GE Healthcare Optima CT580/Discovery CT590RT scanners running software version 13BW16.3. The Discovery CT 590 RT and Optima CT 580 series of Computed Tomography systems are intended to produce cross-sectional images of the body by computer reconstruction of XRay transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), and Gated (Respiratory and Cardiac) acquisitions in patients of all ages.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Mfg Lot or Serial # PA2314840 PA2314841 PA2314842 PA2314843 PA2314844 PA2314845 A1RF3176G A1RF3638GR A2RF1052G B1RF4038GR B3G44377G B3G44511G B3G44515G B3RF4960G C1RF4686G C1RF4793G C1RF4802G C1RF4827GR D2RF2744G D2RF2778GR D2RF2795G D2RF2796G D2RF2797G D2RF2798G D2RF2799G D2RF2814G D2RF2815G D2RF2816G D2RF2817G D2RF2818G D2RF2819G D2RF2820G D2RF2821G D2RF2822G D2RF2823G D2RF2824G D2RF2825G D2RF2826G D2RF2827G D2RF2828G D2RF2829G D2RF2830G D2RF2831G D2RF2832G D2RF2833G D2RF2834G D2RF2835G D2RF2836G D2RF2837G D2RF2838G D2RF2839G D2RF2840G D2RF2841G D2RF2842G D2RF2843G D2RF2844G D2RF2845G D2RF2846G D2RF2847G D2RF2848G D2RF2850G D2RF2851G D3G44977G D3G44983G D3G44984G D3G44985G D43T4071G D43T4072G D43T4073G D43T4074G D43T4075G D43T4076G D43T4077G E1RF5364G E1RF5479GR E1RF5480GR E1RF5599G E3RF5317G E3RF5318G F0RF0134G F0RF0135G F0RF0136G F0RF0140G F0RF0141G F1RF4827GR F1RF5952G F1RF6042GR F1RF6075GR F1RF6079GR F1RF6084GR F1RF6085GR F1RF6086GR F1RF6088GR F1RF6092GR F1RF6097GR F1RF6100GR F1RF6114GR F1RF6118GR F1RF6123GR F1RF6130GR F1RF6227GR F1RF6236G F1RF6236GR F1RF6243GR F1RF6282GR F2G40578G F2G40579G F2G40580G F2G40581G F2G40584G F2G40585G F2G40586G F2G40588G F2G40592G F2G40593G F2G40594G F2G40595G F2G40596G F2G40598G F2G40599G F2G40605G F2G40606G F2G40607G F2G40609G F2G40610G F2G40611G F2G40615G F2G40631G F2G40705G F2G40820G F2G40835G F2RF3572G F2RF3640G F2RF3651G F3G45957G F3G45980G F3G46200G F43S0355G F43S0356G G0RF0516GR G0RF0592GR G0RF0596GR G1RF6518G G1RF6544G G1RF6546G G3G46241G G3G46616G G3G46649G G3G46664G G3G46674G G3G46679G H0RF0607GR H0RF0772G H0RF0845GR H2G41289G H2G41307G H2G41571GR H2G41786G H2G41787G H2RF4002GR J0RF0465G J0RF1103G J1RF7280G J1RF7281G J1RF7282G J1RF7283G J1RF7284G J1RF7285G J1RF7286G J1RF7289G J1RF7290G J1RF7298G J1RF7387G J2G42049G J2G42100GR J33T0400G J43S0499G J43S0503G J43S0504G J43S0506G J43T5719G J43T5720G J43T5727G J43T5731G J43T5799G J43T5801G J43T5810G J43T5812G J43T5813G J43T5814G J43T5815G J43T5827G J43T5831G J43T5832G J43T5877G J43T6078G K1RF7644G K1RF7673G K1RF7674G K1RF7675G K1RF7676G K1RF7678G K1RF7679G K1RF7680G K1RF7681G K1RF7683G K1RF7704G K1RF7925GR K33T0621G K33T1002G K43T6545G L1RF8243G L1RF8254G L1RF8255G L1RF8256G L1RF8258G L1RF8259G L1RF8261G L1RF8263G L1RF8264G L1RF8265G L1RF8267G L1RF8728G L1RF8733G L2G42609G L2G42643G L2G42666G L2G42668G L2G42669G L2G42690G L2G42711G L2G42970G L2RF4373GR L33T1324G M0RF2496G M0RF2562GR M1RF9841G M2G43248G M2RF4686G M33T1729G M33T2293G PA1786357 PA1786438 PA1786439 PA1786440 PA1786441 PA1786442 PA1786443 PA1786444 PA1786445 PA1786446 PA1786447 PA1786448 PA1786449 PA1786450 PA1786451 PA1786452 PA1786453 PA1786454 PA1786456 PA1786457 PA1786458 PA1786461 PA1786462 PA1786494 PA1786495 PA1786496 PA1786497 PA1786498 PA1786499 PA1786500 PA1786501 PA1786502 PA1786503 PA1786504 PA1786505 PA1786530 PA1786531 PA1786532 PA1786533 PA1786534 PA1786535 PA1786536 PA1786537 PA1786538 PA1786539 PA1786540 PA1786541 PA1786542 PA1786543 PA1786544 PA1786545 PA1786546 PA1786547 PA1786548 PA1786549 PA1786677 PA1786678 PA1786679 PA1786680 PA1786681 PA1786682 PA1786683 PA1786684 PA1786685 PA1786686 PA1786687 PA1786688 PA1786689 PA1786690 PA1786691 PA1786692 PA1786693 PA1786694 PA1786695 PA1786696 PA1786697 PA1786698 PA1786699 PA1786700 PA1786701 PA1786702 PA1786703 PA1786704 PA1786705 PA1786706 PA1786707 PA1786708 PA1786709 PA1786710 PA1786711 PA1786712 PA1786713 PA1786714 PA1786715 PA1786716 PA1787000 PA1787001 PA1787002 PA1787003 PA1787004 PA1787005 PA1787006 PA1787007 PA1787008 PA1787009 PA1787010 PA1787011 PA1787079 PA1787080 PA1787081 PA1787082 PA1787083 PA1787084 PA1787085 PA1787086 PA1787087 PA1787088 PA1787351 PA1787352 PA1787353 PA1787354 PA1787355 PA1787356 PA1787357 PA1787478 PA1787479 PA1787480 PA1787481 PA1787482 PA1787483 PA1787484 PA1787485 PA1787561 PA1787562 PA1787563 PA1787564 PA1787565 PA1787566 PA1787567 PA1787568 PA1787569 PA1787570 PA1787571 PA1787572 PA1787573 PA1787574 PA1787575 PA1787576 PA1787577 PA1787578 PA1787579 PA1787580 PA1787581 PA1787582 PA1787583 PA1787584 PA1787585 PA1787586 PA1787587 PA1787588 PA1787589 PA1787590 PA1787591 PA1787592 PA1787593 PA1787594 PA1787595 PA1787596 PA1787597 PA1787598 PA1787599 PA1787600 PA1787601 PA1787602 PA1787603 PA1787604 PA1787605 PA1787606 PA1787607 PA1787608 PA1787609 PA1787610 PA1787611 PA1787612 PA1787613 PA1787614 PA1787615 PA1787616 PA1787617 PA1787618 PA1787619 PA1787620 PA1787621 PA1787622 PA1787623 PA1787624 PA1787625 PA1787626 PA1787627 PA1787628 PA1787629 PA1787630 PA1787631 PA1787632 PA1787633 PA1787634 PA1787635 PA1787636 PA1787637 PA1787638 PA1787639 PA1787640 PA1787641 PA1787642 PA1787643 PA1787644 PA1787645 PA1787646 PA1787647 PA1787648 PA1787649 PA1787650 PA1787651 PA1787652 PA1787653 PA1787654 PA1787655 PA1787656 PA1787657 PA1787658 PA1787659 PA1787660 PA1787661 PA1787662 PA1787663 PA1787664 PA1787665 PA1787666 PA1787667 PA1787668 PA1787669 PA1787670 PA1787671 PA1787672 PA1788113 PA1788114 PA1788115 PA1788116 PA1788117 PA1788146 PA1788147 PA1788148 PA1788149 PA1788150 PA1788151 PA1788152 PA1788153 PA1788154 PA1788155 PA1788156 PA1788157 PA1788158 PA1788159 PA1788160 PA1788161 PA1788162 PA1788163 PA1788164 PA1788165 PA1788166 PA1788167 PA1788168 PA1788169 PA1788170 PA1788171 PA1788172 PA1788173 PA1788174 PA1788175 PA1788176 PA1788177 PA1788178 PA1788179 PA1788180 PA1788310 PA1788311 PA1788312 PA1788313 PA1788314 PA1788315 PA1788316 PA1788317 PA1788318 PA1788319 PA1788320 PA1788321 PA1788322 PA1788323 PA1788324 PA1788325 PA1788326 PA1788327 PA1788328 PA1788329 PA1788330 PA1788331 PA1788332 PA1788333 PA1788334 PA1788335 PA1788336 PA1788337 PA1788338 PA1788339 PA1788360 PA1788361 PA1788362 PA1788363 PA1788364 PA1788365 PA1788366 PA1788367 PA1788368 PA1788370 PA1788371 PA1788372 PA1788373 PA1788375 PA1788378 PA1788379 PA1788380 PA1788381 PA1788382 PA1788384 PA1788407 PA1788408 PA1788409 PA1788410 PA1788411 PA1788412 PA1788413 PA1788414 PA1788415 PA1788416 PA1788417 PA1788418 PA1788419 PA1788420 PA1788421 PA1788422 PA1788423 PA1788424 PA1788425 PA1788426 PA1788427 PA1788428 PA1788429 PA1788430 PA1788431 PA1788432 PA1788433 PA1788434 PA1788435 PA1788436 PA1788471 PA1788472 PA1788473 PA1788474 PA1788475 PA1788476 PA1788477 PA1788478 PA1788479 PA1788480 PA1788481 PA1788482 PA1788483 PA1788484 PA1788485 PA1788486 PA1788487 PA1788488 PA1788489 PA1788490 PA1788491 PA1788492 PA1788493 PA1788494 PA1788495 PA1788496 PA1788497 PA1788498 PA1788499 PA1788500 PA1788501 PA1788502 PA1788503 PA1788504 PA1788505 PA1788506 PA1788507 PA1788508 PA1788509 PA1788510 PA1788511 PA1788512 PA1788513 PA1788514 PA1788515 PA1788516 PA1788517 PA1788518 PA1788519 PA1788520 PA1788521 PA1788522 PA1788523 PA1788524 PA1788525 PA1788526 PA1788527 PA1788528 PA1788529 PA1788530 PA1788531 PA1788532 PA1788533 PA1788534 PA1788535 PA1788625 PA1788626 PA1788627 PA1788628 PA1788629 PA1788630 PA1788631 PA1788632 PA1788633 PA1788634 PA1789287 PA1789288 PA1789289 PA1789290 PA1789291 PA1789292 PA1789293 PA1789294 PA1789295 PA1789296 PA1850116 PA1850117 PA1850118 PA1850119 PA1850120 PA1850121 PA1850122 PA1850123 PA1850124 PA1850125 PA1850126 PA1850127 PA1850128 PA1850129 PA1850130 PA1850131 PA1850132 PA1850133 PA1850134 PA1850135 PA1850136 PA1850137 PA1850138 PA1850139 PA1850140 PA1850141 PA1850142 PA1850143 PA1850144 PA1850145 PA1850146 PA1850147 PA1850391 PA1850392 PA1850393 PA1850394 PA1850395 PA1850396 PA1850397 