Vantage Titan, Model: MRT-1504
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 04987670102400. Serial Numbers: U4B15Y2102, U5B15Z2001, U4B1492080, U4B1582100, U4B1582099, U4B1522091, U4B1512089, U4B1432067, U4B1522093, U4B1452073, U4B1442069, U4B1442068, U4B1642106, U4B1432066, U4B15X2101, U4B14Y2083, U4B1452071, U4B1512087, U4B1652107, U4B1552097, U4B1482078, U4B1512090, U4B1552098, U4B1442070, U4B14Y2082, U4B1482076, U4B14Z2085, U4B14Z2086, U4B1512088, U4B1612103, U4B1472075, U4B1422065, U4B1462074, U4B14Y2084, U4B1632105, U4B1452072, U4B1492081, U4B1482079, U4B1612104, U4B1482077
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Canon Medical System, USA, INC.
- Reason for Recall:
- For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Vantage Titan, Model: MRT-1504
Product Codes/Lot Numbers:
UDI-DI: 04987670102400. Serial Numbers: U4B15Y2102, U5B15Z2001, U4B1492080, U4B1582100, U4B1582099, U4B1522091, U4B1512089, U4B1432067, U4B1522093, U4B1452073, U4B1442069, U4B1442068, U4B1642106, U4B1432066, U4B15X2101, U4B14Y2083, U4B1452071, U4B1512087, U4B1652107, U4B1552097, U4B1482078, U4B1512090, U4B1552098, U4B1442070, U4B14Y2082, U4B1482076, U4B14Z2085, U4B14Z2086, U4B1512088, U4B1612103, U4B1472075, U4B1422065, U4B1462074, U4B14Y2084, U4B1632105, U4B1452072, U4B1492081, U4B1482079, U4B1612104, U4B1482077
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1250-2024
Related Recalls
Alphenix INFX-8000F, interventional fluoroscopic x-ray system
Canon Medical System, USA
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
Canon Medical System, USA
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
Canon Medical System, USA
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.