Reveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Reveal LINQ ICM, Model LNQ11 serial numbers are impacted. Medtronic continues to manufacture and distribute Reveal Ll NQ devices.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc.
- Reason for Recall:
- Medtronic has identified an issue with the sensitivity of an algorithm used in the Reveal LINQ ICM that may prematurely trigger the Recommended Replacement Time (RRT) alert in some devices. Battery capacity is not affected and the device will continue to support data collection and manual data transmissions. As part of the normal behavior of the device, 30 days after RRT status is reached, Reveal
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Reveal LINQ insertable Cardiac Monitor, Model LNQ11. Monitors subcutaneous ECG continuously.
Product Codes/Lot Numbers:
All Reveal LINQ ICM, Model LNQ11 serial numbers are impacted. Medtronic continues to manufacture and distribute Reveal Ll NQ devices.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1249-2016
Related Recalls
There is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur.