J-Plasma(R) Handpiece, Catalog Numbers: GS-018C, GS-270C, BVX-270B
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots manufactured since 01/01/2014
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Bovie Medical Corporation
- Reason for Recall:
- Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
J-Plasma(R) Handpiece, Catalog Numbers: GS-018C, GS-270C, BVX-270B
Product Codes/Lot Numbers:
All lots manufactured since 01/01/2014
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1155-2018
Related Recalls
There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.
There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.
There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.