Focalyx Fusion
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Focalyx Technologies, LLC.
- Reason for Recall:
- Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Focalyx Fusion
Product Codes/Lot Numbers:
UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1243-2026