CentraLink¿ Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location. Siemens Healthcare Diagnostics has determined that there is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product codes: 10485061, 10484212, 10485064, 10483439, 10471393, 10484552, 10814296, 10815474, 11220480, 10481936, 10481937, 10481938, 10482170, 10482936, 10483437, 10483438, 10483440, 10484211, 10484213, 10485060, 10485062, 10485063, 10703007, 10703008, 10711638, 10711639, 10711640, 10711641, 10711643, 10811596, 10811597, 10811598, 10811599, 10811601, 10811741, 10811742, 10811743, 10814298, 10814877, 10814880, 10814881, 10815694, 10816519, 10816520, 10816521, 10816522, 10816523, 10816524, 10817180, 10817181, 10819760, 10819943, 11219840, 11219841.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- There is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type. This can occur when an order is received from the LIS without a sample type, requiring that the sample type be set in CentraLink based on the sample type of the test in the order.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CentraLink¿ Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location. Siemens Healthcare Diagnostics has determined that there is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type.
Product Codes/Lot Numbers:
Product codes: 10485061, 10484212, 10485064, 10483439, 10471393, 10484552, 10814296, 10815474, 11220480, 10481936, 10481937, 10481938, 10482170, 10482936, 10483437, 10483438, 10483440, 10484211, 10484213, 10485060, 10485062, 10485063, 10703007, 10703008, 10711638, 10711639, 10711640, 10711641, 10711643, 10811596, 10811597, 10811598, 10811599, 10811601, 10811741, 10811742, 10811743, 10814298, 10814877, 10814880, 10814881, 10815694, 10816519, 10816520, 10816521, 10816522, 10816523, 10816524, 10817180, 10817181, 10819760, 10819943, 11219840, 11219841.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1204-2017
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.