Zilver¿ PTX¿ Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps Preven the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to:  Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD .  improve blood flow to the extremity.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    REF ZIV6-35-125-6-80-PTX ; REFG24890 UDI 10827002248905 All lots distributed prior to 4/16/2013
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Cook, Inc.
Reason for Recall:
Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Zilver¿ PTX¿ Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps Preven the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to:  Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD .  improve blood flow to the extremity.

Product Codes/Lot Numbers:

REF ZIV6-35-125-6-80-PTX ; REFG24890 UDI 10827002248905 All lots distributed prior to 4/16/2013

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1274-2013