931 MLP-016932 MLP-016938 MLP-016939 MLP-016940 MLP-016941 MLP-016942 MLP-016952 MLP-016955 MLP-016961 MLP-016984 MLP-016985 MLP-016986 MLP-016987 MLP-016988 MLP-016990 MLP-016992 MLP-016994 MLP-017004 MLP-017005 MLP-017007 MLP-017008 MLP-017009 MLP-017010 MLP-017014 MLP-017015 MLP-017016 MLP-017017 MLP-017018 MLP-017021 MLP-017023 MLP-017024 MLP-017025 MLP-017026 MLP-017029 MLP-017030 MLP-017031 MLP-017034 MLP-017035 MLP-017036 MLP-017038 MLP-017039 MLP-017041 MLP-017042 MLP-017043 MLP-017047 MLP-017053 MLP-017054 MLP-017056 MLP-017058 MLP-017062 MLP-017063 MLP-017064 MLP-017065 MLP-017074 MLP-017079 MLP-017080 MLP-017081 MLP-017082 MLP-017083 MLP-017084 MLP-017086 MLP-017087 MLP-017089 MLP-017092 MLP-017093 MLP-017095 MLP-017097 MLP-017098 MLP-017100 MLP-017101 MLP-017102 MLP-017112 MLP-017114 MLP-017115 MLP-017120 MLP-017125 MLP-017127 MLP-017128 MLP-017130 MLP-017133 MLP-017134 MLP-017155 MLP-017158 MLP-017173 MLP-017174 MLP-017176 MLP-017179 MLP-017181 MLP-017182 MLP-017198 MLP-017201 MLP-017203 MLP-017204 MLP-017210 MLP-017213 MLP-017214 MLP-017215 MLP-017218 MLP-017219 MLP-017224 MLP-017225 MLP-017226 MLP-017260 MLP-017261 MLP-017288 MLP-017290 MLP-017291 MLP-017309 MLP-017310 MLP-017311 MLP-017317 MLP-017319 MLP-017321 MLP-017324 MLP-017332 MLP-017333 MLP-017349 MLP-017351 MLP-017353 MLP-017366 MLP-017371 MLP-017375 MLP-017377 MLP-017384 MLP-017385 MLP-017387 MLP-017389 MLP-017393 MLP-017394 MLP-017401 MLP-017402 MLP-017404 MLP-017436 MLP-017437 MLP-017438 MLP-017447 MLP-017449 MLP-017450 MLP-017453 MLP-017455 MLP-017459 MLP-017461 MLP-017462 MLP-017463 MLP-017470 MLP-017472 MLP-017473 MLP-017474 MLP-017475 MLP-017515 MLP-017516 MLP-017517 MLP-017520 MLP-017534 MLP-017540 MLP-017547 MLP-017548 MLP-017549 MLP-017571 MLP-017572 MLP-017573 MLP-017574 MLP-017576 MLP-017579 MLP-017580 MLP-017582 MLP-017583 MLP-017598 MLP-017600 MLP-017602 MLP-017605 MLP-017610 MLP-017613 MLP-017615 MLP-017618 MLP-017619 MLP-017620 MLP-017624 MLP-017625 MLP-017626 MLP-017627 MLP-017628 MLP-017632 MLP-017635 MLP-017639 MLP-017640 MLP-017641 MLP-017642 MLP-017653 MLP-017663 MLP-017664 MLP-017665 MLP-017667 MLP-017684 MLP-017685 MLP-017687 MLP-017690 MLP-017691 MLP-017695 MLP-017697 MLP-017698 MLP-017699 MLP-017711 MLP-017712 MLP-017718 MLP-017719 MLP-017720 MLP-017729 MLP-017735 MLP-017742 MLP-017743 MLP-017746 MLP-017747 MLP-017748 MLP-017749 MLP-017750 MLP-017751 MLP-017752 MLP-017754 MLP-017767 MLP-017770 MLP-017776 MLP-017779 MLP-017784 MLP-017787 MLP-017788 MLP-017789 MLP-017790 MLP-017793 MLP-017799 MLP-017800 MLP-017802 MLP-017803 MLP-017804 MLP-017805 MLP-017807 MLP-017808 MLP-017809 MLP-017816 MLP-017817 MLP-017818 MLP-017833 MLP-017836 MLP-017837 MLP-017840 MLP-017841 MLP-017842 MLP-017843 MLP-017844 MLP-017845 MLP-017849 MLP-017855 MLP-017858 MLP-017860 MLP-017861 