PA1850398 PA1850399 PA1850400 PA1850401 PA1850402 PA1850403 PA1850404 PA1850405 PA1850406 PA1850407 PA1850408 PA1850409 PA1850410 PA1850411 PA1850412 PA1850413 PA1850414 PA1850415 PA1850416 PA1850417 PA1850418 PA1850419 PA1850420 PA1850421 PA1850422 PA1850423 PA1850424 PA1850425 PA1850426 PA1850427 PA1850428 PA1850771 PA1850772 PA1850773 PA1850774 PA1850775 PA1850776 PA1850777 PA1850778 PA1850779 PA1850791 PA1850792 PA1850793 PA1850794 PA1850795 PA1850796 PA1850797 PA1850798 PA1850897 PA1850898 PA1850899 PA1850900 PA1850901 PA1850902 PA1850903 PA1850904 PA1850905 PA1850906 PA1850907 PA1850908 PA1850909 PA1850910 PA1850911 PA1850912 PA1850913 PA1850914 PA1850915 PA1850916 PA1850917 PA1850918 PA1850919 PA1850920 PA1850921 PA1850922 PA1850923 PA1850924 PA1850925 PA1850926 PA1850927 PA1850928 PA1850929 PA1850930 PA1850931 PA1851412 PA1851413 PA1851414 PA1851415 PA1851416 PA1851829 PA1851830 PA1851831 PA1851832 PA1851833 PA1851834 PA1851835 PA1851836 PA1851837 PA1851838 PA1851839 PA1851840 PA1851841 PA1851842 PA1851843 PA1851844 PA1852084 PA1852085 PA1852086 PA1852087 PA1852088 PA1852089 PA1852090 PA1852091 PA1852092 PA1852093 PA1852094 PA1852095 PA1852096 PA1852097 PA1852098 PA1852099 PA1852138 PA1852139 PA1852140 PA1852141 PA1852142 PA1852143 PA1852144 PA1852145 PA1852146 PA1852147 PA1852148 PA1852149 PA1852150 PA1852151 PA1852152 PA1852153 PA1852154 PA1852155 PA1852156 PA1852157 PA1852158 PA1852159 PA1852160 PA1852161 PA1852162 PA1852163 PA1852164 PA1852165 PA1852166 PA1852167 PA1852168 PA1852169 PA1852170 PA1852171 PA1852172 PA1852173 PA1852174 PA1852175 PA1852176 PA1852177 PA1852178 PA1852179 PA1852180 PA1852181 PA1852182 PA1852183 PA1852184 PA1852185 PA1852186 PA1852187 PA1852188 PA1852189 PA1852190 PA1852191 PA1852192 PA1852193 PA1852194 PA1852195 PA1852196 PA1852197 PA1852198 PA1852199 PA1852200 PA1852201 PA1852202 PA1852203 PA1852204 PA1852205 PA1852206 PA1852207 PA1852208 PA1852209 PA1852210 PA1852211 PA1852212 PA1852213 PA1852214 PA1852215 PA1852216 PA1852217 PA1852218 PA1852219 PA1852220 PA1852221 PA1852222 PA1852223 PA1852224 PA1852225 PA1852226 PA1852227 PA1852228 PA1852229 PA1852230 PA1852231 PA1852232 PA1852233 PA1852234 PA1852235 PA1852236 PA1852237 PA1852409 PA1852410 PA1852411 PA1852412 PA1852413 PA1852414 PA1852415 PA1852416 PA1852417 PA1852418 PA1852419 PA1852420 PA1852421 PA1852422 PA1852423 PA1852424 PA1852425 PA1852426 PA1852427 PA1852428 PA1852429 PA1852430 PA1852431 PA1852432 PA1852433 PA1852434 PA1852435 PA1852436 PA1852437 PA1852438 PA1852439 PA1852440 PA1852441 PA1852442 PA1852443 PA1852444 PA1852445 PA1852446 PA1852447 PA1852448 PA1852449 PA1852808 PA1852809 PA1852810 PA1852811 PA1852812 PA1852813 PA1852814 PA1852815 PA1852816 PA1852817 PA1852818 PA1852819 PA1852820 PA1852821 PA1852822 PA1852823 PA1852824 PA1854656 PA1854657 PA1854658 PA1854659 PA1854660 PA1854661 PA1895455 PA1895456 PA1895457 PA1895458 PA1897813 PA1990824 PA1990825 PA1990826 PA1990827 PA1990828 PA1990829 PA1990830 PA1990831 PA1990832 PA1990833 PA1990834 PA1990835 PA1990836 PA1990837 PA1990838 PA1990839 PA1990840 PA1991403 PA1991404 PA1991405 PA1991406 PA1991407 PA1991408 PA1991409 PA1991410 PA1991411 PA1991412 PA1991413 PA1991414 PA1991415 PA1991416 PA1991417 PA1991418 PA1991419 PA1991440 PA1991442 PA1991635 PA1991851 PA1991852 PA1991853 PA1991854 PA1991855 PA1991856 PA1991857 PA1991858 PA1991859 PA1991860 PA1991861 PA1991862 PA1991863 PA1991864 PA1991865 PA1991866 PA1991867 PA1991868 PA1991869 PA1991870 PA1992005 PA1992006 PA1992007 PA1992008 PA1992112 PA1992113 PA1992114 PA1992115 PA1992116 PA1992117 PA1992118 PA1992119 PA1992120 PA1992121 PA1992122 PA1992123 PA1992124 PA1992125 PA1992126 PA1992127 PA1993645 PA1993646 PA1993647 PA1993648 PA1993649 PA1993650 PA1993651 PA1993652 PA1993653 PA1993654 PA1993655 PA1993656 PA1993657 PA1993658 PA1993659 PA1993660 PA1993661 PA1993662 PA1993663 PA1993664 PA1993665 PA1993666 PA1993667 PA1993668 PA1993860 PA1993861 PA1993862 PA1993863 PA1993864 PA1993865 PA1993866 PA1994201 PA1994202 PA1994203 PA1994204 PA1994205 PA2007158 PA2007159 PA2007160 PA2007161 PA2007162 PA2007163 PA2020375 PA2020376 PA2020377 PA2020378 PA2020379 PA2020380 PA2020381 PA2020382 PA2020383 PA2020384 PA2020385 PA2020386 PA2020387 PA2020388 PA2020389 PA2020390 PA2020391 PA2020392 PA2020458 PA2020459 PA2020460 PA2020461 PA2020462 PA2020463 PA2020464 PA2020465 PA2020466 PA2020467 PA2020468 PA2020469 PA2020470 PA2020471 PA2020472 PA2020473 PA2020474 PA2020475 PA2020476 PA2020477 PA2020478 PA2020479 PA2020480 PA2020481 PA2020482 PA2020505 PA2020506 PA2020507 PA2020508 PA2020509 PA2020510 PA2020511 PA2020512 PA2020513 PA2020514 PA2020515 PA2020516 PA2020517 PA2020518 PA2020519 PA2020520 PA2020521 PA2020522 PA2020523 PA2020524 PA2021288 PA2021289 PA2021290 PA2021291 PA2021292 PA2021293 PA2021294 PA2021295 PA2021296 PA2021297 PA2021806 PA2021807 PA2021808 PA2021809 PA2021810 PA2021811 PA2021812 PA2021813 PA2021814 PA2021815 PA2021816 PA2021817 PA2022029 PA2022030 PA2022032 PA2022034 PA2107909 PA2107910 PA2107911 PA2107912 PA2107913 PA2107914 PA2107915 PA2108512 PA2108513 PA2108514 PA2109592 PA2109593 PA2109594 PA2109595 PA2109596 PA2109597 PA2109598 PA2109599 PA2109600 PA2109601 PA2115014 PA2115015 PA2115016 PA2115017 PA2115018 PA2115019 PA2115020 PA2115021 PA2115022 PA2115023 PA2115024 PA2115025 PA2115026 PA2115027 PA2115028 PA2115029 PA2115030 PA2115031 PA2115032 PA2115033 PA2115034 PA2115035 PA2115036 PA2115037 PA2115038 PA2115039 PA2115040 PA2115041 PA2115042 PA2115043 PA2115044 PA2115045 PA2115046 PA2115049 PA2115212 PA2115213 PA2115214 PA2115215 PA2115216 PA2115217 PA2115218 PA2115219 PA2115220 PA2115221 PA2115222 PA2115232 PA2115233 PA2115234 PA2115235 PA2115236 PA2115237 PA2115309 PA2115310 PA2115311 PA2115312 PA2115313 PA2115314 PA2115315 PA2115316 PA2115317 PA2115318 PA2115319 PA2115320 PA2115321 PA2115322 PA2115323 PA2115327 PA2115329 PA2115330 PA2115331 PA2115332 PA2115333 PA2115334 PA2115335 PA2115336 PA2115337 PA2115338 PA2115347 PA2115348 PA2116026 PA2116027 PA2116028 PA2116029 PA2116030 PA2116031 PA2116032 PA2116033 PA2116034 PA2116035 PA2116036 PA2116037 PA2130461 PA2130462 PA2130463 PA2130464 PA2130465 PA2130466 PA2130467 PA2130468 PA2130469 PA2130470 PA2130471 PA2130472 PA2130473 PA2130474 PA2130475 PA2130476 PA2130477 PA2130478 PA2131319 PA2131320 PA2131321 PA2131322 PA2131323 PA2131324 PA2131325 PA2131326 PA2131327 PA2131328 PA2136382 PA2136383 PA2136384 PA2136385 PA2136386 PA2136387 PA2136388 PA2136391 PA2136398 PA2136399 PA2136408 PA2136409 PA2136549 PA2136550 PA2136551 PA2136552 PA2136553 PA2136558 PA2136559 PA2136560 PA2136561 PA2136562 PA2136563 PA2136584 PA2136585 PA2136586 PA2136587 PA2136588 PA2136589 PA2136590 PA2136591 PA2136592 PA2136593 PA2137621 PA2138188 PA2138189 PA2138190 PA2138191 PA2138208 PA2138248 PA2138280 PA2138337 PA2138338 PA2138339 PA2138340 PA2138341 PA2138342 PA2138343 PA2138344 PA2138345 PA2138346 PA2138347 PA2138348 PA2138349 PA2138388 PA2138389 PA2138390 PA2138391 PA2138392 PA2138393 PA2138394 PA2138395 PA2138396 PA2138397 PA2138398 PA2138399 PA2138400 PA2138401 PA2138402 PA2138403 PA2138404 PA2138405 PA2138406 PA2138407 PA2138408 PA2138409 PA2138410 PA2138411 PA2138412 PA2138413 PA2138414 PA2138415 PA2138416 PA2138417 PA2138418 PA2138419 PA2138420 PA2138421 PA2138422 PA2138423 PA2138424 PA2138425 PA2138427 PA2138428 PA2138429 PA2138430 PA2138431 PA2138432 PA2138433 PA2138434 PA2138435 PA2138436 PA2138437 PA2138438 PA2138439 PA2138440 PA2138441 PA2138442 PA2138443 PA2138444 PA2138445 PA2138446 PA2138447 PA2138448 PA2138449 PA2138450 PA2138451 PA2138452 PA2138453 PA2138514 PA2138515 PA2138516 PA2138517 PA2138518 PA2138519 PA2138520 PA2138521 PA2138522 PA2138822 PA2138823 PA2138824 PA2138825 PA2138826 PA2138827 PA2138828 PA2138829 PA2139335 PA2139336 PA2139337 PA2139338 PA2139339 PA2139340 PA2139341 