Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 50195594381965(cs) 10195594381967(ea) Lot Number: 706265 Exp. Date: 5/1/2024
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Cardinal Health 200, LLC
Reason for Recall:
Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A

Product Codes/Lot Numbers:

UDI-DI: 50195594381965(cs) 10195594381967(ea) Lot Number: 706265 Exp. Date: 5/1/2024

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1158-2023

Related Recalls

The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.