Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA Assays. This recall is for any Phadia Prime software version up to and including the current version, 2.1.4.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Phadia Prime software 12-4101-00 Phadia 250 12-3900-01 Listed software: EliA dsDNA Well 14-5500-01 EliA U1RNP Well 14-5501-01 EliA Sm Well 14-5502-01 EliA Ro Well 14-5503-01 EliA La Well 14-5504-01 EliA CENP Well 14-5505-01 EliA Scl-70 Well 14-5506-01 EliA Jo-1 Well 14-5507-01 EliA Symphony Well 14-5508-01 EliA RNP70 Well 14-5511-01 EliA GBM Well 14-5514-01 EliA CCP Well 14-5515-01 EliA Celikey IgA Well 14-5517-01 EliA Celikey IgG Well 14-5518-01 EliA Gliadin IgA Well 14-5519-01 EliA Gliadin IgG Well 14-5520-01 EliA Cardiolipin IgA Well 14-5528-01 EliA Cardiolipin IgG Well 14-5529-01 EliA Cardiolipin IgM Well 14-5530-01 EliA B2-Glycoprotein I IgA Well 14-5531-01 EliA B2-Glycoprotein I IgG Well 14-5532-01 EliA B2-Glycoprotein I IgM Well 14-5533-01 EliA PR3S Well 14-5536-01 EliA MPOS Well 14-5537-01 EliA GliadinDP IgA Well 14-5538-01 EliA GliadinDP IgG Well 14-5539-01 EliA RF IgM Well 14-5600-01 EliA RF IgA Well 14-5601-01 EliA SmDP Well 14-5624-01 EliA Scl-70S Well 14-5637-01 EliA anti-TPO Well 14-5641-01 EliA anti-TG Well 14-5642-01 EliA M2 Well 14-5649-01
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Phadia Ab
Reason for Recall:
We want to inform all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejecting and retesting samples with any EliA assay (the OK function may be used for rejecting single tests and dilution of samples in accordance with product DfU).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA Assays. This recall is for any Phadia Prime software version up to and including the current version, 2.1.4.

Product Codes/Lot Numbers:

Phadia Prime software 12-4101-00 Phadia 250 12-3900-01 Listed software: EliA dsDNA Well 14-5500-01 EliA U1RNP Well 14-5501-01 EliA Sm Well 14-5502-01 EliA Ro Well 14-5503-01 EliA La Well 14-5504-01 EliA CENP Well 14-5505-01 EliA Scl-70 Well 14-5506-01 EliA Jo-1 Well 14-5507-01 EliA Symphony Well 14-5508-01 EliA RNP70 Well 14-5511-01 EliA GBM Well 14-5514-01 EliA CCP Well 14-5515-01 EliA Celikey IgA Well 14-5517-01 EliA Celikey IgG Well 14-5518-01 EliA Gliadin IgA Well 14-5519-01 EliA Gliadin IgG Well 14-5520-01 EliA Cardiolipin IgA Well 14-5528-01 EliA Cardiolipin IgG Well 14-5529-01 EliA Cardiolipin IgM Well 14-5530-01 EliA B2-Glycoprotein I IgA Well 14-5531-01 EliA B2-Glycoprotein I IgG Well 14-5532-01 EliA B2-Glycoprotein I IgM Well 14-5533-01 EliA PR3S Well 14-5536-01 EliA MPOS Well 14-5537-01 EliA GliadinDP IgA Well 14-5538-01 EliA GliadinDP IgG Well 14-5539-01 EliA RF IgM Well 14-5600-01 EliA RF IgA Well 14-5601-01 EliA SmDP Well 14-5624-01 EliA Scl-70S Well 14-5637-01 EliA anti-TPO Well 14-5641-01 EliA anti-TG Well 14-5642-01 EliA M2 Well 14-5649-01

Distribution:

Distributed in: US, TX, VA, NJ, MA, IN, GA, UT, TN, CA, MS, NC, ME, OR

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1276-2018

Related Recalls