PA2139342 PA2139343 PA2139344 PA2139345 PA2139346 PA2139347 PA2139348 PA2139349 PA2139350 PA2139351 PA2139352 PA2139353 PA2139354 PA2139355 PA2139356 PA2139357 PA2139358 PA2139359 PA2139360 PA2139361 PA2139493 PA2139494 PA2139495 PA2139496 PA2139497 PA2139498 PA2139499 PA2139500 PA2139501 PA2139502 PA2139503 PA2139504 PA2139505 PA2139506 PA2139507 PA2139508 PA2139509 PA2139510 PA2139511 PA2139512 PA2139513 PA2139514 PA2139515 PA2139516 PA2139517 PA2139518 PA2139519 PA2139520 PA2139521 PA2139522 PA2139523 PA2139524 PA2139525 PA2139526 PA2139527 PA2139528 PA2139529 PA2139530 PA2139531 PA2139532 PA2139533 PA2139534 PA2139535 PA2139536 PA2139537 PA2139538 PA2139539 PA2182814 PA2184843 PA2184844 PA2184845 PA2184846 PA2184847 PA2184848 PA2184849 PA2184850 PA2184851 PA2184852 PA2184853 PA2184854 PA2184855 PA2184856 PA2184857 PA2184858 PA2184859 PA2184860 PA2184933 PA2184943 PA2300038 PA2300039 PA2300040 PA2300041 PA2300042 PA2300043 PA2300044 PA2300045 PA2300046 PA2300047 PA2300048 PA2300049 PA2300050 PA2300051 PA2300097 PA2300098 PA2300099 PA2300100 PA2300101 PA2300102 PA2300103 PA2300104 PA2300105 PA2300106 PA2300107 PA2300108 PA2300109 PA2300110 PA2300111 PA2300112 PA2300113 PA2300114 PA2300115 PA2300116 PA2300117 PA2300118 PA2300119 PA2300120 PA2300121 PA2300122 PA2300123 PA2300124 PA2300125 PA2300126 PA2300127 PA2300128 PA2300129 PA2300130 PA2300131 PA2300132 PA2300133 PA2300134 PA2300135 PA2300136 PA2300137 PA2300138 PA2300139 PA2300140 PA2300141 PA2300142 PA2300143 PA2300144 PA2300145 PA2300146 PA2300147 PA2300148 PA2300149 PA2300150 PA2300151 PA2300152 PA2300153 PA2300154 PA2300155 PA2300156 PA2300157 PA2300158 PA2300159 PA2300160 PA2300161 PA2300162 PA2300163 PA2300164 PA2300165 PA2300166 PA2300167 PA2300168 PA2300169 PA2300170 PA2300171 PA2300172 PA2300173 PA2300174 PA2300175 PA2300176 PA2300177 PA2300178 PA2300179 PA2300180 PA2300181 PA2300182 PA2300183 PA2300184 PA2300185 PA2300186 PA2300187 PA2300188 PA2300189 PA2300190 PA2300191 PA2300192 PA2300193 PA2300194 PA2300195 PA2300196 PA2300197 PA2300198 PA2300199 PA2300200 PA2300201 PA2300202 PA2300203 PA2300204 PA2300205 PA2300206 PA2300207 PA2300208 PA2300209 PA2300210 PA2300211 PA2300212 PA2300213 PA2300214 PA2300215 PA2300216 PA2300217 PA2300218 PA2300219 PA2300220 PA2300221 PA2300222 PA2300223 PA2300224 PA2300225 PA2300226 PA2300227 PA2300228 PA2300229 PA2300230 PA2300231 PA2300232 PA2300233 PA2300234 PA2300235 PA2300236 PA2300237 PA2300238 PA2300239 PA2300240 PA2300241 PA2300242 PA2300243 PA2300244 PA2300245 PA2300246 PA2300247 PA2300248 PA2300249 PA2300250 PA2300251 PA2300252 PA2300253 PA2300254 PA2300255 PA2300256 PA2300257 PA2300258 PA2300259 PA2300260 PA2300303 PA2302321 PA2302322 PA2312695 PA2312696 PA2312697 PA2312698 PA2312699 PA2312700 PA2312701 PA2312702 PA2312703 PA2312704 PA2312705 PA2312706 PA2312707 PA2312708 PA2312709 PA2312710 PA2312711 PA2312712 PA2312713 PA2312714 PA2312715 PA2312731 PA2312732 PA2312733 PA2312734 PA2312735 PA2312736 PA2312737 PA2312738 PA2312739 PA2312740 PA2312741 PA2312743 PA2312744 PA2312745 PA2312746 PA2312747 PA2312748 PA2312749 PA2312750 PA2312751 PA2312752 PA2312753 PA2312754 PA2312755 PA2312756 PA2312757 PA2312758 PA2312759 PA2312760 PA2312761 PA2312762 PA2312763 PA2312764 PA2312765 PA2312766 PA2312767 PA2312768 PA2312769 PA2312770 PA2312771 PA2312772 PA2312773 PA2312774 PA2312775 PA2312776 PA2312777 PA2312778 PA2312779 PA2312780 PA2312781 PA2312782 PA2312783 PA2312784 PA2312785 PA2312786 PA2312787 PA2312788 PA2312789 PA2312790 PA2312791 PA2312877 PA2312878 PA2312879 PA2312880 PA2312881 PA2312882 PA2312883 PA2312884 PA2312885 PA2312886 PA2312887 PA2312888 PA2312889 PA2312890 PA2312891 PA2312892 PA2312893 PA2312894 PA2312895 PA2312896 PA2312897 PA2312898 PA2312899 PA2312900 PA2312901 PA2312978 PA2312979 PA2312980 PA2312981 PA2312982 PA2312983 PA2312984 PA2312985 PA2312986 PA2312987 PA2312988 PA2312989 PA2312990 PA2312991 PA2312992 PA2312993 PA2312994 PA2312995 PA2312996 PA2312997 PA2312998 PA2312999 PA2313000 PA2313001 PA2313002 PA2313003 PA2313004 PA2313005 PA2313006 PA2313007 PA2313008 PA2313009 PA2313010 PA2313011 PA2313012 PA2313013 PA2313014 PA2313015 PA2313016 PA2313017 PA2313018 PA2313019 PA2313020 PA2313021 PA2313022 PA2313023 PA2313024 PA2313025 PA2313026 PA2313027 PA2313137 PA2313340 PA2313341 PA2313342 PA2313343 PA2313344 PA2313345 PA2313346 PA2313347 PA2313348 PA2313349 PA2313350 PA2313351 PA2313558 PA2313559 PA2313560 PA2313561 PA2313562 PA2313563 PA2313564 PA2313565 PA2313567 PA2313568 PA2313569 PA2313570 PA2313571 PA2313620 PA2313621 PA2313653 PA2313658 PA2313659 PA2313660 PA2313661 PA2313662 PA2313663 PA2313664 PA2313665 PA2313666 PA2313667 PA2313668 PA2313669 PA2313670 PA2313671 PA2313672 PA2313673 PA2313674 PA2313675 PA2313676 PA2313677 PA2313678 PA2313740 PA2313741 PA2313742 PA2313743 PA2313744 PA2313745 PA2313746 PA2313747 PA2313748 PA2313749 PA2313750 PA2313751 PA2313752 PA2313754 PA2313755 PA2313756 PA2313757 PA2313760 PA2313761 PA2313763 PA2313765 PA2313766 PA2313767 PA2313768 PA2313769 PA2313770 PA2313771 PA2313772 PA2313773 PA2313774 PA2313775 PA2313776 PA2313777 PA2313778 PA2313779 PA2313780 PA2313781 PA2313782 PA2313783 PA2313784 PA2313785 PA2313786 PA2313787 PA2313788 PA2313789 PA2313808 PA2313809 PA2313810 PA2313811 PA2313812 PA2313813 PA2313814 PA2313815 PA2313816 PA2313817 PA2313818 PA2313819 PA2313820 PA2313821 PA2313822 PA2313823 PA2313824 PA2313825 PA2313826 PA2313827 PA2313828 PA2313829 PA2313830 PA2313831 PA2313832 PA2313833 PA2313834 PA2313835 PA2313836 PA2313837 PA2313838 PA2313839 PA2313840 PA2313841 PA2313842 PA2313843 PA2313844 PA2313845 PA2313846 PA2313847 PA2313848 PA2313849 PA2313850 PA2313851 PA2313873 PA2313874 PA2313875 PA2313876 PA2313877 PA2313878 PA2313879 PA2313880 PA2313881 PA2313882 PA2313883 PA2313884 PA2313885 PA2313886 PA2313887 PA2313888 PA2313889 PA2313890 PA2313891 PA2313892 PA2313893 PA2313894 PA2313895 PA2313896 PA2313897 PA2313898 PA2313899 PA2313900 PA2313901 PA2313902 PA2313903 PA2313904 PA2313905 PA2313906 PA2313907 PA2313908 PA2313909 PA2313910 PA2313911 PA2313912 PA2314190 PA2314191 PA2314192 PA2314193 PA2314194 PA2314195 PA2314196 PA2314197 PA2314198 PA2314199 PA2314200 PA2314201 PA2314202 PA2314203 PA2314204 PA2314205 PA2314206 PA2314207 PA2314208 PA2314209 PA2314210 PA2314211 PA2314212 PA2314213 PA2314214 PA2314215 PA2314216 PA2314217 PA2314218 PA2314219 PA2314416 PA2314417 PA2314418 PA2314419 PA2314420 PA2314421 PA2314422 PA2314423 PA2314424 PA2314425 PA2314434 PA2314435 PA2314436 PA2314437 PA2314438 PA2314439 PA2314498 PA2314499 PA2314500 PA2314501 PA2314502 PA2314503 PA2314504 PA2314505 PA2314506 PA2314579 PA2314580 PA2314581 PA2314582 PA2314583 PA2314584 PA2314585 PA2314586 PA2314587 PA2314588 PA2314589 PA2314590 PA2314591 PA2314592 PA2314593 PA2314594 PA2314595 PA2314596 PA2314597 PA2314598 PA2314599 PA2314600 PA2314601 PA2314602 PA2314603 PA2314604 PA2314605 PA2314606 PA2314622 PA2314625 PA2314626 PA2314690 PA2314751 PA2314831 PA2314832 PA2314833 PA2314934 PA2314935 PA2314936 PA2314937 PA2314938 PA2314939 PA2314940 PA2314941 PA2314942 PA2314943 PA2314944 PA2314945 PA2314946 PA2314947 PA2314948 PA2352514 PA2352515 PA2352551 PA2352554 PA2352555 PA2352582 PA2352583 PA2352584 PA2352585 PA2352592 PA2352593 PA2352594 PA2352595 PA2352596 PA2352597 PA2352598 PA2352599 PA2352600 PA2352601 PA2352602 PA2352603 PA2352604 PA2352605 PA2352606 PA2352607 PA2352608 PA2352609 PA2352610 PA2352611 PA2352612 PA2352613 PA2352614 PA2352615 PA2352616 PA2352617 PA2352618 PA2352619 PA2352625 PA2352626 PA2352627 PA2352628 PA2352629 PA2352630 PA2352631 PA2352632 PA2352633 PA2352634 PA2352635 PA2352636 PA2352637 PA2352638 PA2352639 PA2352640 PA2352641 PA2352642 PA2352643 PA2352644 PA2352645 PA2352646 PA2352647 PA2352648 PA2352649 PA2352650 PA2352651 PA2352652 PA2352653 PA2352654 PA2352655 PA2352656 