CAIRE DIAGNOSTICS FenomPro REF 900-0001 SN ***** - Product Usage: a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. Fenom Pro should not be used in critical care, emergency care or in anesthesiology.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All units sold before February 23, 2021 are affected (serial numbers 100107 to 100651).
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
CAIRE DIAGNOSTICS INC
Reason for Recall:
Erroneously high FeNO levels which may contribute to premature discontinuation of effective asthma treatment and persistent underlying disease. Two issues affecting the FeNO readings: 1) Drift in the calibration gases and 2) Software error. Each of these issues can cause FeNO scores/results to be erroneously high.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

CAIRE DIAGNOSTICS FenomPro REF 900-0001 SN ***** - Product Usage: a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. Fenom Pro should not be used in critical care, emergency care or in anesthesiology.

Product Codes/Lot Numbers:

All units sold before February 23, 2021 are affected (serial numbers 100107 to 100651).

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1284-2021