EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Medica s Lot # 16292
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medica Corporation
Reason for Recall:
Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.

Product Codes/Lot Numbers:

Medica s Lot # 16292

Distribution:

Distributed in: MI, FL, CA, TX

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1267-2018

Related Recalls

Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.

Feb 6, 2019 Diagnostic Equipment Nationwide View Details →

The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.