PA2352657 PA2352658 PA2352659 PA2352660 PA2352661 PA2352662 PA2352663 PA2352664 PA2352665 PA2352666 PA2352667 PA2352668 PA2352669 PA2352670 PA2352671 PA2352672 PA2352673 PA2352674 PA2352675 PA2352676 PA2352677 PA2352678 PA2352679 PA2352680 PA2352681 PA2352682 PA2352683 PA2352684 PA2352685 PA2352686 PA2352687 PA2352688 PA2352689 PA2352690 PA2352691 PA2352692 PA2352693 PA2352694 PA2352695 PA2352696 PA2352697 PA2352698 PA2352699 PA2352700 PA2352701 PA2352702 PA2352703 PA2352704 PA2352705 PA2352706 PA2352707 PA2352708 PA2352709 PA2352710 PA2352711 PA2352712 PA2352713 PA2352714 PA2352715 PA2352716 PA2352717 PA2352718 PA2352719 PA2352720 PA2352721 PA2352722 PA2352723 PA2352724 PA2352725 PA2352726 PA2352727 PA2352728 PA2352729 PA2352730 PA2352731 PA2352732 PA2352733 PA2352734 PA2352735 PA2352736 PA2352737 PA2352738 PA2352739 PA2352740 PA2352741 PA2352742 PA2352743 PA2352744 PA2352745 PA2352746 PA2352747 PA2352748 PA2352749 PA2352750 PA2352751 PA2352752 PA2352753 PA2352754 PA2352755 PA2352756 PA2352757 PA2352758 PA2352759 PA2352760 PA2352761 PA2352762 PA2352763 PA2352764 PA2352765 PA2352766 PA2352767 PA2352768 PA2352769 PA2352770 PA2352771 PA2352772 PA2352773 PA2352774 PA2352775 PA2352776 PA2352777 PA2352778 PA2352779 PA2352780 PA2352781 PA2352782 PA2352783 PA2352784 PA2352788 PA2352789 PA2352790 PA2352791 PA2352792 PA2352793 PA2352794 PA2352795 PA2352796 PA2352797 PA2352798 PA2352799 PA2352800 PA2352801 PA2352802 PA2352803 PA2352804 PA2352805 PA2352806 PA2352807 PA2352808 PA2352809 PA2352810 PA2352811 PA2352812 PA2352813 PA2352814 PA2352815 PA2352816 PA2352817 PA2352818 PA2352819 PA2352820 PA2352821 PA2352822 PA2352823 PA2352824 PA2352825 PA2352826 PA2352827 PA2352828 PA2352829 PA2352830 PA2352831 PA2352832 PA2352833 PA2352834 PA2352835 PA2352836 PA2352837 PA2352838 R3S04450035GA R3S04490127GA R3S04490129GA R3S04490131GA R3S04490133GA R3S04490134GA R3S04490137GA R3S04490138GA R3S04490139GA R3S05020263GA R3S05050286GA R3S05140399GA R3T05072105GA R3T05072110GA R3T05072111GA R3T05072114GA R3T05072120GA R3T05072123GA R3T05162726GA R3T05162733GA R3T05162736GA R3T05162737GA R3T05162738GA R3T05182961GA R3T05182962GA R3T05182963GA R3T05182966GA R3T05182967GA R3T05182969GA R3T05182970GA R3T05253351GA SBH05260052GA SBH05260053GA SBH05260056GA SBH05260057GA SBH05260058GA SBH05260081GA SBH05260108GA SBH05260109GA SBH05260111GA SBH05260112GA SBH05260113GA SBH05260117GA SBH05280122GA SBH05280124GA SBH05280207GA SBH05280212GA SBH05300380GA SBH05300387GA SBH05300388GA SBH05300396GA SBH05440948GA SBH05471036GA SBH05471066GA SBH05471073GA SBH06061578GA SBH06061580GA SBH06091715GA SBH06091716GA SBL06527807GR SBL06527809GR SBL06527811GR SBL06527822GR SBL07017823GR SBL07017892GR SBL07017894GR SBL07017895GR SBL07017901GR SBL07017906GR SBL07027934GR SBL07038115GR SBL07038163GR SBL07038169GR SBL07078548GR SBL07149225GR SBL07149226GR SBL07149227GR SBL07149229GR SBL07149230GR SBM05260121GR SBM05340190GR SBM05370247GR SBM05370253GR SBM05380320GA SBM05390382GR SBM05420430GR SBM05440465GR SBM05440472GR SBM05440478GA SBM06010826GR SBM06391194GA SCZ07200023GR SCZ07240337GR SCZ07240355GR SCZ07240358GR SCZ07361465GR SCZ07361467GR SCZ07361468GR SCZ07361472GR SCZ07361478GR SCZ07412925GR SCZ07423152GR SCZ07494359GR SCZ07494360GR SCZ07494361GR SCZ07494362GR SCZ07494364GR SCZ07504273GR SCZ07504370GR SCZ07504371GR SCZ07524793GR SE408100752GR SE408131482GR SE408131483GR SE408131484GR SE408131488GR SE408131489GR SE408131492GR SE408131509GR SE408131514GR SE408131518GR SE408141585GR SE408141602GR SE408232436GR SE408436627GR SE408456813GR SE408456814GR SE408456816GR SE408456817GR SE408456833GR SE408456835GR SE408456836GR SE408456837GR SE408456838GR SE408456839GR SE408456842GR SE408456843GR SE408456847GR SE408456849GR SE408456852GR SE408456858GR SE408456862GR SE408456863GR SE408456865GR SE408456868GR SE408466997GR SE408487343GR SE408487348GR SE408518061GR SE409119163GR SE409250439GR SE409250466GR SE409260470GR SE409260479GR SE409260481GR SE409341418GA SE409351619GR SE409371947GR SE409371950GR SE409432543GR SE409432638GA SE409432638GR SE409442685GA SE409442799GA SE409442815GA SE409442816GA SE409442817GA SE409442820GA SE409442824GA SE409442829GA SE409442830GA SE409442831GA SE409442832GA SE409442833GA SE409452834GA SE409452835GA SE409452836GA SE409452837GA SE409452838GA SE409452840GA SE409452841GA SE409452846GA SE409452848GA SE409452850GA SE409452851GA SE409452853GA SE409452854GA SE409452855GA SE409452856GA SE409452857GA SE409452858GA SE409452958GA SE409452961GA SE409452972GA SE409463005GA SE409463010GA SE409463015GA SE409463016GA SE409463025GA SE409463027GA SE409463030GA SE409463034GA SE409463035GA SE409463036GA SE409463038GA SE409463041GA SE409463043GA SE409463045GA SE409463046GA SE409463049GA SE409463050GA SE409463056GA SE409463058GA SE409463081GA SE409463089GA SE409463091GA SE409463093GA SE409463094GA SE409463096GA SE409463098GA SE409463103GA SE409463105GA SE409463106GA SE409463107GA SE409463110GA SE409463111GA SE409463116GA SE409463117GA SE409463119GA SE409463131GA SE409473140GA SE409473141GA SE409473142GA SE409473165GA SE409473169GA SE409473177GA SE409473178GA SE409473179GA SE409473182GA SE409473186GA SE409473187GA SE409473188GA SE409473192GA SE409473193GA SE409473196GA SE409473198GA SE409473200GA SE409473211GA SE409473212GA SE409473220GA SE409473222GA SE409473225GA SE409473226GA SE409473227GA SE409473228GA SE409473229GA SE409473230GA SE409473231GA SE409473232GA SE409473233GA SE409473234GA SE409473235GA SE409473236GA SE409473239GA SE409473240GA SE409473241GA SE409473242GA SE409473243GA SE409473244GA SE409473245GA SE409473246GA SE409473252GA SE409473253GA SE409473256GA SE409493288GA SE409493289GA SE409493290GA SE409493291GA SE409493292GA SE409493293GA SE409493294GA SE409493295GA SE409493307GA SE409493315GA SE409493316GA SE409493317GA SE409493318GA SE409493319GA SE409493321GA SE409493322GA SE409493324GA SE409493337GA SE409493338GA SE409493339GA SE409493341GA SE409493342GA SE409493343GA SE409493345GA SE409493346GA SE409493347GA SE409493348GR SE409493349GA SE409493350GA SE409493351GA SE409493352GA SE409493353GA SE409493362GA SE409493363GA SE409493364GA SE409493365GA SE409493371GA SE409493372GA SE409493373GA SE409493374GA SE409493375GR SE409493376GA SE409493378GA SE409493381GA SE409493383GA SE409493384GA SE409493385GR SE409493386GA SE409493388GA SE409493389GA SE409493390GA SE409493391GA SE409493404GA SE409493408GA SE409493409GA SE409493412GA SE409493413GA SE409503416GA SE409503418GA SE409503419GA SE409503420GA SE409503426GA SE409503427GA SE409503450GR SE409503455GA SE409503457GA SE409503459GA SE409503462GA SE409503464GA SE409503467GA SE409503468GA SE409503471GA SE409503473GA SE409503474GA SE409503475GA SE409503476GA SE409503477GA SE409503478GA SE409503479GA SE409503480GA SE409503481GA SE409503483GA SE409503485GA SE409503496GA SE409503499GA SE409503500GA SE409503501GA SE
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE Healthcare, LLC
- Reason for Recall:
- GE Healthcare has recently become aware of a potential safety issue due to a software issue associated with the Head Scan Protocols used on your Optima CT580 or Discovery CT590RT scanner. A potential hazardous situation may occur during a CT head scan causing a novel and rare artifact that may hide pathology. When using the head and large head SFOV, which applies the Iterative Bone Option (IBO)
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
GE Healthcare Optima CT580/Discovery CT590RT scanners running software version 13BW16.3. The Discovery CT 590 RT and Optima CT 580 series of Computed Tomography systems are intended to produce cross-sectional images of the body by computer reconstruction of XRay transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), and Gated (Respiratory and Cardiac) acquisitions in patients of all ages.
