Simplexa COVID-19 Direct, REF: MOL4150; including Reaction Mix, REF: MOL4150; with previous assay definition v2
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 20816101027020. Product REF/Lot/Reaction Mix REF/Lot: MOL4150/US13950/MOL4151/US14091, MOL4150/US14913/MOL4151/US14063, MOL4150/US13952/MOL4151/US14065, MOL4150/US13955/MOL4151/US14086, MOL4150/US13959/MOL4151/US14088, MOL4150/US13956/MOL4151/US14087, MOL4150/US13960/MOL4151/US14093, MOL4150/US13961/MOL4151/US14094, MOL4150/US13973/MOL4151/US13732, MOL4150/US13974/MOL4151/US13736, MOL4150/US14006/MOL4151/US13738, MOL4150/US14004/MOL4151/US15064, MOL4150/US14005/MOL4151/US13737, MOL4150/US14008/MOL4151/US14100, MOL4150/US14007/MOL4151/US14099, MOL4150/US14010/MOL4151/US14102, MOL4150/US14011/MOL4151/US14103, MOL4150/US14012/MOL4151/US15290, MOL4150/US14018/MOL4151/US14111, MOL4150/US14009/MOL4151/US14101, MOL4150/US14017/MOL4151/US14109, MOL4150/15061N/MOL4151/13474N, MOL4150/X15475N/MOL4151/X15479N, MOL4150/13330N/MOL4151/14829N, MOL4150/13331N/MOL4151/13751N, MOL4150/13336N/MOL4151/13752N, MOL4150/13968N/MOL4151/14036N, MOL4150/14346N/MOL4151/14038N, MOL4150/13967N/MOL4151/15082N
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- DiaSorin Molecular LLC
- Reason for Recall:
- Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Simplexa COVID-19 Direct, REF: MOL4150; including Reaction Mix, REF: MOL4150; with previous assay definition v2
Product Codes/Lot Numbers:
UDI-DI: 20816101027020. Product REF/Lot/Reaction Mix REF/Lot: MOL4150/US13950/MOL4151/US14091, MOL4150/US14913/MOL4151/US14063, MOL4150/US13952/MOL4151/US14065, MOL4150/US13955/MOL4151/US14086, MOL4150/US13959/MOL4151/US14088, MOL4150/US13956/MOL4151/US14087, MOL4150/US13960/MOL4151/US14093, MOL4150/US13961/MOL4151/US14094, MOL4150/US13973/MOL4151/US13732, MOL4150/US13974/MOL4151/US13736, MOL4150/US14006/MOL4151/US13738, MOL4150/US14004/MOL4151/US15064, MOL4150/US14005/MOL4151/US13737, MOL4150/US14008/MOL4151/US14100, MOL4150/US14007/MOL4151/US14099, MOL4150/US14010/MOL4151/US14102, MOL4150/US14011/MOL4151/US14103, MOL4150/US14012/MOL4151/US15290, MOL4150/US14018/MOL4151/US14111, MOL4150/US14009/MOL4151/US14101, MOL4150/US14017/MOL4151/US14109, MOL4150/15061N/MOL4151/13474N, MOL4150/X15475N/MOL4151/X15479N, MOL4150/13330N/MOL4151/14829N, MOL4150/13331N/MOL4151/13751N, MOL4150/13336N/MOL4151/13752N, MOL4150/13968N/MOL4151/14036N, MOL4150/14346N/MOL4151/14038N, MOL4150/13967N/MOL4151/15082N
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1209-2023
Related Recalls
There is a potential for the direct amplification disc to malfunction which may result in spillage of liquid from the disc.
Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.
DiaSorin Molecular
Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455
DiaSorin Molecular
A subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV Direct Gen II. Leakage from a well that had a positive sample may cause further cross-contamination in areas of the lab if not properly cleaned.