LFT v40 Femoral Head; Catalog Number 6260-9-032; V40 CoCr LFit Head 32mm/-4. Designed to be assembled onto a femoral hip stem trunnion during surgery for a total hip arthroplasty.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Code 48681201
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Stryker Howmedica Osteonics Corp.
- Reason for Recall:
- Three customer complaints were received for the same batch of V40 LFIT Vitallium femoral heads (part number 6260-9-032, lot 48681201) alleging that the femoral head couldn't be assembled with its corresponding V40 stem trunnion.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
LFT v40 Femoral Head; Catalog Number 6260-9-032; V40 CoCr LFit Head 32mm/-4. Designed to be assembled onto a femoral hip stem trunnion during surgery for a total hip arthroplasty.
Product Codes/Lot Numbers:
Lot Code 48681201
Distribution:
Distributed in: GA, MA, NC, NY, PA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1241-2015
Related Recalls
Stryker is recalling Hoffmann LRF (Limb Reconstruction Frame) System Safety Clip - Bone Transport Strut because there is a potential for loosening of the safety clips from the intended position on the Quick Release mechanism of the Transport Struts.
Reunion TSA Peg Alignment Sound broke during surgery.
GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-5, Schonkirchen, Germany 24232; T2 K-Wire, Sterile R Intended for the temporary stabilization of bone segments or fragments
Stryker Howmedica Osteonics
A review of packaging revealed the seal integrity of the pouch may be compromised. More specifically, there is a potential that the sterile pouch is not sealed at one end due to a manufacturing error.