The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    10275879, 10413009, 04380213, 04380221, 05242826, 05977066, 05977074, 05984005, 05989079, 05989087, 05989095, 05991109, 05991117, 05992099, 07324119, 07324135, 07324143, 07324150, 07760809, 07760932, 07761161, 07823920, 07823938, 07823946, 07823953, 07823961, 07823979, 10151531, 10151532, 10520745
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc.
Reason for Recall:
We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasize the proper collimator change
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

The E.CAM gamma camera system Product Usage: The E.CAM gamma camera system is a SPECT system and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body

Product Codes/Lot Numbers:

10275879, 10413009, 04380213, 04380221, 05242826, 05977066, 05977074, 05984005, 05989079, 05989087, 05989095, 05991109, 05991117, 05992099, 07324119, 07324135, 07324143, 07324150, 07760809, 07760932, 07761161, 07823920, 07823938, 07823946, 07823953, 07823961, 07823979, 10151531, 10151532, 10520745

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1262-2016

Related Recalls

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →
Class I - Dangerous

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA

Class I - Dangerous

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.