Product Codes/Lot Numbers:
Mfg Lot or Serial # PA2314840 PA2314841 PA2314842 PA2314843 PA2314844 PA2314845 A1RF3176G A1RF3638GR A2RF1052G B1RF4038GR B3G44377G B3G44511G B3G44515G B3RF4960G C1RF4686G C1RF4793G C1RF4802G C1RF4827GR D2RF2744G D2RF2778GR D2RF2795G D2RF2796G D2RF2797G D2RF2798G D2RF2799G D2RF2814G D2RF2815G D2RF2816G D2RF2817G D2RF2818G D2RF2819G D2RF2820G D2RF2821G D2RF2822G D2RF2823G D2RF2824G D2RF2825G D2RF2826G D2RF2827G D2RF2828G D2RF2829G D2RF2830G D2RF2831G D2RF2832G D2RF2833G D2RF2834G D2RF2835G D2RF2836G D2RF2837G D2RF2838G D2RF2839G D2RF2840G D2RF2841G D2RF2842G D2RF2843G D2RF2844G D2RF2845G D2RF2846G D2RF2847G D2RF2848G D2RF2850G D2RF2851G D3G44977G D3G44983G D3G44984G D3G44985G D43T4071G D43T4072G D43T4073G D43T4074G D43T4075G D43T4076G D43T4077G E1RF5364G E1RF5479GR E1RF5480GR E1RF5599G E3RF5317G E3RF5318G F0RF0134G F0RF0135G F0RF0136G F0RF0140G F0RF0141G F1RF4827GR F1RF5952G F1RF6042GR F1RF6075GR F1RF6079GR F1RF6084GR F1RF6085GR F1RF6086GR F1RF6088GR F1RF6092GR F1RF6097GR F1RF6100GR F1RF6114GR F1RF6118GR F1RF6123GR F1RF6130GR F1RF6227GR F1RF6236G F1RF6236GR F1RF6243GR F1RF6282GR F2G40578G F2G40579G F2G40580G F2G40581G F2G40584G F2G40585G F2G40586G F2G40588G F2G40592G F2G40593G F2G40594G F2G40595G F2G40596G F2G40598G F2G40599G F2G40605G F2G40606G F2G40607G F2G40609G F2G40610G F2G40611G F2G40615G F2G40631G F2G40705G F2G40820G F2G40835G F2RF3572G F2RF3640G F2RF3651G F3G45957G F3G45980G F3G46200G F43S0355G F43S0356G G0RF0516GR G0RF0592GR G0RF0596GR G1RF6518G G1RF6544G G1RF6546G G3G46241G G3G46616G G3G46649G G3G46664G G3G46674G G3G46679G H0RF0607GR H0RF0772G H0RF0845GR H2G41289G H2G41307G H2G41571GR H2G41786G H2G41787G H2RF4002GR J0RF0465G J0RF1103G J1RF7280G J1RF7281G J1RF7282G J1RF7283G J1RF7284G J1RF7285G J1RF7286G J1RF7289G J1RF7290G J1RF7298G J1RF7387G J2G42049G J2G42100GR J33T0400G J43S0499G J43S0503G J43S0504G J43S0506G J43T5719G J43T5720G J43T5727G J43T5731G J43T5799G J43T5801G J43T5810G J43T5812G J43T5813G J43T5814G J43T5815G J43T5827G J43T5831G J43T5832G J43T5877G J43T6078G K1RF7644G K1RF7673G K1RF7674G K1RF7675G K1RF7676G K1RF7678G K1RF7679G K1RF7680G K1RF7681G K1RF7683G K1RF7704G K1RF7925GR K33T0621G K33T1002G K43T6545G L1RF8243G L1RF8254G L1RF8255G L1RF8256G L1RF8258G L1RF8259G L1RF8261G L1RF8263G L1RF8264G L1RF8265G L1RF8267G L1RF8728G L1RF8733G L2G42609G L2G42643G L2G42666G L2G42668G L2G42669G L2G42690G L2G42711G L2G42970G L2RF4373GR L33T1324G M0RF2496G M0RF2562GR M1RF9841G M2G43248G M2RF4686G M33T1729G M33T2293G PA1786357 PA1786438 PA1786439 PA1786440 PA1786441 PA1786442 PA1786443 PA1786444 PA1786445 PA1786446 PA1786447 PA1786448 PA1786449 PA1786450 PA1786451 PA1786452 PA1786453 PA1786454 PA1786456 PA1786457 PA1786458 PA1786461 PA1786462 PA1786494 PA1786495 PA1786496 PA1786497 PA1786498 PA1786499 PA1786500 PA1786501 PA1786502 PA1786503 PA1786504 PA1786505 PA1786530 PA1786531 PA1786532 PA1786533 PA1786534 PA1786535 PA1786536 PA1786537 PA1786538 PA1786539 PA1786540 PA1786541 PA1786542 PA1786543 PA1786544 PA1786545 PA1786546 PA1786547 PA1786548 PA1786549 PA1786677 PA1786678 PA1786679 PA1786680 PA1786681 PA1786682 PA1786683 PA1786684 PA1786685 PA1786686 PA1786687 PA1786688 PA1786689 PA1786690 PA1786691 PA1786692 PA1786693 PA1786694 PA1786695 PA1786696 PA1786697 PA1786698 PA1786699 PA1786700 PA1786701 PA1786702 PA1786703 PA1786704 PA1786705 PA1786706 PA1786707 PA1786708 PA1786709 PA1786710 PA1786711 PA1786712 PA1786713 PA1786714 PA1786715 PA1786716 PA1787000 PA1787001 PA1787002 PA1787003 PA1787004 PA1787005 PA1787006 PA1787007 PA1787008 PA1787009 PA1787010 PA1787011 PA1787079 PA1787080 PA1787081 PA1787082 PA1787083 PA1787084 PA1787085 PA1787086 PA1787087 PA1787088 PA1787351 PA1787352 PA1787353 PA1787354 PA1787355 PA1787356 PA1787357 PA1787478 PA1787479 PA1787480 PA1787481 PA1787482 PA1787483 PA1787484 PA1787485 PA1787561 PA1787562 PA1787563 PA1787564 PA1787565 PA1787566 PA1787567 PA1787568 PA1787569 PA1787570 PA1787571 PA1787572 PA1787573 PA1787574 PA1787575 PA1787576 PA1787577 PA1787578 PA1787579 PA1787580 PA1787581 PA1787582 PA1787583 PA1787584 PA1787585 PA1787586 PA1787587 PA1787588 PA1787589 PA1787590 PA1787591 PA1787592 PA1787593 PA1787594 PA1787595 PA1787596 PA1787597 PA1787598 PA1787599 PA1787600 PA1787601 PA1787602 PA1787603 PA1787604 PA1787605 PA1787606 PA1787607 PA1787608 PA1787609 PA1787610 PA1787611 PA1787612 PA1787613 PA1787614 PA1787615 PA1787616 PA1787617 PA1787618 PA1787619 PA1787620 PA1787621 PA1787622 PA1787623 PA1787624 PA1787625 PA1787626 PA1787627 PA1787628 PA1787629 PA1787630 PA1787631 PA1787632 PA1787633 PA1787634 PA1787635 PA1787636 PA1787637 PA1787638 PA1787639 PA1787640 PA1787641 PA1787642 PA1787643 PA1787644 PA1787645 PA1787646 PA1787647 PA1787648 PA1787649 PA1787650 PA1787651 PA1787652 PA1787653 PA1787654 PA1787655 PA1787656 PA1787657 PA1787658 PA1787659 PA1787660 PA1787661 PA1787662 PA1787663 PA1787664 PA1787665 PA1787666 PA1787667 PA1787668 PA1787669 PA1787670 PA1787671 PA1787672 PA1788113 PA1788114 PA1788115 PA1788116 PA1788117 PA1788146 PA1788147 PA1788148 PA1788149 PA1788150 PA1788151 PA1788152 PA1788153 PA1788154 PA1788155 PA1788156 PA1788157 PA1788158 PA1788159 PA1788160 PA1788161 PA1788162 PA1788163 PA1788164 PA1788165 PA1788166 PA1788167 PA1788168 PA1788169 PA1788170 PA1788171 PA1788172 PA1788173 PA1788174 PA1788175 PA1788176 PA1788177 PA1788178 PA1788179 PA1788180 PA1788310 PA1788311 PA1788312 PA1788313 PA1788314 PA1788315 PA1788316 PA1788317 PA1788318 PA1788319 PA1788320 PA1788321 PA1788322 PA1788323 PA1788324 PA1788325 PA1788326 PA1788327 PA1788328 PA1788329 PA1788330 PA1788331 PA1788332 PA1788333 PA1788334 PA1788335 PA1788336 PA1788337 PA1788338 PA1788339 PA1788360 PA1788361 PA1788362 PA1788363 PA1788364 PA1788365 PA1788366 PA1788367 PA1788368 PA1788370 PA1788371 PA1788372 PA1788373 PA1788375 PA1788378 PA1788379 PA1788380 PA1788381 PA1788382 PA1788384 PA1788407 PA1788408 PA1788409 PA1788410 PA1788411 PA1788412 PA1788413 PA1788414 PA1788415 PA1788416 PA1788417 PA1788418 PA1788419 PA1788420 PA1788421 PA1788422 PA1788423 PA1788424 PA1788425 PA1788426 PA1788427 PA1788428 PA1788429 PA1788430 PA1788431 PA1788432 PA1788433 PA1788434 PA1788435 PA1788436 PA1788471 PA1788472 PA1788473 PA1788474 PA1788475 PA1788476 PA1788477 PA1788478 PA1788479 PA1788480 PA1788481 PA1788482 PA1788483 PA1788484 PA1788485 PA1788486 PA1788487 PA1788488 PA1788489 PA1788490 PA1788491 PA1788492 PA1788493 PA1788494 PA1788495 PA1788496 PA1788497 PA1788498 PA1788499 PA1788500 PA1788501 PA1788502 PA1788503 PA1788504 PA1788505 PA1788506 PA1788507 PA1788508 PA1788509 PA1788510 PA1788511 PA1788512 PA1788513 PA1788514 PA1788515 PA1788516 PA1788517 PA1788518 PA1788519 PA1788520 PA1788521 PA1788522 PA1788523 PA1788524 PA1788525 PA1788526 PA1788527 PA1788528 PA1788529 PA1788530 PA1788531 PA1788532 PA1788533 PA1788534 PA1788535 PA1788625 PA1788626 PA1788627 PA1788628 PA1788629 PA1788630 PA1788631 PA1788632 PA1788633 PA1788634 PA1789287 PA1789288 PA1789289 PA1789290 PA1789291 PA1789292 PA1789293 PA1789294 PA1789295 PA1789296 PA1850116 PA1850117 PA1850118 PA1850119 PA1850120 PA1850121 PA1850122 PA1850123 PA1850124 PA1850125 PA1850126 PA1850127 PA1850128 PA1850129 PA1850130 PA1850131 PA1850132 PA1850133 PA1850134 PA1850135 PA1850136 PA1850137 PA1850138 PA1850139 PA1850140 PA1850141 PA1850142 PA1850143 PA1850144 PA1850145 PA1850146 PA1850147 PA1850391 PA1850392 PA1850393 PA1850394 PA1850395 PA1850396 PA1850397 PA1850398 PA1850399 PA1850400 PA1850401 PA1850402 PA1850403 PA1850404 PA1850405 PA1850406 PA1850407 PA1850408 