Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- R & D Systems, Inc.
- Reason for Recall:
- R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Quantikine IVD Human sTfR Immunoassay (Human sTransferrin ELISA Kit (sTfR immunological test system)). An enzyme linked immunosorbent assay (ELISA) for the quantitative determination of Soluble Transferrin Receptor (sTfR) concentration in human serum and plasma as an aid in the diagnosis of anemia and polycythemia.
Product Codes/Lot Numbers:
Catalog #: DTFR1; Unique Device Identification (UDI) #: 815762021657; Lot #: P252478; P263156; P256349; P263156; P263160; P264604
Distribution:
Distributed in: FL, LA, CA, OH
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1255-2021
Related Recalls
For one lot of CBC-3D Hematology Control, the Normal Level may exhibit hemolysis or deterioration due to microbial contamination.
The Instructions for Use (IFU, Part Number 750095) provided with the affected lots by the firm contains an error associated with product s CE marking. The CE mark contained in the IFU is followed by the Notified Body designation, British Standards Institution number 2797 (BSI-2797) for product inserts 750095.28 and 750095.29. The erroneous label CE2797 were in the IFU insert revision 750095.28 for the 408 affected lots that were shipped. No product was packed with erroneous insert revision 750095.29. Since the product is currently CE marked as a Class I - self-certified product under the EU In Vitro Diagnostics Directive, the Notified Body designation should not be specified. As part of the corrective action, the firm has notified the customers who received the products with erroneous labeling, and is conducting the replacements with the correct product insert containing the Insert Revision 750095.30
The Human sTfR Quantikine IVD ELISA Kit, catalog # DTFR1, lot P180151 was packed with the incorrect control data card number 750228.4.