ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV13F
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lots: N18573, N18687-1
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- OriGen Biomedical, Inc.
- Reason for Recall:
- Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV13F
Product Codes/Lot Numbers:
Lots: N18573, N18687-1
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1230-2019
Related Recalls
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
It was identified that tissue culture bags were incorrectly labeled as free of phthalates.
3/8" FloPump, 32mL, (Sterile), REF 6400S, UDI: (01) 081432102013 1, Sterile EO, Rx Only
OriGen Biomedical
The firm has become aware that they have distributed product that was expired.