Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Code: UDI: 0735000201006820171130, 0735000201011220180608 and 0735000201013620180928 GTIN: 07350002010068, 07350002010112 and 07350002010136 Serial Numbers: 7.0.0.19, 8.0.0.61 and 8.0.1.10 Software Revision: RayStation 7, 8A and 8A Service Pack 1 Expiration Date: 2023-10-04
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
RAYSEARCH LABORATORIES AB
Reason for Recall:
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1

Product Codes/Lot Numbers:

Lot Code: UDI: 0735000201006820171130, 0735000201011220180608 and 0735000201013620180928 GTIN: 07350002010068, 07350002010112 and 07350002010136 Serial Numbers: 7.0.0.19, 8.0.0.61 and 8.0.1.10 Software Revision: RayStation 7, 8A and 8A Service Pack 1 Expiration Date: 2023-10-04

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1185-2025

Related Recalls

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Nov 28, 2025 Other Medical Devices View Details →

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Nov 28, 2025 Other Medical Devices View Details →

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Nov 28, 2025 Other Medical Devices View Details →