NeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computed X-ray, Tomography system. The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120020, N16E120022, N16E120023, N16E120024, N16E120052, N16E120053, and N16E130002.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips And Neusoft Medical Systems Co., Ltd.
- Reason for Recall:
- User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
NeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computed X-ray, Tomography system. The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients
Product Codes/Lot Numbers:
Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120020, N16E120022, N16E120023, N16E120024, N16E120052, N16E120053, and N16E130002.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1165-2013
Related Recalls
NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Philips and Neusoft Medical Systems Co.
The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be
NeuViz Dual series CT Scanner System, Part number (PN): 989605651321. This is a computed X-ray, Tomography system. NeuViz Dual Multi-slice CT Scanner System is intended to produce cross-section images of head and body by computer reconstruction of X-ray transmission data taken at different angles.
Philips And Neusoft Medical Systems Co.
User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Philips And Neusoft Medical Systems Co.
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.