VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
Class I - DangerousWhat Should You Do?
- Check if you have this product: VITROS System Software Version 3.8.1 Regular UDI-DI 10758750002740 Refurbished UD-DI 10758750007110 Serial Numbers US: J56001885 J56003484 J56001886 J56003541 J56001910 J56003570 J56002308 J56003583 J56002461 J56003646 J56002474 J56003669 J56002608 J56003670 J56002698 J56004068 J56002704 J56004137 J56003342 J56004174 J56003356 J56004491 J56003377 J56004506 J56003433 J56004549 J56003477 J56004551 J56003478 J56004577 OUS: J56000425 J56002518 J56000512 J56002615 J56000533 J56002625 J56000674 J56002637 J56000680 J56002926 J56000719 J56002929 J56000809 J56003061 J56000831 J56003260 J56000836 J56003285 J56001051 J56003401 J56001111 J56003464 J56001114 J56003508 J56001208 J56003523 J56001731 J56003538 J56001765 J56004346 J56001826 J56004355 J56001853 J56004415 J56002085 J56004531 J56002116 J56004695
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ortho-Clinical Diagnostics, Inc.
- Reason for Recall:
- A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
VITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
Product Codes/Lot Numbers:
VITROS System Software Version 3.8.1 Regular UDI-DI 10758750002740 Refurbished UD-DI 10758750007110 Serial Numbers US: J56001885 J56003484 J56001886 J56003541 J56001910 J56003570 J56002308 J56003583 J56002461 J56003646 J56002474 J56003669 J56002608 J56003670 J56002698 J56004068 J56002704 J56004137 J56003342 J56004174 J56003356 J56004491 J56003377 J56004506 J56003433 J56004549 J56003477 J56004551 J56003478 J56004577 OUS: J56000425 J56002518 J56000512 J56002615 J56000533 J56002625 J56000674 J56002637 J56000680 J56002926 J56000719 J56002929 J56000809 J56003061 J56000831 J56003260 J56000836 J56003285 J56001051 J56003401 J56001111 J56003464 J56001114 J56003508 J56001208 J56003523 J56001731 J56003538 J56001765 J56004346 J56001826 J56004355 J56001853 J56004415 J56002085 J56004531 J56002116 J56004695
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1226-2024
Related Recalls
VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
Ortho-Clinical Diagnostics
Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.