Barco MDNC-3421 DISPLAY, DIAGNOSTIC RADIOLOGY

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: MDNC-3421, Serial number 2590263533, UDI/DI 05415334028388.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Barco N.V.
Reason for Recall:
A number of MDNC-3421 displays were shipped while not being DICOM-compliant. These displays will not pass DICOM acceptance tests.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Barco MDNC-3421 DISPLAY, DIAGNOSTIC RADIOLOGY

Product Codes/Lot Numbers:

Model Number: MDNC-3421, Serial number 2590263533, UDI/DI 05415334028388.

Distribution:

Distributed in: WI

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1172-2021

Related Recalls

Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures

May 9, 2025 Other Medical Devices Nationwide View Details →

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Apr 24, 2024 Other Medical Devices View Details →
Class I - Dangerous

Barco became aware of a problem with I2C cables in the manufacturing process of a limited batch of MDSC-85xx NXF products. The functionality of the built-in MNA decoder is blocked.

Dec 20, 2023 Other Medical Devices Nationwide View Details →