MLP-017862 MLP-017864 MLP-017865 MLP-017867 MLP-017872 MLP-017883 MLP-017884 MLP-017885 MLP-017886 MLP-017887 MLP-017888 MLP-017889 MLP-017904 MLP-017905 MLP-017907 MLP-017910 MLP-017912 MLP-017933 MLP-017947 MLP-017949 MLP-017950 MLP-017952 MLP-017959 MLP-017960 MLP-017961 MLP-017962 MLP-017977 MLP-017979 MLP-017980 MLP-017982 MLP-017984 MLP-017992 MLP-017993 MLP-017997 MLP-017999 MLP-018002 MLP-018003 MLP-018006 MLP-018007 MLP-018009 MLP-018010 MLP-018013 MLP-018014 MLP-018015 MLP-018016 MLP-018017 MLP-018018 MLP-018020 MLP-018023 MLP-018034 MLP-018035 MLP-018047 MLP-018049 MLP-018052 MLP-018054 MLP-018055 MLP-018057 MLP-018059 MLP-018060 MLP-018061 MLP-018062 MLP-018069 MLP-018071 MLP-018089 MLP-018091 MLP-018095 MLP-018097 MLP-018105 MLP-018108 MLP-018120 MLP-018122 MLP-018124 MLP-018132 MLP-018135 MLP-018137 MLP-018138 MLP-018139 MLP-018140 MLP-018143 MLP-018155 MLP-018158 MLP-018160 MLP-018161 MLP-018162 MLP-018175 MLP-018180 MLP-018183 MLP-018187 MLP-018188 MLP-018191 MLP-018194 MLP-018195 MLP-018197 MLP-018201 MLP-018205 MLP-018213 MLP-018215 MLP-018218 MLP-018220 MLP-018221 MLP-018223 MLP-018226 MLP-018235 MLP-018237 MLP-018240 MLP-018245 MLP-018246 MLP-018258 MLP-018259 MLP-018261 MLP-018262 MLP-018263 MLP-018264 MLP-018266 MLP-018269 MLP-018272 MLP-018274 MLP-018276 MLP-018277 MLP-018278 MLP-018299 MLP-018320 MLP-018323 MLP-018327 MLP-018332 MLP-018334 MLP-018337 MLP-018340 MLP-018341 MLP-018351 MLP-018359 MLP-018360 MLP-018364 MLP-018367 MLP-018370 MLP-018376 MLP-018377 MLP-018383 MLP-018386 MLP-018401 MLP-018408 MLP-018413 MLP-018415 MLP-018416 MLP-018418 MLP-018421 MLP-018439 MLP-018441 MLP-018443 MLP-018445 MLP-018446 MLP-018447 MLP-018465 MLP-018466 MLP-018467 MLP-018470 MLP-018483 MLP-018484 MLP-018485 MLP-018486 MLP-018487 MLP-018489 MLP-018490 MLP-018492 MLP-018493 MLP-018494 MLP-018495 MLP-018496 MLP-018498 MLP-018500 MLP-018501 MLP-018511 MLP-018513 MLP-018514 MLP-018515 MLP-018517 MLP-018523 MLP-018524 MLP-018525 MLP-018527 MLP-018528 MLP-018531 MLP-018536 MLP-018539 MLP-018541 MLP-018542 MLP-018543 MLP-018544 MLP-018546 MLP-018547 MLP-018549 MLP-018550 MLP-018551 MLP-018553 MLP-018555 MLP-018560 MLP-018563 MLP-018564 MLP-018565 MLP-018566 MLP-018567 MLP-018585 MLP-018587 MLP-018588 MLP-018589 MLP-018602 MLP-018604 MLP-018605 MLP-018606 MLP-018608 MLP-018610 MLP-018636 MLP-018637 MLP-018638 MLP-018644 MLP-018646 MLP-018649 MLP-018651 MLP-018652 MLP-018653 MLP-018654 MLP-018655 MLP-018656 MLP-018660 MLP-018663 MLP-018666 MLP-018667 MLP-018668 MLP-018675 MLP-018678 MLP-018679 MLP-018680 MLP-018683 MLP-018685 MLP-018686 MLP-018715 MLP-018718 MLP-018719 MLP-018722 MLP-018729 MLP-018748 MLP-018750 MLP-018751 MLP-018752 MLP-018756 MLP-018758 MLP-018760 MLP-018765 MLP-0187

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1301-2024