S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 2258046 2863490
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Stryker Communications
- Reason for Recall:
- The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
Product Codes/Lot Numbers:
Serial Numbers: 2258046 2863490
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1297-2021
Related Recalls
Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
Stryker Communications
Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
Chromophare Surgical Light System , REF CH00000001
Stryker Communications
Due to insufficient mounting force the stability of surgical light system cannot be guaranteed.
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Stryker Communications
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.