PA1850409 PA1850410 PA1850411 PA1850412 PA1850413 PA1850414 PA1850415 PA1850416 PA1850417 PA1850418 PA1850419 PA1850420 PA1850421 PA1850422 PA1850423 PA1850424 PA1850425 PA1850426 PA1850427 PA1850428 PA1850771 PA1850772 PA1850773 PA1850774 PA1850775 PA1850776 PA1850777 PA1850778 PA1850779 PA1850791 PA1850792 PA1850793 PA1850794 PA1850795 PA1850796 PA1850797 PA1850798 PA1850897 PA1850898 PA1850899 PA1850900 PA1850901 PA1850902 PA1850903 PA1850904 PA1850905 PA1850906 PA1850907 PA1850908 PA1850909 PA1850910 PA1850911 PA1850912 PA1850913 PA1850914 PA1850915 PA1850916 PA1850917 PA1850918 PA1850919 PA1850920 PA1850921 PA1850922 PA1850923 PA1850924 PA1850925 PA1850926 PA1850927 PA1850928 PA1850929 PA1850930 PA1850931 PA1851412 PA1851413 PA1851414 PA1851415 PA1851416 PA1851829 PA1851830 PA1851831 PA1851832 PA1851833 PA1851834 PA1851835 PA1851836 PA1851837 PA1851838 PA1851839 PA1851840 PA1851841 PA1851842 PA1851843 PA1851844 PA1852084 PA1852085 PA1852086 PA1852087 PA1852088 PA1852089 PA1852090 PA1852091 PA1852092 PA1852093 PA1852094 PA1852095 PA1852096 PA1852097 PA1852098 PA1852099 PA1852138 PA1852139 PA1852140 PA1852141 PA1852142 PA1852143 PA1852144 PA1852145 PA1852146 PA1852147 PA1852148 PA1852149 PA1852150 PA1852151 PA1852152 PA1852153 PA1852154 PA1852155 PA1852156 PA1852157 PA1852158 PA1852159 PA1852160 PA1852161 PA1852162 PA1852163 PA1852164 PA1852165 PA1852166 PA1852167 PA1852168 PA1852169 PA1852170 PA1852171 PA1852172 PA1852173 PA1852174 PA1852175 PA1852176 PA1852177 PA1852178 PA1852179 PA1852180 PA1852181 PA1852182 PA1852183 PA1852184 PA1852185 PA1852186 PA1852187 PA1852188 PA1852189 PA1852190 PA1852191 PA1852192 PA1852193 PA1852194 PA1852195 PA1852196 PA1852197 PA1852198 PA1852199 PA1852200 PA1852201 PA1852202 PA1852203 PA1852204 PA1852205 PA1852206 PA1852207 PA1852208 PA1852209 PA1852210 PA1852211 PA1852212 PA1852213 PA1852214 PA1852215 PA1852216 PA1852217 PA1852218 PA1852219 PA1852220 PA1852221 PA1852222 PA1852223 PA1852224 PA1852225 PA1852226 PA1852227 PA1852228 PA1852229 PA1852230 PA1852231 PA1852232 PA1852233 PA1852234 PA1852235 PA1852236 PA1852237 PA1852409 PA1852410 PA1852411 PA1852412 PA1852413 PA1852414 PA1852415 PA1852416 PA1852417 PA1852418 PA1852419 PA1852420 PA1852421 PA1852422 PA1852423 PA1852424 PA1852425 PA1852426 PA1852427 PA1852428 PA1852429 PA1852430 PA1852431 PA1852432 PA1852433 PA1852434 PA1852435 PA1852436 PA1852437 PA1852438 PA1852439 PA1852440 PA1852441 PA1852442 PA1852443 PA1852444 PA1852445 PA1852446 PA1852447 PA1852448 PA1852449 PA1852808 PA1852809 PA1852810 PA1852811 PA1852812 PA1852813 PA1852814 PA1852815 PA1852816 PA1852817 PA1852818 PA1852819 PA1852820 PA1852821 PA1852822 PA1852823 PA1852824 PA1854656 PA1854657 PA1854658 PA1854659 PA1854660 PA1854661 PA1895455 PA1895456 PA1895457 PA1895458 PA1897813 PA1990824 PA1990825 PA1990826 PA1990827 PA1990828 PA1990829 PA1990830 PA1990831 PA1990832 PA1990833 PA1990834 PA1990835 PA1990836 PA1990837 PA1990838 PA1990839 PA1990840 PA1991403 PA1991404 PA1991405 PA1991406 PA1991407 PA1991408 PA1991409 PA1991410 PA1991411 PA1991412 PA1991413 PA1991414 PA1991415 PA1991416 PA1991417 PA1991418 PA1991419 PA1991440 PA1991442 PA1991635 PA1991851 PA1991852 PA1991853 PA1991854 PA1991855 PA1991856 PA1991857 PA1991858 PA1991859 PA1991860 PA1991861 PA1991862 PA1991863 PA1991864 PA1991865 PA1991866 PA1991867 PA1991868 PA1991869 PA1991870 PA1992005 PA1992006 PA1992007 PA1992008 PA1992112 PA1992113 PA1992114 PA1992115 PA1992116 PA1992117 PA1992118 PA1992119 PA1992120 PA1992121 PA1992122 PA1992123 PA1992124 PA1992125 PA1992126 PA1992127 PA1993645 PA1993646 PA1993647 PA1993648 PA1993649 PA1993650 PA1993651 PA1993652 PA1993653 PA1993654 PA1993655 PA1993656 PA1993657 PA1993658 PA1993659 PA1993660 PA1993661 PA1993662 PA1993663 PA1993664 PA1993665 PA1993666 PA1993667 PA1993668 PA1993860 PA1993861 PA1993862 PA1993863 PA1993864 PA1993865 PA1993866 PA1994201 PA1994202 PA1994203 PA1994204 PA1994205 PA2007158 PA2007159 PA2007160 PA2007161 PA2007162 PA2007163 PA2020375 PA2020376 PA2020377 PA2020378 PA2020379 PA2020380 PA2020381 PA2020382 PA2020383 PA2020384 PA2020385 PA2020386 PA2020387 PA2020388 PA2020389 PA2020390 PA2020391 PA2020392 PA2020458 PA2020459 PA2020460 PA2020461 PA2020462 PA2020463 PA2020464 PA2020465 PA2020466 PA2020467 PA2020468 PA2020469 PA2020470 PA2020471 PA2020472 PA2020473 PA2020474 PA2020475 PA2020476 PA2020477 PA2020478 PA2020479 PA2020480 PA2020481 PA2020482 PA2020505 PA2020506 PA2020507 PA2020508 PA2020509 PA2020510 PA2020511 PA2020512 PA2020513 PA2020514 PA2020515 PA2020516 PA2020517 PA2020518 PA2020519 PA2020520 PA2020521 PA2020522 PA2020523 PA2020524 PA2021288 PA2021289 PA2021290 PA2021291 PA2021292 PA2021293 PA2021294 PA2021295 PA2021296 PA2021297 PA2021806 PA2021807 PA2021808 PA2021809 PA2021810 PA2021811 PA2021812 PA2021813 PA2021814 PA2021815 PA2021816 PA2021817 PA2022029 PA2022030 PA2022032 PA2022034 PA2107909 PA2107910 PA2107911 PA2107912 PA2107913 PA2107914 PA2107915 PA2108512 PA2108513 PA2108514 PA2109592 PA2109593 PA2109594 PA2109595 PA2109596 PA2109597 PA2109598 PA2109599 PA2109600 PA2109601 PA2115014 PA2115015 PA2115016 PA2115017 PA2115018 PA2115019 PA2115020 PA2115021 PA2115022 PA2115023 PA2115024 PA2115025 PA2115026 PA2115027 PA2115028 PA2115029 PA2115030 PA2115031 PA2115032 PA2115033 PA2115034 PA2115035 PA2115036 PA2115037 PA2115038 PA2115039 PA2115040 PA2115041 PA2115042 PA2115043 PA2115044 PA2115045 PA2115046 PA2115049 PA2115212 PA2115213 PA2115214 PA2115215 PA2115216 PA2115217 PA2115218 PA2115219 PA2115220 PA2115221 PA2115222 PA2115232 PA2115233 PA2115234 PA2115235 PA2115236 PA2115237 PA2115309 PA2115310 PA2115311 PA2115312 PA2115313 PA2115314 PA2115315 PA2115316 PA2115317 PA2115318 PA2115319 PA2115320 PA2115321 PA2115322 PA2115323 PA2115327 PA2115329 PA2115330 PA2115331 PA2115332 PA2115333 PA2115334 PA2115335 PA2115336 PA2115337 PA2115338 PA2115347 PA2115348 PA2116026 PA2116027 PA2116028 PA2116029 PA2116030 PA2116031 PA2116032 PA2116033 PA2116034 PA2116035 PA2116036 PA2116037 PA2130461 PA2130462 PA2130463 PA2130464 PA2130465 PA2130466 PA2130467 PA2130468 PA2130469 PA2130470 PA2130471 PA2130472 PA2130473 PA2130474 PA2130475 PA2130476 PA2130477 PA2130478 PA2131319 PA2131320 PA2131321 PA2131322 PA2131323 PA2131324 PA2131325 PA2131326 PA2131327 PA2131328 PA2136382 PA2136383 PA2136384 PA2136385 PA2136386 PA2136387 PA2136388 PA2136391 PA2136398 PA2136399 PA2136408 PA2136409 PA2136549 PA2136550 PA2136551 PA2136552 PA2136553 PA2136558 PA2136559 PA2136560 PA2136561 PA2136562 PA2136563 PA2136584 PA2136585 PA2136586 PA2136587 PA2136588 PA2136589 PA2136590 PA2136591 PA2136592 PA2136593 PA2137621 PA2138188 PA2138189 PA2138190 PA2138191 PA2138208 PA2138248 PA2138280 PA2138337 PA2138338 PA2138339 PA2138340 PA2138341 PA2138342 PA2138343 PA2138344 PA2138345 PA2138346 PA2138347 PA2138348 PA2138349 PA2138388 PA2138389 PA2138390 PA2138391 PA2138392 PA2138393 PA2138394 PA2138395 PA2138396 PA2138397 PA2138398 PA2138399 PA2138400 PA2138401 PA2138402 PA2138403 PA2138404 PA2138405 PA2138406 PA2138407 PA2138408 PA2138409 PA2138410 PA2138411 PA2138412 PA2138413 PA2138414 PA2138415 PA2138416 PA2138417 PA2138418 PA2138419 PA2138420 PA2138421 PA2138422 PA2138423 PA2138424 PA2138425 PA2138427 PA2138428 PA2138429 PA2138430 PA2138431 PA2138432 PA2138433 PA2138434 PA2138435 PA2138436 PA2138437 PA2138438 PA2138439 PA2138440 PA2138441 PA2138442 PA2138443 PA2138444 PA2138445 PA2138446 PA2138447 PA2138448 PA2138449 PA2138450 PA2138451 PA2138452 PA2138453 PA2138514 PA2138515 PA2138516 PA2138517 