Cardiosave Hybrid and rescue IABP
Class I - DangerousWhat Should You Do?
- Check if you have this product: Cardiosave Hybrid IABP 0998-00-0800-53,0998-UC-0800-53, Cardiosave Rescue IABP 0998-00-0800-83 & 0998-UC-0800-83
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Maquet Datascope Corp - Cardiac Assist Division
- Reason for Recall:
- Maquet has recieved information that in some Cardiosave IABPS, the scroll compressor did not meet the specifications for output pressure or vacuum at specific flow rates.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Cardiosave Hybrid and rescue IABP
Product Codes/Lot Numbers:
Cardiosave Hybrid IABP 0998-00-0800-53,0998-UC-0800-53, Cardiosave Rescue IABP 0998-00-0800-83 & 0998-UC-0800-83
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1248-2016
Related Recalls
Cardiosave Hybrid IABP, Part Number 0998-00-0800-53
Maquet Datascope Corp - Cardiac Assist Division
There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.
Cardiosave Rescue IABP, Part Number 0998-UC-0800-83
Maquet Datascope Corp - Cardiac Assist Division
There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.
CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 & 0998-00-0800-75) and cart 0997-00-1179
Maquet Datascope Corp - Cardiac Assist Division
Ingress of fluids into the device can affect various electronic circuit boards which would prevent initiation or continuation of therapy.