Basic Pack, part number AMS1868(B Basic Pack, part number AMS2608(A Basic Pack, part number AMS3861 Basic Pack, part number AMS4309 Basic Pack, part number PSS1833 Basic Pack, part number PSS1833(A Basic Pack, part number PSS1848(A
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot number and expiration date 100394, 9/15/2019 100593, 9/21/2019 100594, 10/5/2019 101063, 10/15/2019 101304, 10/23/2019 83725, 6/8/2018 84749, 9/11/2018 85159, 8/1/2018 85635, 8/6/2018 85810, 7/19/2018 86252, 9/2/2018 86440, 8/7/2018 86859, 8/18/2018 86939, 11/22/2018 87105, 9/15/2018 87269, 9/30/2018 87351, 9/15/2018 87447, 11/11/2018 87942, 9/23/2018 88172, 12/3/2018 88282, 9/7/2018 88368, 9/5/2018 88846, 12/11/2018 88887, 12/7/2018 89004, 10/8/2018 89160, 12/1/2018 89438, 11/22/2018 89512, 10/9/2018 89716, 1/9/2019 89743, 12/3/2018 89811, 12/1/2018 94572, 12/15/2018 94705, 12/15/2018 94751, 1/4/2019 94853, 1/27/2019 94935, 1/25/2019 95293, 1/26/2019 95412, 1/25/2019 95936, 1/11/2019 95974, 5/29/2019 96023, 1/16/2019 96255, 1/2/2019 96410, 3/7/2019 96566, 3/9/2019 96678, 5/24/2019 96913, 5/22/2019 97124, 5/12/2019 97410, 6/3/2019 97577, 5/1/2019 97780, 6/15/2019 98017, 6/29/2019 98312, 6/24/2019 98478, 6/29/2019 98502, 6/6/2019 98663, 6/28/2019 98633, 6/30/2019 99172, 6/12/2019 99581, 6/27/2019 99703, 9/27/2019 99843, 9/30/2019 99887, 9/11/2019 99913, 9/23/2019
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Windstone Medical Packaging, Inc.
- Reason for Recall:
- The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES of the custom Aligned Medical Solutions (AMS) kits were recalled because gloves may contain splits or holes compromising sterility.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Basic Pack, part number AMS1868(B Basic Pack, part number AMS2608(A Basic Pack, part number AMS3861 Basic Pack, part number AMS4309 Basic Pack, part number PSS1833 Basic Pack, part number PSS1833(A Basic Pack, part number PSS1848(A
Product Codes/Lot Numbers:
Lot number and expiration date 100394, 9/15/2019 100593, 9/21/2019 100594, 10/5/2019 101063, 10/15/2019 101304, 10/23/2019 83725, 6/8/2018 84749, 9/11/2018 85159, 8/1/2018 85635, 8/6/2018 85810, 7/19/2018 86252, 9/2/2018 86440, 8/7/2018 86859, 8/18/2018 86939, 11/22/2018 87105, 9/15/2018 87269, 9/30/2018 87351, 9/15/2018 87447, 11/11/2018 87942, 9/23/2018 88172, 12/3/2018 88282, 9/7/2018 88368, 9/5/2018 88846, 12/11/2018 88887, 12/7/2018 89004, 10/8/2018 89160, 12/1/2018 89438, 11/22/2018 89512, 10/9/2018 89716, 1/9/2019 89743, 12/3/2018 89811, 12/1/2018 94572, 12/15/2018 94705, 12/15/2018 94751, 1/4/2019 94853, 1/27/2019 94935, 1/25/2019 95293, 1/26/2019 95412, 1/25/2019 95936, 1/11/2019 95974, 5/29/2019 96023, 1/16/2019 96255, 1/2/2019 96410, 3/7/2019 96566, 3/9/2019 96678, 5/24/2019 96913, 5/22/2019 97124, 5/12/2019 97410, 6/3/2019 97577, 5/1/2019 97780, 6/15/2019 98017, 6/29/2019 98312, 6/24/2019 98478, 6/29/2019 98502, 6/6/2019 98663, 6/28/2019 98633, 6/30/2019 99172, 6/12/2019 99581, 6/27/2019 99703, 9/27/2019 99843, 9/30/2019 99887, 9/11/2019 99913, 9/23/2019
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1267-2017
Related Recalls
Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes) Angio Pack Part numbers AMS6908E, AMS6908F
Windstone Medical Packaging
Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.
Custom Convenience Kits Pre Op Kit - NS, Part Number AMS14994
Windstone Medical Packaging
The product correction was initiated because B Braun is issuing a voluntary Urgent Medical Device Correction for gravity IV administration sets as well as pump administration sets utilized with BBMI s Infusomat Space Large Volume Pump, Outlook Pump and Vista Basic Pump due to potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers or other connections and the ability to prime (occlusion).
Custom Convenience Kits Pre Op Kit - NS, Part Number AMS14994A
Windstone Medical Packaging
The product correction was initiated because B Braun is issuing a voluntary Urgent Medical Device Correction for gravity IV administration sets as well as pump administration sets utilized with BBMI s Infusomat Space Large Volume Pump, Outlook Pump and Vista Basic Pump due to potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers or other connections and the ability to prime (occlusion).