PA2138518 PA2138519 PA2138520 PA2138521 PA2138522 PA2138822 PA2138823 PA2138824 PA2138825 PA2138826 PA2138827 PA2138828 PA2138829 PA2139335 PA2139336 PA2139337 PA2139338 PA2139339 PA2139340 PA2139341 PA2139342 PA2139343 PA2139344 PA2139345 PA2139346 PA2139347 PA2139348 PA2139349 PA2139350 PA2139351 PA2139352 PA2139353 PA2139354 PA2139355 PA2139356 PA2139357 PA2139358 PA2139359 PA2139360 PA2139361 PA2139493 PA2139494 PA2139495 PA2139496 PA2139497 PA2139498 PA2139499 PA2139500 PA2139501 PA2139502 PA2139503 PA2139504 PA2139505 PA2139506 PA2139507 PA2139508 PA2139509 PA2139510 PA2139511 PA2139512 PA2139513 PA2139514 PA2139515 PA2139516 PA2139517 PA2139518 PA2139519 PA2139520 PA2139521 PA2139522 PA2139523 PA2139524 PA2139525 PA2139526 PA2139527 PA2139528 PA2139529 PA2139530 PA2139531 PA2139532 PA2139533 PA2139534 PA2139535 PA2139536 PA2139537 PA2139538 PA2139539 PA2182814 PA2184843 PA2184844 PA2184845 PA2184846 PA2184847 PA2184848 PA2184849 PA2184850 PA2184851 PA2184852 PA2184853 PA2184854 PA2184855 PA2184856 PA2184857 PA2184858 PA2184859 PA2184860 PA2184933 PA2184943 PA2300038 PA2300039 PA2300040 PA2300041 PA2300042 PA2300043 PA2300044 PA2300045 PA2300046 PA2300047 PA2300048 PA2300049 PA2300050 PA2300051 PA2300097 PA2300098 PA2300099 PA2300100 PA2300101 PA2300102 PA2300103 PA2300104 PA2300105 PA2300106 PA2300107 PA2300108 PA2300109 PA2300110 PA2300111 PA2300112 PA2300113 PA2300114 PA2300115 PA2300116 PA2300117 PA2300118 PA2300119 PA2300120 PA2300121 PA2300122 PA2300123 PA2300124 PA2300125 PA2300126 PA2300127 PA2300128 PA2300129 PA2300130 PA2300131 PA2300132 PA2300133 PA2300134 PA2300135 PA2300136 PA2300137 PA2300138 PA2300139 PA2300140 PA2300141 PA2300142 PA2300143 PA2300144 PA2300145 PA2300146 PA2300147 PA2300148 PA2300149 PA2300150 PA2300151 PA2300152 PA2300153 PA2300154 PA2300155 PA2300156 PA2300157 PA2300158 PA2300159 PA2300160 PA2300161 PA2300162 PA2300163 PA2300164 PA2300165 PA2300166 PA2300167 PA2300168 PA2300169 PA2300170 PA2300171 PA2300172 PA2300173 PA2300174 PA2300175 PA2300176 PA2300177 PA2300178 PA2300179 PA2300180 PA2300181 PA2300182 PA2300183 PA2300184 PA2300185 PA2300186 PA2300187 PA2300188 PA2300189 PA2300190 PA2300191 PA2300192 PA2300193 PA2300194 PA2300195 PA2300196 PA2300197 PA2300198 PA2300199 PA2300200 PA2300201 PA2300202 PA2300203 PA2300204 PA2300205 PA2300206 PA2300207 PA2300208 PA2300209 PA2300210 PA2300211 PA2300212 PA2300213 PA2300214 PA2300215 PA2300216 PA2300217 PA2300218 PA2300219 PA2300220 PA2300221 PA2300222 PA2300223 PA2300224 PA2300225 PA2300226 PA2300227 PA2300228 PA2300229 PA2300230 PA2300231 PA2300232 PA2300233 PA2300234 PA2300235 PA2300236 PA2300237 PA2300238 PA2300239 PA2300240 PA2300241 PA2300242 PA2300243 PA2300244 PA2300245 PA2300246 PA2300247 PA2300248 PA2300249 PA2300250 PA2300251 PA2300252 PA2300253 PA2300254 PA2300255 PA2300256 PA2300257 PA2300258 PA2300259 PA2300260 PA2300303 PA2302321 PA2302322 PA2312695 PA2312696 PA2312697 PA2312698 PA2312699 PA2312700 PA2312701 PA2312702 PA2312703 PA2312704 PA2312705 PA2312706 PA2312707 PA2312708 PA2312709 PA2312710 PA2312711 PA2312712 PA2312713 PA2312714 PA2312715 PA2312731 PA2312732 PA2312733 PA2312734 PA2312735 PA2312736 PA2312737 PA2312738 PA2312739 PA2312740 PA2312741 PA2312743 PA2312744 PA2312745 PA2312746 PA2312747 PA2312748 PA2312749 PA2312750 PA2312751 PA2312752 PA2312753 PA2312754 PA2312755 PA2312756 PA2312757 PA2312758 PA2312759 PA2312760 PA2312761 PA2312762 PA2312763 PA2312764 PA2312765 PA2312766 PA2312767 PA2312768 PA2312769 PA2312770 PA2312771 PA2312772 PA2312773 PA2312774 PA2312775 PA2312776 PA2312777 PA2312778 PA2312779 PA2312780 PA2312781 PA2312782 PA2312783 PA2312784 PA2312785 PA2312786 PA2312787 PA2312788 PA2312789 PA2312790 PA2312791 PA2312877 PA2312878 PA2312879 PA2312880 PA2312881 PA2312882 PA2312883 PA2312884 PA2312885 PA2312886 PA2312887 PA2312888 PA2312889 PA2312890 PA2312891 PA2312892 PA2312893 PA2312894 PA2312895 PA2312896 PA2312897 PA2312898 PA2312899 PA2312900 PA2312901 PA2312978 PA2312979 PA2312980 PA2312981 PA2312982 PA2312983 PA2312984 PA2312985 PA2312986 PA2312987 PA2312988 PA2312989 PA2312990 PA2312991 PA2312992 PA2312993 PA2312994 PA2312995 PA2312996 PA2312997 PA2312998 PA2312999 PA2313000 PA2313001 PA2313002 PA2313003 PA2313004 PA2313005 PA2313006 PA2313007 PA2313008 PA2313009 PA2313010 PA2313011 PA2313012 PA2313013 PA2313014 PA2313015 PA2313016 PA2313017 PA2313018 PA2313019 PA2313020 PA2313021 PA2313022 PA2313023 PA2313024 PA2313025 PA2313026 PA2313027 PA2313137 PA2313340 PA2313341 PA2313342 PA2313343 PA2313344 PA2313345 PA2313346 PA2313347 PA2313348 PA2313349 PA2313350 PA2313351 PA2313558 PA2313559 PA2313560 PA2313561 PA2313562 PA2313563 PA2313564 PA2313565 PA2313567 PA2313568 PA2313569 PA2313570 PA2313571 PA2313620 PA2313621 PA2313653 PA2313658 PA2313659 PA2313660 PA2313661 PA2313662 PA2313663 PA2313664 PA2313665 PA2313666 PA2313667 PA2313668 PA2313669 PA2313670 PA2313671 PA2313672 PA2313673 PA2313674 PA2313675 PA2313676 PA2313677 PA2313678 PA2313740 PA2313741 PA2313742 PA2313743 PA2313744 PA2313745 PA2313746 PA2313747 PA2313748 PA2313749 PA2313750 PA2313751 PA2313752 PA2313754 PA2313755 PA2313756 PA2313757 PA2313760 PA2313761 PA2313763 PA2313765 PA2313766 PA2313767 PA2313768 PA2313769 PA2313770 PA2313771 PA2313772 PA2313773 PA2313774 PA2313775 PA2313776 PA2313777 PA2313778 PA2313779 PA2313780 PA2313781 PA2313782 PA2313783 PA2313784 PA2313785 PA2313786 PA2313787 PA2313788 PA2313789 PA2313808 PA2313809 PA2313810 PA2313811 PA2313812 PA2313813 PA2313814 PA2313815 PA2313816 PA2313817 PA2313818 PA2313819 PA2313820 PA2313821 PA2313822 PA2313823 PA2313824 PA2313825 PA2313826 PA2313827 PA2313828 PA2313829 PA2313830 PA2313831 PA2313832 PA2313833 PA2313834 PA2313835 PA2313836 PA2313837 PA2313838 PA2313839 PA2313840 PA2313841 PA2313842 PA2313843 PA2313844 PA2313845 PA2313846 PA2313847 PA2313848 PA2313849 PA2313850 PA2313851 PA2313873 PA2313874 PA2313875 PA2313876 PA2313877 PA2313878 PA2313879 PA2313880 PA2313881 PA2313882 PA2313883 PA2313884 PA2313885 PA2313886 PA2313887 PA2313888 PA2313889 PA2313890 PA2313891 PA2313892 PA2313893 PA2313894 PA2313895 PA2313896 PA2313897 PA2313898 PA2313899 PA2313900 PA2313901 PA2313902 PA2313903 PA2313904 PA2313905 PA2313906 PA2313907 PA2313908 PA2313909 PA2313910 PA2313911 PA2313912 PA2314190 PA2314191 PA2314192 PA2314193 PA2314194 PA2314195 PA2314196 PA2314197 PA2314198 PA2314199 PA2314200 PA2314201 PA2314202 PA2314203 PA2314204 PA2314205 PA2314206 PA2314207 PA2314208 PA2314209 PA2314210 PA2314211 PA2314212 PA2314213 PA2314214 PA2314215 PA2314216 PA2314217 PA2314218 PA2314219 PA2314416 PA2314417 PA2314418 PA2314419 PA2314420 PA2314421 PA2314422 PA2314423 PA2314424 PA2314425 PA2314434 PA2314435 PA2314436 PA2314437 PA2314438 PA2314439 PA2314498 PA2314499 PA2314500 PA2314501 PA2314502 PA2314503 PA2314504 PA2314505 PA2314506 PA2314579 PA2314580 PA2314581 PA2314582 PA2314583 PA2314584 PA2314585 PA2314586 PA2314587 PA2314588 PA2314589 PA2314590 PA2314591 PA2314592 PA2314593 PA2314594 PA2314595 PA2314596 PA2314597 PA2314598 PA2314599 PA2314600 PA2314601 PA2314602 PA2314603 PA2314604 PA2314605 PA2314606 PA2314622 PA2314625 PA2314626 PA2314690 PA2314751 PA2314831 PA2314832 PA2314833 PA2314934 PA2314935 PA2314936 PA2314937 PA2314938 PA2314939 PA2314940 PA2314941 PA2314942 PA2314943 PA2314944 PA2314945 PA2314946 PA2314947 PA2314948 PA2352514 PA2352515 PA2352551 PA2352554 PA2352555 PA2352582 PA2352583 PA2352584 PA2352585 PA2352592 PA2352593 PA2352594 PA2352595 PA2352596 PA2352597 PA2352598 PA2352599 PA2352600 PA2352601 