Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    ALL LOT #s 2013010311 and lower
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Smith & Nephew Inc.
Reason for Recall:
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Product Codes/Lot Numbers:

ALL LOT #s 2013010311 and lower

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1157-2014

Related Recalls

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

Mar 18, 2025 Other Medical Devices Nationwide View Details →