Interventional Fluoroscopic X-Ray System - Product Usage: developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    ARTIS Icono biplane- 11327600 ARTIS Icono floor- 11327700
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
Siemens has become aware of a potential issue in which the imaging system UPS sporadically forces a shutdown of the imaging system PC without a true power supply problem.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Interventional Fluoroscopic X-Ray System - Product Usage: developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Product Codes/Lot Numbers:

ARTIS Icono biplane- 11327600 ARTIS Icono floor- 11327700

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1288-2021

Related Recalls

MAMMOMAT Inspiration;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Fusion;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →

MAMMOMAT Revelation;

Siemens Medical Solutions USA

Class I - Dangerous

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Oct 24, 2025 Other Medical Devices Nationwide View Details →