PA2352602 PA2352603 PA2352604 PA2352605 PA2352606 PA2352607 PA2352608 PA2352609 PA2352610 PA2352611 PA2352612 PA2352613 PA2352614 PA2352615 PA2352616 PA2352617 PA2352618 PA2352619 PA2352625 PA2352626 PA2352627 PA2352628 PA2352629 PA2352630 PA2352631 PA2352632 PA2352633 PA2352634 PA2352635 PA2352636 PA2352637 PA2352638 PA2352639 PA2352640 PA2352641 PA2352642 PA2352643 PA2352644 PA2352645 PA2352646 PA2352647 PA2352648 PA2352649 PA2352650 PA2352651 PA2352652 PA2352653 PA2352654 PA2352655 PA2352656 PA2352657 PA2352658 PA2352659 PA2352660 PA2352661 PA2352662 PA2352663 PA2352664 PA2352665 PA2352666 PA2352667 PA2352668 PA2352669 PA2352670 PA2352671 PA2352672 PA2352673 PA2352674 PA2352675 PA2352676 PA2352677 PA2352678 PA2352679 PA2352680 PA2352681 PA2352682 PA2352683 PA2352684 PA2352685 PA2352686 PA2352687 PA2352688 PA2352689 PA2352690 PA2352691 PA2352692 PA2352693 PA2352694 PA2352695 PA2352696 PA2352697 PA2352698 PA2352699 PA2352700 PA2352701 PA2352702 PA2352703 PA2352704 PA2352705 PA2352706 PA2352707 PA2352708 PA2352709 PA2352710 PA2352711 PA2352712 PA2352713 PA2352714 PA2352715 PA2352716 PA2352717 PA2352718 PA2352719 PA2352720 PA2352721 PA2352722 PA2352723 PA2352724 PA2352725 PA2352726 PA2352727 PA2352728 PA2352729 PA2352730 PA2352731 PA2352732 PA2352733 PA2352734 PA2352735 PA2352736 PA2352737 PA2352738 PA2352739 PA2352740 PA2352741 PA2352742 PA2352743 PA2352744 PA2352745 PA2352746 PA2352747 PA2352748 PA2352749 PA2352750 PA2352751 PA2352752 PA2352753 PA2352754 PA2352755 PA2352756 PA2352757 PA2352758 PA2352759 PA2352760 PA2352761 PA2352762 PA2352763 PA2352764 PA2352765 PA2352766 PA2352767 PA2352768 PA2352769 PA2352770 PA2352771 PA2352772 PA2352773 PA2352774 PA2352775 PA2352776 PA2352777 PA2352778 PA2352779 PA2352780 PA2352781 PA2352782 PA2352783 PA2352784 PA2352788 PA2352789 PA2352790 PA2352791 PA2352792 PA2352793 PA2352794 PA2352795 PA2352796 PA2352797 PA2352798 PA2352799 PA2352800 PA2352801 PA2352802 PA2352803 PA2352804 PA2352805 PA2352806 PA2352807 PA2352808 PA2352809 PA2352810 PA2352811 PA2352812 PA2352813 PA2352814 PA2352815 PA2352816 PA2352817 PA2352818 PA2352819 PA2352820 PA2352821 PA2352822 PA2352823 PA2352824 PA2352825 PA2352826 PA2352827 PA2352828 PA2352829 PA2352830 PA2352831 PA2352832 PA2352833 PA2352834 PA2352835 PA2352836 PA2352837 PA2352838 R3S04450035GA R3S04490127GA R3S04490129GA R3S04490131GA R3S04490133GA R3S04490134GA R3S04490137GA R3S04490138GA R3S04490139GA R3S05020263GA R3S05050286GA R3S05140399GA R3T05072105GA R3T05072110GA R3T05072111GA R3T05072114GA R3T05072120GA R3T05072123GA R3T05162726GA R3T05162733GA R3T05162736GA R3T05162737GA R3T05162738GA R3T05182961GA R3T05182962GA R3T05182963GA R3T05182966GA R3T05182967GA R3T05182969GA R3T05182970GA R3T05253351GA SBH05260052GA SBH05260053GA SBH05260056GA SBH05260057GA SBH05260058GA SBH05260081GA SBH05260108GA SBH05260109GA SBH05260111GA SBH05260112GA SBH05260113GA SBH05260117GA SBH05280122GA SBH05280124GA SBH05280207GA SBH05280212GA SBH05300380GA SBH05300387GA SBH05300388GA SBH05300396GA SBH05440948GA SBH05471036GA SBH05471066GA SBH05471073GA SBH06061578GA SBH06061580GA SBH06091715GA SBH06091716GA SBL06527807GR SBL06527809GR SBL06527811GR SBL06527822GR SBL07017823GR SBL07017892GR SBL07017894GR SBL07017895GR SBL07017901GR SBL07017906GR SBL07027934GR SBL07038115GR SBL07038163GR SBL07038169GR SBL07078548GR SBL07149225GR SBL07149226GR SBL07149227GR SBL07149229GR SBL07149230GR SBM05260121GR SBM05340190GR SBM05370247GR SBM05370253GR SBM05380320GA SBM05390382GR SBM05420430GR SBM05440465GR SBM05440472GR SBM05440478GA SBM06010826GR SBM06391194GA SCZ07200023GR SCZ07240337GR SCZ07240355GR SCZ07240358GR SCZ07361465GR SCZ07361467GR SCZ07361468GR SCZ07361472GR SCZ07361478GR SCZ07412925GR SCZ07423152GR SCZ07494359GR SCZ07494360GR SCZ07494361GR SCZ07494362GR SCZ07494364GR SCZ07504273GR SCZ07504370GR SCZ07504371GR SCZ07524793GR SE408100752GR SE408131482GR SE408131483GR SE408131484GR SE408131488GR SE408131489GR SE408131492GR SE408131509GR SE408131514GR SE408131518GR SE408141585GR SE408141602GR SE408232436GR SE408436627GR SE408456813GR SE408456814GR SE408456816GR SE408456817GR SE408456833GR SE408456835GR SE408456836GR SE408456837GR SE408456838GR SE408456839GR SE408456842GR SE408456843GR SE408456847GR SE408456849GR SE408456852GR SE408456858GR SE408456862GR SE408456863GR SE408456865GR SE408456868GR SE408466997GR SE408487343GR SE408487348GR SE408518061GR SE409119163GR SE409250439GR SE409250466GR SE409260470GR SE409260479GR SE409260481GR SE409341418GA SE409351619GR SE409371947GR SE409371950GR SE409432543GR SE409432638GA SE409432638GR SE409442685GA SE409442799GA SE409442815GA SE409442816GA SE409442817GA SE409442820GA SE409442824GA SE409442829GA SE409442830GA SE409442831GA SE409442832GA SE409442833GA SE409452834GA SE409452835GA SE409452836GA SE409452837GA SE409452838GA SE409452840GA SE409452841GA SE409452846GA SE409452848GA SE409452850GA SE409452851GA SE409452853GA SE409452854GA SE409452855GA SE409452856GA SE409452857GA SE409452858GA SE409452958GA SE409452961GA SE409452972GA SE409463005GA SE409463010GA SE409463015GA SE409463016GA SE409463025GA SE409463027GA SE409463030GA SE409463034GA SE409463035GA SE409463036GA SE409463038GA SE409463041GA SE409463043GA SE409463045GA SE409463046GA SE409463049GA SE409463050GA SE409463056GA SE409463058GA SE409463081GA SE409463089GA SE409463091GA SE409463093GA SE409463094GA SE409463096GA SE409463098GA SE409463103GA SE409463105GA SE409463106GA SE409463107GA SE409463110GA SE409463111GA SE409463116GA SE409463117GA SE409463119GA SE409463131GA SE409473140GA SE409473141GA SE409473142GA SE409473165GA SE409473169GA SE409473177GA SE409473178GA SE409473179GA SE409473182GA SE409473186GA SE409473187GA SE409473188GA SE409473192GA SE409473193GA SE409473196GA SE409473198GA SE409473200GA SE409473211GA SE409473212GA SE409473220GA SE409473222GA SE409473225GA SE409473226GA SE409473227GA SE409473228GA SE409473229GA SE409473230GA SE409473231GA SE409473232GA SE409473233GA SE409473234GA SE409473235GA SE409473236GA SE409473239GA SE409473240GA SE409473241GA SE409473242GA SE409473243GA SE409473244GA SE409473245GA SE409473246GA SE409473252GA SE409473253GA SE409473256GA SE409493288GA SE409493289GA SE409493290GA SE409493291GA SE409493292GA SE409493293GA SE409493294GA SE409493295GA SE409493307GA SE409493315GA SE409493316GA SE409493317GA SE409493318GA SE409493319GA SE409493321GA SE409493322GA SE409493324GA SE409493337GA SE409493338GA SE409493339GA SE409493341GA SE409493342GA SE409493343GA SE409493345GA SE409493346GA SE409493347GA SE409493348GR SE409493349GA SE409493350GA SE409493351GA SE409493352GA SE409493353GA SE409493362GA SE409493363GA SE409493364GA SE409493365GA SE409493371GA SE409493372GA SE409493373GA SE409493374GA SE409493375GR SE409493376GA SE409493378GA SE409493381GA SE409493383GA SE409493384GA SE409493385GR SE409493386GA SE409493388GA SE409493389GA SE409493390GA SE409493391GA SE409493404GA SE409493408GA SE409493409GA SE409493412GA SE409493413GA SE409503416GA SE409503418GA SE409503419GA SE409503420GA SE409503426GA SE409503427GA SE409503450GR SE409503455GA SE409503457GA SE409503459GA SE409503462GA SE409503464GA SE409503467GA SE409503468GA SE409503471GA SE409503473GA SE409503474GA SE409503475GA SE409503476GA SE409503477GA SE409503478GA SE409503479GA SE409503480GA SE409503481GA SE409503483GA SE409503485GA SE409503496GA SE409503499GA SE409503500GA SE409503501GA SE
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1245-2014
Related Recalls
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.