DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- DeRoyal Industries Inc
- Reason for Recall:
- The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
Product Codes/Lot Numbers:
GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1283-2023
Related Recalls
Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement
DeRoyal Industries
The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.
DeRoyal CHILDREN'S LAPTRACEPACK, 89-9362.06
DeRoyal Industries
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.