RayStation 11B
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- RAYSEARCH LABORATORIES AB
- Reason for Recall:
- This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
RayStation 11B
Product Codes/Lot Numbers:
UDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1299-2022
Related Recalls
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
RAYSEARCH LABORATORIES AB
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
RAYSEARCH LABORATORIES AB
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.