RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Product Usage: The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software revisions of affected RT Elements applications: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Unique Identifier (UDI): GTIN for affected product(s): Cranial SRS 1.0 -7 04056481000554 Cranial SRS 1.5 -704056481140403 Spine SRS 1.0 - 04056481000547 Spine SRS 1.5 - 04056481140380 Multiple Brain Mets SRS 1.5 7 04056481006853 RT QA 1.0 -7 04056481007119 RT QA 1.5 -7 04056481140397 Model/catalogue numbers: 21702A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21702-04C ELEMENTS MULTIPLE BRAIN METS SRS INST 21702-0 6 ELEMENTS MULTIPLE BRAIN METS SRS LICENSE 21702-12A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21702-36A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21702-60A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21712A ELEMENTS MULTIPLE BRAIN METS SRS 21722A ELEMENTS MULTIPLE BRAIN METS SRS SUB FOC 21722-12A ELEMENTS MULTIPLE BRAIN METS SRS SUB FOC 21741A UPD TO ELEMENTS BRAIN MET PLAN 21742 ELEMENTS CRANIAL SRS 21742A ELEMENTS CRANIAL SRS 21743 ELEMENTS CRANIAL SRS SUBSCRIPTION 21743A ELEMENTS CRANIAL SRS SUBSCRIPTION 21743-01 ELEMENTS CRANIAL SRS LICENSE ELEMENTS CRANIAL SRS LICENSE 1.5 ELEMENTS CRANIAL SRS INSTALLER 21743-02A ELEMENTS CRANIAL SRS INSTALLER 1.5 21743-12 ELEMENTS CRANIAL SRS SUBSCRIPTION 21743-36 ELEMENTS CRANIAL SRS SUBSCRIPTION 21743-60 ELEMENTS CRANIAL SRS SUBSCRIPTION 21744 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744A ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744-12 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744-36 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744-60 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21745 ELEMENTS CRANIAL SRS (UPD) 21745A ELEMENTS CRANIAL SRS (UPD) 21747 ELEMENTS SPINE SRS 21747A ELEMENTS SPINE SRS 21748 ELEMENTS SPINE SRS SUBSCRIPTION 21748A ELEMENTS SPINE SRS SUBSCRIPTION 21748-01 ELEMENTS SPINE SRS LICENSE ELEMENTS SPINE SRS LICENSE 1.5 ELEMENTS SPINE SRS INSTALLER 21748-02A ELEMENTS SPINE SRS INSTALLER 1.5 21748-12 ELEMENTS SPINE SRS SUBSCRIPTION 21748-36 ELEMENTS SPINE SRS SUBSCRIPTION 21748-60 ELEMENTS SPINE SRS SUBSCRIPTION 21749 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749A ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749-12 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749-36 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749-60 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21750 ELEMENTS SPINE SRS (UPD) 21750A ELEMENTS SPINE SRS (UPD) 21762 RT QA 1.0 21762A RT QA 1.5 21762-01 RT QA LICENSE RT QA LICENSE 1.5 RT QA INSTALLER 21762-02A RT QA INSTALLER 1.5
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Brainlab AG
- Reason for Recall:
- There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Product Usage: The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions.
Product Codes/Lot Numbers:
Software revisions of affected RT Elements applications: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Unique Identifier (UDI): GTIN for affected product(s): Cranial SRS 1.0 -7 04056481000554 Cranial SRS 1.5 -704056481140403 Spine SRS 1.0 - 04056481000547 Spine SRS 1.5 - 04056481140380 Multiple Brain Mets SRS 1.5 7 04056481006853 RT QA 1.0 -7 04056481007119 RT QA 1.5 -7 04056481140397 Model/catalogue numbers: 21702A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21702-04C ELEMENTS MULTIPLE BRAIN METS SRS INST 21702-0 6 ELEMENTS MULTIPLE BRAIN METS SRS LICENSE 21702-12A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21702-36A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21702-60A ELEMENTS MULTIPLE BRAIN METS SRS SUB. 21712A ELEMENTS MULTIPLE BRAIN METS SRS 21722A ELEMENTS MULTIPLE BRAIN METS SRS SUB FOC 21722-12A ELEMENTS MULTIPLE BRAIN METS SRS SUB FOC 21741A UPD TO ELEMENTS BRAIN MET PLAN 21742 ELEMENTS CRANIAL SRS 21742A ELEMENTS CRANIAL SRS 21743 ELEMENTS CRANIAL SRS SUBSCRIPTION 21743A ELEMENTS CRANIAL SRS SUBSCRIPTION 21743-01 ELEMENTS CRANIAL SRS LICENSE ELEMENTS CRANIAL SRS LICENSE 1.5 ELEMENTS CRANIAL SRS INSTALLER 21743-02A ELEMENTS CRANIAL SRS INSTALLER 1.5 21743-12 ELEMENTS CRANIAL SRS SUBSCRIPTION 21743-36 ELEMENTS CRANIAL SRS SUBSCRIPTION 21743-60 ELEMENTS CRANIAL SRS SUBSCRIPTION 21744 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744A ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744-12 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744-36 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21744-60 ELEMENTS CRANIAL SRS SUBSCRIPTION FOC 21745 ELEMENTS CRANIAL SRS (UPD) 21745A ELEMENTS CRANIAL SRS (UPD) 21747 ELEMENTS SPINE SRS 21747A ELEMENTS SPINE SRS 21748 ELEMENTS SPINE SRS SUBSCRIPTION 21748A ELEMENTS SPINE SRS SUBSCRIPTION 21748-01 ELEMENTS SPINE SRS LICENSE ELEMENTS SPINE SRS LICENSE 1.5 ELEMENTS SPINE SRS INSTALLER 21748-02A ELEMENTS SPINE SRS INSTALLER 1.5 21748-12 ELEMENTS SPINE SRS SUBSCRIPTION 21748-36 ELEMENTS SPINE SRS SUBSCRIPTION 21748-60 ELEMENTS SPINE SRS SUBSCRIPTION 21749 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749A ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749-12 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749-36 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21749-60 ELEMENTS SPINE SRS SUBSCRIPTION FOC 21750 ELEMENTS SPINE SRS (UPD) 21750A ELEMENTS SPINE SRS (UPD) 21762 RT QA 1.0 21762A RT QA 1.5 21762-01 RT QA LICENSE RT QA LICENSE 1.5 RT QA INSTALLER 21762-02A RT QA INSTALLER 1.5
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1205-2019
Related Recalls
Under certain circumstances, it can occur that patient records from unrelated individuals are automatically merged into one patient record without user notification.
Cirq Arm System 2.0
Brainlab AG
Risk of mechanical instability due to potential manufacturing error of the device
Deep Inspiration Breath Hold (DIBH) functionality may not work as specified when based on a standard-workflow Treatment Template