HeartWare HVAD Pump Implant Kit, a) Model Number 1104JP, b) Model Number MCS1705PU

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a) Model Number 1104JP: GTIN: 00763000170967 Serial numbers: HW41784, HW41825, HW41829, HW41849, HW41852, HW41853, HW41863, HW42312, HW42352, HW42775, HW42818, HW42879, HW42882, and HW42884. b) Model Number MCS1705PU: GTIN 00888707005364, Serial Numbers: HW42448 and HW42571; GTIN 00888707007139, Serial Numbers: HW42955 and HW42962; GTIN 00888707009041, Serial Numbers HW43483 and HW43484; and GTIN 00888707003070, Serial number HW42153.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Inc
Reason for Recall:
The pump may have a welding defect that can lead the pump to malfunction. The issue is under investigation by the firm. It is suspected that a weld defect on the center post cap may allow moisture into the center post and corrode the magnets that keep the impeller rotating concentrically, leading to unintended contact between the center post and impeller. If this occurs, it may affect blood flow through the pump. The issue may present clinical signs and symptoms that resemble pump thrombosis.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

HeartWare HVAD Pump Implant Kit, a) Model Number 1104JP, b) Model Number MCS1705PU

Product Codes/Lot Numbers:

a) Model Number 1104JP: GTIN: 00763000170967 Serial numbers: HW41784, HW41825, HW41829, HW41849, HW41852, HW41853, HW41863, HW42312, HW42352, HW42775, HW42818, HW42879, HW42882, and HW42884. b) Model Number MCS1705PU: GTIN 00888707005364, Serial Numbers: HW42448 and HW42571; GTIN 00888707007139, Serial Numbers: HW42955 and HW42962; GTIN 00888707009041, Serial Numbers HW43483 and HW43484; and GTIN 00888707003070, Serial number HW42153.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1207-2022

Related Recalls

ventricular fibrillation (VF) during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.

Mar 11, 2026 Surgical Instruments View Details →

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

Jun 18, 2025 Infusion Pumps Nationwide View Details →