Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code 0042-0000-US UDI-DI 00813502010022; Product Code 0042-0010-US UDI-DI 00813502010985; Product Code 0042-0040-US UDI-DI 00813502011401; Software Versions V8.4 and V8.4.1
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abiomed, Inc.
- Reason for Recall:
- Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
Product Codes/Lot Numbers:
Product Code 0042-0000-US UDI-DI 00813502010022; Product Code 0042-0010-US UDI-DI 00813502010985; Product Code 0042-0040-US UDI-DI 00813502011401; Software Versions V8.4 and V8.4.1
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1175-2024
Related Recalls
Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.